Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K963761

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1997-04-18

    (212 days)

    Product Code
    Regulation Number
    868.5905
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Quartet Clinical System delivers continuous positive airway pressure (CPAP) and bi-level positive airway pressure therapies for the treatment of adult OSA. The Quartet Clinical System is intended for use by a qualified clinician (any individual trained to perform sleep study diagnoses) in clinical (hospital or sleep laboratory) settings where adult patients suffering from OSA are diagnosed and treated. The device is used in conjunction with masks, headgear, and various combinations of patient circuit accessories.

    Device Description

    The Quartet Clinical System is intended to apply continuous positive airway pressure, known as CPAP, bi-level positive airway pressure, and self-titrating CPAP therapies to adult patients who suffer from OSA. The device can operate on either 115 or 230 Volts AC. As shown in Figure 12-1, the System consists of the Quartet device, a clinical remote control device called the Maestro CRC, and the recommended patient circuit that is used to direct the air to the patient. The tubing, exhalation port, and mask are referred to as the patient circuit.

    The primary components of Quartet device are a blower motor assembly and a microprocessor control system. The blower motor is capable of generating positive airway pressures from 3 to 20 cm H2O. The level of pressure is dependent on motor speed, which is controlled by a microprocessor-controlled feedback system.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the Respironics Quartet Clinical System:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria provided primarily relate to the maintenance of pressure within specified ranges and the clinical effectiveness of the Auto-CPAP mode in reducing AHI and ADI.

    Acceptance Criteria (Claims)Reported Device Performance (as stated in the document)
    General
    Provides four modes of operation.Confirmed: Manual CPAP Titration Mode, Bi-level Titration Mode, Auto-CPAP Titration Mode, Split Night with Auto-CPAP Titration Mode.
    Provides data monitoring capabilities.Confirmed: Maestro CRC provides alphanumeric and graphical representations of data.
    Presents alphanumeric and graphical representation of data.Confirmed: Maestro CRC provides alphanumeric and graphical representations of data.
    Interfaces (optionally) with a polysomnograph.Confirmed: Maestro CRC provides DC outputs that connect to a polysomnograph for data recording.
    CPAP Mode
    Maintains constant pressure within ± 1.5 cm H2O at patient interface for flows of -35 to +85 L/min during both inspiratory and expiratory phase.Not explicitly stated as "met" but implied through the statement "The Quartet Clinical System was found compliant and has been certified to the standards referenced in the 'FDA Reviewer Guidance for Premarket Notifications'" after performance-related testing, and the statement that it is 'equivalent' to predicate devices with similar claims. The provided document does not contain an explicit table or statement directly validating this for the Quartet itself within this submission, but rather relies on equivalency and general compliance for the non-Auto-CPAP modes.
    Bi-level Mode
    Senses the patient's breathing efforts by monitoring airflow.Claimed functionality.
    Increases pressure upon sensing inspiratory effort. Decreases pressure upon sensing expiratory effort.Claimed functionality.
    Trigger sensitivity adjusts automatically.Claimed functionality.
    Compensates for most leaks.Claimed functionality.
    Maintains IPAP and EPAP within ± 1.5 cm H2O at patient interface for flows of -35 to +85 L/min.Not explicitly stated as "met" but implied through equivalency to predicate devices and general compliance.
    Auto-CPAP Mode
    Effective in treating OSA (reduces AHI to ≤ 10 events/hour).Met: Virtuoso System Auto-CPAP (which is the same as Quartet's) reduced AHI from 51.81 to 8.39 (mean).
    Reduces AHI and improves Arterial Desaturation Index (ADI).Met: Virtuoso System Auto-CPAP reduced AHI from 51.81 to 8.39 and ADI from 61.59 to 16.39 (mean).
    Improves sleep architecture (percentage of sleep time for four stages of sleep).Claimed functionality, but no specific data presented to confirm this particular outcome. The study data focuses on AHI and ADI.
    Maintains CPAP pressure within ± 1.5 cm H2O at patient interface for flows of -35 to +85 L/min.Not explicitly stated as "met" but implied through equivalency and general compliance. Bench testing showing the Quartet responds to apneic events like the Virtuoso, and no significant differences in frequency response.
    Split Night with Auto-CPAP Mode
    Delivers minimum CPAP pressure for a preset Delay Time.Claimed functionality.
    Provides opportunity to study patient at minimal pressures during Delay Time.Claimed functionality.
    Responds to airway instability by increasing/decreasing pressure to meet patient needs within designated limits once Delay Time ends.Claimed functionality. This is essentially the Auto-CPAP mode after the Delay Time. The effectiveness is covered under the Auto-CPAP claims.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 39 clinical subjects.
    • Data Provenance: The study was "Respironics-sponsored clinical studies." No specific country of origin is mentioned, but "clinical subjects" implies human participants. Given the context of a 510(k) in the US, it's highly likely to be US data. The study was prospective in nature, as it involved conducting tests with active intervention (Auto-CPAP vs. manual CPAP) on these subjects.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not explicitly state the number of experts or their specific qualifications for establishing the ground truth for AHI and ADI in the clinical study. However, the study involved "manually titrated CPAP" which would have been overseen by qualified clinicians (likely sleep specialists/pulmonologists). The "ground truth" for the effectiveness of the therapy (reduction in AHI/ADI) would have been determined through standard polysomnography conducted and interpreted by trained personnel in a clinical setting.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for the test set data. The clinical study involved comparing outcomes (AHI, ADI, pressure) between two therapeutic approaches (manual CPAP and Auto-CPAP). The outcomes themselves would be measured via polysomnography, which has a standardized interpretation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. The study described is a clinical trial comparing different therapeutic modes on subjects, not a study evaluating human reader performance with or without AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, the clinical study essentially evaluates the standalone performance of the Auto-CPAP algorithm compared to manual CPAP settings. The "Virtuoso System Auto-CPAP" (which is identical to the Quartet's Auto-CPAP mode) functioned independently, adjusting pressure based on its internal algorithms. The performance metrics (AHI, ADI reduction) were directly attributed to the device's algorithmic function.

    7. The Type of Ground Truth Used

    The ground truth for the clinical study was based on clinical outcomes data, specifically:

    • Polysomnography results: AHI (Apnea-Hypopnea Index) and ADI (Arterial Desaturation Index) are physiological measurements derived from polysomnography, which is the gold standard for diagnosing and assessing the severity of sleep apnea.
    • Expert titration/diagnosis: While not explicitly called "ground truth," the comparison to "manually titrated CPAP" implies that the benchmark was set by human clinical experts determining optimal pressure.

    8. The Sample Size for the Training Set

    The document does not provide information on the sample size for the training set. The clinical study described served as a validation set for the Auto-CPAP mode that was already developed and present in the Virtuoso system.

    9. How the Ground Truth for the Training Set Was Established

    Since no information on a training set is provided, the method for establishing its ground truth is also not described in this document. The Auto-CPAP mode was previously developed and validated in the K953930 submission for the Virtuoso Smart CPAP System. Therefore, any training activities or ground truth establishment would have occurred prior to this submission and are referenced implicitly through the predicate device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1