Search Filters

Search Results

Found 8 results

510(k) Data Aggregation

    K Number
    K190080

    Validate with FDA (Live)

    Date Cleared
    2019-06-27

    (162 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Black) is a specialty medical glove which is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between examiner and patient. In addition, these gloves are worn to protect the wearer against exposure to chemotherapy drugs. Tested for use with chemotherapy drugs.

    Device Description

    Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Black) meets all the requirements of ASTM D6319 - 10(2015) Standard Specification for Nitrile Examination Gloves for Medical Application.

    AI/ML Overview

    This document is a 510(k) premarket notification for Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Black). It describes the device and compares it to a predicate device (Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green), K180476).

    Since this document is for a medical glove, not an AI/ML powered medical device, the information requested in the prompt related to AI/ML device studies (e.g., sample size for test/training sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, ground truth establishment) is not applicable. The device's performance is demonstrated through non-clinical testing against recognized standards.

    Here's a breakdown of the acceptance criteria and performance as gathered from the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The device, "Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Black)," meets established standards for medical examination gloves. The key acceptance criteria and reported performance, especially concerning chemotherapy drug permeation, are detailed below. The comparison is made against the predicate device (K180476) and relevant ASTM/ISO standards.

    CharacteristicStandard/Acceptance CriteriaPredicate Device (K180476) PerformanceSubject Device (K190080) PerformanceComparison
    GeneralASTM D6319-10(2015) Standard Specification for Nitrile Examination Gloves for Medical ApplicationMeets ASTM D6319-10(2015) requirementsMeets ASTM D6319-10(2015) requirementsSame
    DimensionsASTM 6319-10 (2015): Length-Min 240mm; Thickness palm and finger-Min 0.05mmLength-Min 240mm; Thickness palm and finger-Min 0.05mmLength-Min 230mm; Thickness palm and finger-Min 0.05mmSimilar
    Physical PropertiesASTM 6319-10 (2015): Min Tensile Strength (e.g., 20.23 MPa); Min Elongation (e.g., 445%)Min - 20.23 MPa; Min - 445%Min - 22.31 MPa; Min - 579%Similar
    Thickness (Finger)ASTM 6319-10 (2015)0.09 mm - 0.10 mm0.09 mm - 0.10 mmSimilar
    Thickness (Palm)ASTM 6319-10 (2015)0.12 mm - 0.14 mm0.11 mm - 0.13 mmSimilar
    Powder ContentASTM 6124-06 (2011): ≤ 2 mg/gloveMax - 2 mg/gloves (0.20 mg/glove reported)Max - 2 mg/gloves (0.60 mg/glove reported)Same
    Chemotherapy Permeation TestASTM D6978-05: Breakthrough Detection Time (for most listed drugs, ≥ 240 minutes)Similar
    * Cisplatin 1.0 mg/ml≥ 240 min>240 min>240 minSame
    * Cyclophosphamide 20 mg/ml≥ 240 min>240 min>240 minSame
    * Dacarbazine 10.0 mg/ml≥ 240 min>240 min>240 minSame
    * Doxorubicin 2.0 mg/ml≥ 240 min>240 min>240 minSame
    * Etoposide 20.0 mg/ml≥ 240 min>240 min>240 minSame
    * Fluorouracil 50.0 mg/ml≥ 240 min>240 min>240 minSame
    * Paclitaxel 6.0 mg/ml≥ 240 min>240 min>240 minSame
    * Carmustine 3.3 mg/ml*Warning for low permeation time reported23.4 min54.1 minImprovement
    * Thiotepa 10.0 mg/ml*Warning for low permeation time reported16.2 min16.0 minSimilar
    BiocompatibilityISO 10993-10:2010 (1. Primary Skin Irritation; 2. Skin Sensitization)1. Not irritating; 2. Not sensitizing1. Not irritating; 2. Not sensitizingSame
    * CytotoxicityISO 10993-5:2009 (Tests for in vitro cytotoxicity)Mild cytotoxicity reactivity (score of 2) with neat extract (100%)Mild cytotoxicity reactivity (score of 2) with neat extract (100%)Same
    Watertight21 CFR 800.20, ASTM D5151PassesPassesSame
    Indications for UseIntended for medical purpose to prevent contamination and protect against chemotherapy drugs (listed above).As described for predicate deviceAs described for subject deviceSame
    MaterialASTM D6319-10(2015)NitrileNitrileSame
    SizeMedical Glove Guidance Manual - Labeling: Extra Small, Small, Medium, Large, Extra LargeExtra Small, Small, Medium, Large, Extra LargeExtra Small, Small, Medium, Large, Extra LargeSame
    Single UseMedical Glove Guidance Manual - LabelingSingle UseSingle UseSame

