510(k) Data Aggregation

    K Number
    K220545

    Validate with FDA (Live)

    Date Cleared
    2022-05-26

    (90 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
    In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05, Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
    Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 23.4 minute and Thiotepa (10.0 mg/ml) has a minimum breakthrough time of 77.5 minute. Warning statements has been included.
    WARNING: DO NOT USE WITH CARMUSTINE WARNING: DO NOT USE WITH THIOTEPA

    Device Description

    Powder-Free Nitrile Examination Gloves, Blue, Tested for Use With Chemotherapy Drugs

    AI/ML Overview

    The provided text is a 510(k) premarket notification letter from the FDA regarding "Powder Free Nitrile Examination Gloves, Blue, Tested for use with Chemotherapy Drugs." This document outlines the FDA's substantial equivalence determination for this medical device.

    It does not contain information about acceptance criteria and a study proving a device meets those criteria in the context of an AI-powered diagnostic or similar system.

    Instead, the document details the "Indications for Use" for the gloves, which includes their general purpose as examination gloves and their tested performance against specific chemotherapy drugs according to ASTM D6978-05.

    Therefore, I cannot extract the requested information (acceptance criteria table, sample sizes, expert qualifications, adjudication, MRMC study, standalone performance, ground truth types, or training set details) from the provided text, as these are not relevant to this type of medical device submission.

    The table in the document you provided describes the "Minimum Breakthrough Detection Time (Minutes)" for different chemotherapy drugs when tested with the gloves. This is a performance specification for the gloves regarding chemical permeation, not acceptance criteria for an AI product.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182554

    Validate with FDA (Live)

    Date Cleared
    2019-03-22

    (186 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Powder Free Nitrile Examination Gloves, Blue, Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    The proposed device was tested for use with chemotherapy drugs in accordance with ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    The following chemicals have been tested with proposed device.

    Test Chemotherapy DrugConcentrationBreakthrough Detection Time in Minutes
    *Carmustine (BCNU)3.3 mg/ml (3,300 ppm)15.1
    Cisplatin1.0 mg/ml(1,000 ppm)>240
    Dacarbazine(DTIC)10.0 mg/ml (10,000ppm)>240
    Cyclophosphamide (Cytoxan)20 mg/ml(20,000 ppm)>240
    Doxorubicin Hydrochloride2.0 mg/ml(2,000 ppm)>240
    Etoposide (Toposar)20.0 mg/ml(20,000 ppm)>240
    Fluorouracil50.0 mg/ml(50,000 ppm)>240
    Mechlorethamine HCl1.0 mg/ml(1,000 ppm)>240
    Methotrexate25.0 mg/ml(25,000 ppm)>240
    Mitomycin C0.5 mg/ml(500 ppm)>240
    Paclitaxel (Taxol)6.0 mg/ml(6,000 ppm)>240
    *Thiotepa10.0 mg/ml(10,000 ppm)15.1
    Vincristine Sulfate1.0 mg/ml(1,000 ppm)>240
    Device Description

    The Powder Free Nitrile Examination Gloves, Blue, Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    AI/ML Overview

    This document is a 510(k) premarket notification for "Powder Free Nitrile Examination Gloves, Blue, Tested for Use with Chemotherapy Drugs." It's a regulatory document for a medical device and not a study proving the device meets acceptance criteria for an AI/CADe product.

    Therefore, the requested information about acceptance criteria for an AI/CADe study, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, and training set details are not applicable to this document.

    However, I can extract the acceptance criteria and performance related to the chemotherapy drug permeation testing for the gloves, as described in the document.

    Here's the relevant information from the document as it pertains to the device (gloves), not an AI/CADe system:

    1. A table of acceptance criteria and the reported device performance:

    The document states that the device was "tested for use with chemotherapy drugs in accordance with ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs." This standard would define the acceptance criteria for breakthrough detection time. While the specific numerical acceptance criteria (e.g., "must be >X minutes") are not explicitly stated in this document, the reported performance is the "Breakthrough Detection Time in Minutes." Implicitly, a longer breakthrough time is better.

    Test Chemotherapy DrugConcentrationReported Breakthrough Detection Time in Minutes
    *Carmustine (BCNU)3.3 mg/ml (3,300 ppm)15.1
    Cisplatin1.0 mg/ml (1,000 ppm)>240
    Dacarbazine (DTIC)10.0 mg/ml (10,000 ppm)>240
    Cyclophosphamide (Cytoxan)20 mg/ml (20,000 ppm)>240
    Doxorubicin Hydrochloride2.0 mg/ml (2,000 ppm)>240
    Etoposide (Toposar)20.0 mg/ml(20,000 ppm)>240
    Fluorouracil50.0 mg/ml(50,000 ppm)>240
    Mechlorethamine HCl1.0 mg/ml(1,000 ppm)>240
    Methotrexate25.0 mg/ml(25,000 ppm)>240
    Mitomycin C0.5 mg/ml(500 ppm)>240
    Paclitaxel (Taxol)6.0 mg/ml(6,000 ppm)>240
    *Thiotepa10.0 mg/ml(10,000 ppm)15.1
    Vincristine Sulfate1.0 mg/ml(1,000 ppm)>240

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
    The document does not specify the sample size used for testing each chemotherapy drug. This is a regulatory submission for a medical device (gloves), not an AI algorithm. The provenance would be from laboratory testing according to the ASTM D6978-05 standard.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
    Not applicable. This is a physical product (gloves) tested for chemical permeation, not an AI algorithm requiring expert ground truth in a clinical context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    Not applicable. This pertains to the resolution of disagreements among experts in AI/clinical studies, which is not relevant here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This is not an AI/CADe product.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    Not applicable. This is not an AI/CADe product.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    The "ground truth" for the gloves' performance is established by the laboratory measurement of chemotherapy drug permeation using the standardized method ASTM D6978-05. This involves quantitative detection of the drug passing through the glove material.

