Found 3 results

510(k) Data Aggregation

    K Number
    K180533

    Validate with FDA (Live)

    Date Cleared
    2018-12-10

    (285 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Rx Indications for Use: For professional use, Procellera Composite Antibacterial Wound Dressing is intended for the management of wounds to provide a moist wound environment and is indicated for partial and full-thickness wounds such as pressure ulcers, venous ulcers, diabetic ulcers, first and second degree burns, surgical incisions, donor and recipient graft sites, etc. OTC Indications for Use: For over-the-counter use, Procellera Composite Antibacterial Wound Dressing is intended for the management of wounds to provide a moist wound environment and is indicated for superficial wounds such as minor cuts, scrapes, irritations, abrasions, blisters, etc.

    Device Description

    Procellera Composite Antibacterial Wound Dressing is a three-layer dressing comprising a broadspectrum antibacterial contact layer (Procellera), a polyester-based absorbent layer, and a polyurethane semi-occlusive outer adhesive layer to keep the dressing in place and help maintain a moist wound environment. The dressing is offered in various shapes and elbow/shoulder dressings are prescription use only, while the square, rectangular and round shaped dressings are for both prescription and over-the-counter use. Embedded in the dressing are microcell batteries made of elemental zinc applied in a dot-matrix pattern to a polyester substrate. In the presence of a conductive medium such as wound exudate, water-based wound hydrogels, saline, or water, microcurrents are generated at the dressing surface due to its inherent design. Silver and zinc in the dressing help to preserve it and minimize or prevent the growth of bacteria within the dressing. Procellera Composite Antibacterial Wound Dressing may be used with other common wound treatment products such as sutures, staples, liquid skin adhesives, or steri-strips as an adjunct to local clinical protocols.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device (Procellera Composite Antibacterial Wound Dressing) and as such, it focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study of acceptance criteria and device performance in the way a clinical trial or a more extensive performance study might.

    Based on the available information, here's a breakdown:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with specific numerical targets and then report corresponding device performance against those targets (e.g., "Device must achieve X% sensitivity; device achieved Y% sensitivity").

    Instead, the acceptance for this 510(k) submission is implicitly based on demonstrating that the new device is substantially equivalent to legally marketed predicate devices through a comparison of technological characteristics, intended use, and safety/performance data. The performance is assessed by showing compliance with relevant standards and established methods, implying that meeting these standards is the "acceptance criteria" for the device's characteristics.

    Here's an interpretation of the implied acceptance criteria and reported performance:

    CharacteristicImplied Acceptance Criteria (Demonstrated Equivalence/Compliance)Reported Device Performance/Evidence
    BiocompatibilityCompliance with ISO 10993: Biological Evaluation of Medical Devices standard."Procellera Composite Antibacterial Wound Dressing was tested in accordance with applicable parts of the ISO 10993: Biological Evaluation of Medical Devices standard and found to be biocompatible. Additional tests recommended by FDA guidance Use of International Standard ISO 10993-1 were also performed. Testing included: cytotoxicity, intracutaneous irritation, sensitization, acute systemic toxicity, material-mediated pyrogenicity, subchronic toxicity and implantation."
    Antibacterial EffectivenessDemonstrated broad-spectrum antibacterial effectiveness within the dressing (as per AATCC TM 100 or equivalent)."Procellera Composite Antibacterial Wound Dressing was tested per AATCC TM 100: Assessment of Antibacterial Finishes on Textile Materials and demonstrated broad-spectrum antibacterial effectiveness within the dressing."
    Intended UseIdentical to primary predicate device."The intended use... of Procellera Composite Antibacterial Wound Dressing are identical to the primary predicate single layer Procellera Antimicrobial Wound Dressing cleared under K160783 and its predecessors."
    Wound Contact LayerIdentical to primary predicate device."The wound contact layer of the subject device is identical to the Procellera Antimicrobial Wound Dressing."
    Antibacterial TechnologyIdentical to primary predicate device."The antibacterial technology... of Procellera Composite Antibacterial Wound Dressing are identical to the primary predicate single layer Procellera Antimicrobial Wound Dressing cleared under K160783 and its predecessors."
    MaterialsComparable composition to predicate devices, with new materials evaluated.New materials evaluated through standardized test methods (implicitly showing no new safety/effectiveness questions). See "Substantial Equivalence Comparison of Key Characteristics" table for comparison.
    ConfigurationComparable to predicate devices (multi-layered like secondary predicate, contact layer like primary).Comparisons provided in the "Substantial Equivalence Comparison of Key Characteristics" table.
    SterilityProvided sterile by irradiation."Provided sterilized by irradiation (gamma)." (Comparable to predicates.)
    Duration of UseUp to 7 days."Up to 7 days." (Identical to predicate devices.)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "sample size for a test set" in the context of clinical data. The studies conducted are primarily benchtop and in-vitro (biocompatibility, antibacterial effectiveness) rather than human clinical trials.