    Study Information for the Device:

    As mentioned, much of the requested information is for AI/ML devices and is not applicable to a physical medical device like examination gloves. The provided document is a 510(k) submission, confirming substantial equivalence to a predicate device based on non-clinical performance testing.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not explicitly stated but implied to be sufficient for compliance with the referenced ASTM and ISO standards for glove testing (e.g., ASTM D6978 for chemotherapy permeation, ASTM D6319 for physical properties, ISO 10993 for biocompatibility). These standards define sampling plans for lot release testing. For instance, ISO 2859 for "Sampling Procedures and Tables for Inspection by Attributes" is referenced, indicating a statistical sampling approach.
    • Data Provenance: The manufacturing company is Comfort Rubber Gloves Industries Sdn. Bhd., located in Matang, Perak, Malaysia. The testing would presumably have been conducted in laboratories, potentially in Malaysia or certified testing facilities elsewhere, in accordance with the specified international standards. The data is prospective in the sense that the tests were conducted specifically for this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. The "ground truth" for a medical glove is determined by its physical and chemical properties as measured against established performance standards, not by expert interpretation of data. Laboratory testing provides the objective "ground truth."

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. Adjudication methods are relevant for subjective interpretations, typically in clinical readings or image analysis for AI/ML. For glove testing, objective measurements are taken according to standard protocols.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is a physical medical device, not an AI-powered diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" is established through objective laboratory testing results against predetermined pass/fail criteria defined by international standards (ASTM, ISO). These tests measure properties such as physical dimensions, tensile strength, elongation, powder content, integrity (watertightness), chemotherapy drug permeation time, skin irritation, sensitization, and cytotoxicity.

    8. The sample size for the training set

    • Not applicable. This is a physical medical device, not an AI/ML device requiring a training set.

    9. How the ground truth for the training set was established

    • Not applicable. There is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K180476

    Validate with FDA (Live)

    Date Cleared
    2018-05-15

    (82 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green) is a specialty medical glove which is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between examiner and patient. In addition, these gloves are worn to protect the wearer against exposure to chemotherapy drugs. Tested for use chemotherapy drugs. Tested drugs are as follows:

    Average Breakthrough Detection Time (minutes)
    Cisplatin 1.0 mg/ml≥ 240
    Cyclophosphamide (Cytoxan) 20 mg/ml≥ 240
    Dacarbazine (DTIC) 10.0 mg/ml≥ 240
    Doxorubicin Hydrochloride 2.0 mg/ml≥ 240
    Etoposide (Toposar) 20.0 mg/ml≥ 240
    Fluorouracil 50.0 mg/ml≥ 240
    Paclitaxel (Taxol) 6.0 mg/ml≥ 240

    Please note that the following drugs have extremely low permeation time for
    *Carmustine (BCNU) 3.3mg.ml - 23.4 (mins)
    *Thiotepa 10.0 mg/ml - 16.2 (mins)

    Device Description

    The proposed device is the Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green). The proposed device is for single use and is provided non-sterile. The device meets the requirements for dimensions, physical properties, and thickness in ASTM D6319 - 10(2015) Standard Specification for Nitrile Examination Gloves for Medical Application. The device accomplishes its intended function by creating a physical barrier between the examiner and patient and creates a barrier for the chemotherapy drugs listed in the Indications for Use.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for Powder Free Nitrile Examination Gloves tested for use with chemotherapy drugs. It focuses on demonstrating substantial equivalence to a legally marketed predicate device.