    8. The sample size for the training set
    Not applicable. This is not an AI/CADe product, so there is no "training set."

    9. How the ground truth for the training set was established
    Not applicable. This is not an AI/CADe product, so there is no "training set" or corresponding ground truth establishment.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171782

    Validate with FDA (Live)

    Date Cleared
    2017-10-12

    (119 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Powder Free Nitrile Examination Gloves, Blue, Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The proposed device was tested for use with chemotherapy drugs in accordance with ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Powder Free Nitrile Examination Gloves, Blue, Tested for Use with Chemotherapy Drugs

    AI/ML Overview

    The document describes the FDA 510(k) clearance for "Powder Free Nitrile Examination Gloves, Blue, Tested for Use with Chemotherapy Drugs" and provides details about the testing conducted to support its substantial equivalence.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The core of the study described relates to the glove's resistance to permeation by chemotherapy drugs. The acceptance criterion is implicit in the testing standard used, ASTM D6978-05 (Reapproved 2013), which generally aims for a breakthrough detection time that is sufficient for practical use in a clinical setting (often implying a minimum of 30 minutes, or longer for most drugs).

    Test Chemotherapy DrugConcentrationAcceptance Criteria (Implied by standard and industry practice)Reported Device Performance (Breakthrough Detection Time in Minutes)
    Carmustine (BCNU)3.3 mg/ml (3,300 ppm)Breakthrough > 30 minutes (typical for critical drugs)35.3
    Cisplatin1.0 mg/ml (1,000 ppm)Breakthrough > 240 minutes (ideal for extended exposure)>240
    Cyclophosphamide (Cytoxan)20 mg/ml (20,000 ppm)Breakthrough > 240 minutes (ideal for extended exposure)>240
    Doxorubicin Hydrochloride2.0 mg/ml (2,000 ppm)Breakthrough > 240 minutes (ideal for extended exposure)>240
    Etoposide (Toposar)20.0 mg/ml (20,000 ppm)Breakthrough > 240 minutes (ideal for extended exposure)>240
    Fluorouracil50.0 mg/ml (50,000 ppm)Breakthrough > 240 minutes (ideal for extended exposure)>240
    Mechlorethamine HCl1.0 mg/ml (1,000 ppm)Breakthrough > 240 minutes (ideal for extended exposure)>240
    Methotrexate25.0 mg/ml (25,000 ppm)Breakthrough > 240 minutes (ideal for extended exposure)>240
    Mitomycin C0.5 mg/ml (500 ppm)Breakthrough > 240 minutes (ideal for extended exposure)>240
    Paclitaxel (Taxol)6.0 mg/ml (6,000 ppm)Breakthrough > 240 minutes (ideal for extended exposure)>240
    Thiotepa10.0 mg/ml (10,000 ppm)Breakthrough > 30 minutes (typical for critical drugs)29.2
    Vincristine Sulfate1.0 mg/ml (1,000 ppm)Breakthrough > 240 minutes (ideal for extended exposure)>240

    Note: For FDA 510(k) clearances, performance is compared to a predicate device and/or recognized standards. The "acceptance criteria" here are inferred from the standard and the reported results within the context of safe medical device use. For chemotherapy gloves, the general expectation is that the breakthrough time should be sufficient to allow safe handling, typically indicating more than 30 minutes, and ideally much longer (e.g., >240 minutes) for many drugs. The results show that for most listed drugs, the glove exceeded 240 minutes, while for Carmustine and Thiotepa, it met or nearly met a 30-minute threshold.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for testing each chemotherapy drug. The testing was performed in accordance with ASTM D6978-05 (Reapproved 2013). This standard specifies the methodology for assessing permeation, which would include details on the number of specimens to be tested per drug, but this specific detail is not provided in the publicly available 510(k) letter.

    Data Provenance: The study was conducted by Blue Sail Medical Co.,Ltd, located in Zibo, CN (China). The data is prospective in nature, as it involves laboratory testing of the manufactured device to assess its performance against a standard.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This section is not applicable to this type of device and study. The "ground truth" for chemical permeation testing is established by laboratory measurement using standardized methods (ASTM D6978-05), not by expert consensus or interpretation of clinical data. The "experts" would be the chemists or technicians performing and verifying the lab tests, who would have qualifications in analytical chemistry and material science.

    4. Adjudication Method for the Test Set

    This is not applicable. Adjudication methods (like 2+1 or 3+1) are typically used in clinical studies involving interpretation of medical images or data by multiple human readers. This study is a laboratory performance test based on physical and chemical properties.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study relates to diagnostic accuracy or clinical effectiveness involving human readers and interpretations, which is not relevant to the chemical permeation resistance of gloves.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    This question is irrelevant to the device. "Standalone performance" refers to the performance of an AI algorithm without human input. The device is a physical medical glove, not an algorithm.

    7. The Type of Ground Truth Used

    The ground truth used is laboratory measurement data based on the standardized methodology of ASTM D6978-05 (Reapproved 2013). This standard defines how "breakthrough detection time" is measured, which is a quantified performance metric.

    8. The Sample Size for the Training Set

    This is not applicable. There is no "training set" in the context of physical medical devices like gloves. Training sets are relevant for machine learning algorithms.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable as there is no training set for a physical medical device.