    • Biocompatibility testing: This typically involves a range of in-vitro and in-vivo tests using animal models or cell cultures, adhering to ISO 10993 standards. Specific sample sizes for these tests are not provided but are assumed to follow standard regulatory guidelines for such testing.
    • Antibacterial effectiveness testing (AATCC TM 100): This is an in-vitro laboratory test. The number of samples tested is not specified but would follow the methodology of the standard.

    There is no mention of human subject data, and therefore no information on data provenance (country of origin, retrospective/prospective).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This information is not applicable to the studies described. The ground truth for biocompatibility and antibacterial effectiveness is established by the standardized methods and accepted scientific principles of those tests, not by expert consensus on clinical cases.

    4. Adjudication Method for the Test Set

    This is not applicable. There is no mention of a human-read test set requiring adjudication in this 510(k) summary.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No, an MRMC comparative effectiveness study was not done. The document focuses on demonstrating substantial equivalence through technical characteristics and compliance with benchtop/in-vitro standards, not on human reader performance with or without AI assistance. This device is a wound dressing, not an AI-powered diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, this is not a software/AI device, so this concept is not applicable.

    7. The type of ground truth used

    For the safety and performance data presented:

    • Biocompatibility: Ground truth is established by adherence to ISO 10993 series standards and FDA guidance, meaning the biological response observed (e.g., lack of cytotoxicity, irritation) is the "ground truth" relative to the standard's criteria.
    • Antibacterial effectiveness: Ground truth is established by the quantifiable reduction in bacterial growth as measured by the AATCC TM 100 method.

    There is no clinical "ground truth" (pathology, expert consensus, outcomes data) derived from human subjects for this submission.

    8. The sample size for the training set

    This is not applicable. The device is a physical wound dressing, not an algorithm that requires a "training set."

    9. How the ground truth for the training set was established

    This is not applicable as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160783

    Validate with FDA (Live)

    Device Name
    Procellera
    Date Cleared
    2016-09-07

    (169 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For professional use. Procellera® antimicrobial wound dressing is intended for the management of wounds to provide a moist wound environment and is indicated for partial and full-thickness wounds such as pressure ulcers, diabetic ulcers, first and second degree burns, surgical incisions, donor and recipient graft sites, etc.

    For over-the-counter use, Procellera® antimicrobial wound dressing is intended for the management of wounds to provide a moist wound environment and is indicated for superficial wounds such as minor cuts, scrapes, blisters, etc.

    Device Description

    Procellera® is a single layer, broad-spectrum antimicrobial wound dressing: it consists of a polyester substrate containing elemental silver and elemental zinc bound to the surface by a biocompatible binder in a well-characterized dot matrix pattern.

    In the presence of a conductive fluid, such as wound exudate or moisture, a small amount of current is produced at the surface of the device, due to its inherent design. The device is self-contained and has no accessories.