    Here's the information extracted and organized according to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (Standard)Reported Device Performance (New Device - K180476)Comparison
    Chemotherapy Drugs Permeation (Minimum Breakthrough Detection Time)ASTM D6978-05
    Cisplatin 1.0 mg/mlN/A (Comparative to predicate)>240 minutesSimilar to predicate
    Cyclophosphamide (Cytoxan) 20 mg/mlN/A (Comparative to predicate)>240 minutesSimilar to predicate
    Dacarbazine (DTIC) 10.0 mg/mlN/A (Comparative to predicate)>240 minutesSimilar to predicate
    Doxorubicin Hydrochloride 2.0 mg/mlN/A (Comparative to predicate)>240 minutesSimilar to predicate
    Etoposide (Toposar) 20.0 mg/mlN/A (Comparative to predicate)>240 minutesSimilar to predicate
    Fluorouracil 50.0 mg/mlN/A (Comparative to predicate)>240 minutesSimilar to predicate
    Paclitaxel (Taxol) 6.0 mg/mlN/A (Comparative to predicate)>240 minutesSimilar to predicate
    *Carmustine (BCNU) 3.3 mg/mlN/A (Comparative to predicate, with warning)23.4 minutesSimilar to predicate, with warning
    *Thiotepa 10.0 mg/mlN/A (Comparative to predicate, with warning)16.2 minutesSimilar to predicate, with warning
    DimensionsASTM D6319 - 10(2015)Length - Min 240mm; Thickness palm and finger - Min 0.05mmSame
    Physical PropertiesASTM D6319 - 10(2015)MeetsSame
    Thickness (Finger & Palm)ASTM D6319 - 10(2015)MeetsSame
    Powder ContentASTM D6124 - 06(2011) (≤ 2 mg/glove)MeetsSame
    Biocompatibility: Primary Skin IrritationISO 10993-10:2010Passes (non-irritating)Same
    Biocompatibility: Dermal SensitizationISO 10993-10:2010Passes (non-sensitization)Same
    Biocompatibility: Cytotoxicity and acute systemic toxicityISO 10993-5:2009 & ISO 10993-5:2017Passes (Certificate of Analysis for color imparting compound indicates non-cytotoxic, not systemically toxic)Different (Predicate N/A for this test)
    Watertight (1000ml)21 CFR 800.20, ASTM D5151PassesSame
    Indication for UseN/A (Comparative to predicate)Specialty medical glove for medical purpose, worn on examiner's hand/finger to prevent contamination between examiner and patient, and to protect against exposure to listed chemotherapy drugs.Similar
    MaterialASTM D6319 - 10(2015)NitrileSame
    ColorN/AGreenDifferent (Predicate: Blue or White)
    SizeMedical Glove Guidance Manual - LabelingExtra Small, Small, Medium, Large, Extra LargeSame
    Single UseMedical Glove Guidance Manual - LabelingSingle UseSame

    The acceptance criterion for the chemotherapy drug permeation test appears to be established by comparison to the predicate devices. For most drugs, the criterion is a minimum breakthrough detection time of ">240 minutes". For Carmustine and Thiotepa, the low permeation times observed for both the new device and the predicate lead to a warning statement, indicating that these drugs have "extremely low permeation times" and are not held to the ">240 minutes" criterion.

