    Intended Use

    A Patient examination glove is a disposable device intended for medical purposes that is worn on the examiners hand or finger to prevent contamination between patient and examiner.

    Device Description

    Powder Free Blue Nitrile Examination Glove

    AI/ML Overview

    This document is an FDA 510(k) clearance letter for a medical device: "Powder Free Blue Nitrile Examination Glove". It primarily details the regulatory clearance for marketing this device, stating that it is substantially equivalent to legally marketed predicate devices.

    Crucially, this document does NOT contain information about acceptance criteria for an AI/ML powered medical device, nor does it describe a study proving such a device meets acceptance criteria.

    The context of the original request (asking for information about "acceptance criteria and the study that proves the device meets the acceptance criteria" for an AI/ML device) is not addressed by the provided text. The document is strictly a regulatory clearance for a physical medical glove.

    Therefore, I cannot extract the requested information from the provided text. The questions posed in the prompt relate to the performance validation of an AI/ML algorithm, which is entirely absent from this 510(k) clearance for a nitrile glove.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213678

    Validate with FDA (Live)

    Date Cleared
    2022-03-22

    (120 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Powder Free Blue Nitrile Examination Gloves are disposable device intended for medical purposes that are worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    The nitrile rubber is watertight under normal conditions of use. Its tensile properties cause it to conform to the hand, allowing movements necessary for the medical properties. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D6319-19 and ASTM D5151-19. This is a disposable device, intended for medical purposes, that is worn on the examiner's hand to prevent contamination between patient and examiner.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for "Powder Free Blue Nitrile Examination Gloves." This document focuses on demonstrating substantial equivalence to a predicate device, rather than providing a detailed study proving the device meets specific acceptance criteria for an AI/ML-driven medical device. Therefore, much of the requested information regarding AI/ML device performance and testing (e.g., sample sizes for test/training sets, expert ground truth establishment, MRMC studies, standalone performance, effect size of human reader improvement with AI assistance) is not applicable to this document.

    However, the document does contain information about performance testing for the physical and biological properties of the gloves, which can be interpreted as "acceptance criteria" and "reported device performance."

    Here's the information extracted from the document that is relevant to "acceptance criteria" and "study that proves the device meets the acceptance criteria" for these physical and biological properties:

    1. A table of acceptance criteria and the reported device performance:

    TestAcceptance CriteriaReported Device Performance
    Biocompatibility Testing:
    Guinea Pig SensitizationNo deviations noted, observed, nor requiring clarification. Product is not a sensitizer.No reaction/positive allergic reaction observed. Conforms to requirements.
    Primary Dermal Irritation in RabbitsNo deviations noted, observed, nor requiring clarification. Product is not a primary irritant. Primary Irritation Index (PII) of 0.PII was 0. Conforms to requirements.
    Acute Systemic Cytotoxicity in RatsNo adverse reaction noted, observed, nor requiring clarification. No adverse biological reaction.All animals survived, healthy, active, gained weight. No abnormalities in gross necropsies. Conforms to requirements.
    Physical Properties Testing (ASTM D6319-19, D5151-19, D6124-06):
    Watertight (Pinhole Test) per ASTM D5151-19Free from holes, AQL 1.5 (G1: AQL 2.5 for 1000mL test)Pass (G1: AQL 2.5 for 1000mL test)
    Length (mm)Size S: Min 220; M, L, XL: Min 230240mm minimum for all sizes
    Palm Width (mm)Size S: 80 ± 10; M: 95 ± 10; L: 110 ± 10; XL: 120 ± 10S: 84-86; M: 95-96; L: 108-109; XL: 114-115
    Thickness (mm) - Single LayerFinger: Min 0.05; Palm: Min 0.05; Cuff: Min 0.05Finger: Min 0.06; Palm: Min 0.06; Cuff: Min 0.05
    Tensile Strength (MPa) - Before AgingMin 1417-23
    Ultimate Elongation (%) - Before AgingMin 500540-607
    Tensile Strength (MPa) - After AgingMin 1421-27
    Ultimate Elongation (%) - After AgingMin 400480-565
    Powder Content per ASTM D6124-06Max 2.0 mg/gloveBelow 2.0 mg/glove