    Procellera® contains Silver and Zinc as preservatives to the dressing, to minimize or prevent the growth of microorganisms within the dressing, not at the wound site, and to help preserve the dressing.

    Procellera® is a primary contact layer dressing and it should be used under a secondary dressing or bandage, which keeps it in place and helps maintain a moist wound environment. Procellera® may be used with other common wound treatment products such as, sutures, staples, liquid skin adhesives, or steri-strips as an adjunct to the local clinical protocols.

    AI/ML Overview

    This document is a 510(k) premarket notification for the Procellera Antimicrobial Wound Dressing, primarily focusing on demonstrating substantial equivalence to predicate devices rather than establishing novel safety and effectiveness through clinical trials with specific acceptance criteria. Therefore, much of the requested information regarding detailed acceptance criteria, specific study designs for performance, number of experts for ground truth, and training set details are not explicitly present in the provided text.

    Based on the information provided, here's an attempt to answer your questions:

    1. A table of acceptance criteria and the reported device performance

    The document does not specify quantitative acceptance criteria in terms of performance metrics (sensitivity, specificity, etc.) typically seen for diagnostic or AI-driven devices. Instead, the "acceptance criteria" are implied by the demonstration of:

    • Substantial Equivalence: The primary objective is to show the device is as safe and effective as legally marketed predicate devices.
    • Biocompatibility: Meeting ISO 10993 standards.
    • Antimicrobial Efficacy (in-vitro within the dressing): Previously cleared for prescription use.
    • Voltage Potential: Generation of sustained, measurable voltage.
    Acceptance Criteria (Implied)Reported Device Performance
    Substantial Equivalence to Predicate DevicesConcluded to be as safe, as effective, and performs as well as the predicate devices (Procellera® K130350, Procellera® K081977, CMBTM Antimicrobial Dressing K060237), which are identical to the current device prior to labeling for OTC use.
    BiocompatibilityTested per ISO 10993 and found to be biocompatible (including cytotoxicity, irritation, sensitization, pyrogenicity, and 28-day subcutaneous implantation).
    Broad-Spectrum in-vitro Antimicrobial EfficacyPreviously submitted to FDA and resulted in clearance for prescription use with an antimicrobial claim. Silver and zinc in the dressing inhibit growth of microorganisms within the dressing. The same data applies to OTC use as products are identical.
    Sustained Voltage Generation (Device Mechanism)Voltage potential testing confirmed that a sustained, measurable voltage is generated on the surface of the device for up to 30 days when immersed in a conductive fluid. (Note: Device use is for up to 7 days, or longer by clinician discretion, and safety beyond 28 days not studied).

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document refers to "in-vitro antimicrobial efficacy testing" and "voltage potential testing." These are laboratory/benchtop tests, not clinical studies involving human patient samples. Therefore, the concept of "sample size for a test set" in the context of human data is not applicable here. The "data provenance" would be laboratory data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not applicable. The evaluations are based on laboratory testing (in-vitro efficacy, biocompatibility, voltage potential), not expert interpretation of clinical data where a ground truth would be established by experts.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not applicable. Adjudication methods are typically used in clinical studies with expert reviewers, which is not the type of study described here.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was done, as this device is a wound dressing, not an AI-assisted diagnostic tool or an imaging device that involves human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a wound dressing, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance evaluation in this context is based on:

    • Laboratory measurements/standards: For biocompatibility (ISO 10993), in-vitro antimicrobial efficacy (presumably against standard microbial cultures), and electrical properties (voltage measurement).
    • Predicate devices: The "ground truth" for substantial equivalence is the established safety and effectiveness of the existing predicate devices.

    8. The sample size for the training set

    This information is not applicable. There is no mention of a "training set" as this device is not an AI/machine learning model. The studies described are performance tests of the physical device.

