Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K162900

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-12-14

    (58 days)

    Product Code
    Regulation Number
    880.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PowerMidline™ Catheters are indicated for short term access to the peripheral venous system for selected intravenous therapies, blood sampling, and power injection of contrast may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The PowerMidline™ Catheters are suitable for use with power injectors.

    Device Description

    The dual lumen PowerMidline™ catheters are a family of peripherally placed catheters made from radiopaque body-softening polyurethane materials. Each dual lumen PowerMidline™ catheter is designed with kink-resistant, reverse taper design. The dual lumen PowerMidline™ catheters are offered in a 4 F dual lumen (DL) and 5 F dual lumen (DL) configuration for reliable short term (less than 30 days) vascular access. These catheters are offered in 20 cm trimmable lengths. The dual lumen PowerMidline™ catheters are suitable for use with power injectors.

    AI/ML Overview

    This document is a 510(k) summary for the PowerMidline Catheter (K162900) by Bard Access Systems, Inc. It describes the device, its intended use, indications for use, and a comparison to its predicate device (K153393). The document primarily focuses on demonstrating substantial equivalence to a previously cleared device, rather than defining and proving new acceptance criteria for an innovative device feature or algorithm.

    Therefore, the requested information about acceptance criteria and a study proving a device-specific, newly established acceptance criteria cannot be fully extracted from this document in the way it's usually applied to a new AI/medical device. The document explicitly states: "The performance tests completed on the subject device were limited to those tests required to support a determination of substantial equivalence to the predicate device." and "when technological characteristics between the subject and predicate device were found to be identical, results of the performance testing conducted on the predicate device were applied to the subject device."

    The "acceptance criteria" here are essentially "meets the performance requirements of the predicate device" or "meets the requirements of the specified ISO/FDA guidance documents."

    However, I can extract the types of performance tests conducted and the standards/guidances used, which implicitly define what was considered acceptable performance.

    1. Table of Acceptance Criteria and Reported Device Performance

    Instead of explicit numerical acceptance criteria, the document states that the device "met all predetermined acceptance criteria" derived from various performance tests. The "reported device performance" is the confirmation that these implicit criteria were met, leading to a determination of substantial equivalence.