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not explicitly state the sample sizes used for each specific test (e.g., number of gloves for watertightness, physical properties, or number of animals for biocompatibility). It references ASTM and ISO standards, which would define the required sample sizes for such tests.
    • Data Provenance: Not explicitly stated, but the tests were performed on the "proposed device" (Powder Free Blue Nitrile Examination Gloves manufactured by Lienteh Technology Sdn Bhd in Malaysia). The tests are described as non-clinical "performance data of gloves based on animal studies, biocompatibility studies" and "performance data of gloves based on ASTM D6319-19 and FDA Watertight Test." This implies prospective laboratory testing of the manufactured product.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. This is not an AI/ML device relying on human expert annotation for ground truth. The "ground truth" for the physical and biological properties is established by the specified ASTM/ISO standard testing methodologies and their defined criteria.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not Applicable. This document does not describe a study requiring human adjudication for test results; the outcomes are based on standardized laboratory measurements and observations according to specific protocols.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This device is an examination glove, not an AI/ML diagnostic aid influencing human reader performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is not an AI algorithm. The performance described is the standalone performance of the physical glove itself.

    7. The type of ground truth used:

    • The "ground truth" is based on standardized testing protocols and criteria defined by:
      • ASTM D6319-19 (Standard Specification for Nitrile Examination Gloves for Medical Application)
      • ASTM D5151-19 (Standard Test Method for Detection of Holes in Medical Gloves)
      • ASTM D6124-06 (Reapproved 2017), (Standard test method for Residual powder on medical gloves)
      • ASTM D412-16 (for physical properties like tensile strength and elongation)
      • ISO 10993-10:2010 (Biological evaluation of medical devices- Part 10: Tests for Irritation and Skin Sensitization)
      • ISO 10993-11 (Biological Evaluation of Medical devices- Part 11: Tests for Systemic toxicity)
      • FDA watertight test AQL criteria (21 CFR 800.20)

    8. The sample size for the training set:

    • Not Applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established:

    • Not Applicable. As above, it's not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K213107

    Validate with FDA (Live)

    Date Cleared
    2022-02-17

    (146 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    0 - 21
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner

    Device Description

    The nitrile rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical properties. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D6319 and ASTM D5151. This is a disposable device, intended for medical purposes, that is worn on the examiner's hand to prevent contamination between patient and examiner.

    AI/ML Overview

    The provided document is a 510(k) summary for "Powder Free Blue Nitrile Examination Gloves" (K213107) and does not describe an AI medical device. Therefore, many of the requested fields are not applicable. This summary focuses on the safety and performance of physical medical gloves, primarily through non-clinical testing against established standards.

    Here's an analysis based on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test Method / PropertyAcceptance CriteriaReported Device Performance
    ASTM D 5151 (Pinhole)Free from holes, AQL 1.5Pass
    ASTM D6319 (Physical)Before Aging: Tensile Strength: Min 14 MPa, Elongation: Min 500%Pass
    After Aging: Tensile Strength: Min 14 MPa, Elongation: Min 400%Pass
    Watertight (1000mL)Single Sampling in accordance with ISO 2859 G1: AQL 2.5Pass G1: AQL 2.5
    Length (mm)Min 220 (S), Min 230 (M, L, XL)240mm minimum for all sizes
    Palm width (mm)80 ± 10 (S), 95 ± 10 (M), 110 ± 10 (L), 120 ± 10 (XL)84-86 (S), 95-96 (M), 108-109 (L), 114-115 (XL)
    Thickness (mm)Min 0.05 (Finger, Palm, Cuff)Min 0.06 (Finger), Min 0.06 (Palm), Min 0.05 (Cuff)
    ASTM D412 (Physical)Before Aging: Tensile Strength: Min 14 MPa, Elongation: Min 500%17-23 (Tensile Strength), 540-607 (Elongation)
    After Aging: Tensile Strength: Min 14 MPa, Elongation: Min 400%21-27 (Tensile Strength), 480-565 (Elongation)
    ASTM D6124 (Powder)Max 2.0 mg/gloveBelow 2.0 mg/glove
    Guinea Pig SensitizationNo deviations noted, no positive allergic reactionNo reaction observed, no positive allergic reaction. Meets conformance requirements.
    Primary Dermal Irritation (Rabbits)No deviations noted, identify primary irritantsPrimary Irritation Index was 0. Meets conformance requirements.
    Acute Systemic Cytotoxicity (Rats)No adverse reaction notedAll animals survived, gained weight, no abnormalities in necropsies. Meets conformance requirements.

