Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K241970

    Validate with FDA (Live)

    Date Cleared
    2024-08-02

    (28 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Powder Free Nitrile Examination Gloves (Blue) , Tested for Use with Chemotherapy Drugs and Fentanyl Citrate is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The tested chemotherapy drugs are: Carmustine 3.3 mg/ml(3,300 ppm) 27.7 (35.5,27.7,27.7) Cisplatin 1 mg/ml (1,000 ppm) >240min Cyclophosphamide 20 mg/ml (20,000 ppm) >240min Dacarbazine 10mg/ml (10,000 ppm) >240min Doxorubicin HCl 2 mg/ml(2,000 ppm) >240min Etoposide 20 mg/ml (20,000 ppm) >240min Fluorouracil 50mg/ml(50,000 ppm) >240min Gemcitabine HCl 38 mg/ml (38,000 ppm) >240min Ifosfamide 50mg/ml (50,000 ppm) >240min Methotrexate 25mg/ml (25,000 ppm) >240min Mitomycin C 0.5 mg/ml(500 ppm) >240min Mitoxantrone HC1 2 mg/ml (2,000 ppm) >240min Paclitaxel 6 mg/ml (6,000 ppm) >240min Thiotepa 10 mg/ml (10,000 ppm) 59.4(67.5,59.4,65.9) Vincristine Sulfate 1 mg/ml (1,000 ppm) > 240min The tested non-chemotherapy drugs are: Fentanyl Citrate injection 100mcg/2ml >240min Note: Carmustine and Thiotepa have extremely low permeation times of 27.7 and 59.4 minutes respectively. Warning: Do Not Use with Carmustine, Thiotepa

    Device Description

    The Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs and Fentanyl Citrate are non-sterile, single use only, disposable examination gloves intended for medical purposes to be worn by examiners to prevent contamination between the examiner. The gloves are blue color, powder free, nitrile gloves. The gloves are offered in six sizes: XS、S、M、L、XL、XXL, packed in a paper box.

    The proposed gloves are designed and manufactured in accordance with the ASTM D6319-19 standard and are tested for use with chemotherapy drugs and Fentanyl per ASTM D6978-05. The proposed device was modification from the legally marketed (existing) device "Powder free Nitrile Examination Gloves (Blue, Violet Blue, White, Cobalt Blue)"/K202356, manufactured by "KANGLONGDA VIETNAM PROTECTION TECHNOLOGY COMPANY LIMITED", same as the sponsor of this submission.

    There are two modifications, one is to add a large model XXL, and the other is to add the "Tested for Use with Chemotherapy Drugs and Fentanyl Citrate" indications for use than the legally marketed (existing) device. Testing has been conducted on the proposed device passed as per ASTM D6978-05, ASTM D6319-19, ASTM D5151-19, ASTMD 6124-06.

    The materials and the manufacturing process technology are the same.

    AI/ML Overview

    The provided document (a 510(k) summary for medical gloves) does not contain information about a study that proves a device meets acceptance criteria related to AI/ML or human reader performance. The document is for "Powder Free Nitrile Examination Gloves" and pertains to mechanical and material properties, as well as resistance to chemical permeation.

    Therefore, I cannot provide details on:

    1. A table of acceptance criteria and reported device performance for an AI/ML device.
    2. Sample size used for a test set or data provenance for an AI/ML device.
    3. Number of experts used to establish ground truth or their qualifications for an AI/ML device.
    4. Adjudication method for an AI/ML device test set.
    5. Multi-reader multi-case (MRMC) comparative effectiveness study for human readers with and without AI.
    6. Standalone performance of an algorithm.
    7. Type of ground truth used (expert consensus, pathology, outcomes data, etc.) for an AI/ML device.
    8. Sample size for the training set of an AI/ML device.
    9. How the ground truth for an AI/ML training set was established.

    However, based on the document, I can present the acceptance criteria and performance for the glove device as non-clinical testing.


