510(k) Data Aggregation

    K Number
    K202733

    Validate with FDA (Live)

    Date Cleared
    2020-10-16

    (28 days)

    Product Code
    Regulation Number
    870.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Quantum Ventilation Module is intended for the continuous monitoring of critical clinical parameters during procedures that require extracorporeal circulation. The Quantum Ventilation Module is an accessory that only works with the Quantum Workstation. Parameters provided by the Quantum Ventilation Module include:

    • Measurement of up to three blood flow channels and arterial and venous flow differential and gas bubbles
    • Extracorporeal gas flow measurements that includes O2 & CO2 and calculated CO2 removal
    • Predicted PO2 and PCO2
    • Up to three temperature channels
    • Up to three circuit pressure channels
    • Reservoir level indication
    • Two channels of vacuum
    • Blend and control gas flow (air/O2/CO2)

    The Quantum Ventilation Module is to only be used by an experienced and trained clinician. The device is not intended to be used by the patient or other untrained personnel.

    Device Description

    The Quantum Ventilation Module is an on-line, cardiopulmonary bypass, blood gas monitor that is used for extracorporeal monitoring of blood oxygen (arterial and venous) saturation, hematocrit, and hemoglobin levels. The Quantum Ventilation Module provides gas blending and continuous non-invasive monitoring of critical clinical parameters in extracorporeal circuits used in cardiopulmonary bypass (CPB) or extracorporeal membrane oxygenation (ECMO) procedures. The Quantum Ventilation Module is an accessory to the Quantum Workstation. When paired with the Quantum Workstation, the combination of the Quantum Workstation and Quantum Ventilation Module (QVM2) is known as the Quantum Ventilation System.

    The Quantum Ventilation Module performs five functions:

      1. Provides measurements from embedded and attached sensors to monitor gases into and out of a blood oxygenator.
      1. Provides measurements from attached sensors for blood flow, bubble detection, pressure, level, and temperature to monitor an extracorporeal blood loop.
      1. Provides gas blending to ensure the precision delivery of FiO2, CO2 and sweep flow rates.
      1. Provides regulation of vacuum supply to provide two channels of vacuum.
      1. Sends these physiological measurements to the Quantum Workstation for display to the user.

    The Quantum Ventilation Module, with its attached sensors, can measure flow, pressure, reservoir level, temperature and gas diagnostics, in addition to performing electronic gas blending of up to three gases and built-in vacuum management for the removal of waste anesthetic gas. The primary interface for controlling and displaying measurements is the Quantum Workstation; however, the Quantum Ventilation Module also contains a touchscreen display with control knobs. The Quantum Ventilation Module only works with the Quantum Workstation.

    AI/ML Overview

    This FDA 510(k) summary for the Spectrum Medical Quantum Ventilation Module (K202733) indicates that it is a Class II device intended for the continuous monitoring of critical clinical parameters during extracorporeal circulation. The submission claims substantial equivalence to a legally marketed predicate device, Spectrum Medical Ltd.'s Quantum Ventilation Module (K181942).

    Here's an analysis of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state acceptance criteria in a quantitative table format for specific performance metrics (e.g., accuracy, precision for flow, pressure, temperature). Instead, it states that the proposed Quantum Ventilation Module (QVM2) has "equivalent sensor performance" to the predicate device.

    Acceptance Criteria (Implied)Reported Device Performance
    Functional Equivalence: Provide measurements from embedded and attached sensors to monitor gases, blood flow, bubble detection, pressure, level, and temperature for an extracorporeal blood loop.The QVM2 performs these functions, including various types of measurements (blood flow, bubble detection, circuit pressure, blood reservoir level, temperature, and gas diagnostics).
    Gas Blending: Provide gas blending to ensure precision delivery of FiO2, CO2, and sweep flow rates.The QVM2 performs gas blending of air, oxygen, and carbon dioxide (air/O2/CO2).
    Vacuum Regulation: Provide regulation of vacuum supply with two channels.The QVM2 provides regulation of vacuum supply with two channels.
    Sensor Performance: Equivalent sensor performance to the predicate device."equivalent sensor performance" to the predicate (K181942).
    Electrical Safety: Compliance with relevant standards."Electrical safety" testing was performed.
    Electromagnetic Compatibility (EMC): Compliance with relevant standards."Electromagnetic compatibility (EMC)" testing was performed.
    Hardware Functionality: Proper operation of hardware components."Hardware testing" was performed.
    Software Verification and Validation: Proper functioning and reliability of software."Software verification and validation" was performed.

