510(k) Data Aggregation

    K Number
    K130911

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-05-24

    (52 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RapidCross™ PTA Rapid Exchange Balloon Dilatation Catheter is intended to dilate stenoses in the iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature.

    Device Description

    The RapidCross PTA Rapid Exchange Balloon Dilatation Catheter (RapidCross catheter) is a rapid exchange (RX) coaxial catheter compatible with 0.014" guidewires, with a distally mounted semi-compliant inflatable balloon and an atraumatic, tapered tip. The distal portion of the catheter has a lubricious coating. The manifold includes an inflation lumen, which is used to inflate and deflate the dilatation balloon with a mixture of

    AI/ML Overview

    The provided text describes a 510(k) summary for the RapidCross™ PTA Rapid Exchange Balloon Dilatation Catheter, a medical device. This document focuses on demonstrating substantial equivalence to predicate devices through bench testing and biocompatibility testing, not on clinical acceptance criteria or a study proving performance against such criteria in the context of diagnostic or AI-driven systems.

    Therefore, most of the requested information regarding acceptance criteria, study details, sample sizes, ground truth establishment, expert involvement, and MRMC studies is not applicable to this type of device submission. This is a premarket notification for a physical medical device, not a diagnostic algorithm or image analysis software.

    Here's a breakdown of the relevant (and non-relevant) information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: The document states that "The RapidCross catheter met all acceptance criteria for the bench testing with results similar to the predicates." However, specific numerical acceptance criteria for each test (e.g., "Balloon Burst Strength must be > X psi") are not detailed in this summary.
    • Reported Device Performance: The document lists the types of tests performed. It doesn't provide specific numerical results, but rather states that the device performed "similar to the predicates" and "met all acceptance criteria."
    Test PerformedReported Device Performance
    Crossing ProfileMet acceptance criteria; similar to predicates
    Balloon Burst StrengthMet acceptance criteria; similar to predicates
    Balloon ComplianceMet acceptance criteria; similar to predicates
    Balloon ODMet acceptance criteria; similar to predicates
    Inflation/Deflation TimeMet acceptance criteria; similar to predicates
    Balloon FatigueMet acceptance criteria; similar to predicates
    Bond Tensile StrengthMet acceptance criteria; similar to predicates
    KinkMet acceptance criteria; similar to predicates
    Device TrackingMet acceptance criteria; similar to predicates
    Insertion ForceMet acceptance criteria; similar to predicates
    Balloon Pull-back ForceMet acceptance criteria; similar to predicates
    Repeat Inflations (In Stent)Met acceptance criteria; similar to predicates
    Torque StrengthMet acceptance criteria; similar to predicates
    RadiopacityMet acceptance criteria; similar to predicates
    Presence of CoatingMet acceptance criteria; similar to predicates
    Coating DurabilityMet acceptance criteria; similar to predicates
    Particle GenerationMet acceptance criteria; similar to predicates
    PushabilityMet acceptance criteria; similar to predicates
    Support Wire SecurementMet acceptance criteria; similar to predicates
    Tip ID / RX Port IDMet acceptance criteria; similar to predicates
    Tip / Lesion Entry ProfileMet acceptance criteria; similar to predicates
    Re-Insertion ForceMet acceptance criteria; similar to predicates
    Catheter Working LengthMet acceptance criteria; similar to predicates
    RX Port LengthMet acceptance criteria; similar to predicates
    RX Port ODMet acceptance criteria; similar to predicates
    Biocompatibility (cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemolysis, pyrogen, complement activation, thromboresistance, partial thromboplastin time, platelet/leukocyte count)Met acceptance criteria (per ISO 10993-1 for short duration contact); similar to predicates

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not specified for any of the bench tests.
    • Data Provenance: The testing was "Bench testing" and "biocompatibility testing." This is laboratory/engineering data, not clinical data from patients. Country of origin is not relevant for this type of testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. This relates to physical device performance and biocompatibility, not diagnostic interpretation requiring expert ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. This applies to clinical data interpretation, not bench testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is a physical medical device, not an AI or diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Ground Truth: For bench testing, the "ground truth" is typically defined by engineering specifications, validated test methods, and industry standards (e.g., ISO, ASTM). For biocompatibility, it's defined by ISO 10993-1.

    8. The sample size for the training set

    • Not applicable. This is a physical medical device, not an algorithm that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable. This is a physical medical device.