Found 2 results

510(k) Data Aggregation

    K Number
    K162441

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-04-24

    (236 days)

    Product Code
    Regulation Number
    880.5970
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PowerPICC Provena Catheters with SOLO² Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, power injection of contrast media, and allows for central venous pressure monitoring. For central venous pressure monitoring, it is recommended that a catheter lumen of 20 gauge or larger be used.

    Device Description

    Bard Access Systems, Inc.'s PowerPICC Provena Catheters with SOLO² Valve Technology are sterile, single use devices designed to provide access to the patient's vascular system. The devices are intended for long or short-term use, as clinically indicated, to sample blood and administer fluids intravenously. The catheters are capable of central venous pressure monitoring, and can withstand power injection of contrast media. The catheters are peripherally inserted central catheters (PICC) and utilize the same placement technique as the predicate device. The subject devices include a silicone valve on the proximal end. The subject devices included in this notification are of varying French size and catheter configuration types, as summarized in the table below.

    Summary of Subject Devices
    Catheter Configuration | French size (Number of Lumens)
    PowerPICC Provena Catheters with SOLO² Valve Technology | 3 French Single Lumen (SL)
    PowerPICC Provena Catheters with SOLO² Valve Technology | 4 French Dual Lumen (DL)

    The following device descriptors apply to all French sizes and configurations of the subject catheters:
    ● Catheters are open-ended, radiopaque polyurethane;
    ● Catheters incorporate a silicone valve on the proximal end;
    ● Catheters have a reverse taper design;
    ● Catheter shaft tubing is marked with depth indicators, with "0" indicated to serve as a reference for the catheter insertion point;
    ● Purple colorant is included in the catheter material to provide the catheter with an appearance that allows the end user to differentiate Bard's power injectable catheters from other manufacturers' power injectable catheters;
    ● Catheter extension legs, luer hubs, and junction are printed with markings to identify the catheter as PowerPICC Provena Catheters with SOLO² Valve Technology, and include information to facilitate proper use of the device.

    The subject devices are provided sterile in basic interventional radiology (IR) kits, as well as basic, full, and max barrier nursing PICC kits with legally marketed components to assist in the placement procedure. These kits are available in both standard and small patient versions.

    AI/ML Overview

    The provided text describes specific performance tests conducted for the PowerPICC Provena Catheters with SOLO² Valve Technology to demonstrate substantial equivalence to predicate devices, rather than an AI/ML medical device. Therefore, the information provided does not align with the request to describe acceptance criteria and a study proving an AI/ML device meets those criteria.

    However, based on the provided text, I can infer the "acceptance criteria" from the "Risk Acceptability Criteria (Acceptance Criteria of Test)" column and the reported performance from the "Verification / Validation Method" column. The study described is a series of in-house and standard-based performance tests, not an MRMC or standalone AI study.

    Here's an interpretation based on the provided content, acknowledging its limitations regarding AI/ML:

    This document describes the testing and verification of a medical catheter, not an AI/ML-driven device. Therefore, many of the requested criteria specific to AI/ML systems (sample size for test/training sets, data provenance, expert ground truth, MRMC studies, effect size, standalone performance) are not applicable or cannot be extracted from the provided text.

    However, I can extract the acceptance criteria and stated performance for the catheter device as presented:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Risk Acceptability Criteria)Reported Device Performance (Verification / Validation Method)
    Mechanical Hemolysis Test: Testing to determine the hemolytic properties when blood is aspirated through the catheter assembly. • Bard internal standards and proceduresTesting to determine the hemolytic properties when blood is aspirated through the catheter assembly. (Implies satisfactory performance as per the concluding statement: "The subject devices met all predetermined acceptance criteria...")
    Dimensional Test: Test to measure OD and ID for single lumen catheters and OD and lumen area for dual lumen catheters to ensure compliance with dimensional specification. • Bard internal standards and procedures and ISO 10555-1:2013 - Sterile Single-Use Intravascular Catheters - Part 1: General requirementsTest to measure OD and ID for single lumen catheters and OD and lumen area for dual lumen catheters to ensure compliance with dimensional specification. (Implies satisfactory performance as per the concluding statement: "The subject devices met all predetermined acceptance criteria...")
    Catheter Collapse Test: Test to measure the flow rate of aspiration and demonstrate that the catheter will not collapse under a vacuum. • Bard internal standards and procedures and FDA guidance, Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters (March 16, 1995)Test to measure the flow rate of aspiration and demonstrate that the catheter will not collapse under a vacuum. (Implies satisfactory performance as per the concluding statement: "The subject devices met all predetermined acceptance criteria...")
    Luer to Extension Leg Tensile Test: Test to demonstrate the peak tensile force of each test piece exceeds the minimum peak tensile force. • ISO 10555-1:2013 – Sterile Single-Use Intravascular Catheters - Part 1: General requirementsTest to demonstrate the peak tensile force of each test piece exceeds the minimum peak tensile force. (Implies satisfactory performance as per the concluding statement: "The subject devices met all predetermined acceptance criteria...")
    Priming Volume: Test to measure the volume required to prime a full length catheter. • FDA guidance, Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters (March 16, 1995)Test to measure the volume required to prime a full length catheter. (Implies satisfactory performance as per the concluding statement: "The subject devices met all predetermined acceptance criteria...")
    Pump Flow: Test to determine the maximum pressure generated by the catheter when infusing water through it at a maximum pump flow rate. • Bard internal standards and proceduresTest to determine the maximum pressure generated by the catheter when infusing water through it at a maximum pump flow rate. (Implies satisfactory performance as per the concluding statement: "The subject devices met all predetermined acceptance criteria...")
    Gravity Flow: Test to measure the gravity flow rate. • Bard internal standards and procedures and FDA guidance, Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters (March 16, 1995) • ISO 10555-1:2013 – Sterile Single-Use Intravascular Catheters - Part 1: General requirementsTest to measure the gravity flow rate. (Implies satisfactory performance as per the concluding statement: "The subject devices met all predetermined acceptance criteria...")