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state the exact sample sizes for each specific test (e.g., number of gloves tested for pinholes, physical properties, or biocompatibility). However, it refers to standards such as "Single Sampling in accordance with ISO 2859 G1" for the watertight test, which implies a statistically determined sample size based on the Acceptable Quality Level (AQL). The tests appear to be performed on the final and finished devices.

    • Data Provenance: Not explicitly stated, but given the applicant's location (Novid PPE Sdn Bhd, Malaysia), the testing was likely conducted in Malaysia or a lab contracted by them. The tests are non-clinical, meaning they involve physical product testing and animal studies, not human patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a non-AI medical device (examination gloves). Ground truth for these tests is established by adherence to recognized international and national standards (e.g., ASTM, ISO) and laboratory protocols, not by expert consensus on clinical data.

    4. Adjudication method for the test set

    Not applicable. The tests are objective measurements against predefined criteria in established standards, not subjective assessments requiring expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI medical device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI medical device.

    7. The type of ground truth used

    For the performance and physical properties, the "ground truth" is defined by the specified minimum or maximum values and methodologies outlined in the referenced ASTM and ISO standards (e.g., ASTM D6319 for physical properties, ASTM D5151 for pinhole detection, ASTM D6124 for powder content). For biocompatibility, the ground truth is established by the absence of adverse biological reactions using standard animal testing protocols (ISO 10993-10, ISO 10993-11).

    8. The sample size for the training set

    Not applicable. This is not an AI medical device, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210375

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-06-21

    (133 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Powder Free Blue Nitrile Examination Glove meets all the current specifications listed under the ASTM Specification D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application. The principle operation and mechanism of this device is to prevent contamination between patient and examiner and this principle is achieved through testing of barrier, physical properties and other testing stated in the performance data. This device is for over-the counter single use.

    AI/ML Overview

    This response is tailored to a medical device's performance evaluation, specifically a Nitrile Examination Glove, based on the provided FDA 510(k) summary. The document describes the testing performed to demonstrate substantial equivalence to a predicate device, rather than a novel AI/software medical device. Therefore, aspects related to AI (e.g., human-in-the-loop, training/test sets, expert adjudication, MRMC studies) are not applicable and are addressed as such.

    Here's the breakdown of the acceptance criteria and study proving device performance:

    Device Name: Powder Free Blue Nitrile Examination Glove (K210375)
    Predicate Device: Powder Free Blue Nitrile Examination Glove (K153562)


    1. Table of Acceptance Criteria and Reported Device Performance

    The device performance is evaluated against established ASTM and ISO standards for examination gloves.

    Test Method (Reference Standard)PurposeAcceptance CriteriaReported Device Performance (Results)
    ASTM D6124To check the amount of powder residue from glove surfaceLess than 2mg/glovePass (Average powder residue for each size: S: 0.61 mg/glove, M: 0.59 mg/glove, L: 0.57 mg/glove)
    ASTM D5151To detect the presence of hole in gloveMeet AQL 2.5 from sampling of the lotPass
    ASTM D6319To measure the following physical parameters of the glove: a) Length b) Palm Thickness c) Tensile Strength before age d) Tensile Strength after age e) Elongation before age f) Elongation after agea) Length at minimum 230mm b) Palm Thickness at minimum 0.05mm c) Tensile strength before age at minimum 14MPa d) Tensile Strength after age at minimum 14MPa e) Elongation before age at minimum 500% f) Elongation after age at minimum 400%Pass (Reported averages meet or exceed criteria for all sizes)
    ISO 10993-10To determine the irritation potential of glove when exposed to skin surface (Primary Skin Irritation Test) and to evaluate potential for Type IV allergy (Skin Sensitization Test)a) Not an Irritant b) Not a SensitizerPass (Not an irritant, Not a sensitizer)
    ISO 10993-11To evaluate the human hazard potential from glove (Acute Systemic Toxicity)Not induce systemic toxicityPass (Did not induce systemic toxicity)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a distinct "test set" in the context of an AI/ML device validating against a separate dataset. Instead, it refers to non-clinical tests performed on the "proposed device" (the Nitrile Examination Glove).

