    Intended Use

    Temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), lower extremities (leg), abdomen and bottom due to strain from exercise or normal household work activities.

    Device Description

    The Progenix Select Stim (Model: PRO1205-OTC1) device is a portable, battery powered (9 V battery) two channel device designed to provide Transcutaneous Electrical Nerve Stimulation (TENS) for the purpose of the symptomatic relief of pain.

    The device consists of the TENS stimulator, two lead wires and set of four Skin Electrodes. It provides 10 preset TENS treatment programmes. The programs include four modes of operation: Conventional TENS(Continuous), Burst Mode, Modulation mode and HAN TENS. The device LCD screen shows the information of program, intensity level of each channel and operating time.

    The electrode pads are cleared by FDA (510(k) number K160138). They are used as an accessory to the TENS device unit, to transmit electrical current to patient skin. The electrodes consist of biocompatible self-adhesive conductive application use. These electrodes are sized 50x50mm and are sold under the model number PRO5050.

    AI/ML Overview

    The provided FDA 510(k) clearance letter for the Progenix Select Stim (K242460) is for a Transcutaneous Electrical Nerve Stimulator (TENS) device. Based on the document, this device is cleared as substantially equivalent to a predicate device (HIVOX OTC Electrical Stimulator, K190347).

    It is crucial to understand that a 510(k) clearance, especially for devices like TENS units, primarily relies on demonstrating substantial equivalence to a previously cleared predicate device, rather than requiring extensive de novo clinical trials or complex AI performance studies. The provided document explicitly states under "9. Clinical Performance Data" that it is "Not applicable." This means the clearance was granted based on non-clinical (bench) testing demonstrating that the device performs similarly to the predicate and meets relevant safety standards.

    Therefore, the request to describe "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of AI performance metrics (like sensitivity, specificity, MRMC studies, ground truth establishment by experts, etc.) is not directly applicable to this specific device and clearance. The device is a TENS unit, not an AI/ML-driven diagnostic or interpretative system.

    However, I can extract and structure the information related to the device's performance relative to its substantial equivalence claim and the non-clinical testing performed, which serves as its "proof of meeting acceptance criteria" in this context.

    Here's a breakdown based on the provided document, addressing the closest relevant points:


    Acceptance Criteria and Study that Proves the Device Meets the Acceptance Criteria (for a TENS Device)

    For the Progenix Select Stim (K242460), the "acceptance criteria" are primarily defined by its substantial equivalence to the predicate device and compliance with applicable voluntary safety standards. The "study that proves the device meets the acceptance criteria" refers to the non-clinical (bench) testing performed to demonstrate this equivalence and safety.

    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this is not an AI/ML device, the "performance" is about meeting electrical and mechanical specifications, and adherence to safety standards, rather than diagnostic accuracy metrics. The table below summarizes key comparative parameters that demonstrate substantial equivalence, which serve as "acceptance criteria" for the device's design and function.