    The document states: "The subject devices met all predetermined acceptance criteria derived from the above listed references and demonstrated substantial equivalence as compared to the cited predicate device."


    AI/ML Specific Questions (Cannot be answered from the provided text as it's not an AI/ML device)

    1. Sample sizes used for the test set and data provenance: Not applicable. This is a physical device.
    2. Number of experts used to establish the ground truth for the test set and qualifications: Not applicable. Ground truth for a physical device is established through physical and mechanical testing against standards.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Ground truth for this physical device is established through meeting material, mechanical, and functional performance specifications defined by internal standards, ISO standards (e.g., ISO 10555-1:2013), and FDA guidance documents.
    7. The sample size for the training set: Not applicable. This is a physical device, not an AI model.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K162443

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-10-25

    (55 days)

    Product Code
    Regulation Number
    880.5970
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PowerPICC Provena Catheters are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, power injection of contrast media, and allows for central venous pressure monitoring. For central venous pressure monitoring, it is recommended that a catheter lumen of 20 gauge or larger be used.

    Device Description

    Bard Access Systems, Inc.'s PowerPICC Provena Catheters are sterile, single use devices designed to provide access to the patient's vascular system. The devices are intended for short- or long-term use (>30 days) to sample blood and administer fluids intravenously. The catheters are capable of central venous pressure monitoring, and can withstand power injection of contrast media. The catheters are peripherally inserted central catheters (PICC) and utilize the same placement technique as the predicate devices.

    The subject devices included in this notification are of varying French size and catheter configuration types, as summarized in the table below.

    Summary of Subject Devices
    French size (Number of Lumens)
    Catheter Configuration
    PowerPICC Provena Catheters
    3 French Single Lumen (SL)
    4 French Dual Lumen (DL)

    The following device descriptors apply to all French sizes and configurations of the subject catheters:

    • Catheters are open-ended, radiopaque polyurethane;
    • Catheters have a reverse taper design;
    • Catheter shaft tubing is marked with depth indicators, with "0" indicated to serve as a reference for the catheter insertion point;
    • Purple colorant is included in the catheter material to provide the catheter with an appearance that allows the end user to differentiate Bard's power injectable catheters from other manufacturers' power injectable catheters; and
    • Catheter extension leg, luer hub, junction, and clamp ID tags are printed with markings to identify the catheter as PowerPICC Provena Catheters, and include information to facilitate proper use of the device.

    The subject devices are provided sterile in basic interventional radiology (IR) as well as basic, full, and max barrier nursing PICC kits with legally marketed components to assist in the placement procedure. These kits are available in both standard and small patient versions.

    AI/ML Overview

    The provided document describes the acceptance criteria and a detailed study for the C.R. Bard, Inc. PowerPICC Provena Catheters (K162443). This is a medical device, and the study involves a series of performance tests to demonstrate substantial equivalence to a predicate device, rather than a clinical study evaluating AI performance.

    Here's the information extracted from the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document provides a comprehensive list of verification/validation methods and their corresponding risk acceptability criteria. The reported performance for all these tests is a general statement that the device met all predetermined acceptance criteria.