    2. Sample size used for the test set and the data provenance:

    • The document does not explicitly state the sample size used for the test set for properties like chemotherapy drug permeation, physical properties, or biocompatibility. It only reports "meets" or specific breakthrough times, implying that tests were conducted according to relevant standards (e.g., ASTM D6978-05 for chemotherapy drug permeation and ASTM D6319-10(2015) for physical properties and dimensions). These standards typically define sampling plans.
    • Data Provenance: The manufacturer is Comfort Rubber Gloves Industries Sdn. Bhd., located in Matang, Perak, Malaysia. The document does not specify if the testing was performed in Malaysia or another country. The data is non-clinical performance data, likely from prospective testing conducted to support the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not applicable as the device is a medical glove, and its performance is evaluated through objective, standardized laboratory tests against established physical, chemical, and biological criteria, not through expert interpretation of images or patient data to establish a ground truth.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • This is not applicable. The evaluation of medical gloves against performance standards (e.g., ASTM, ISO) involves quantitative measurements and predefined pass/fail criteria, not subjective interpretation requiring adjudication by experts.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This is not applicable. The device is a physical medical glove, not an AI-powered diagnostic or assistive technology. Therefore, no MRMC study or AI-related effectiveness assessment was conducted or reported.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • This is not applicable. The device is a physical medical glove and does not involve any algorithm or AI component.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for this device is based on objective, standardized test methods and criteria defined by international and national standards (e.g., ASTM D6319, ASTM D6124, ASTM D6978, ISO 10993, 21 CFR 800.20). These standards establish the acceptable performance characteristics for medical gloves. For biocompatibility, it's based on "passes" criteria of ISO standards. For chemotherapy permeation, it's the objectively measured breakthrough detection time.

    8. The sample size for the training set:

    • This is not applicable. The device is a physical medical glove and does not utilize a "training set" in the context of machine learning or AI. Its design and manufacturing are based on engineering principles and material science, not data-driven model training.

    9. How the ground truth for the training set was established:

    • This is not applicable for the reasons stated in point 8.
    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (White) is a specialty medical glove which is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between examiner and patient. In addition these gloves are worn to protect the wearer against exposure to chemotherapy drugs. Tested for use with chemotherapy drugs.
    The Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Blue) is a specialty medical glove which is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between examiner and patient. In addition these gloves are worn to protect the wearer against exposure to chemotherapy drugs. Tested for use with chemotherapy drugs.

    Device Description

    The proposed devices are Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim. The gloves are available in two colors: blue and white. The proposed devices are non-sterile. The Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Blue and White) meet all the requirements of ASTM Specification D6319-10 (2015) - Standard Specification for Nitrile Examination Gloves for Medical Application.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for "Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (White)" and "Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Blue)". This document is for a medical device (gloves) and not for an AI/ML-driven medical device or software. Therefore, the questions related to AI/ML device testing (such as sample size for test set, number of experts, MRMC studies, standalone performance, training set details) are not applicable to the content provided.

    However, I can extract the acceptance criteria and the study results for the glove's performance as presented in the document, especially regarding its resistance to chemotherapy drugs and other physical properties.

    Here's the information based on the provided document, addressing the relevant points and indicating "N/A" for those that do not apply to this type of device:


    Device Description: Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Blue and White)

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the gloves are primarily based on ASTM standards and comparisons to a predicate device (K141982). The key performance characteristic tested for this submission is the "Breakthrough Detection Time" for various chemotherapy drugs, which reflects the glove's ability to protect against exposure.

    Acceptance Criteria & Device Performance Comparison (Current Device vs. Standards/Predicate)