    • Sample Size: The document does not explicitly state the sample sizes for each test (e.g., number of gloves tested for holes, tensile strength, etc.). However, it refers to standard practices like "sampling of the lot" for ASTM D5151 (Holes), and implies multiple samples for calculating averages for physical properties (D6319) and powder residue (D6124) across different glove sizes (S, M, L). For biocompatibility tests (ISO 10993), specific animal or in-vitro sample sizes are common for such assessments but are not detailed here. ASTM and ISO standards typically outline appropriate sampling plans.
    • Data Provenance: The data is generated from performance testing of the "Powder Free Blue Nitrile Examination Glove" from Duramitt Sdn Bhd, located in Kulim, Kedah, Malaysia. These are prospective test results generated for the purpose of this 510(k) submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This section is not applicable as the described device is a physical medical glove, not an AI/software device that requires human expert annotation for establishing ground truth on a dataset. The "ground truth" for glove performance is established by standardized physical, chemical, and biological testing methods (e.g., measuring dimensions, tensile strength, conducting irritation tests in a lab setting) as per ASTM and ISO standards.


    4. Adjudication Method for the Test Set

    This section is not applicable. Since the device is a physical product and the "ground truth" is determined by objective, standardized laboratory tests, there is no need for expert adjudication of results as would be the case for subjective interpretations in an expert-based AI/ML ground truth dataset. Test results are quantified and compared to predefined criteria.


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    This section is not applicable. An MRMC study is relevant for evaluating the impact of an AI system on human reader performance for tasks involving interpretation (e.g., radiology image reading). This submission pertains to a physical medical glove, which does not involve human "readers" or AI assistance in its primary function or evaluation.


    6. Standalone Performance (Algorithm Only Without Human-in-the-Loop)

    This section is not applicable. This concept applies to AI algorithms. The performance of the medical glove is evaluated as a standalone product based on its physical properties and biocompatibility as per the listed standards. There is no "algorithm" to evaluate.


    7. Type of Ground Truth Used

    The "ground truth" for the device's performance is established by compliance with recognized international consensus standards (ASTM, ISO). These standards define objective, measurable criteria for physical properties (e.g., length, thickness, tensile strength, elongation), barrier integrity (freedom from holes), and biological safety (biocompatibility).

    • Physical Properties: Measured directly (e.g., length, thickness) or through standardized mechanical testing (tensile strength, elongation).
    • Barrier Integrity: Determined by a standardized water leak test (ASTM D5151).
    • Biocompatibility: Assessed through standardized biological evaluation methods (ISO 10993-10 for irritation/sensitization, ISO 10993-11 for systemic toxicity), typically involving in-vitro or animal models as per the standard's guidelines.

    8. Sample Size for the Training Set

    This section is not applicable. There is no "training set" as this is not an AI/ML device.


    9. How the Ground Truth for the Training Set Was Established

    This section is not applicable. Since there is no training set (as this is not an AI/ML device), no ground truth establishment for such a set was performed.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203512

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-04-26

    (147 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    POWDER FREE BLUE NITRILE EXAMINATION GLOVE

    AI/ML Overview

    I am sorry, but based on the provided text, there is no information about a study proving that a device meets acceptance criteria, nor is there any mention of a device that requires such a study.

    The document is an FDA 510(k) clearance letter for a "Powder Free Blue Nitrile Examination Glove." This type of device (a Class I medical device) typically does not require extensive clinical studies or AI algorithm performance evaluations to demonstrate substantial equivalence. The clearance is based on comparison to a predicate device and adherence to general controls, good manufacturing practices, and labeling requirements.

