Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K202874

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-01-09

    (103 days)

    Product Code
    Regulation Number
    868.5750
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CPM is intended to continuously measure and automatically maintain the cuff pressure of an endotracheal tube or tracheostomy tube that has been set by the user/clinician during mechanical ventilation. The CPM is intended for use with adult and pediatric patients during mechanical ventilation in the listed areas:

    Intensive care unit (ICU)
    Recovery room
    Emergency medical care
    Medical emergency vehicles when the patient is being transported
    During transport within and outside of the hospital

    Device Description

    The Puritan Bennett™ Cuff Pressure Manager (CPM) continuously measures and automatically maintains cuff pressure during ventilation of patients using a cuffed endotracheal tube or tracheostomy tube. It is an integrated and continuous cuff pressure control solution that secures airway management in intensive care units, recovery room and during interhospital and intrahospital transport.

    It operates in range of 5-50 cm H2O of user-set cuff pressures for various cuffed endotracheal tubes to provide suitable solutions for various clinical patient situations. The Puritan Bennett™ Cuff Pressure Manager introduces some new small features and convenient user interface to maximize safe use and usability. Both manual adjusting and one-click inflation function are provided alternatively for user's preference to start inflation. Measure mode allows user to observe the cuff pressure without device intervening before taking actions. Auto Key-lock feature help to avoid mis-operation when there's no user intervene for a long time. For inflation, room air is used and no contact to the respiratory gas system of a patient occurs. It is not intended to connect with ventilator.

    The cuff pressure manager is designed with closed loop mechanism which is Pl controlled, and the software level of concern is Class C/Major concern.

    The Puritan Bennett™ Cuff Pressure Manager is provided non-sterile and no re-processing is required. A sterile extension tube is equipped with the Puritan Bennett™ Cuff Pressure Manager for inflating/deflating or measuring cuff pressure. The associated accessories include:

    • Extension Tube with filter Rechargeable Li-ion Battery Mounting Rack AC/DC Adaptor
    AI/ML Overview

    The provided document describes the Puritan Bennett™ Cuff Pressure Manager (CPM) and its substantial equivalence to a predicate device (IntelliCuff Cuff Pressure Controller, K150893). The information mainly focuses on technical specifications, safety, and performance evaluations, primarily through bench testing against recognized standards. There is no mention of a study involving human or animal subjects, ground truth, expert adjudication, or multi-reader multi-case studies in the provided text.

    Based on the available information, the acceptance criteria and device performance relate to technical specifications and compliance with various international standards for medical devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily defined by the specified technical characteristics and compliance with the referenced industry standards. The reported device performance is indicated by its adherence to these specifications and successful completion of the described tests.

    CharacteristicAcceptance Criteria (from Proposed Device Specs)Reported Device Performance
    Operation Altitude-411m to 4000m above sea levelComplies with specified range
    Operation Temperature0°C to 40°C / 32°F to 104°F (operating)Complies with specified range
    Operation Relative Humidity10% to 95% noncondensing (operating)Complies with specified range
    Electrical SafetyClass-II Type BF in accordance with IEC 60601-1 and IEC 60601-1-12Complies with specified standards
    Electromagnetic CompatibilityIn accordance with IEC 60601-1-2Complies with specified standard
    AlarmIn accordance with IEC 60601-1-8Complies with specified standard
    AC Power Input85 to 264 VAC / 47 to 63 Hz; 7.5 VA typical, 15 VA maximumOperates within specified range
    Battery Requirement3.6 V / 2300mAh Li-ion rechargeable battery; supports 4 hours operationComplies with specified battery type and operation duration
    Default Inflation Pressure Settings25 cm H2O25 cm H2O
    Pressure Hold Default Pressure Increase5 cm H2O5 cm H2O
    Pressure Hold Maximum Allowable Pressure50 cm H2O50 cm H2O
    Pressure Hold Time Setting Range5 minutes to 30 minutes5 minutes to 30 minutes
    Pressure Hold Time Factory Default Setting5 minutes5 minutes
    Pressure Setting Range5 to 50 cm H2O5 to 50 cm H2O
    Pressure Unit Switchingcm H2O, hPa, mbarSupports all specified units
    Resolution (setting/display)+/-1cm H2O+/-1cm H2O
    Pressure Accuracy (Display)±1.5 cm H2OAchieved ±1.5 cm H2O
    Pressure Accuracy (Adjustment)±1 cm H2OAchieved ±1 cm H2O
    Alarm Priority (High Priority)Low Pressure (including cuff leakage), High Pressure, Battery is depleted, Technical ErrorsFunctions as specified
    Alarm Volume50 dB(A) ±5 dB(A) within 1 m distanceAchieved within specified range
    Applied Pressure Range in Deflation Mode-45 cm H2OAchieved -45 cm H2O
    Pressure Monitoring Resolution±0.01 cm H2O (0.01 mbar)Achieved ±0.01 cm H2O
    Essential PerformanceApplied cuff pressure must be maintained and monitored. If higher or lower than set limits (±1cm H2O), must be detected and alarm activated.Demonstrated conformance to this requirement
    Extension Tube with FilterYesYes
    Connector CompatibilityCompatible with endotracheal tube (ETT) or tracheostomy tube (TT)Compatible as specified
    Software Class of ConcernClass C/Major concern (as indicated in device description)Verified and validated for this level of concern
    Sterilization (Extension tube)Sterile by EtO validation per ISO 11135:2014 and 10993-7: 2008 AC:2009Validated
    Shelf Life (Extension tube)5 years shelf life per ASTM F1980:16Validated

    2. Sample size used for the test set and the data provenance

    The document does not specify a "test set" in the context of clinical data or human-derived data. The testing described is primarily bench testing of the device's hardware and software components. For these types of tests, samples would refer to the number of devices or components tested. This information is not provided in the document.

    Data provenance is not applicable in the sense of country of origin of patient data, as no clinical or patient data was used. All described tests are non-clinical, likely performed by the manufacturer, Covidien llc, in China or its designated testing facilities. The tests are retrospective in the sense that they are conducted on manufactured devices before market release.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This is not applicable as there is no "test set" requiring expert ground truth in the context of clinical images or patient data. The ground truth for the technical performance tests is established by the engineering specifications and recognized international standards themselves.

    4. Adjudication method for the test set

    This is not applicable as there is no clinical test set requiring expert adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done as the device is a cuff pressure manager, not an AI-assisted diagnostic or interpretative tool for human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This concept is not directly applicable to this device in the traditional sense of an "algorithm only" performance study for image analysis or diagnosis. The device itself is an automated system for managing cuff pressure, and its performance is inherently a "standalone" evaluation of its functional accuracy and safety. The performance tests ("Function and performance Test-Cuff Pressure Manager") demonstrate the algorithm (closed-loop PI control) and device's ability to maintain pressure automatically without human intervention.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the device's performance is established by:

    • Technical Specifications: The design parameters and target performance values set by the manufacturer.
    • International Standards: Compliance with recognized standards like IEC 60601 series for electrical safety, EMC, alarms, usability, and software, as well as ISO standards for sterilization, biological evaluation, and connectors.
    • Benchmark Testing: Comparison against the predicate device's known performance characteristics (though the specific "ground truth" for this comparison is the predicate's established performance).

    8. The sample size for the training set

    This is not applicable as the device is not an AI/ML model that requires a training set of data in the typical sense. Its control mechanism is a PI controller, a traditional engineering control system, not a learned AI model.

    9. How the ground truth for the training set was established

    This is not applicable as there is no training set for an AI/ML model. The "ground truth" for the device's control logic is based on engineering and physiological principles for maintaining appropriate cuff pressure.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1