    2. Sample size used for the test set and the data provenance

    The document states that "No animal testing was submitted" and "No clinical data were submitted" to support the substantial equivalence. The non-clinical testing mentioned (electrical safety, EMC, hardware, software) refers to engineering verification and validation, not a test set of clinical or animal samples. Therefore, information regarding sample size and data provenance in the context of clinical or animal testing is not applicable here as such studies were not performed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. Since no clinical or animal testing with a "test set" and "ground truth" established by experts was performed or submitted, this information is not provided. The assessment was based on non-clinical engineering testing and comparison to a predicate device.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. No test set requiring expert adjudication was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a medical monitoring and control unit, not an AI-assisted diagnostic imaging or interpretation device. Therefore, an MRMC study is not relevant to its type of performance evaluation.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a hardware and software system for continuous monitoring and control during extracorporeal circulation, not a standalone algorithm. Its performance is evaluated through engineering verification and validation of its sensors and control mechanisms.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The concept of "ground truth" derived from expert consensus, pathology, or outcomes data is typically associated with diagnostic or prognostic devices that interpret patient data. For this device, the "ground truth" for its performance evaluation would be based on engineering standards, calibrated reference instruments, and defined physical parameters. For example, when testing flow measurement, the "ground truth" would be established by a known, accurately measured flow rate from a reference system.

    8. The sample size for the training set

    Not applicable. This device is not an AI/ML device that requires a "training set" in the conventional sense for model development. Its software verification and validation would involve testing against requirements, but not training data for a learning algorithm.

    9. How the ground truth for the training set was established

    Not applicable, as there is no "training set" for an AI/ML model for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181942

    Validate with FDA (Live)

    Date Cleared
    2018-10-18

    (90 days)

    Product Code
    Regulation Number
    870.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Quantum Ventilation Module is intended for the continuous monitoring of critical clinical parameters during procedures that require extracorporeal circulation. The Quantum Ventilation Module is an accessory that only works with the Quantum Workstation. Parameters provided by the Quantum Ventilation Module include:

    • · Measurement of up to three blood flow channels and arterial and venous flow differential
    • Indication of gas bubbles
    • · Extracorporeal gas flow measurements (02, CO2, gas flow, and CO2 removal)
    • · Predicted PO2 and PCO2
    • · Temperature
    • · Up to three circuit pressure channels
    • · Reservoir level indication
    • Two channels of vacuum
    • · Blend and control gas flow (air/O2/CO2)

    The Quantum Ventilation Module is to only be used by an experienced and trained clinician. The device is not intended to be used by the patient or other untrained personnel.

    Device Description

    The Quantum Ventilation Module provides gas blending and continuous non-invasive monitoring of critical clinical parameters in extracorporeal circuits used in cardiopulmonary bypass (CPB) or extracorporeal membrane oxygenation (ECMO) procedures. The Quantum Ventilation Module is an accessory to the Quantum Workstation or can be used in place of the Quantum Diagnostic Module as part of the Quantum Pump Console. When paired with the Quantum Workstation, the combination of the Quantum Workstation and Quantum Ventilation Module is known as the Quantum Ventilation System.

    The Quantum Ventilation Module performs five functions:

      1. Provides measurements from embedded and attached sensors to monitor gases into and out of a blood oxygenator.
      1. Provides measurements from attached sensors for blood flow, bubble detection, pressure, level and temperature to monitor an extracorporeal blood loop.
      1. Provides gas blending to ensure the precision delivery of FiO₂ (21 to 100%), CO₂ and sweep flow rates.
      1. Provides regulation of vacuum supply to provide two channels of vacuum. One is flow-regulated to remove waste anesthesia gas, the other pressure-regulated for applications including Vacuum-Assisted Venous Drainage (VAVD) and hemoconcentration.
      1. Sends these physiological measurements to the Quantum Workstation for display to the user.

    The Quantum Ventilation Module, with its attached sensors, can measure flow, pressure, reservoir level, temperature and gas diagnostics, in addition to performing electronic gas blending of up to three gases and built-in vacuum management for the removal of waste anesthetic gas. The primary interface for controlling and displaying measurements is the Quantum Workstation: however, the Quantum Ventilation Module also contains a display with control knobs. The Quantum Ventilation Module only works with the Quantum Workstation.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for a medical device called the Quantum Ventilation Module. This type of submission is for demonstrating "substantial equivalence" to a legally marketed predicate device, rather than proving safety and effectiveness through extensive clinical trials as would be required for a Premarket Approval (PMA) application.