    Therefore, I cannot provide the requested information, which includes:

    1. A table of acceptance criteria and reported device performance (not applicable for this type of device and clearance).
    2. Sample size and data provenance for a test set (no such test set or study described).
    3. Number of experts and their qualifications (not relevant for this clearance).
    4. Adjudication method (not applicable).
    5. MRMC comparative effectiveness study (not applicable).
    6. Standalone algorithm performance (not applicable, as it's a physical glove, not an algorithm).
    7. Type of ground truth used (not applicable).
    8. Sample size for the training set (not applicable).
    9. How ground truth for the training set was established (not applicable).

    The document is purely a regulatory letter confirming that the device is substantially equivalent to legally marketed predicate devices, allowing it to be marketed.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182089

    Validate with FDA (Live)

    Date Cleared
    2019-07-30

    (362 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A Patient examination glove is a disposable device intended for medical purposes that worn on the examiner's hand or finger to prevent contamination between patient and examiner. Tested for use with chemotherapy drugs as listed below.

    Device Description

    Powder Free Blue Nitrite Examination Gloves- Tested for Use with Chemotherapy Drugs.

    AI/ML Overview

    This document describes the FDA's 510(k) clearance for "Powder Free Blue Nitrile Examination Gloves - Tested for Use with Chemotherapy Drugs" (K182089). The primary focus of the performance data in this document relates to the gloves' resistance to permeation by chemotherapy drugs.

    Here's an analysis based on your request, using the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The "acceptance criteria" for gloves used with chemotherapy drugs are typically defined by regulatory bodies or consensus standards (like ASTM D6978, which is often referenced for chemotherapy glove testing) as a minimum breakthrough time. While explicit "acceptance criteria" values are not stated in this document as a separate column, the reported device performance clearly indicates whether the gloves meet the implied standard of extended protection (often >240 minutes for many drugs).

    Chemotherapy DrugConcentrationAcceptance Criteria (Implied by standard practice, e.g., ASTM D6978)Reported Device Performance (Breakthrough Detection Time)Meets Implied Criteria?
    Carmustine (BCNU)3.3 mg/mlTypically >240 min desired9.0 minNo (specifically cautioned against)
    Cisplatin1.0 mg/mlTypically >240 min desired>240 minYes
    Cyclophosphamide (Cytoxan)20mg/mlTypically >240 min desired>240 minYes
    Dacarbazine (DTIC)10.0 mg/mlTypically >240 min desired>240 minYes
    Doxorubicin Hydrochloride2.0mg/mlTypically >240 min desired>240 minYes
    Etoposide (Toposar)20.0 mg/mlTypically >240 min desired>240 minYes
    Fluorouracil50.0 mg/mlTypically >240 min desired>240 minYes
    Ifosfamide50.0 mg/mlTypically >240 min desired>240 minYes
    Methotrexate25 mg/mlTypically >240 min desired>240 minYes
    Mitomycin C0.5 mg/mlTypically >240 min desired>240 minYes
    Mitoxantrone2.0 mg/mlTypically >240 min desired>240 minYes
    Paclitaxel (Taxol)6.0 mg/mlTypically >240 min desired>240 minYes
    Thio-Tepa10.0 mg/mlTypically >240 min desired16.2 minNo (specifically cautioned against)
    Vincristine Sulfate1.0 mg/mlTypically >240 min desired>240 minYes

    Note: The document explicitly states: "Please Note that Carmustine and Thio-Tepa have low permeation time of less than 240 minutes. Do not use with Carmustine and Thio-Tepa." This indicates that for these two drugs, the gloves did not meet the standard expectation for chemotherapy glove use, and this limitation is clearly communicated.


    The subsequent questions (2-9) are primarily relevant to clinical studies involving AI or diagnostics that interpret medical data. This document describes the clearance of a medical device (gloves) based on laboratory testing for chemical permeation, not a clinical study involving human patients, expert readers, or AI algorithms. Therefore, most of these questions are not applicable to the provided information.