    Acceptance Criteria and Study for Powder Free Nitrile Examination Gloves

    This document describes the non-clinical testing performed on "Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs and Fentanyl Citrate" to demonstrate its safety and effectiveness.

    1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical Testing of Gloves):

    Test MethodPurposeAcceptance CriteriaReported Device Performance
    ASTM D6319-19Physical Dimensions TestLength (mm): XS/S/M/L/XL/XXL: ≥230Pass
    Width (mm): XS: 70±10mm; S: 85±10mm; M: 95±10mm; L: 110±10mm; XL: 120±10mm; XXL: 130±10mmPass
    Thickness (mm): Finger: ≥0.05; Palm: ≥0.05Pass
    ASTM D5151-19Watertightness Test for Detection of HolesBe free from holes when tested in accordance with ASTM D5151Pass
    ASTM D412Physical Properties (Before Aging)Tensile Strength: 14MPa, minPass
    Ultimate Elongation: 500% minPass
    ASTM D412Physical Properties (After Aging)Tensile Strength: 14MPa, minPass
    Ultimate Elongation: 400% minPass
    ASTM D6124Powder ContentMeet the requirements of Less than 2mg per glovePass
    ASTM D6978-05 (2019)Detect the Permeation time by Chemotherapy Drugs and Fentanyl of the gloveMeet the requirements of ASTM D6978-05. (Specific breakthrough times listed elsewhere in document)Pass
    ISO 10993-10Irritation and SensitizationNon-irritating; Non-sensitizerUnder the conditions of the study, not an irritant and not a sensitizer
    ISO 10993-5CytotoxicityNon CytotoxicUnder conditions of the study, not cytotoxic

    2. Sample size used for the test set and the data provenance:

    • The document does not specify the exact sample sizes (e.g., number of gloves tested for each attribute) for the physical and chemical resistance tests. It only states that tests were conducted according to the listed ASTM and ISO standards.
    • Data Provenance: The testing was conducted to verify that the device meets design specifications and complies with international standards. The origin of the testing data is not explicitly stated beyond being "non-clinical tests."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is not applicable as the device is a medical glove, not an AI/ML diagnostic or prognostic tool. Ground truth is established by objective physical and chemical measurements following established standards, not by expert consensus or physician reads.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable for this type of device testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable as this is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable as this is not an AI/ML device.

    7. The type of ground truth used:

    • For the physical properties (dimensions, watertightness, tensile strength, powder content), the ground truth is based on direct physical measurement and standard test methods outlined in ASTM standards.
    • For chemical permeation (chemotherapy drugs and Fentanyl), the ground truth is established by analytical measurement of breakthrough time according to ASTM D6978-05.
    • For biocompatibility (irritation, sensitization, cytotoxicity), the ground truth is established by biological assays according to ISO 10993 standards.

    8. The sample size for the training set:

    • Not applicable as this is not an AI/ML device. The "training" for glove manufacturing relates to process control and material specifications, not a data-driven model.

    9. How the ground truth for the training set was established:

    • Not applicable as this is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K240051

    Validate with FDA (Live)

    Date Cleared
    2024-02-29

    (52 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves have been tested for use with chemotherapy drugs, Fentanyl Citrate, and select other drugs using ASTM D6978-05(2019)

    Device Description

    Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, and select other drugs are Class I Patient Examination Gloves and Specialty Chemotherapy Gloves. They are ambidextrous and come in different sizes - Extra Small, Small, Medium, Large, Extra Large and XXL. Gloves meet the specification of ASTM D6319-19 and have been tested for resistance to permeation by chemotherapy drugs, Fentanyl Citrate, and select other drugs as per ASTM D6978-05(2019). The gloves are single use, disposable, and non-sterile.

    AI/ML Overview

    The provided document describes the acceptance criteria and performance of "Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, and select other drugs" (K240051).

