Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K202681

    Validate with FDA (Live)

    Date Cleared
    2020-12-31

    (107 days)

    Product Code
    Regulation Number
    880.5570
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Promisemed Covered Safety Pen Needle is intended for subcutaneous injection of insulin in the treatment of diabetes.

    Device Description

    Promisemed Covered Safety Pen Needle is manufactured by Promisemed Hangzhou Meditech Co., Ltd, which is designed for use with a pen injector for the subcutaneous injection of insulin. The user proceeds with inserting the needle into the skin manually.

    The Promisemed Covered Safety Pen Needle consists of needle tube, trigger shield, spring, fixer, needle hub, needle container, UV glue and silicone oil. UV glue is used to glue needle tube and needle hub and the silicone oil is used to needle tube lubrication.

    Promisemed Covered Safety Pen Needle is the modification of Verifine Safety Type Insulin Pen Needle cleared in K161950. The baseline location of trigger shield is adjusted to allow the needle tube be covered by trigger shield thoroughly before use. This modification prevents accidental needle sticks before inserting the needle into the skin while the needle container is removed. As the user proceeds with inserting the needle into the skin the shield will retract. After the injection is completed and needle is removed from the skin, the shield will automatically extend to cover the needle aqain and lock in place. Once the Covered Safety Insulin Pen Needle is in the locked mode, it can no longer be used.

    Promisemed Covered Safety Pen Needle is sterility Assurance Level (SAL) of 10-6, non-pyrogenic and single-use devices. It is supplied with several models. Different models are distinquished by needle gauge and length.

    AI/ML Overview

    This document is a 510(k) Pre-market Notification for a medical device called "Promisemed Covered Safety Pen Needle". It focuses on establishing substantial equivalence to a predicate device, not on presenting a standalone study of the device's performance against specific acceptance criteria in a clinical or AI context. Therefore, much of the requested information regarding AI/clinical studies, expert ground truth, sample sizes for test/training sets, and MRMC studies is not applicable or not provided in this regulatory document.

    However, based on the provided text, I can extract information related to non-clinical testing and the acceptance criteria established by adherence to recognized international standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    The device performance is demonstrated by compliance with international standards. The acceptance criteria are implicit in the requirements of these standards.

    Acceptance Criteria (Defined by Standard)Reported Device Performance
    ISO 9626:2016 (Needle material)Complied
    ISO 11608-2:2012Complied
    Needle DimensionsComplied
    Flow rateComplied
    Bond between hub and needle tubeComplied
    Needle pointsComplied
    Freedom from defectsComplied
    LubricationComplied
    Dislocation of measuring point patient endComplied
    Functional compatibility with needle-based injected systemsComplied
    Ease of assembly and disassemblyComplied
    ISO 7864:2016 (Pierce Force)Complied
    ISO 23908:2011Complied
    Accidental access to a sharp once in safe modeComplied
    Safety mechanism activationComplied
    Safety overriding and unlocking force after activationComplied
    ISO 10993-1:2018, ISO 10993-5:2009, ISO 10993-10:2013, ISO 10993-11:2006, ISO 10993-11:2017 & ASTM F756-13 (Biocompatibility)Complied
    CytotoxicityNo cytotoxicity
    SensitizationNo evidence of sensitization
    IrritationNo evidence of skin irritation
    Acute Systemic ToxicityNo systemic toxicity
    Material Mediated PyrogenicityNo pyrogenic
    HemocompatibilityNo evidence of hemolysis
    ISO 11135:2014, ISO 11737-1:2018, ISO 11737-2:2009 & ISO 10993-7:2008 (Sterilization)Complied
    Sterility Assurance Level (SAL)10^-6

    2. Sample size used for the test set and the data provenance

    The document does not specify exact sample sizes for each non-clinical test. The testing was conducted to verify compliance with various ISO standards, which typically involve specific sample size requirements for each test. The data provenance is presumed to be from tests performed by the manufacturer, Promisemed Hangzhou Meditech Co., Ltd., in China, as they are the submitter. The tests are non-clinical (bench testing), not derived from patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a non-clinical device safety and performance evaluation based on adherence to engineering and biological safety standards, not a study requiring expert ground truth for diagnostic purposes.

    4. Adjudication method for the test set

    Not applicable. The tests are based on objective measurements and compliance with standard specifications, not subjective expert judgment or adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a pen needle for insulin injection; it does not involve human readers, AI, or diagnostic interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a mechanical medical device, not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the non-clinical tests is defined by the objective metrics and specifications outlined in the referenced ISO and ASTM standards (e.g., specific dimensions, force thresholds, biological response, sterility levels).

    8. The sample size for the training set

    Not applicable. This is a medical device approval based on non-clinical testing against standards, not an AI model requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1