Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K101790

    Validate with FDA (Live)

    Date Cleared
    2011-03-07

    (252 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Quantum-MIS Cannulated System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and /or lordosis), spinal tumor, and failed previous fusion (pseudoarthrosis). In addition, this device is a pedicle screw system indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3-S1) with removal of the implants after the attainment of a solid fusion.

    The Quantum Spinal System components are non-cervical spinal fixation devices intended as an adjunct to fusion for use as a pedicle screw system (T1 - S2), posterior hook (TI-L5) sacral/iliac screw fixation or as an anterolateral fixation system (T8 - L5). Pedicle screw fixation is limited to skeletally mature patients. These devices are indicated for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, pseudoarthrosis, and failed previous fusion.

    Device Description

    The Quantum Spinal System (including MIS) consists of a variety of rods, fixed and polyaxial screws, transverse connectors, and locking caps used to build a spinal construct. Instrumentation is also available to facilitate implantation of the device components.

    The Quantum Spinal System (including MIS) are intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic and/or lumbar spine. The implant components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.

    The Quantum Spinal System (including MIS) components are comprised of Titanium Alloy Ti6A14V per ASTM F136. A subcomponent of the locking cap, a spring clip, is comprised of nitinol (NiTi) per ASTM F 2063. All components of the Quantum Spinal System (including MIS) HA Coated Pedicle Screws (Non-Cannulated and Cannulated) are comprised of Titanium Alloy Ti6Al4V per ASTM F136 and have a hydroxyapatite (HA) coating.

    The purpose of this submission is to include HA coated pedicle screws to the Quantum Spinal System (including MIS).

    AI/ML Overview

    This document describes the acceptance criteria and the study conducted for the Pioneer Quantum Spinal System (including MIS) HA Coated Pedicle Screws (Non-Cannulated and Cannulated). This is a medical device, and the evaluation for substantial equivalence to a predicate device is based on non-clinical mechanical tests, not an AI/algorithm-based system. Therefore, many standard AI/ML study components (like sample sizes for test/training sets, ground truth by experts, MRMC studies, standalone algorithm performance) are not applicable.

    Here's the breakdown based on the provided text, focusing on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test PerformedAcceptance Criteria (Implicit)Reported Device Performance
    Pedicle Screw Insertion and Removal TorquesPerformance comparable to or better than predicate devices, and within acceptable industry standards.Performed. Results were sufficient to demonstrate substantial equivalence.
    Static Flexion-Extension TestPerformance comparable to or better than predicate devices, and within acceptable industry standards.Performed. Results were sufficient to demonstrate substantial equivalence.
    Dynamic Flexion-Extension of HA Coated and Uncoated Cannulated ScrewsPerformance comparable to or better than predicate devices, and within acceptable industry standards.Performed. Results were sufficient to demonstrate substantial equivalence.
    Compliance with FDA Guidance for HA Coated ImplantsMeet "510(k) Information Needed for Hydroxyapatite Coated Implants" requirements.Existing testing provided and deemed compliant.

    Note: The document implicitly states that the performance met the criteria by concluding "the subject device demonstrates substantial equivalence to the listed predicate devices" based on the performed mechanical tests. Specific numerical acceptance criteria are not detailed in this summary.

    2. Sample Size Used for the Test Set and Data Provenance

    Since this is a mechanical device evaluation, the "test set" refers to the physical devices and components subjected to testing, not a dataset of patient information.

    • Sample Size for Test Set: Not explicitly stated as a number of devices/components, but the described tests would have involved multiple samples to ensure reproducibility and statistical significance.
    • Data Provenance: The tests were conducted by the manufacturer, Pioneer Surgical Technology. The data is based on laboratory mechanical testing (non-clinical) of prototype or production-representative devices. There is no information about country of origin of data or retrospective/prospective in the context of patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Not Applicable. This is a mechanical device. "Ground truth" in this context would be physical measurements and engineering analyses, not expert medical opinion on clinical data.

    4. Adjudication Method for the Test Set

    • Not Applicable. As per point 3, there's no "adjudication" in the sense of reconciling expert opinions on medical cases. The evaluation is based on objective mechanical testing results against engineering standards and comparison with predicate devices.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    • No. This is a medical device, not an AI/algorithm. MRMC studies are used for evaluating diagnostic software/algorithms.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    • Not Applicable. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    • Mechanical Performance Standards and Predicate Device Comparison: The "ground truth" for the device's performance is its ability to withstand specific forces, torques, and cycles as demonstrated by the mechanical tests (Pedicle Screw Insertion and Removal Torques, Static Flexion-Extension Test, Dynamic Flexion-Extension) and its equivalence to the previously cleared predicate devices in terms of material properties and performance. Compliance with FDA guidance for hydroxyapatite-coated implants also serves as a "ground truth" for material safety and performance.

    8. The Sample Size for the Training Set

    • Not Applicable. This is a medical device, not an AI/algorithm.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. This is a medical device, not an AI/algorithm. Design and manufacturing processes are iteratively developed and tested against engineering specifications, but there isn't a "training set" in the machine learning sense.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1