Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K170158

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-06-01

    (134 days)

    Product Code
    Regulation Number
    880.5200
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PowerGlide ST™ Midline Catheter is inserted into a patient's vascular system for short term use to sample blood or administer fluids intravenously. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of the procedure. The PowerGlide ST™ Midline Catheter is suitable for use with power injectors.

    Device Description

    Bard Access Systems, Inc.'s PowerGlide ST™ Midline Catheter is a sterile, single use device designed to provide access to the patient's vascular system. The device is intended for short term use (<30 days) to sample blood and administer fluids intravenously. The subject device consists of a single lumen, radiopaque, body softening polyurethane catheter rated for power injection, and a dilator (when applicable). Additional procedural kit accessories (e.g. introducer needle and guidewire) are included to facilitate catheter placement. The PowerGlide ST™ Midline Catheter is offered in 18, 20, and 22 gauge sizes. The 18 and 20 gauge devices are offered in 8 cm or 10 cm lengths with an integrated dilator. The 22 gauge device is offered in only an 8 cm length and does not include a dilator.

    AI/ML Overview

    This document describes the premarket notification for a medical device, the PowerGlide ST™ Midline Catheter. It focuses on demonstrating substantial equivalence to a predicate device, not on proving clinical effectiveness through novel AI or diagnostic algorithms. Therefore, many of the requested criteria related to AI performance, ground truth establishment, expert readers, and MRMC studies are not applicable to this type of regulatory submission.

    Here's an analysis based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document lists performance tests and states that the "subject device met all predetermined acceptance criteria". However, the specific quantitative acceptance criteria values and the exact reported performance results for each test are not disclosed in this summary. It only indicates that the device passed.

    Performance TestAcceptance Criteria (Not Explicitly Stated)Reported Performance
    Dilator Outer DiameterWithin design specifications (Bard internal stds)Met criteria
    Dilator Effective LengthWithin design specifications (BS EN ISO 11070:2014)Met criteria
    Dilator Luer ConnectorComply with ISO 594-1 and 594-2 (6% luer taper)Met criteria
    Catheter/Dilator Tip Adhesion Break ForceForce required to remove dilator (Bard internal stds)Met criteria
    Dilator Assembly TensileDoes not fail in tension (BS EN ISO 11070:2014)Met criteria
    Catheter/Dilator Assembly Burst PressureGreater than maximum use pressure (ISO 10555-1:2013)Met criteria

    Biocompatibility Tests:

    Biological TestAcceptance Criteria (Not Explicitly Stated)Reported Performance
    CytotoxicityPer ISO 10993-5 (2009)Met criteria
    SensitizationPer ISO 10993-10 (2010)Met criteria
    Intracutaneous ReactivityPer ISO 10993-10 (2010)Met criteria
    Acute Systemic ToxicityPer ISO 10993-11 (2006)Met criteria
    PyrogenicityPer ISO 10993-11 (2006)Met criteria
    HemocompatibilityPer ISO 10993-4 (2002, amended 2006)Met criteria
    GenotoxicityPer ISO 10993-3 (2014), ISO/TR 10993-33 (2015)Met criteria
    Sterilization ResidualsPer ISO 10993-7Met criteria

    2. Sample size used for the test set and the data provenance
    The document does not specify the sample sizes used for the described physical and biological performance tests. It also does not discuss data provenance (e.g., country of origin, retrospective/prospective) as these are non-clinical, in-vitro/bench tests on device components, not clinical data sets.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    This question is not applicable. The "ground truth" for this medical device submission is based on engineering specifications, physical/chemical properties, and biological safety standards (ISO standards). It does not involve expert interpretations of medical images or diagnoses that would require human expert consensus.

    4. Adjudication method for the test set
    This question is not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies, particularly for diagnostic accuracy, where there's a need to resolve discrepancies between human readers or between AI and human reads. This document describes bench testing and biocompatibility assessments which do not involve this type of adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    This question is not applicable. This submission is for a physical medical device (catheter) and does not involve AI or human readers for diagnostic purposes. Therefore, an MRMC study and effects on human reader performance are irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    This question is not applicable. This device is a physical medical catheter, not an algorithm.

    7. The type of ground truth used
    The "ground truth" for this device's acceptance is based on:

    • Engineering design specifications: The physical dimensions and properties of the dilator and catheter.
    • Industry Standards (ISO/BS EN ISO): Established international and national standards for medical devices, specifically for intravascular catheters, luer connectors, and biocompatibility.
    • Bard internal standards and procedures: Proprietary testing methods and acceptance criteria used by the manufacturer.

    8. The sample size for the training set
    This question is not applicable. This is not an AI/machine learning device, so there is no concept of a "training set."

    9. How the ground truth for the training set was established
    This question is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1