    CharacteristicStandard / Predicate Acceptance Criteria (Predicate 1: Dermagrip Powder Free Blue Nitrile Patient Examination Gloves Tested for Use with Chemotherapy Drugs, K141982)Current Device Performance (White Gloves)Current Device Performance (Blue Gloves)
    Physical Properties
    LengthMin 240mm (ASTM D6319-10(2015))Min 240mmMin 240mm
    Width (medium glove)Min 85 mm (ASTM D6319-10(2015))Min 85 mmMin 85 mm
    Tensile Strength (before aging)Meets min 14Mpa (ASTM D6319-10(2015))Meets min 14MpaMeets min 14Mpa
    Tensile Strength (after aging)Meets min 14Mpa (ASTM D6319-10(2015))Meets min 14MpaMeets min 14Mpa
    Elongation (before aging)Meets min 500% (ASTM D6319-10(2015))Meets min 500%Meets min 500%
    Elongation (after aging)Meets min 400% (ASTM D6319-10(2015))Meets min 400%Meets min 400%
    Thickness – FingerMeets min 0.05 mm (ASTM D6319-10(2015))Meets min 0.05 mmMeets min 0.05 mm
    Thickness – PalmMeets min 0.05 mm (ASTM D6319-10(2015))Meets min 0.05 mmMeets min 0.05 mm
    Powder Content≤ 2 mg/glove (ASTM D6124-06(2011))MeetsMeets
    Watertight (1000ml)Passes (21 CFR 800.20, ASTM D5151)PassesPasses
    Biocompatibility
    Primary Skin IrritationNon-irritating (ISO 10993-10:2010)Non-irritatingNon-irritating
    Dermal SensitizationNon-sensitizing (ISO 10993-10:2010)Non-sensitizingNon-sensitizing
    Chemotherapy Drug Permeation Test (ASTM D6978-05)Predicate Breakthrough Detection Time (min)Current Device White Gloves Breakthrough Detection Time (min)Current Device Blue Gloves Breakthrough Detection Time (min)
    Cisplatin, 1.0 mg/ml>240≥ 240≥ 240
    Cyclophosphamide, 20.0 mg/ml>240≥ 240≥ 240
    Dacarbazine (DTIC), 10.0 mg/ml>240≥ 240≥ 240
    Doxorubicin HCl, 2.0 mg/ml>240≥ 240≥ 240
    Etoposide (Toposar), 20.0 mg/ml>240≥ 240≥ 240
    Fluorouracil, 50.0 mg/ml>240≥ 240≥ 240
    Paclitaxel (Taxol), 6.0 mg/ml>240≥ 240≥ 240
    Ifosfamide, 50.0 mg/ml>240 (Predicate only, current device not tested for this)N/AN/A
    Mitoxantrone, 2.0 mg/ml>240 (Predicate only, current device not tested for this)N/AN/A
    Vincristine Sulfate, 1.0 mg/ml>240 (Predicate only, current device not tested for this)N/AN/A
    Carmustine (BCNU) 3.3mg/ml15.024.018.2
    Thiotepa 10.0 mg/ml2.054.957.3
    Warning StatementIdentical warning regarding low permeation times for Carmustine and Thiotepa.Identical warning, updated times based on testing.Identical warning, updated times based on testing.

    Note: For Carmustine and Thiotepa, while the breakthrough times might seem low, the predicate also had low times (Thiotepa as low as 2 minutes). The current device shows improved or comparable performance for these challenging drugs, and the critical acceptance criterion is that the warning statement regarding these drugs' properties is present and updated to reflect the tested performance.

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample size for each specific test (e.g., number of gloves tested for chemotherapy permeation or physical properties). However, the testing would have been conducted in accordance with the cited ASTM standards, which define the necessary sample sizes and testing methodologies for such evaluations.

    • Data Provenance: The device is manufactured by Comfort Rubber Gloves Industries Sdn Bhd in Matang, Perak, Malaysia. The testing data would originate from laboratory tests conducted on the manufactured gloves.
    • Retrospective or Prospective: These tests are prospective (pre-market) evaluations conducted specifically to support the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    N/A. This is a physical device where "ground truth" is established through standardized laboratory testing protocols (e.g., ASTM D6978 for chemotherapy drug permeation, ASTM D6319 for physical properties, ISO 10993 for biocompatibility) that yield objective measurements, rather than expert interpretation of complex data (like medical images).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    N/A. Not applicable for physical device performance testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    N/A. Not applicable for this physical device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    N/A. Not applicable for this physical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device is based on objective measurements derived from standardized laboratory test methods (e.g., ASTM standards for glove performance and ISO standards for biocompatibility). For instance, the breakthrough detection time for chemotherapy drugs is a direct measurement based on chemical analysis, not expert consensus or clinical outcomes.

    8. The sample size for the training set

    N/A. This document pertains to a physical medical device (gloves), not an AI/ML system requiring a training set.