    Here's the breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test PerformedAcceptance CriteriaReported Device Performance
    Physical Dimensions
    LengthMinimum 220mm for size XS and S, 230mm for size M, L, XL, XXL (ASTM D6319-19)Pass
    Palm Width (size)XS: 70±10 mm, S: 80±10 mm, M: 95±10 mm, L: 110±10 mm, XL: 120±10 mm, XXL: 130±10 mm (ASTM D6319-19)Pass
    ThicknessFinger: 0.05mm (min), Palm: 0.05mm (min) (ASTM D6319-19)Pass
    Physical Properties
    Tensile Strength & Elongation (Before Aging)Tensile Strength: 14MPa, min; Elongation: 500%, min (ASTM D6319-19, ASTM D412-16(2021))Pass
    Tensile Strength & Elongation (After Accelerated Aging)Tensile Strength: 14MPa, min; Elongation: 400%, min (ASTM D6319-19, ASTM D412-16(2021))Pass
    Barrier Integrity
    Watertight (1000ml)G-I, AQL 2.5 (21 CFR 800.20, ASTM D5151-19)Pass
    Powder ResidueMax 2mg/glove (ASTM D6319-19, ASTM D6124-06(2017))Pass
    Chemotherapy Drugs & Opioid Drug PermeationMinimum Breakthrough Detection Time (BDT) for various chemotherapy drugs and Fentanyl Citrate. Accepted values differ by drug, but generally, longer BDT indicates better performance. Specific acceptance criteria are not explicitly listed as a single value but are implied by the reported BDTs (e.g., >240 minutes for many drugs, with lower specific values for Carmustine and Thiotepa which are cautioned against).See specific BDT values in the tables below for each drug. Notably, Carmustine: 29.1 minutes, Thiotepa: 78.3 minutes. All other listed drugs exhibit >240 minutes BDT. Fentanyl Citrate: >240 minutes.
    Biocompatibility
    Irritation and Skin SensitizationNon-sensitization and Non-irritation (ISO 10993-10 & -23)Non-sensitization and Non-irritation
    CytotoxicityCytotoxicity reactivity (ISO 10993-5:2009)Showed potential toxicity to L929 cells, but concern addressed by acute systemic toxicity testing.
    Acute systemic toxicity studyNo adverse biological reaction (ISO 10993-11:2017)No evidence of acute systemic toxicity.

    Chemotherapy, Fentanyl Citrate & other drugs Permeation Comparison (Proposed Device K240051):

    Tested Chemotherapy Drug and ConcentrationMinimum Breakthrough Detection Time in Minutes
    Bleomycin Sulfate 15mg/ml (15000 ppm)>240
    Busulfan 6mg/ml (6,000 ppm)>240
    Carboplatin 10mg/ml (10,000 ppm)>240
    Carmustine 3.3 mg/ml (3,300 ppm)29.1
    Cisplatin 1mg/ml (1,000 ppm)>240
    Cyclophosphamide 20mg/ml (20,000 ppm)>240
    Cytarabine, 100 mg/ml (100,000 ppm)>240
    Dacarbazine 10 mg/ml (10,000 ppm)>240
    Daunorubicin HCL, 5 mg/ml (5,000 ppm)>240
    Docetaxel , 10 mg/ml (10,000 ppm)>240
    Doxorubicin HCL, 2 mg/ml (2,000 ppm)>240
    Epirubicin HCL, 2 mg/ml (2,000 ppm)>240
    Etoposide, 20 mg/ml (20,000 ppm)>240
    Fludarabine Phosphate, 25 mg/ml (25,000 ppm)>240
    Fluorouracil, 50mg/ml (50,000ppm)>240
    Gemcitabine HCL, 38mg/ml (38,000ppm)>240
    Idarubicin HCL, 1mg/ml (1,000ppm)>240
    Ifosfamide, 50mg/ml (50,000ppm)>240
    Irinotecan HCL, 20mg/ml (20,000ppm)>240
    Mechlorethamine HCI, 1mg/ml (1,000ppm)>240
    Melphalan HCL, 5mg/ml (5,000ppm)>240
    Methotrexate, 25mg/ml (25,000ppm)>240
    Mitomycin C, 0.5mg/ml (500ppm)>240
    Mitoxantrone HCL, 2mg/ml (2,000ppm)>240
    Oxaliplatin, 5mg/ml (5,000ppm)>240
    Paclitaxel, 6mg/ml (6,000ppm)>240
    Paraplatin, 10mg/ml (10,000ppm)>240
    Rituximab, 10mg/ml (10,000ppm)>240
    Thiotepa, 10mg/ml (10,000ppm)78.3
    Topotecan HCL, 1mg/ml (1,000ppm)>240
    Trisenox (Arsenic Trioxide), 1mg/ml (1,000ppm)>240
    Velcade (Bortezomib), 1mg/ml (1,000ppm)>240
    Vincristine Sulfate, 1mg/ml (1,000ppm)>240
    Fentanyl Citrate Injection (100 mcg/2ml)>240
    Chloroquine 50mg/ml (50,000ppm)>240
    Cyclosporin A 100 mg/ml (100,000 ppm)>240
    Retrovir, 10mg/ml (10,000ppm)>240

