510(k) Data Aggregation

    K Number
    K180558

    Validate with FDA (Live)

    Device Name
    PressureWire X
    Date Cleared
    2018-03-28

    (27 days)

    Product Code
    Regulation Number
    870.2870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PressureWire™ X guidewire is indicated to direct a catheter through a blood vessel and to measure physiological parameters in the heart and in the coronary and peripheral blood vessels. Physiological parameters include blood pressure. The PressureWire™ X guidewire can also measure blood temperature.

    Device Description

    The PressureWire™ X guidewire has an integrated sensor element at the tip to enable measurements of physiological parameters. The wire is introduced into the patient's blood vessel. A torque device is used to steer the wire and sensor into the required position for pressure measurements according to standard clinical practice. PressureWire™ X guidewire is available in two different lengths. The guidewire is uniquely paired with a specific connection cable or with a specific transmitter. Both PressureWire™ X guidewire connection configurations connect to a diagnostic computer or a catheter laboratory hemodynamic recording system. The PressureWire™ guidewire is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a vessel. The signal output from the sensor is transmitted to associated equipment for analysis, calculations, and display of physiological parameters or indices based on pressure or temperature, e.g. Fractional Flow Reserve (FFR).

    AI/ML Overview

    The provided text describes the PressureWire™ X guidewire, a medical device used to direct catheters and measure physiological parameters like blood pressure and temperature. The submission is a Special 510(k), indicating a minor design modification to an already cleared device (K161171). As such, a detailed comparative effectiveness study with human readers or extensive clinical trials are not typically required for this type of submission. The focus is on demonstrating that the modified device remains substantially equivalent to the predicate device and that the modification does not raise new safety or effectiveness concerns.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    For a Special 510(k) like this, the "acceptance criteria" are primarily based on demonstrating that the modified device performs comparably to the predicate device and meets established performance standards for its type. The reported device performance centers on various bench tests.

    Acceptance Criterion (Based on Device Type and Modification)Reported Device Performance (Summary of Bench Testing)
    Maintain Guidewire Tensile StrengthSuccessfully demonstrated
    Maintain Torque StrengthSuccessfully demonstrated
    Maintain Fatigue ResistanceSuccessfully demonstrated
    Maintain Fracture ResistanceSuccessfully demonstrated
    Maintain Signal Quality in Severe BendSuccessfully demonstrated
    Maintain Signal Quality during Pullback ProcedureSuccessfully demonstrated
    Maintain Signal Drift within acceptable limitsSuccessfully demonstrated

    Note: The text states "Tests were conducted to evaluate the following: ... Additionally, simulated-use testing was completed as part of design validation to demonstrate the subject device met user needs and the intended use. The following test was performed: Signal Drift." It implies all these items were successfully met to demonstrate safety and effectiveness.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated in terms of a specific number of units for each test. The text outlines categories of tests (e.g., Tensile Strength, Torque Strength) and generally states "Performance bench testing was completed as part of design verification" and "simulated-use testing was completed as part of design validation." Typically, for bench testing, a statistically relevant number of samples would be tested to ensure reliability. However, this specific number is not disclosed in the summary.
    • Data Provenance: The testing was "Performance bench testing" and "simulated-use testing." This indicates the data was generated in a controlled laboratory environment, not from human or animal subjects. Thus, there's no "country of origin of the data" in the typical sense, nor is it retrospective or prospective clinical data. It's pre-market engineering and product verification/validation data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    • Number of Experts: Not applicable or mentioned. The "ground truth" for bench testing is defined by engineering specifications, validated test methods, and industry standards (e.g., for tensile strength, torque strength).
    • Qualifications of Experts: Not applicable. The tests are designed to objectively measure physical and electrical properties according to predefined methods and pass/fail criteria.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. Bench tests typically involve objective measurements and comparisons against predetermined specifications rather than subjective expert adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the Effect Size of human readers improve with AI vs without AI assistance

    • MRMC Comparative Effectiveness Study: No. The device is a physical guidewire with measurement capabilities, not an AI-powered diagnostic imaging tool that assists human readers in interpreting medical cases. Therefore, an MRMC study comparing human reader performance with and without AI assistance is not relevant or applicable here.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Not applicable in the context of an "algorithm only" device. The PressureWire™ X is a hardware device with an integrated sensor. Its "standalone performance" is demonstrated through the various bench tests ensuring the physical and measurement properties meet specifications.