    9. How the ground truth for the training set was established

    N/A. This document pertains to a physical medical device (gloves), not an AI/ML system.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151390

    Validate with FDA (Live)

    Date Cleared
    2016-05-02

    (342 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTMD6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:

    Test Chemotherapy Drug and ConcentrationMinimum Breakthrough Detection Time (Minutes)
    Cisplatin, 1.0mg/ml>240
    Cyclophosphamide (Cytoxan), 20mg/ml>240
    Carmustine (BCNU), 3.3mg/ml1.0
    Dacarbazine (DTIC), 10.0mg/ml>240
    Doxorubicin Hydrochloride, 2.0mg/ml>240
    Etoposide (Toposar), 20.0mg/ml>240
    Fluorouracil, 50.0mg/ml>240
    Paclitaxel (Taxol), 6.0mg/ml>240
    Thiotepa, 10.0mg/ml31.3
    Methotrexate ,25mg/ml>240
    Mitomycin C, 0.5mg/ml>240
    Vincristine Sulfate, 1.0mg/ml>240
    Ifosfamide, 50.0mg/ml>240

    Please note that the following drugs have low permeation time of less than 240 minutes: Carmustine (BCNU), 3.3mg/ml: 1.0 minute Thiotepa, 10.0mg/ml: 31.3 minutes

    Device Description

    These patient examination gloves are formulated using nitrile, non-sterile, powder free, meet all the requirements of ASTM D6319 and tested for use with chemotherapy drugs.

    AI/ML Overview

    This document does not describe a device that uses Artificial Intelligence (AI) or Machine Learning (ML). It is an FDA 510(k) premarket notification for Powder Free Nitrile Examination Gloves Tested For Use with Chemotherapy Drugs.

    Therefore, the requested information regarding acceptance criteria, study details, sample sizes, ground truth, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance for an AI/ML device is not applicable to this document.

    The document focuses on the physical and chemical properties of the gloves, primarily their resistance to permeation by chemotherapy drugs, and compares them to established ASTM standards and a predicate device.

    However, to address the acceptance criteria and performance data as presented in the document for the physical device (gloves):

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the gloves are primarily based on ASTM standards. The relevant performance data is provided in "Table 1" on page 4 of the document, comparing the applicant's device to a predicate device and relevant standards.

    CharacteristicsStandardsAcceptance Criteria (from Standards or Predicate)Reported Device Performance (Current Device)
    DimensionsASTM D6319-10Min 240mm (Predicate)Min 270mm
    Physical PropertiesASTM D6319-10MeetMeet
    Thickness - FingerASTM D6319-100.07-0.10mm (Predicate)Min 0.10mm
    Thickness - PalmASTM D6319-100.07-0.09mm (Predicate)Min 0.10mm
    Freedom From HolesASTM D6319-10 and ASTM D5151PassPass
    Powder Free ResidueASTM D6319-10 and ASTM D6124MeetMeet
    Resistance to Permeation by Chemotherapy Drugs (ASTM D6978-05)Minimum Breakthrough Detection Time (min)Minimum Breakthrough Detection Time (min)
    Cisplatin, 1.0mg/ml>240 minutes>240 minutes
    Cyclophosphamide (Cytoxan), 20mg/ml>240 minutes>240 minutes
    Carmustine (BCNU), 3.3mg/ml15.0 minutes1.0 minute (with warning)
    Dacarbazine (DTIC), 10.0mg/ml>240 minutes>240 minutes
    Etoposide (Toposar), 20.0mg/ml>240 minutes>240 minutes
    Doxorubicin Hydrochloride, 2.0mg/ml>240 minutes>240 minutes
    Fluorouracil, 50.0mg/ml>240 minutes>240 minutes
    Paclitaxel (Taxol), 6.0mg/ml>240 minutes>240 minutes
    Thiotepa, 10.0mg/ml2.0 minutes31.3 minutes (with warning)
    Methotrexate, 25mg/ml- (Not explicitly listed for predicate)>240 minutes
    Mitomycin C, 0.5mg/ml- (Not explicitly listed for predicate)>240 minutes
    Vincristine Sulfate, 1.0mg/ml>240 minutes>240 minutes
    Ifosfamide, 50.0mg/ml>240 minutes>240 minutes
    Biocompatibility - Primary Skin IrritationISO 10993PassPass
    Biocompatibility - Dermal SensitizationISO 10993PassPass
    MaterialASTM D6319-10NitrileNitrile
    Texture-Finger TexturedFinger Textured

