Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K103539

    Validate with FDA (Live)

    Date Cleared
    2011-01-20

    (49 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nd:YAG (1064nm): The DUOLITE Q-Switched laser is intended for treatment of vascular lesions, pigmented lesions, and for hair removal and the incision, excision, ablation, vaporization of soft tissue for general dermatology. The DUOLITE Q-Switched laser is intended for tattoo removal (blue, black and green tattoo)

    Nd: YAG (532nm): The DUOLITE Q-Switched laser is intended for treatment of vascular lesions, pigmented lesions, and for hair removal and the incision, excision, ablation, vaporization of soft tissue for general dermatology. The DUOLITE Q-Switched laser is intended for tattoo removal (red, violet, orange, yellow and brown tattoo)

    Device Description

    The device includes a Q-Switched Nd: YAG laser source with 900mJ max energy at 1064 nm and 450mJ max energy at 532 nm wavelengths. The optical delivery system for the two wavelengths is the articulated arm. In addition, the DUOLITE includes a power supply; a cooling system; an optical delivery system; a microprocessor based controller; and safety features to ensure use of the appropriate laser, wavelength and hand piece.

    AI/ML Overview

    The provided text is a 510(k) summary for the Quanta System DUOLITE laser, which is a predicate device application. This type of submission relies on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting new clinical study data with detailed acceptance criteria and performance metrics. Therefore, many of the requested details about acceptance criteria, clinical study specifics, and ground truth establishment are not typically found or required in a 510(k) summary for a predicate device of this nature.

    However, I can extract the relevant information that is present in the document.

    1. A table of acceptance criteria and the reported device performance

    Since this is a 510(k) for a predicate device, explicit "acceptance criteria" in the sense of predefined thresholds for a clinical study endpoint are not presented. Instead, the device's performance is demonstrated through its "Technological Characteristics" being substantially equivalent to the predicate device. The performance comparison is based on technical specifications rather than clinical outcomes.

    CharacteristicQuanta System DUOLITE Reported PerformancePredicate Device (Cynosure Affinity QS K050382) PerformanceAcceptance Criteria (Implied by Substantial Equivalence)
    Nd:YAG Wavelength1064 nm1064 nmMatch/be equivalent to predicate
    532 nm Wavelength532 nm532 nmMatch/be equivalent to predicate
    Pulse Width6 ns (for both 1064 nm and 532 nm)6 ns (for both 1064 nm and 532 nm)Match/be equivalent to predicate
    Fluence (1064 nm)
    @ 2mm spot size28 J/cm²28 J/cm²Match/be equivalent to predicate
    @ 3mm spot size12 J/cm²12 J/cm²Match/be equivalent to predicate
    @ 6mm spot size3 J/cm²3 J/cm²Match/be equivalent to predicate
    Fluence (532 nm)
    @ 2mm spot size14 J/cm²14 J/cm²Match/be equivalent to predicate
    @ 3mm spot size6 J/cm²6 J/cm²Match/be equivalent to predicate
    @ 6mm spot size1.5 J/cm²1.5 J/cm²Match/be equivalent to predicate
    Spot Sizes2, 3, and 6mm2, 3, 4, and 6mmSimilar spot sizes (DUOLITE offers a subset of the predicate's)
    Repetition Rate1, 2, 5, and 10Hz1, 2, 5, and 10HzMatch/be equivalent to predicate
    Biocompatible Handpiece TipYesYesMatch/be equivalent to predicate
    Sterilization MethodSame as predicateSame as predicateMatch/be equivalent to predicate
    Intended Use / IndicationsSame as predicate, with specific tattoo colors listed.Same as predicate, with specific tattoo colors listed.Similar intended uses and indications
    Principles of OperationSame as predicateSame as predicateMatch/be equivalent to predicate

    The "acceptance criteria" for a 510(k) in this context is that the device is "as safe and effective as the predicate devices" and that "Minor technological differences...raise no new issue of safety or effectiveness," leading to a determination of "Substantial Equivalence."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not describe a clinical "test set" with a sample size or data provenance in the way one would for a new clinical study. The equivalence is demonstrated through comparison of technical specifications, not patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. There was no clinical test set requiring expert ground truth establishment.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There was no clinical test set requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a laser surgical instrument, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a laser surgical instrument.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. No clinical ground truth was established for this 510(k) submission. The "ground truth" for demonstrating substantial equivalence lies in the technical specifications, intended uses, and principles of operation being comparable to the predicate device.

    8. The sample size for the training set

    Not applicable. There is no "training set" as this is not a machine learning device.

    9. How the ground truth for the training set was established

    Not applicable. There is no "training set."

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1