510(k) Data Aggregation

    K Number
    K211770

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-12-25

    (200 days)

    Product Code
    Regulation Number
    888.3040
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The R3ACT™ Stabilization System is intended as an adjunct in fracture repair and ligamentous injuries of small bones of the feet and ankles including the distal tibia, distal fibula, talus, and calcaneus, and as an adjunct in external and intramedullary fixation systems involving plates and rods. Specifically, the R3ACT™ Stabilization System is intended to provide fixation during the healing process following a syndesmotic trauma, such as fixation of syndesmosis (syndesmosis disruptions) in connection with Weber B and C ankle fractures.

    Device Description

    The R3ACT™ Stabilization System is a fixation device comprised of a titanium alloy screw, UHMWPE suture, and a polyurethane component. It is provided in various sizes to accommodate patient anatomy. The device maintains fixation prior to weight bearing and then allows motion after weight bearing. The implants can be used in conjunction with the Baby Gorilla®/Gorilla® Plating System (K203511).

    AI/ML Overview

    This document describes the R3ACT™ Stabilization System, a medical device for fracture repair and ligamentous injuries of the feet and ankles. It is a 510(k) premarket notification to the FDA. The information provided is for a conventional medical device, not an AI/ML powered medical device. Therefore, the questions related to AI/ML powered devices cannot be fully answered.

    Here's the information extracted from the provided text regarding the device and its testing:

    1. A table of acceptance criteria and the reported device performance

    The provided text states: "All performance testing conducted for the R3ACTTM Stabilization System met the predetermined acceptance criteria or were otherwise considered acceptable." However, specific quantitative acceptance criteria and detailed performance results for each test are not provided in the document.

    The document lists the following performance tests conducted:

    Test TypeReported Device Performance
    Torque to failureMet predetermined acceptance criteria
    Insertion and removal torqueMet predetermined acceptance criteria
    Static pulloutMet predetermined acceptance criteria
    Static bendingMet predetermined acceptance criteria
    Static axial dissociationMet predetermined acceptance criteria
    Dynamic axial dissociationMet predetermined acceptance criteria
    Bacterial endotoxin testingTest results meet acceptance criteria

    2. Sample sizes used for the test set and the data provenance

    The document does not specify the sample sizes (number of devices/components) used for each performance test. It also does not discuss data provenance in terms of country of origin or retrospective/prospective as these are typically relevant for clinical studies, which were not performed for this device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable. The device underwent performance testing (mechanical and material tests) on the device itself, not a clinical study involving human or image data with ground truth established by experts.

    4. Adjudication method for the test set

    This question is not applicable for the same reason as point 3. Performance tests on a physical device do not involve adjudication by experts.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. This is a conventional medical device, not an AI/ML powered device. No MRMC study was performed, and thus no effect size related to AI assistance is provided.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This question is not applicable. This is a conventional medical device, not an AI/ML powered device. No algorithm-only performance was evaluated.

    7. The type of ground truth used

    For the performance testing mentioned (e.g., torque to failure, static pullout), the "ground truth" would be the engineered specifications and expected physical behavior of the device components under various loads, as defined by engineering standards and design requirements. It's not "expert concensus, pathology, or outcomes data" in the context of clinical evaluation.

    8. The sample size for the training set

    This question is not applicable. This device is not an AI/ML powered device and therefore does not have a "training set" in the machine learning sense. The testing performed was for product performance validation.

    9. How the ground truth for the training set was established

    This question is not applicable, as there is no training set for this conventional medical device.