    Note on Chemotherapy Drug Permeation: The "acceptance criteria" here are comparison points from the predicate device and the standard. The device performance for Carmustine and Thiotepa for the current device are lower than >240 minutes, but still considered acceptable with explicit warnings regarding their low permeation times, similar to how the predicate device also had drugs with low permeation times. For Thiotepa, the current device performance (31.3 minutes) is significantly better than the predicate (2.0 minutes). For Carmustine, the current device performance (1.0 minute) is lower than the predicate (15.0 minutes). These differences are noted and addressed by a warning statement.

    This summary is based on the information provided, which details the device's compliance with established industry standards for patient examination gloves and specific testing for chemotherapy drug permeation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K113100

    Validate with FDA (Live)

    Date Cleared
    2012-02-06

    (110 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTMD6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:

    Test Chemotherapy Drug and ConcentrationMinimum Breakthrough Detection Time (Minutes)
    Cisplatin, 1.0mg/ml (1,000ppm)>240
    Cyclophosphamide (Cytoxan), 20mg/ml (20,000ppm)>240
    Carmustine (BCNU), 3.3mg/ml (3,300ppm)0.53
    Dacarbazine (DTIC), 10.0mg/ml (10,000ppm)>240
    Doxorubicin Hydrochloride, 2.0mg/ml (2,000ppm)>240
    Etoposide (Toposar), 20.0mg/ml (20,000ppm)>240
    Fluorouracil, 50.0mg/ml (50,000ppm)>240
    Paclitaxel (Taxol), 6.0mg/ml (6,000ppm)>240
    Thiotepa, 10.0mg/ml (10,000ppm)75.54

    Please note that the following drugs have low permeation time of less than 240 minutes: Carmustine (BCNU), 3.3mg/ml (3,300ppm) Thiotepa, 10.0mg/ml (10,000ppm)

    Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)

    Device Description

    Not Found

    AI/ML Overview

    This document describes a 510(k) premarket notification for "Blue Powder Free Nitrile Examination Gloves Tested For Use With Chemotherapy Drugs (Non-Sterile)". The primary purpose of this submission is to demonstrate substantial equivalence to legally marketed predicate devices for general patient examination glove use, and additionally, to provide data on the gloves' resistance to permeation by chemotherapy drugs.

    As such, the information provided focuses on the testing for chemotherapy drug permeation.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the test standard ASTM D6978-05, which is a standard practice for assessing medical gloves for permeation by chemotherapy drugs. While the document doesn't explicitly state a "pass/fail" threshold for breakthrough time (e.g., all drugs must be > X minutes), it does highlight drugs with permeation times less than 240 minutes, implying that longer breakthrough times are generally more desirable. The reported performance is the minimum breakthrough detection time for each drug.

    Test Chemotherapy Drug and ConcentrationAcceptance Criteria (Implicit)Reported Device Performance (Minimum Breakthrough Detection Time)
    Cisplatin, 1.0mg/ml (1,000ppm)Longer breakthrough time is desirable.>240 minutes
    Cyclophosphamide (Cytoxan), 20mg/ml (20,000ppm)Longer breakthrough time is desirable.>240 minutes
    Carmustine (BCNU), 3.3mg/ml (3,300ppm)Longer breakthrough time is desirable.0.53 minutes
    Dacarbazine (DTIC), 10.0mg/ml (10,000ppm)Longer breakthrough time is desirable.>240 minutes
    Doxorubicin Hydrochloride, 2.0mg/ml (2,000ppm)Longer breakthrough time is desirable.>240 minutes
    Etoposide (Toposar), 20.0mg/ml (20,000ppm)Longer breakthrough time is desirable.>240 minutes
    Fluorouracil, 50.0mg/ml (50,000ppm)Longer breakthrough time is desirable.>240 minutes
    Paclitaxel (Taxol), 6.0mg/ml (6,000ppm)Longer breakthrough time is desirable.>240 minutes
    Thiotepa, 10.0mg/ml (10,000ppm)Longer breakthrough time is desirable.75.54 minutes

