510(k) Data Aggregation

    K Number
    K180140

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-04-20

    (92 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ProFlex® Laser Fibers are intended for use in laser-based surgical applications including, but not limited to endoscopic, laparoscopic and open surgical procedures involving vaporization, ablation and fragmentation of calculi (urinary and biliary) and surgical procedures involving vaporization, ablation, hemostasis, excision, resection and incision and of soft and cartilaginous tissues. While designed primarily for holmium (Ho:Y AG) lasers, ProFlex laser fibers may be used with any laser wavelength between 500 nm and 2200 nm that have been cleared for surgical use including, but not limited to frequency doubled Nd: Y AG (KTP) lasers, argon lasers, alexandrite lasers, ruby lasers, dye lasers, Nd:YAG lasers and Tm:YAG lasers.

    Device Description

    Subject devices and primary predicate devices are fiber optic energy delivery devices consisting of a stainless steel laser connector, e.g. SMA 905, an anodized aluminum expansion nut, strain relief, doubly step index clad optical fiber (fluoroacrylate over fluorine doped fused silica) with an ethylene tetrafluorethylene (ETFE, Tefzel™) protective jacket and are available in five nominal fiber core diameters (200 um, 273 µm, 365 µm, 550 um, 910 um, and 940 um), three nominal lengths (3 meters, 3.5 meters and 4.25 meters) and two laser-formed working tips (flat and orb). Subject devices and primary predicate devices are equipped with spatial and angular laser launch overfill protection -- a protective and guiding quartz ferrule, fused about the fiber input face upon which an input lens is laser-formed for high laser and physical damage threshold and for collimation of the laser focus energy within the optical fiber - and ruggedized output tips designed for ease of transit in flexible ureteroscopes and stone basket channels. Smaller fibers (where base fiber cores are smaller than laser foci) are equipped with tapered inputs where larger fibers are straight terminations.

    Subject devices are packaged in a coiled high density polyethylene (HDPE) tube carrier providing ease of dispensing within the surgical field while maintaining sterility. All materials of construction are USP Class VI biocompatible and are compatible with flash autoclave and ethylene oxide (EtO) sterilization.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA for a medical device called "ProFlex® Laser Fibers." It's a regulatory clearance letter, not a study report demonstrating the device meets acceptance criteria.

    Therefore, the requested information regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" cannot be fully provided from this document. This letter states that the device is substantially equivalent to a predicate device, meaning it has similar indications, technological characteristics, and performs as well as a legally marketed device. It does not contain a detailed report of a study designed to prove specific acceptance criteria against a defined performance metric.

    However, I can extract the relevant information that is present regarding performance and comparison:

    1. A table of acceptance criteria and the reported device performance:

    This document does not specify quantitative acceptance criteria. Instead, it relies on substantial equivalence. The reported performance is a comparative statement.

    Acceptance Criteria (Implied by Substantial Equivalence to Predicate)Reported Device Performance
    Safe and Effective (at least as good as predicate)"Subject devices performed as well than the predicate devices in all cases." "Subject devices are as safe, as effective and perform as well the predicate devices."
    Laser Power Delivery (relaxed configuration)Performed as well as predicate
    Laser Power Delivery (strained/bent configuration)Performed as well as predicate
    Shelf LifeDetermined to be 2 years (per accelerated aging tests)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size: Not explicitly stated. The document mentions "Subject and predicate fibers were laser power tested," implying a sample of fibers was used, but the quantity is not provided.
    • Data Provenance: The testing was conducted by InnovaQuartz, whose manufacturing facility is in Phoenix, Arizona, USA. The document doesn't specify if the testing was prospective or retrospective, but typically, premarket testing for substantial equivalence would be prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    Not applicable. This is a technical performance study of a physical device (laser fibers), not an assessment requiring expert interpretation of medical images or patient outcomes for ground truth establishment.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. This is a technical performance study, not a ground truth establishment process involving multiple readers.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a laser fiber, not an AI-powered diagnostic tool. No MRMC study was conducted or is relevant for this type of device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is a physical medical instrument, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" here is the measured performance of the laser fibers (e.g., laser power delivery) compared to a predicate device. This is based on bench testing/empirical measurements of physical characteristics, not medical "ground truth" typically established in diagnostic studies.

    8. The sample size for the training set:

    Not applicable. This is a physical device, not a machine learning algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable. (See #8)

    Ask a Question

    Ask a specific question about this device

    K Number
    K142638

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-12-19

    (93 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Proflex™ Laser Fibers are intended for use in laser-based surgical applications including, but not limited to endoscopic, laparoscopic and open surgical procedures involving vaporization, ablation and fragmentation of calculi (urinary and biliary) and surgical procedures involving vaporization, ablation, hemostasis, excision, resection and incision and of soft and cartilaginous tissues. While designed primarily for holmium (Ho: YAG) lasers, ProFlex™ fibers may be used with any laser wavelength between 500nm and 2200mm that have been cleared for surgical use including, but not limited to frequency doubled Nd: YAG (KTP) lasers, argon lasers, alexandrite lasers, ruby lasers, dye lasers, Nd:YAG lasers and Tm:YAG lasers.

    Device Description

    The ProFlex™ Laser Fibers are fiber optic energy delivery devices consisting of a stainless steel laser connector, e.g. SMA 905, an anodized aluminum expansion nut, strain relief, doubly step index clad optical fiber (fluoroacrylate over fluorine doped fused silica) with an ethylene tetrafluorethylene (ETFE, Tefzel™) protective jacket and are available in two fiber core diameters: 200µm & 273µm. The ProFlex™ Laser Fibers are equipped with laser launch overfill protection (spatial and angular) -- a protective, transmissive quartz ferrule, fused about the fiber input face and laser polished for high damage threshold (laser and physical), and ruggedized output tips designed for ease of transit in flexible ureteroscopes. All materials of construction are USP Class VI biocompatible and compatible with flash autoclave and EtO sterilization.

    AI/ML Overview

    The given text is a 510(k) premarket notification for a medical device (ProFlex™ Laser Fibers). It describes the device, its intended use, technological characteristics, and a comparison to predicate devices, but it does not contain information about acceptance criteria or a study proving that the device meets specific performance criteria in terms of clinical outcomes or diagnostic accuracy.

    The "Performance Testing" section (page 7) briefly mentions:

    • Bench and User Evaluation: "subject and predicate fibers were power tested in relaxed and strained (bending) configurations using a cleared surgical holmium laser."
    • Results: "ProFlex™ Laser Fibers performed better than the predicate devices in all four cases... individually, on average and in consistency."
    • OEM evaluation: "ProFlex has also been evaluated by one OEM holmium laser manufacturer and found to perform well."

    However, this is bench testing of the device's physical properties (power transmission under stress), not a clinical study to establish diagnostic performance, human reader improvement, or stand-alone AI performance. The document focuses on demonstrating substantial equivalence to existing predicate devices based on materials, intended use, technological characteristics, and some performance characteristics (power output, damage resistance).

    Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving that the device meets those criteria, as the provided text does not contain the necessary details for a clinical performance or diagnostic accuracy study. It specifically states that "ProFlex™ Laser Fibers are as safe, as effective and perform as well as the predicate devices," which refers to technical and material performance, not a clinical trial with human subjects or an AI component.

    To answer your prompt, the document would need to describe a study with a test set, ground truth, human readers (if applicable), and diagnostic metrics like accuracy, sensitivity, specificity, AUC, etc., along with predefined acceptance criteria for these metrics. This information is absent.

    Ask a Question

    Ask a specific question about this device