    Parameter DescriptionAcceptance Criteria (Predicate Device K190347/K171803)Reported Device Performance (Progenix Select Stim K242460)Met Criteria? (SE Justification)
    Regulation Number21 CFR 882.589021 CFR 882.5890Met (SE) - Same regulatory classification.
    ClassificationClass IIClass IIMet (SE) - Same device class.
    General ClassificationOTCOTCMet (SE) - Both are Over-The-Counter devices.
    Intended UseTemporary pain relief (TENS)Temporary pain relief (TENS)Met (SE) - The intended use for TENS is identical. The predicate also has EMS function, which the new device does not, but this is acceptable for substantial equivalence for the TENS function.
    Power Source(s)4.5V (3xAAA)9V (1xPP3 alkaline)Met (SE Note 1) - Difference in voltage and battery type. Justified as not affecting safety/efficacy, as both deliver low-voltage electrical impulses and the device passed IEC 60601-1 tests.
    Patient Leakage Current (Normal)6.0 μA3 μAMet (SE) - Proposed device has lower leakage, indicating better safety performance.
    Patient Leakage Current (Single Fault)5.6 μA1 μAMet (SE) - Proposed device has lower leakage, indicating better safety performance.
    Output Waveform/ShapeBiphasic, RectangularBiphasic, RectangularMet (SE) - Identical.
    Number of Output Channels22Met (SE) - Identical.
    Regulated OutputRegulated CurrentRegulated CurrentMet (SE) - Identical.
    Maximum Output Voltage@500Ω 100V, @2kΩ 180V, @10kΩ 250V@500Ω 40V, @2kΩ 100V, @10kΩ 145VMet (SE Note 1) - Lower maximum output voltages. Justified as safer without compromising effectiveness, and device compliant with IEC 60601-1 and 60601-2-10.
    Maximum Output Current@500Ω 200mA, @2kΩ 90mA, @10kΩ 25mA@500Ω 80mA, @2kΩ 50mA, @10kΩ 19mAMet (SE Note 1) - Lower maximum output currents. Justified as safer without compromising effectiveness, and device compliant with IEC 60601-1 and 60601-2-10.
    Pulse Width (μs)50-45050-200Met (SE Note 2) - Narrower range. Justified as still within effective range for stimulating sensory/motor nerves and potentially minimizing discomfort.
    Frequency (Hz)1-1502-100Met (SE Note 3) - Narrower range. Justified as still within safe and effective range for therapeutic stimulation, covering low- and medium-frequency applications.
    Maximum Phase Charge (μC @ 500 Ω)4516Met (SE Note 4) - Lower. Justified as not exceeding safety limits (<0.25 W/cm²) and related to lower max output.
    Maximum Current Density (mA/cm²@ 500)9.873.2Met (SE Note 4) - Lower. Justified as not exceeding safety limits (<0.25 W/cm²) and related to lower max output and larger electrode surface area.
    Maximum Power Density (W/cm²@ 500)0.00460.0025Met (SE Note 4) - Lower. Justified as not exceeding safety limits (<0.25 W/cm²) and related to lower max output.
    Compliance with Voluntary StandardsES60601-1, IEC 60601-1-2, 60601-1-11, 60601-2-10IEC 60601-1, 60601-1-2, 60601-1-6, 60601-1-11, 60601-2-10, 62304Met (SE) - New device complies with all relevant standards of the predicate, plus additional standards (e.g., Usability, Software Life Cycle).

    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size for Test Set: Not applicable in the context of diagnostic/AI performance. The "test set" for this device involved non-clinical bench testing to verify electrical outputs, functional performance, and safety parameters against established engineering specifications and voluntary standards. The sample size would typically refer to the number of device units tested according to a quality control or design verification plan, but this specific number is not disclosed in the 510(k) summary.
    • Data Provenance: The data provenance is from non-clinical laboratory studies and safety testing conducted by the manufacturer (Progenix LLC) to demonstrate compliance with recognized consensus standards. This is not patient or clinical data. There is no mention of country of origin of data in this context. It is inherently prospective as it's part of the pre-market design verification and validation process.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Not applicable. This device is a TENS unit, not an AI/ML-driven diagnostic device requiring expert interpretation for ground truth establishment. The "ground truth" for this device's performance is adherence to engineering specifications and safety standards as assessed through physical and electrical measurements.

    4. Adjudication Method for the Test Set

    • Not applicable. No expert adjudication process like 2+1 or 3+1 is relevant for the non-clinical testing of a TENS device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No. The document explicitly states "9. Clinical Performance Data: Not applicable." MRMC studies are typically performed for imaging or diagnostic AI devices to assess the impact of AI assistance on human reader performance. This is a therapeutic device for pain relief.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Not applicable. There is no AI algorithm in this TENS device that performs a diagnostic or interpretative function. Its "standalone" performance is its ability to generate electrical impulses according to its specifications, which was verified through non-clinical testing.

    7. The Type of Ground Truth Used

    • Engineering Specifications and Voluntary Safety Standards: The "ground truth" for this device is its adherence to the design specifications (e.g., measured voltage, current, pulse width, frequency within specified ranges) and compliance with recognized international safety standards (e.g., IEC 60601-1, IEC 60601-2-10 for electrical safety, ISO 10993 for biocompatibility of electrodes).

    8. The Sample Size for the Training Set

    • Not applicable. This is not an AI/ML device, so there is no "training set" in the machine learning sense.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no training set for an AI/ML algorithm.