    Warning: Do not use with Carmustine and Thiotepa due to "extremely low permeation times" (29.1 minutes and 78.3 minutes, respectively).

    2. Sample size used for the test set and the data provenance:

    • The document primarily refers to standard ASTM and ISO test methodologies for device performance (e.g., ASTM D6319-19, ASTM D6978-05(2019), ISO 10993 series). These standards typically define the sample sizes required for each specific test to ensure statistical validity.
    • Data Provenance: The document states "Non-clinical tests were conducted to verify that the proposed device met all design specifications." The tests were performed by Lingshi Hongruida Health Protection Technology Co., Ltd. located in China. The data is retrospective in the sense that the testing was performed and then submitted with the 510(k) application. Specific details on exact sample sizes (e.g., number of gloves tested for each characteristic) within each standard are not explicitly detailed in this summary for every test. However, the chemotherapy drug permeation testing likely involved multiple samples per drug as per ASTM D6978.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not applicable as the device is a physical medical device (examination glove) and the testing involved objective, quantitative, and standardized measurements according to established consensus standards (ASTM, ISO). There were no "experts" establishing a "ground truth" in the way they would for, say, image interpretation or clinical diagnosis. The ground truth for performance relied on the specified chemical and physical property measurements.

    4. Adjudication method for the test set:

    • This is not applicable. As mentioned above, the testing involved objective physical and chemical measurements governed by established standards, not subjective interpretations requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This is not applicable. This device is an examination glove, not an AI-powered diagnostic tool or imaging software. Therefore, no MRMC study or AI-related comparative effectiveness study was performed or is relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • This is not applicable. This device is an examination glove and does not involve any algorithm or AI for standalone performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth for this device's performance is based on objective measurements against established performance standards and chemical permeation resistance thresholds.
      • For physical properties (length, thickness, tensile strength, elongation, watertightness, powder content): The ground truth is the quantitative measurement compared against the specified ASTM standard values.
      • For chemical permeation (chemotherapy drugs, Fentanyl Citrate): The ground truth is the experimentally determined Minimum Breakthrough Detection Time (BDT) in minutes, measured according to ASTM D6978-05(2019).
      • For biocompatibility (irritation, sensitization, cytotoxicity, acute systemic toxicity): The ground truth is the experimental findings and conclusions derived from tests conducted under ISO 10993 series standards.

    8. The sample size for the training set:

    • This is not applicable. This device is a physical product, not a machine learning model, so there is no training set in the context of AI/ML.

    9. How the ground truth for the training set was established:

    • This is not applicable as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K223320

    Validate with FDA (Live)

    Date Cleared
    2022-11-02

    (5 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K211604

    Validate with FDA (Live)

    Date Cleared
    2021-10-27

    (155 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves have been tested for use with chemotherapy drugs using ASTM D6978 and will be labeled with a statement of compliance and a summary of the testing results.