    Acceptance Criteria (Implicit)Reported Device Performance
    Priming Volume: The priming volume of the dual lumen PowerMidline™ Catheter should be measured and reported in accordance with FDA Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters, March 16, 1995, and BAS Internal Protocols/Procedures. (Implicit criterion: The device must have a measured priming volume that is safe and acceptable for its intended use, comparable to similar devices if applicable)."Test and Report Priming Volume" - Met (as implied by the statement "the subject device met all predetermined acceptance criteria"). Specific values are not provided in this summary.
    Shaft Tensile Strength: The catheter shaft must meet tensile strength requirements as specified in ISO 10555-1: 2013, Sterile, single-use intravascular catheters, Part 1: General requirements. (Implicit criterion: The catheter shaft must be strong enough to withstand anticipated tensile forces during clinical use without failure)."Shaft Tensile" - Met (as implied by the statement "the subject device met all predetermined acceptance criteria").
    Modulus and Elongation: The device must meet modulus and elongation requirements as specified in FDA Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters, March 16, 1995 (Implicit criterion: The material properties of the catheter must allow for flexibility and durability appropriate for intravascular use)."Test and Report Modulus and Elongation Requirements" - Met (as implied by the statement "the subject device met all predetermined acceptance criteria").
    Assembly Burst (Burst Pressure with Power Injection): The catheter assembly must withstand burst pressure, particularly during power injection, within acceptable limits. (Implicit criterion: The catheter must not burst or leak under the pressures encountered during power injection)."Assembly Burst (Burst Pressure with Power Injection)" - Met (as implied by the statement "the subject device met all predetermined acceptance criteria").
    Catheter Assembly Leak: The catheter assembly must be leak-proof. (Implicit criterion: No fluid leakage should occur from the catheter assembly under normal operating conditions)."Catheter Assembly Leak" - Met (as implied by the statement "the subject device met all predetermined acceptance criteria").
    Catheter Assembly Tensile Strength: The catheter assembly must meet tensile strength requirements as specified in ISO 10555-1: 2013, Sterile, single-use intravascular catheters, Part 1: General requirements. (Implicit criterion: The assembled catheter must be strong enough to withstand anticipated tensile forces during clinical use without separation or failure)."Catheter Assembly Tensile" - Met (as implied by the statement "the subject device met all predetermined acceptance criteria").
    Gravity Flow: The device must exhibit acceptable gravity flow characteristics. (Implicit criterion: The catheter must allow for adequate fluid flow rates under gravity)."Test and Report Gravity Flow" - Met (as implied by the statement "the subject device met all predetermined acceptance criteria"). Specific values are not provided in this summary.
    Device Dimensional Characterization: The dimensions of the device must be accurately characterized and meet design specifications. (Implicit criterion: The physical dimensions of the catheter, including French size, length, and lumen dimensions, must conform to engineering drawings and be suitable for its intended use)."Device Dimensional Characterization" - Met (as implied by the statement "the subject device met all predetermined acceptance criteria"). Specific values are provided in the comparison table regarding 4 F DL x 20 cm and 5 F DL x 20 cm usable lengths, and Dual Lumen, two "D" shaped lumens.
    Catheter Cyclic Kink: The catheter must resist kinking under cyclic stress. (Implicit criterion: The catheter must maintain patency and functionality when subjected to bending forces encountered during patient movement or catheter placement)."Catheter Cyclic Kink" - Met (as implied by the statement "the subject device met all predetermined acceptance criteria").
    Tip Displacement During Power Injection: The catheter tip displacement during power injection must remain within acceptable limits. (Implicit criterion: The catheter tip should not migrate excessively or pose a risk of vascular damage during high-pressure contrast injection)."Tip Displacement During Power Injection" - Met (as implied by the statement "the subject device met all predetermined acceptance criteria").
    Power Injection Testing (Maximum Flow Rate): The device must safely accommodate specified maximum power injection flow rates per ISO 10555-1: 2013 and BAS Internal Protocols/Procedures. (Implicit criterion: The device must be able to withstand power injection at specified flow rates without failure or compromise to patient safety. Specific flow rates are stated as 4 F DL = 4 mL/s and 5 F DL = 7 mL/s)."Power Injection Testing" - Met, specifically: - 4 F DL = 4 mL/s - 5 F DL = 7 mL/s (as implied by the statement "the subject device met all predetermined acceptance criteria" and the values provided in the comparison table.)
    Biocompatibility: The device materials must be biocompatible as per ISO 10993-1:2009. (Implicit criterion: The device must not cause adverse biological reactions when in contact with the body)."Per ISO 10993-1:2009, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process, a biocompatibility evaluation was performed based upon the modifications made to develop the subject device." - Met.
    Risk Management: The device must comply with risk management procedures as per BS EN ISO 14971:2012. (Implicit criterion: Identified risks must be acceptably managed and mitigated)."Risk management, including a failure modes and effects analysis (FMEA), of the subject device was conducted in accordance with BS EN ISO 14971:2012, Medical Devices – Application of Risk Management to Medical Devices." - Met.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify an exact "sample size" for the performance tests in terms of number of catheters tested for each criterion. It simply states that "performance tests completed on the subject device were limited to those tests required to support a determination of substantial equivalence."

    • Sample Size: Not explicitly stated as a numerical count for each test. The tests were performed on "the subject device" which includes 4 F DL and 5 F DL configurations.
    • Data Provenance: The tests were conducted internally by Bard Access Systems, Inc. ("BAS Internal Protocols/Procedures"). This implies the data is prospective in nature, generated specifically for this submission, rather than retrospective clinical data. The country of origin is implicitly the USA, where the manufacturer is located.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable in the context of this device. This is a physical medical device (catheter), and the "ground truth" for its performance is established through objective engineering and biological performance testing against recognized standards (ISO, FDA guidance). There is no "ground truth" derived from expert interpretation of data in the way one would for an AI diagnostic algorithm.

    4. Adjudication Method for the Test Set

    Not applicable. This concept is relevant for studies involving human interpretation or clinical outcomes where discrepancies need to be resolved. Performance testing of a physical device against standards does not typically involve adjudication in this sense. The acceptance relies on the measured performance meeting the predefined (or implicitly defined by the standard) thresholds.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a physical medical device (catheter), not an AI algorithm intended for interpretation by human readers. Therefore, an MRMC study and analysis of human reader improvement with AI assistance is not relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is established by:

    • Compliance with recognized international standards (e.g., ISO 10555-1:2013, ISO 10993-1:2009, BS EN ISO 14971:2012).
    • Compliance with FDA Guidance documents (e.g., "Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters, March 16, 1995").
    • Internal protocols and procedures developed by the manufacturer, which would be designed to ensure the device meets its design specifications and is safe and effective for its intended use.