    Therefore, the document does not describe a study involving an AI algorithm or meeting the typical acceptance criteria for AI/ML device performance. Instead, it focuses on non-clinical performance data to demonstrate the device's functionality and safety as a medical instrument.

    Given this context, I will address the questions to the best of what can be inferred from the provided text, noting where the information is absent due to the nature of a 510(k) submission for a non-AI hardware device.


    Acceptance Criteria and Device Performance (Based on Non-Clinical Testing for a Hardware Device)

    The document primarily discusses non-clinical performance data, which are typically tests to ensure the device performs as intended and is safe. The "acceptance criteria" here are implied by the successful completion of these tests.

    1. A table of acceptance criteria and the reported device performance

    Since this is a hardware device (Ventilation Module) and not an AI algorithm, the acceptance criteria are not in terms of common AI metrics like sensitivity, specificity, or AUC, nor is there comparative effectiveness data against human readers. The criteria are related to the device's physical and electronic performance, along with its software validation. The document states:

    Acceptance Criteria Category (Implied)Reported Device Performance Summary (as per document)
    Electrical SafetyNon-clinical testing performed to support substantial equivalence. (Implies successful completion.)
    Electromagnetic Compatibility (EMC)Non-clinical testing performed to support substantial equivalence. (Implies successful completion.)
    Electrosurgery InterferenceNon-clinical testing performed to support substantial equivalence. (Implies successful completion.)
    Hardware Testing (Printed Circuit Boards)Non-clinical testing performed to support substantial equivalence. (Implies successful completion.)
    Software Verification and ValidationNon-clinical testing performed to support substantial equivalence. (Implies successful completion.)
    Usability ValidationNon-clinical testing performed to support substantial equivalence. (Implies successful completion.)
    Diagnostic Measurements Accuracy (e.g., flow, pressure, temperature, gas)"Equivalent sensor performance" to the predicate device (Quantum Diagnostic Module K173591) is claimed, implying accuracy meets established standards for these parameters.
    Gas Blending Precision (FiO₂, CO₂, sweep flow rates)Claimed to "ensure the precision delivery" of these parameters.
    Vacuum RegulationProvides "regulation of vacuum supply."

    2. Sample size used for the test set and the data provenance

    This information is not applicable in the context of an "AI test set" here. The "test set" would refer to the physical units of the device subjected to non-clinical tests. The tests performed ("Electrical safety", "Electromagnetic compatibility", "Electrosurgery interference", "Hardware testing of printed circuit boards", "Software verification and validation", "Usability validation") are typically laboratory-based engineering and software validation tests. The document does not specify the number of units tested, the conditions, or the specific "data provenance" (e.g., country of origin) beyond the manufacturer being in the UK. These are not data-driven performance studies on patient cohorts.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This is not applicable as this is a hardware device. Ground truth, in the context of AI, refers to annotated data. For a hardware device, "ground truth" might refer to established measurement standards or known physical properties used for calibration and validation of sensors. The document does not specify the number or qualifications of experts involved in these engineering validation processes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 are used for establishing consensus among human readers for image labeling or clinical decision-making ground truth in AI studies. For hardware testing, performance is measured against engineering specifications and industry standards, not through expert adjudication in this manner.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This is a hardware medical device, not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study is not relevant, and the concept of human readers improving with AI assistance does not apply.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to an AI algorithm's performance. The Quantum Ventilation Module is a hardware device with embedded software, but it's not an AI algorithm that makes diagnostic decisions or interpretations in the way this question implies. Its performance is the measurement and control capabilities of the instrument itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. For a hardware device, the "ground truth" for non-clinical performance validation would be derived from:

    • Engineering specifications and design documents.
    • International and national standards for electrical safety, EMC, and medical device performance (e.g., IEC standards).
    • Calibration standards for sensors (e.g., precision gas mixes, flow simulators, temperature baths, pressure gauges).
    • Bench testing and physical measurements.

    8. The sample size for the training set

    Not applicable. This device is not an AI/ML model trained on a dataset. It's a hardware device with firmware.

    9. How the ground truth for the training set was established

    Not applicable. As there is no "training set" in the AI sense, this question is not relevant. The "ground truth" for developing the device's functionality would stem from engineering principles, clinical requirements for extracorporeal circulation, and established medical device design practices.