    Note on Acceptance Criteria: The document explicitly notes: "Please note that the following drugs have low permeation time of less than 240 minutes: Carmustine (BCNU), 3.3mg/ml (3,300ppm) Thiotepa, 10.0mg/ml (10,000ppm)". This indicates that while the device was tested, a breakthrough time of less than 240 minutes for these specific drugs is acknowledged and disclosed, rather than being an outright "failure". The acceptance criterion is therefore the performance according to the specified ASTM standard, with the understanding that users would be informed about the permeation characteristics.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size (number of gloves or individual tests) used for the chemotherapy drug permeation testing. It only lists the results for each drug.

    The data provenance is not explicitly stated as "country of origin" or "retrospective/prospective". However, the manufacturer is Maxter Glove Manufacturing Sdn. Bhd. in Malaysia, suggesting the testing likely occurred in or was commissioned by the manufacturer, possibly in Malaysia or a testing lab accredited to perform ASTM D6978-05. The testing itself is a prospective performance test of the device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of device (patient examination gloves) and the specific test (chemotherapy drug permeation) do not involve "experts" establishing a ground truth in the traditional sense of clinical diagnosis or interpretation. The "ground truth" is established by the standardized laboratory test procedure itself (ASTM D6978-05), which objectively measures the breakthrough time of the chemotherapy drugs through the glove material. The "truth" is the measured time according to the standard.

    4. Adjudication Method for the Test Set

    Not applicable. This is a laboratory performance test, not a subjective assessment requiring adjudication. The results are quantitative measurements.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a performance test for a patient examination glove, not an AI-assisted diagnostic or interpretive device involving human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is a physical device (glove), not an algorithm or AI. The testing performed is an objective measurement of its physical properties.

    7. The type of ground truth used

    The ground truth used is objective laboratory measurement based on a standardized test method (ASTM D6978-05). This standard details how to measure the permeation of chemotherapy drugs through medical glove materials, providing a quantifiable breakthrough time.

    8. The sample size for the training set

    Not applicable. This is a physical device undergoing performance testing, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. As this is not a machine learning model, there is no training set or associated ground truth for a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K081315

    Validate with FDA (Live)

    Date Cleared
    2008-08-11

    (94 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. This device also provides protection against exposure to certain chemotherapy drugs.

    Device Description

    Powder Free Nitrile Examination Gloves Tested For Use With Chemotherapy Drugs (Pink, Black, Grey, and Blue Peppermint Scented)

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for "Powder Free Nitrile Examination Gloves Tested For Use With Chemotherapy Drugs." It does not contain information on acceptance criteria for a device's performance or a study proving that a device meets such criteria. Therefore, I cannot answer your request based on the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K071119

    Validate with FDA (Live)

    Date Cleared
    2007-07-24

    (95 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K063627

    Validate with FDA (Live)

    Date Cleared
    2007-03-20

    (104 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. This device also provides protection against exposure to certain chemotherapy drugs.

    Device Description

    Powder Free Nitrile Examination Gloves Tested For Use With Chemotherapy (Blue)

    AI/ML Overview

    This document is a 510(k) premarket notification approval letter for "Powder Free Nitrile Examination Gloves Tested for use with Chemotherapy (Blue)". It does not contain information about the acceptance criteria or a study that proves the device meets acceptance criteria in the context of an AI/ML device.

    Therefore, I cannot answer the question based on the provided text. The document is about medical gloves, not an AI/ML device, and thus doesn't discuss performance metrics like sensitivity, specificity, or typical study designs for AI systems.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1