    Essentially, the "ground truth" is adherence to established engineering, materials, and biological performance specifications and regulatory requirements.

    8. The sample size for the training set

    Not applicable. This is a physical device submission; there is no "training set" in the context of machine learning or AI models. The design and manufacturing processes are analogous to a training process, leading to the final device design.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set in the AI sense, there is no ground truth establishment for it. The development of the device's design (analogous to "training") is guided by medical necessity, engineering principles, material science, regulatory requirements, and feedback from previous predicate devices. The "ground truth" for the predicate device would have been its demonstrated safety and effectiveness in previous clearances.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153393

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-06-28

    (217 days)

    Product Code
    Regulation Number
    880.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PowerMidline™ Catheters are indicated for short term access to the peripheral venous system for selected intravenous therapies, blood sampling, and power injection of contrast media. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The PowerMidline™ Catheters are suitable for use with power injectors.

    Device Description

    Bard Access Systems, Inc.'s (BAS) PowerMidline™ catheters are a family of peripherally placed catheters made from radiopaque body-softening polyurethane materials. Each PowerMidline™ catheter is designed with kink-resistant, reverse taper design. The PowerMidline™ catheters are offered in a 3 F Single Lumen (SL) and 4 F Single Lumen (SL) configuration for reliable short term (less than 30 days) vascular access. These catheters are offered in 20 cm trimmable lengths. The PowerMidline™ catheters are suitable for use with power injectors.

    AI/ML Overview

    I am sorry, but the provided text does not contain the detailed information necessary to fully answer your request regarding acceptance criteria and the study that proves the device meets them.

    Specifically, the text does not directly state:

    • A table of acceptance criteria and the reported device performance: While it lists performance tests, it does not provide the specific quantitative acceptance criteria or the results achieved.
    • Sample sized used for the test set and the data provenance: The document mentions performance tests were completed but does not specify the sample sizes used for these tests or the origin of any data (e.g., country, retrospective/prospective).
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: This information is not present as the studies described are primarily engineering performance tests of the catheter itself, not clinical studies involving expert evaluation of outcomes or data.
    • Adjudication method: Not applicable to the type of performance tests described.
    • If a multi reader multi case (MRMC) comparative effectiveness study was done: The document does not describe any MRMC studies.
    • If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable, as this is a physical medical device, not an algorithm.
    • The type of ground truth used: For the performance tests, the "ground truth" would be established by the test specifications and standards themselves (e.g., ISO, ASTM standards, BAS internal protocols), but not in the sense of clinical ground truth like pathology or expert consensus.
    • The sample size for the training set: Not applicable, as this is a physical medical device.
    • How the ground truth for the training set was established: Not applicable, as this is a physical medical device.

    What the document does provide regarding performance:

    The document states that "The subject device met all predetermined acceptance criteria derived from the above listed tests and demonstrated substantially equivalent performance as compared to the cited predicate and reference devices."

    It lists the following Performance Tests that were completed on the subject device, along with the standards utilized:

    • Device Dimensional Characterization
    • Catheter Assembly Tensile (ISO 10555-1: 2013)
    • Catheter Assembly Leak
    • Burst Pressure Without Power Injection
    • Burst Pressure Post Power Injection Gravity Flow (ISO 10555-1: 2013)
    • Tensile, Modulus and Elongation Requirements (FDA Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters, March 16, 1995)
    • Priming Volume (FDA Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters, March 16, 1995)
    • Catheter Radiopacity (BAS Internal Protocols/Procedures, ASTM F640-79 (reapproved 2000))
    • Power Injection Testing (ISO 10555-1: 2013)
    • Tip Displacement During Power Injection (BAS Internal Protocols/Procedures)
    • Catheter Cyclic Kink (BAS Internal Protocols/Procedures)

    Additionally, the following sterilization and biological tests were performed:

    • Sterilant Residual Testing (ISO 10993-7: 2008)
    • Cytotoxicity (ISO 10993-5: 2009)
    • Sensitization/Irritation (ISO 10993-10: 2010)

    The document concludes that the device met these criteria and demonstrated substantially equivalent performance to predicate and reference devices. However, the specific quantitative values for the acceptance criteria and the device's performance are not detailed in this summary.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1