    Verification / Validation MethodRisk Acceptability Criteria (Acceptance Criteria of Test)Reported Device Performance
    Biocompatibility TestingTests to confirm that the catheter is free from biological hazard. Bard internal standards and procedures and ISO 10555-1:2013 - Sterile Single-Use Intravascular Catheters – 1: General requirements. Testing Performed includes: Sensitization, Cytotoxicity, Irritation/Intracutaneous Reactivity, Acute Systemic Injection, Material-mediated Pyrogenicity, Subacute 14 day IV toxicity (Saline), Subacute 14 day IP toxicity (Cottonseed oil), Ames Genotoxicity, Mouse Lymphoma, Rodent Blood Micronucleus, 2 week and 6 week implantation, Hemolysis, Complement Activation.Met all predetermined acceptance criteria
    Cantilever Stiffness TestTest to characterize the catheter shaft stiffness at the distal section using a cantilever bend technique. Bard internal standards and procedures.Met all predetermined acceptance criteria
    Clamp EngagementTest to confirm that the catheter assembly will not leak when the clamp is engaged. Bard internal standards and procedures and ISO 10555-1:2013 – Sterile Single-Use Intravascular Catheters – 1: General requirements.Met all predetermined acceptance criteria
    Mechanical Hemolysis TestTesting to determine the hemolytic properties when blood is aspirated through the catheter assembly. Bard internal standards and procedures.Met all predetermined acceptance criteria
    Leak TestTest to confirm that the catheter assembly will not leak when the distal end of the catheter is occluded. Bard internal standards and procedures and ISO 10555-1:2013 – Sterile Single-Use Intravascular Catheters – 1: General requirements.Met all predetermined acceptance criteria
    Dimensional TestTest to measure OD and ID for single lumen catheters and OD and lumen area for dual lumen catheters to ensure compliance with dimensional specification. Bard internal standards and procedures and ISO 10555-1:2013 - Sterile Single-Use Intravascular Catheters - 1: General requirements.Met all predetermined acceptance criteria
    Implantable LengthTest to measure useful length for catheters to ensure compliance with dimensional specification. Bard internal standards and procedures and ISO 10555-1:2013 – Sterile Single-Use Intravascular Catheters – 1: General requirements.Met all predetermined acceptance criteria
    Extension Leg LengthTest to measure and confirm extension leg length compliance with dimensional specification. Bard internal standards and procedures.Met all predetermined acceptance criteria
    Catheter Collapse TestTest to measure the flow rate of aspiration and demonstrate that the catheter will not collapse under a vacuum. Bard internal standards and procedures and ISO 10555-1:2013 – Sterile Single-Use Intravascular Catheters – 1: General requirements.Met all predetermined acceptance criteria
    Burst TestBurst pressure test to confirm the catheter burst pressure exceeds the peak pressure present in the catheter at maximum flow conditions when the distal end is occluded. Bard internal standards and procedures and ISO 10555-1:2013 - Sterile Single-Use Intravascular Catheters - Part 1: General requirements.Met all predetermined acceptance criteria
    OD SwellTest to confirm that the catheter does not swell beyond twice the size of the labeled OD during power injection. Bard internal standards and procedures.Met all predetermined acceptance criteria
    Hydraulic Catheter Burst TestBurst pressure test to confirm the catheter burst pressure exceeds the peak pressure present in the catheter at maximum flow conditions when the distal end is occluded. Bard internal standards and procedures and ISO 10555-1:2013 - Sterile Single-Use Intravascular Catheters - Part 1: General requirements.Met all predetermined acceptance criteria
    Extension Leg Burst TestBurst pressure test to confirm the extension leg burst pressure exceeds the peak pressure present in the catheter at maximum flow conditions when the extension leg is occluded. Bard internal standards and procedures and ISO 10555-1:2013 - Sterile Single-Use Intravascular Catheters - Part 1: General requirements.Met all predetermined acceptance criteria
    Power Injection ConditioningTest to confirm the catheter does not leak or burst as a result of power injections at maximum indicated flow rate. Bard internal standards and procedures and ISO 10555-1:2013 - Sterile Single-Use Intravascular Catheters - Part 1: General requirements.Met all predetermined acceptance criteria
    Tip Stability TestTest to confirm that the catheter tip remains in the same orientation during power injection (tip pointing in direction of venous flow) at the maximum indicated flow rate. Bard internal standards and procedures.Met all predetermined acceptance criteria
    Assembly Tensile TestTest to demonstrate the peak tensile force of each test piece exceeds the minimum peak tensile force.Met all predetermined acceptance criteria
    Shaft Tensile Test Luer to Extension Leg Tensile TestISO 10555-1:2013 – Sterile Single-Use Intravascular Catheters – Part 1: General requirements.Met all predetermined acceptance criteria
    Shaft Tensile TestTest to evaluate the maximum catheter strain and modulus at break. Bard internal standards and procedures and FDA guidance, Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters (March 16, 1995).Met all predetermined acceptance criteria
    RadiopacityTest to demonstrate catheter radio-detectability. Bard internal standards and procedures and ISO 10555-1:2013 - Sterile Single-Use Intravascular Catheters - Part 1: General requirements.Met all predetermined acceptance criteria
    Suture Wing Integrity TestTest to measure the maximum force a catheter junction suture wing can withstand prior to break. Bard internal standards and procedures and FDA guidance, Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters (March 16, 1995).Met all predetermined acceptance criteria
    Priming VolumeTest to measure the volume required to prime a full length catheter. FDA guidance, Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters (March 16, 1995).Met all predetermined acceptance criteria
    Gravity FlowTest to measure the gravity flow performance of a full length catheter. Bard internal standards and procedures and ISO 10555-1:2013 - Sterile Single-Use Intravascular Catheters - Part 1: General requirements.Met all predetermined acceptance criteria
    Pump FlowTest to determine the maximum pressure generated by the catheter when infusing water through it at a maximum pump flow rate. Bard internal standards and procedures.Met all predetermined acceptance criteria
    Kink Diameter TestTest to quantitatively evaluate kink characteristics associated with the catheter shafts. Bard internal standards and procedures.Met all predetermined acceptance criteria
    Stylet Drag TestTest to ensure that the stylets used to place the catheter can be removed without difficulty. Bard internal standards and procedures.Met all predetermined acceptance criteria
    StatLock CompatibilityTest to ensure the catheter is compatible with StatLock for catheter securement. Bard internal standards and procedures.Met all predetermined acceptance criteria
    Taper LengthTest to measure and confirm taper length compliance with dimensional specification. Bard internal standards and procedures.Met all predetermined acceptance criteria
    Catheter PrintingTest is to confirm ink on the catheter exhibits proper adherence. Bard internal standards and procedures.Met all predetermined acceptance criteria