    Device Description

    Powder Free Nitrile Examination Gloves(Blue), Tested for Use with Chemotherapy Drugs are Class I Patient Examination Gloves and Specialty Chemotherapy Gloves. They are ambidextrous and come in different sizes-Extra Small, Small, Medium, Large and Extra Large. Gloves meet the specification of ASTM D6319-19 and have been tested for resistance to permeation by chemotherapy drugs as per ASTM D6978-05(2019).

    AI/ML Overview

    The provided document describes the acceptance criteria and performance data for "Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs." This is a Class I medical device, and the testing focuses on its physical properties, biocompatibility, and resistance to chemotherapy drug permeation. It is crucial to note that this document pertains to a physical medical device (gloves) and not an AI/Software as a Medical Device (SaMD). Therefore, questions related to AI/SaMD specific aspects like expert data labeling, MRMC studies, training set data, etc., are not applicable to this submission. The information provided aligns with typical performance testing for physical devices.

    Here's the breakdown of the acceptance criteria and reported performance, tailored to the available information:


    Acceptance Criteria and Reported Device Performance

    1. Table of Acceptance Criteria and the Reported Device Performance

    Test MethodPurposeAcceptance CriteriaReported Device Performance
    ASTM D6319-19LengthMinimum 230mmAll size ≥230mm
    ASTM D6319-19Palm WidthXS: 70 ± 10mmS: 80 ± 10mmM: 95 ± 10mmL: 110 ± 10mmXL: 120 ± 10mmXS: 77-78mmS: 86-88 mmM: 96-98mmL: 108-110 mmXL: 116-117 mm
    ASTM D6319-19ThicknessFinger: 0.05mm (min)Palm: 0.05mm (min)≥0.05mm
    ASTM D6319-19, ASTM D412-16Tensile Strength, Before Aging14MPa, min≥14 MPa
    ASTM D6319-19, ASTM D412-16Tensile Strength, After Accelerated Aging14MPa, min≥14 MPa
    ASTM D6319-19, ASTM D412-16Ultimate Elongation, Before Aging500%, min≥500%
    ASTM D6319-19, ASTM D412-16Ultimate Elongation, After Accelerated Aging400%, min≥400%
    ASTM D5151-19Freedom from holesG-I, AQL 2.5Meet and above AQL2.5 requirements
    ASTM D6124-06 (2017)Powder-Content≤2 mg per glove≤2 mg, meet requirements
    ISO 10993-10:2010Skin Irritation StudyUnder the conditions of the study, not an irritantUnder the conditions of the study, not an irritant
    ISO 10993-10:2010Maximization Sensitization StudyUnder the conditions of the study, not a sensitizerUnder the conditions of the study, not a sensitizer
    ISO 10993-5:2009Cytotoxicity TestUnder the conditions of this study, the test article extract showed potential toxicityUnder the conditions of this study, the test article extract showed potential toxicity
    ISO 10993-11:2017Systemic toxicityUnder the conditions of this study, there was no evidence of systemic toxicity.Under the conditions of this study, there was no evidence of systemic toxicity.
    ASTM D6978-05 (2019)Tested for Use with Chemotherapy Drugs (Minimum BDT)No explicit numerical acceptance criteria for each drug are listed as "acceptance criteria" in the table. The table provides the reported values for the device. The "Acceptance Criteria" column for this test indicates "/", implying that the objective is to report the measured breakthrough times. Note: For Carmustine (BCNU) and Thio Tepa, the document explicitly states a warning due to "extremely low permeation times."Bleomycin (15.0mg/ml): >240Bortezomib (Velcade) 1mg/ml: >240Busulfan (6.0mg/ml): >240Carboplatin, (10.0mg/ml): >240Carmustine(BCNU) (3.3 mg/ml): 7.0 minutesChloroquine, (50.0mg/ml): >240Cisplatin (1mg/ml): >240Cyclophosphamide (Cytoxan) (20mg/ml): >240Cyclosporin, (100.0mg/ml): >240Cytarabine, (100.0mg/ml): >240Dacarbazine (10mg/ml): >240Daunorubicin, HCL (5.0mg/ml): >240Docetaxel, (10.0mg/ml): >240Doxorubicin HCL (2mg/ml): >240Epirubicin HCL (Ellence), (2.0mg/ml): >240Etoposide (20mg/ml): >240Fludarabine, (25.0mg/ml): >240Fluorouracil (50mg/ml): >240Gemcitabine (38.0mg/ml): >240Idarubicin HCL (1.0mg/ml): >240Ifosfamide, (50.0mg/ml): >240Irinotecan, (20.0mg/ml): >240Mechlorethamine HCI, (1.0mg/ml): >240Melphalan, (5.0mg/ml): >240Methotrexate (25mg/ml): >240Mitomycin C, (0.5mg/ml): >240Mitoxantrone, (2.0mg/ml): >240Oxaliplatin, (5.0mg/ml): >240Paclitaxel (Taxol) (6mg/ml): >240Paraplatin, (10.0mg/ml): >240Retrovir, (10.0mg/ml): >240Rituximab, (10.0mg/ml): >240Thio Tepa (10mg/ml): 23.0 minutesTopotecan HCL, (1.0mg/ml): >240Trisonex, (1.0mg/ml): >240Vincristine, (1.0mg/ml): >240

