Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K233623

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-02-27

    (106 days)

    Product Code
    Regulation Number
    880.5200
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For vascular access devices flushing only. Sterile fluid path only. Not for use on a sterile field.

    Device Description

    The Praxiject™ SP 0.9% NaCl prefilled syringe is a single use plastic piston syringe with a Luer lock connection fitting, prefilled to labeled volume of 10mL with 0.9% Sodium Chloride Injection, USP, with no preservatives (normal saline), and capped with a plastic tip cap. The Praxiject™ SP 0.9% NaCl prefilled syringe is designed to maintain a sterile fluid path only and is individually packaged in a clear flexible plastic film, flow-wrapped heatsealed pouch that functions only as a protective barrier (dust cover). The device is terminally sterilized by gamma irradiation.

    AI/ML Overview

    The provided text is a 510(k) Pre-market Notification for a medical device called Praxiject™ SP 0.9% NaCl, a prefilled saline flush syringe. The document describes the device, its intended use, comparison to a predicate device, and non-clinical testing performed to demonstrate substantial equivalence. It explicitly states that "Clinical testing was not required for device evaluation."

    Therefore, based on the information provided, it's not possible to answer questions related to clinical studies, human reader performance, multi-reader multi-case (MRMC) studies, or the establishment of ground truth for a test set or training set in the context of an AI/algorithm-driven device. This document describes a medical device, not an AI/ML algorithm.

    However, I can extract information related to the acceptance criteria and the non-clinical study that proves the device meets those criteria, as detailed in the "Summary of Non-Clinical Testing" section.

    Here's the breakdown of the acceptance criteria and reported device performance from the non-clinical testing:

    Acceptance Criteria and Reported Device Performance (Non-Clinical Testing)

    1. A table of acceptance criteria and the reported device performance

    The document provides a table outlining performance characteristics, the applicable performance standards, and the acceptance criteria. It then states, "The design and manufacturing of the Praxiject™ SP 0.9% NaCl prefilled syringe is subject to risk assessment per ISO 14971 and verification and validation testing in conformance with regulatory guidance and consensus standards applicable to this device type." This implies that the device successfully met all listed acceptance criteria through this testing, although specific "reported device performance" values are not explicitly given in the table beyond the acceptance limits.

    Performance CharacteristicPerformance StandardAcceptance Criteria
    0.9% Sodium Chloride Solution
    AssayUSP Sodium Chloride Injection0.855 – 0.945% (w/v)
    IdentificationUSP <191> Sodium and ChlorideMeets USP requirements
    pHUSP <791>4.5 – 7.0
    Bacterial EndotoxinsUSP <85>, USP <161>≤ 0.5 USP EU/mL
    Particulate MatterUSP <788>Particles ≥ 10 µm: ≤ 6000/syringe;Particles ≥ 25 µm: ≤ 600/syringe
    Elemental ImpuritiesUSP <232> and USP <233>Arsenic: ≤ 1.5 µg/gCadmium: ≤ 0.2 µg/gLead: ≤ 0.5 µg/gMercury: ≤ 0.3 µg/g
    IronUSP <241>≤ 2 ppm
    AppearanceUSP <790> / Visual inspectionClear and colorless.Free of foreign solid particles.
    VolumeISO 7886-1, USP <1151>10.0 - 11.0 mL
    SterilityUSP <71>Sterile solution
    Sterilization Method ValidationISO 11137-1, ISO 11137-2, USP <61>(Terminal sterilization by gamma radiation)SAL 10-6
    Shelf-life (Stability)FDA Guidance: Shelf Life of Medical Devices (1991); FDA Guidance: Q1D Bracketing and Matrixing Designs for Stability Testing of New Drug Substances and Products (2003)Device specifications must be maintainedfor the labeled shelf life
    Syringe Design and Container closure integrityISO 7886-1, ISO 80369-7 / ISO 594-1 and ISO 594-2 (Adapted solution leakage test)Conformity to standards and devicespecifications; Label integrity; Syringe andtip cap seal integrity (no damage, noleakage past piston and tip cap.)
    Package integrityISO 11607-1, ISO 11607-2, ASTM F1980, ASTM D4169, ISO 7886-1 and ISO 594-2 (Syringe and tip cap integrity verification)No structural damage to shipping carton;Label integrity; Syringe and tip cap sealintegrity (no damage, no leakage pastpiston and tip cap.)
    Biological Safety EvaluationISO 10993-1, FDA Guidance: Use of International Standard ISO 10993-1, Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process (2020)Compliant process for the evaluation ofexternal communicating devices intendedfor indirect blood path contact with limitedduration
    Chemical Characterization (Extractables/Leachables)ISO 10993-18, ISO 10993-17USP <467>Acceptable extractables/leachables profile;Negligible risk of health hazard
    CytotoxicityISO 10993-5Non-cytotoxic
    SensitizationISO 10993-10Non-sensitizer
    IrritationISO 10993-23Non-irritant
    HemolysisISO 10993-4 (ASTM F756)Non-hemolytic
    Acute Systemic ToxicityISO 10993-11Non-toxic
    PyrogenicityISO 10993-11 (USP <151>)Non-pyrogenic

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify sample sizes for each non-clinical test beyond implication from standard compliance. These tests are typically conducted on batches of manufactured devices following established quality control procedures. Given the manufacturing details, data provenance would be MedXL Inc., Canada. The testing is prospective in nature, as it's part of the pre-market validation for the device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is not an AI/ML device requiring human expert annotation or ground truth establishment in the traditional sense. The "ground truth" for non-clinical testing of a physical medical device refers to the objective, measurable properties and performance standards themselves, verified by laboratory testing protocols and validated analytical methods (e.g., chemical assays, sterility tests, physical integrity tests).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as this is not a study involving subjective human interpretation or multi-reader scenarios. The testing involves objective, measurable criteria.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The device is a prefilled saline flush syringe, not an AI-assisted diagnostic or therapeutic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The "device" itself is a physical medical product, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For non-clinical testing of this device, the "ground truth" is defined by established consensus standards and pharmacopeial specifications (e.g., USP standards, ISO standards) for chemical composition, sterility, physical integrity, and biological safety. These are objective, measurable parameters.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1