    9. How the ground truth for the training set was established

    This information is not applicable, as there is no training set for this type of device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K130350

    Validate with FDA (Live)

    Device Name
    PROCELLERA
    Date Cleared
    2013-07-02

    (140 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For over-the-counter use, Procellera® wound dressing is intended for the management of wounds to provide a moist wound environment and is indicated for superficial wounds such as minor cuts, scrapes, irritations, abrasions, blisters, etc.

    For prescription use, Procellera® antimicrobial wound dressing is intended for the management of wounds to provide a moist wound environment and is indicated for partial and full-thickness wounds such as pressure ulcers, venous ulcers, diabetic ulcers, first and second degree burns, surgical incisions, donor and/or recipient graft sites, etc.

    Device Description

    Procellera® is a single layer dressing; it consists of a polyester substrate containing elemental silver and elemental zinc bound to the surface by a biocompatible binder in a well-characterized dot matrix pattern.

    In the presence of a conductive fluid, such as wound exudate or moisture, a small amount of current is produced at the surface of the device, and it occurs because it is inherent to its design. The device is self-contained and has no accessories.

    Silver and zinc in the dressing produce microcurrent which helps the dressing to be preserved and to minimize or prevent growth of microorganisms within the dressing.

    Procellera® is a primary contact layer dressing and it should be used under a secondary dressing or bandage, which keeps it in place and helps maintain a moist wound environment.

    AI/ML Overview

    The provided text describes a 510(k) submission for the Procellera® Wound Dressing (K130350), which is a re-submission for over-the-counter use of a device previously cleared for prescription use (K081977). This submission primarily focuses on demonstrating substantial equivalence to predicate devices, drawing heavily on the previous clearances.

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Safety:
    (Implicit by reference to biocompatibility and substantial equivalence)The device was tested per ISO 10993 and found to be biocompatible, covering cytotoxicity, irritation, sensitization, pyrogenicity, and 28-day subcutaneous implantation. No new safety concerns were identified in comparison to predicate devices.
    Effectiveness:
    Ability to provide a moist wound environment for superficial wounds (minor cuts, scrapes, irritations, abrasions, blisters, etc.)Based on substantial equivalence to predicate devices (K081977, CMB™ dressing K060237), which are also wound dressings intended to provide a moist wound environment. The device's design (single layer, polyester substrate, used under a secondary dressing) supports its ability to maintain a moist environment.
    Mechanical/Electrical Performance:
    Sustained measurable voltage generation (inherent to design with elemental silver and zinc)Voltage potential testing (in-house) confirmed a sustained measurable voltage is generated on the surface of the device for up to 30 days when immersed in a conductive fluid.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not describe a clinical test set with a specific sample size for demonstrating performance and safety for this over-the-counter clearance. The submission relies on demonstrating substantial equivalence to previously cleared devices.

    The "Performance Data" section mentions "in-house" testing for voltage potential, but this is a technical verification of the device's inherent characteristic, not a clinical study on human subjects or a test set as typically understood in AI/diagnostic device validation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This information is not applicable as there was no clinical test set or diagnostic algorithm requiring ground truth established by experts. The substantiation is based on similarity to predicate devices and inherent material characteristics.

    4. Adjudication Method for the Test Set

    This information is not applicable as there was no clinical test set requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The device is a wound dressing, not an AI-powered diagnostic or assistive technology. Therefore, no MRMC study or AI performance evaluation was conducted.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a wound dressing and does not involve an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    This information is not applicable in the context of clinical performance evaluation for this 510(k). The primary "ground truth" or basis for clearance is demonstrating substantial equivalence to predicate devices (K081977, K060237), based on similar materials, technological characteristics, and intended use. The biocompatibility testing followed ISO standards, which provide their own established criteria (a form of 'ground truth' for safety).

    8. The sample size for the training set

    This information is not applicable. This is not an AI/ML device, and therefore, no training set was used.

    9. How the ground truth for the training set was established

    This information is not applicable. As there is no training set mentioned, no ground truth needed to be established for it.