    7. The type of ground truth used

    • Type of Ground Truth: For the "test set" (bench and simulated-use testing), the ground truth is based on engineering specifications, established physical properties, and validated measurement techniques. For example, a tensile strength test would have a pre-defined acceptable range of force the wire must withstand, and the measured force is compared against that objective standard.

    8. The sample size for the training set

    • Sample Size for Training Set: Not applicable. This device is a physical hardware product. It does not employ machine learning or AI algorithms that require a "training set" of data in the conventional sense. The "training" for such devices involves design, prototyping, and iterative testing, not algorithmic data training.

    9. How the ground truth for the training set was established

    • How Ground Truth for Training Set was Established: Not applicable, as there is no "training set" for an AI algorithm in this context. The "ground truth" during the device's development (design and iterative testing) would similarly be based on engineering principles, material science, and performance specifications derived from intended use and predicate device characteristics.
    Ask a Question

    Ask a specific question about this device

    K Number
    K161171

    Validate with FDA (Live)

    Date Cleared
    2016-08-15

    (111 days)

    Product Code
    Regulation Number
    870.2870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PressureWire™ X guidewire is indicated to direct a catheter through a blood vessel and to measure physiological parameters in the heart and in the coronary and peripheral blood vessels. Physiological parameters include blood pressure. The PressureWire™ X Guidewire can also measure blood temperature.

    Device Description

    The PressureWire™ X guidewire has an integrated sensor element at the tip to enable measurements of physiological parameters. The wire is introduced into the patient's blood vessel. A torque device is used to steer the wire and sensor into the required position for pressure measurements according to standard clinical practice. PressureWire™ X guidewire is available in two different lengths. The guidewire is uniquely paired with a specific connection cable or with a specific transmitter. Both PressureWire™ X guidewire connection configurations connect to a diagnostic computer or a catheter laboratory hemodynamic recording system.

    AI/ML Overview

    This document is a 510(k) premarket notification for the PressureWire™ X device, which is a catheter tip pressure transducer and guidewire. The content primarily focuses on demonstrating substantial equivalence to a predicate device and does not contain detailed information about acceptance criteria, device performance metrics, or specific study designs as would be found in a clinical trial report.

    Therefore, many of the requested elements for describing "acceptance criteria and the study that proves the device meets the acceptance criteria" cannot be fully provided from this document. The document describes design verification and validation testing rather than acceptance criteria for a diagnostic algorithm studied in a clinical setting.

    Here's what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance

    This document does not provide a table of acceptance criteria for a diagnostic performance study (e.g., sensitivity, specificity, accuracy). Instead, it refers to compliance with various standards and internal design control procedures for physical, mechanical, and coating integrity.

    Acceptance Criteria (General Categories)Reported Device Performance (Summary statement from document)
    BiocompatibilityIn compliance with ISO 10993-1:2009 with C1:2010
    SterilizationIn compliance with ISO TS 11135-2:2008 (Ethylene Oxide)
    Particulate MatterIn compliance with USP 33-NF 28 General chapter <788>
    Container MaterialsIn compliance with USP 36 <661>
    Physical, Mechanical, and Coating Integrity (e.g., fractures, friction force, wire diameter, straightness)Design verification and validation performed in compliance with internal design control procedures. Conclusion: "determined to be safe and effective and is substantially equivalent to the predicate device PressureWire™ Certus TM and PressureWire™ Aeris cleared in K140466."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The testing described is primarily bench testing and compliance with standards, not a clinical study on a patient test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided because the document describes verification and validation against pre-defined engineering and material standards, not the establishment of ground truth for a diagnostic performance test set by clinical experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not provided. The device (PressureWire™ X) is a physical guidewire with sensing capabilities, not an AI or imaging diagnostic algorithm that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A standalone performance evaluation for a diagnostic algorithm is not applicable here as the device is a medical instrument (guidewire) for measuring physiological parameters, not an algorithm. Its performance relates to its ability to accurately measure pressure and temperature, as well as its physical and mechanical properties. The document states a "testing summary" was done, concluding it is "safe and effective."

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For the physical and mechanical tests, the "ground truth" would be the established engineering specifications and manufacturing standards (e.g., specific friction force values, wire diameter tolerances). For biocompatibility and sterilization, the ground truth is compliance with the cited ISO and USP standards.

    8. The sample size for the training set

    This information is not provided. This device is not an AI/ML algorithm that requires a training set in the conventional sense.

    9. How the ground truth for the training set was established

    This information is not provided as it's not applicable to this type of device.