    Intended Use

    'RapidOne-OXY' Test is a one-step, lateral flow immunoassay for the detection of oxycodone in urine.

    'RapidOne-Oxy' Test is intended for the qualitative detection of oxycodone in human urine at 100 ng/ml.

    'RapidOne-OXY' Test is intended for professional use. It is not intended for over the counter sales to nonprofessionals. The assay is easy to perform, but should not be used without proper supervision. This immunoassay is a simplified qualitative screening method that provides only a preliminary result for use in determining the need for additional or confirmatory testing, i.e. gaschromatography/mass spectrometry (GC/MS.)

    'RapidOne-OXY' Test provides only a preliminary analytical result. A more specific alternate chemical method must be used in order to obtain a more confirmed result. GC/MS is the preferred confirmatory method. Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary results are used.

    Device Description

    The assay employed in the 'RapidOne-OXY' Test is based on the same principle of highly specific reaction between antigens and antibodies.

    This assay is a one-step, immunoassay in which a specially labeled drug (drug conjugate) competes with drug that may be present in the sample for the limited number of binding sites on the antibody. The test device consists of a membrane strip onto which a drug conjugate has been immobilized. A colloidal gold-antibody complex is dried at one end of a membrane. In the absence of any drug in the urine sample, the colloidal goldantibody moves with the urine by capillary action to contact the immobilized drug conjugate. An antibody-antigen reaction occurs forming a visible line in the 'test' area. The formation of a visible line in the 'test' area occurs when the test is negative.

    When drug is present in the urine sample, the drug or metabolite will compete with the immobilized drug conjugate in the test area for the limited antibody sites on the colloidal gold-antibody complex. If sufficient amount of drug is present, it will fill all of the available binding sites, thus preventing attachment of the labeled antibody to the drug conjugate. An absence of a color band (line) in the 'test' area is indicative of a positive result.

    A control band (line), comprised of a different antibody/antigen reaction, is present on the membrane strip. The 'control; line is not influenced by the presence or absence of drug in the urine, and therefore, should be present in all reactions.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study details for the 'RapidOne-OXY' Test based on the provided text:

    Acceptance Criteria and Device Performance for 'RapidOne-OXY' Test

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Detection of OxycodoneDetects 100 ng/ml of oxycodone in urine.
    Qualitative ScreeningProvides a preliminary qualitative result for oxycodone in human urine.
    Agreement with GC/MS (Negative Samples)Correctly identified all 50 drug-free samples as negative.
    Agreement with GC/MS (Positive Samples)Identified positive specimens (ranging from 54 to 1025 ng/ml) as positive.
    Reproducibility with No Drug Present80/80 (100%) negative results.
    Reproducibility at 50 ng/ml8/80 (10%) positive, 72/80 (90%) negative results.
    Reproducibility at 75 ng/ml65/80 (81.25%) positive, 15/80 (18.75%) negative results.
    Reproducibility at 100 ng/ml (Cut-off)80/80 (100%) positive results.
    Reproducibility at 125 ng/ml80/80 (100%) positive results.

    Note on Acceptance Criteria: The document primarily describes the intended performance and target detection limit, rather than explicit, pre-defined quantitative acceptance criteria with specific thresholds for sensitivity, specificity, or agreement. The "Reproducibility" table shows data around the cut-off, which implicitly demonstrates expected performance at and around that threshold. However, specific percentage targets for "acceptable" performance at these concentrations are not stated. The overall conclusion states "substantially equivalent to GC/MS," which is the overarching acceptance criterion for 510(k) clearances.


    2. Sample Size and Data Provenance

    • Test Set Sample Size: 90 samples (50 drug-free, 40 positive).
    • Data Provenance: Not explicitly stated (e.g., country of origin). The study is retrospective in nature, as samples were "selected for evaluation" and "tested as positive by Syva Emit II" and "confirmed as positive and quantified by GC/MS" prior to testing with the 'RapidOne-OXY' device.

    3. Number and Qualifications of Experts for Ground Truth

    • Not applicable. The ground truth was established by laboratory methods (Syva Emit II and GC/MS), not by human experts.

    4. Adjudication Method for the Test Set

    • Not applicable. Ground truth was established by laboratory methods (Syva Emit II and GC/MS), not requiring human adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a MRMC comparative effectiveness study was not done. This device is an in-vitro diagnostic (IVD) test, not a medical imaging or diagnostic aid device that typically involves human readers assisted by AI. The study directly compares the device's performance to a reference standard (GC/MS).

    6. Standalone Performance Study

    • Yes, a standalone performance study was done. The described performance characteristics directly evaluate the 'RapidOne-OXY' Test's ability to detect oxycodone in urine independently against a reference standard (GC/MS) and its reproducibility.

    7. Type of Ground Truth Used

    • Laboratory Reference Standard:
      • GC/MS (Gas Chromatography/Mass Spectrometry): This was the primary confirmatory and quantitative method used to establish the true oxycodone concentrations in the positive samples.
      • Syva Emit II: Used to initially identify the 40 positive samples before GC/MS confirmation and quantification.

    8. Sample Size for the Training Set

    • Not applicable / Not explicitly stated. This device is an immunoassay (an antibody-based test), not a machine learning or AI-driven algorithm that typically requires a distinct "training set" in the context of device performance studies. The development of the assay itself would involve optimization and testing, but the provided documentation focuses on the validation of the finished product against a reference standard.

    9. How Ground Truth for the Training Set Was Established

    • Not applicable. As noted above, there isn't a traditional "training set" in the context of AI/ML for this immunoassay. The development of the specific antibodies and the assay format would have been based on established biochemical principles and experimental validation during product development, but this isn't detailed as a specific "training set" for performance evaluation.