510(k) Data Aggregation

    K Number
    K221868

    Validate with FDA (Live)

    Date Cleared
    2023-01-27

    (214 days)

    Product Code
    Regulation Number
    892.2080
    Age Range
    21 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    QOCA® image Smart CXR Image Processing System is a software as medical device (SaMD) used, through artificial intelligence/deep learning technology, to analyze chest X-ray images of adult patient, and then identify cases with suspected pneumothorax. This product shall be used in conjunction with Picture Archiving and Communication System (PACS) at the hospital. This product will automatically analyze the DICOM files automatically pushed from PACS, and then make a notation next to the cases with suspected pneumothorax. This product is only used to remind radiologists to prioritize reviewing cases with suspected pneumothorax. Its results cannot be used as a substitute for a diagnosis by a radiologist, nor can it be used on a stand-alone basis for clinical decision-making.

    Device Description

    This product, QOCA® image Smart CXR Image Processing System, is a web-based medical device using a locked artificial intelligence algorithm. It provides features such as cases sorting and image viewing, and supports multiple users at a time.

    After connecting to Picture Archiving and Communication System (PACS) at the hospital, this product is capable of automatically analyzing either posteroanterior (PA) view or anteroposterior (AP) erect view chest X-ray images automatically pushed from PACS. Once a case with suspected pneumothorax is identified, a notation will be made next to the case in question, so the radiologist can prioritize to review cases with suspected pneumothorax in the Viewer Page. This product will not directly indicate, however, the specific portions or anomalies on the image.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the QOCA® image Smart CXR Image Processing System:


    1. Acceptance Criteria and Reported Device Performance

    MetricAcceptance Criteria (Predicate Device K190362 Performance)Reported Device Performance (QOCA® image Smart CXR)Overall Performance
    AUC98.3% (95% CI: [97.40%, 99.02%])97.8% (95% CI: [97.0%, 98.5%])Met
    Sensitivity93.15% (95% CI: [87.76%, 96.67%])92.5% (95% CI: [90.5%, 94.2%])Met
    Specificity92.99% (95% CI: [90.19%, 95.19%])94.0% (95% CI: [93.9%, 94.6%])Met
    Average Performance Time22.1 seconds4.94 secondsMet

    Note: The reported device performance is an overall performance across both the MIMIC and Taiwanese datasets. Individual performance for each dataset is also provided in the document.


    2. Sample Size Used for the Test Set and Data Provenance

    The device's performance was assessed using two separate pivotal studies/datasets:

    • MIMIC Dataset:

      • Sample Size: 3,105 radiographs (336 positive pneumothorax cases, 2,769 negative pneumothorax cases).
      • Data Provenance: US patient population (MIMIC dataset). This was an independent medical institution from the training dataset.
    • Taiwanese Dataset:

      • Sample Size: 2,947 radiographs (472 positive pneumothorax cases, 2,475 negative pneumothorax cases).
      • Data Provenance: Taiwanese hospital. This was an independent medical institution from the training dataset.

    Overall Test Set: 6,052 radiographs (3,105 from MIMIC + 2,947 from Taiwan). Both datasets were retrospective.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    For both the MIMIC dataset and the Taiwanese dataset:

    • Number of Experts: Three radiologists.
    • Qualifications: The document states "truthed by three radiologists" without specifying their years of experience or sub-specialty.

    4. Adjudication Method for the Test Set

    The document does not explicitly state the adjudication method (e.g., 2+1, 3+1). It only mentions that the datasets were "truthed by three radiologists," implying a consensus-based approach, but the specific process for resolving disagreements is not detailed.


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    There is no mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study being performed to assess how much human readers improve with AI vs. without AI assistance. The study focused on the standalone performance of the AI algorithm.


    6. Standalone Performance Study

    Yes, a standalone performance study was done. The document explicitly states: "Bases on the results of the standalone performance assessment, this product achieves, identification accuracy of AUC > 95% with Sensitivity > 91% and Specificity > 92%." The performance metrics provided in section 1 (AUC, sensitivity, specificity) reflect the algorithm's performance without human-in-the-loop.


    7. Type of Ground Truth Used

    The ground truth for the test sets (MIMIC and Taiwanese) was established by "three radiologists," which indicates expert consensus diagnosis.


    8. Sample Size for the Training Set

    The document states: "The training dataset is used to train the model, and divided into three sets: the training set, the validation set, and the test set." However, the specific sample size for the entire training dataset (including training, validation, and its own internal test set used during development) is not provided in the summary. It only indicates that it was "collected from two hospitals, and additional data from the US National Institutes of Health (NIH) was added to the test set to improve its US patient population representativeness during training."


    9. How the Ground Truth for the Training Set Was Established

    The document states that the "model training dataset was collected from two hospitals, and additional data from the US National Institutes of Health (NIH) was added to the test set." While it implies the data was labeled for training, it does not explicitly describe how the ground truth for the training set was established (e.g., whether it was also by expert radiologists, pathology, etc.).