    Regarding the AI/SaMD specific questions, the following points are not applicable as this document describes performance testing for a physical medical glove.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • The document does not specify the exact sample sizes for each physical test (e.g., how many gloves were tested for tensile strength, or how many samples for biocompatibility). Standard test methods (ASTM, ISO) typically define minimum sample sizes. The data provenance is implied to be from testing labs in relation to the manufacturer (Shanxi Hongjin Plastic Technology Co., Ltd. in China) and their US representative. This is a prospective testing of the manufactured product.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. Ground truth for physical device performance is established by standardized laboratory testing methods (ASTM, ISO) rather than expert human interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. Adjudication methods are typically relevant for human-in-the-loop assessments or expert consensus in image interpretation, not for physical material testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI/SaMD.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is not an AI/SaMD. The "standalone" performance here refers to the physical properties of the glove itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The 'ground truth' for this device is based on objective measurements derived from standardized testing protocols (ASTM and ISO standards) measuring physical properties (like length, thickness, tensile strength, freedom from holes, powder content) and chemical resistance (breakthrough time for chemotherapy drugs), as well as biological responses (irritation, sensitization, cytotoxicity, systemic toxicity).

    8. The sample size for the training set

    • Not applicable. This is not an AI/SaMD, so there is no "training set."

    9. How the ground truth for the training set was established

    • Not applicable. As there is no "training set" for physical device testing, there is no ground truth establishment method for it.

    Summary of the Study Proving Acceptance Criteria:

    The study proving the device meets the acceptance criteria is a series of non-clinical performance tests conducted in accordance with recognized international (ISO) and national (ASTM) standards. These tests assess the physical, chemical, and biological properties of the nitrile examination gloves.

    The "Non-Clinical Performance Data" section (Page 9 of the document) explicitly lists the test methods used, their purpose, the acceptance criteria, and the results, confirming that the proposed device complies with these standards. For example, ASTM D6319-19 was used for dimensions and tensile properties, ASTM D5151-19 for freedom from holes, ASTM D6124-06 for powder content, and ASTM D6978-05 (Reapproved 2019) for chemotherapy drug permeation. Biocompatibility was assessed using ISO 10993 series standards (Parts 5, 10, and 11) for cytotoxicity, irritation, sensitization, and systemic toxicity.

    The conclusion states, "The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device." This indicates that the results met the established benchmarks (which are often derived from the requirements of the standards themselves or comparison to a predicate device).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1