    The document explicitly states: "The subject devices met all predetermined acceptance criteria derived from the above listed references and demonstrated substantial equivalence as compared to the cited predicate device."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes (number of devices tested) for each individual test. It generally references "Bard internal standards and procedures" and relevant ISO standards (e.g., ISO 10555-1:2013) for the testing methodology. These standards typically define acceptable sample sizes for device testing.

    The data provenance is implied to be from in-house testing conducted by Bard Access Systems, Inc. There is no mention of country of origin for data if it were to be from external sources, nor does it specify if the data is retrospective or prospective, as this is a device performance study rather than a clinical trial.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information (number and qualifications of experts) is not relevant for this device performance study. The ground truth for these tests is established by objective, measurable engineering and material science standards and physical properties, not by expert interpretation.

    4. Adjudication Method for the Test Set

    Adjudication methods (like 2+1, 3+1, none) are typically used in clinical studies or studies involving human assessment of data. This document describes a battery of engineering and material science tests. Therefore, no adjudication method as described for clinical data interpretation is applicable or mentioned. The "adjudication" is inherent in the objective measurement against established criteria.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case comparative effectiveness study was not done. This document describes the safety and performance testing of a medical device (intravascular catheter), not an AI system. Therefore, there is no mention of human readers, AI assistance, or effect sizes related to AI.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    No, a standalone algorithm performance study was not done. This document pertains to the physical performance of a medical catheter, not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the tests described is based on objective, measurable engineering and material properties, and established industry standards. Examples include:

    • Physical measurements: OD, ID, length, flow rates, burst pressure, tensile strength.
      • Reference: Dimensional Test, Implantable Length, Extension Leg Length, Burst Test, Hydraulic Catheter Burst Test, Extension Leg Burst Test, Power Injection Maximum Flow Rate, Assembly Tensile Test, Shaft Tensile Test, Suture Wing Integrity Test, Priming Volume, Gravity Flow, Pump Flow, Kink Diameter Test, Taper Length.
    • Chemical/Biological properties: Biocompatibility outcomes (e.g., absence of biological hazard, lack of sensitization, cytotoxicity, irritation, hemolysis).
      • Reference: Biocompatibility Testing, Mechanical Hemolysis Test.
    • Functional performance: Absence of leaks, tip stability during power injection, ability to remove stylets easily, adherence of printing.
      • Reference: Clamp Engagement, Leak Test, Catheter Collapse Test, Power Injection Conditioning, Tip Stability Test, Stylet Drag Test, Catheter Printing.
    • Compliance with published standards: ISO 10555-1:2013 and FDA guidance documents.

    8. The sample size for the training set

    This is not applicable. This is a device performance study, not an AI model development or validation study. There is no concept of a "training set" in this context.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for an AI model in this document.