    However, I will address those that can be inferred or directly commented on:

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: Not explicitly stated in terms of number of gloves tested, but it would involve multiple samples of the glove material subjected to permeation testing according to a standard like ASTM D6978. For each chemotherapy drug listed, a sample of the glove material would be exposed, and the time until breakthrough is detected would be measured.
    • Data Provenance: The document implies laboratory testing conducted by the manufacturer or a certified lab. The country of origin of the data is not specified beyond the manufacturer being "Platinum Glove Industries Sdn Bhd" from "Malaysia." This is retrospective testing of the glove material properties.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. This is not a study involving human interpretation of medical data where "ground truth" is established by experts. The "ground truth" in this context is the physical property of the glove's resistance to chemical permeation, measured by laboratory instrumentation and protocols (e.g., using a permeation cell and analytical detection methods).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable. See point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is not a clinical or AI-based diagnostic study.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not Applicable. This is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • The "ground truth" in this context is the measured chemical permeation time of specific chemotherapy drugs through the glove material, determined through standardized laboratory testing (e.g., in accordance with ASTM D6978 or an equivalent standard).

    8. The sample size for the training set:

    • Not Applicable. This is not an AI or machine learning study, so there is no "training set."

    9. How the ground truth for the training set was established:

    • Not Applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K183354

    Validate with FDA (Live)

    Date Cleared
    2019-06-20

    (199 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A Patient examination glove is a disposable device intended for medical purposes that worn on the examiner's hand or finger to prevent contamination between patient and examiner. Tested for use with chemotherapy drugs as listed below.

    Device Description

    Powder Free Blue Nitrile Examination Gloves - Tested for Use with 32 Chemotherapy Drugs and Fentanyl Citrate

    AI/ML Overview

    The provided document is an FDA 510(k) clearance letter for "Powder Free Blue Nitrile Examination Gloves - Tested for use with 32 Chemotherapy Drugs and Fentanyl Citrate." This document primarily focuses on the substantial equivalence determination for a medical device (gloves) and its performance against chemical permeation for certain drugs.

    The request asks for information typically found in the clinical validation of AI/ML-based medical devices or diagnostic tools. The provided document does not contain any information related to AI/ML device performance, test sets, ground truth establishment, expert adjudication, or MRMC studies. It is a regulatory clearance for a physical medical device (gloves) based on chemical permeation testing, not an AI system.

    Therefore, I cannot provide the requested information. The document details chemical resistance and breakthrough times, not AI algorithm performance.

    However, if we were to interpret the document's content as "acceptance criteria" and "device performance" in the context of the gloves' chemical resistance, the information could be presented as follows:


    Acceptance Criteria and Device Performance (Based on Chemical Permeation Testing for Gloves)

    Given the nature of the device (examination gloves) and the provided data, the "Acceptance Criteria" are implicitly derived from the testing standards for medical gloves' resistance to chemotherapy drugs and fentanyl citrate. The "Device Performance" refers to the measured breakthrough times. For regulated medical gloves, the expectation is generally to demonstrate sufficient resistance (i.e., a "breakthrough detection time" that meets or exceeds a certain threshold, often >240 minutes, indicating prolonged protection).

    1. Table of Acceptance Criteria and Reported Device Performance

    Chemotherapy Drugs:Acceptance Criteria (e.g., Target Breakthrough Time)Reported Device Performance (Breakthrough Detection Time in Minutes)
    1. Carmustine (BCNU)3.3 mg/ml(3,300 ppm)To be defined by standard (e.g., >x min)54.8 min
    2. Cisplatin 1.0 mg/ml (1,000ppm)To be defined by standard (e.g., >x min)>240 min
    3. Cyclophosphamide (Cytoxan) 20mg/ml(20,000ppm)To be defined by standard (e.g., >x min)>240 min
    4. Dacarbazine (DTIC)10.0 mg/ml(10,000ppm)To be defined by standard (e.g., >x min)>240 min
    5. Doxorubicin Hydrochloride 2.0 mg/ml (2,000ppm)To be defined by standard (e.g., >x min)>240 min
    6. Etoposide (Toposar) 20.0 mg/ml (20,000ppm)To be defined by standard (e.g., >x min)>240 min
    7. Fluorouracil 50.0 mg/ml (50,000ppm)To be defined by standard (e.g., >x min)>240 min
    8. Ifosfamide 50.0 mg/ml (50,000ppm)To be defined by standard (e.g., >x min)>240 min
    9. Methotrexate 25 mg/ml (25,000ppm)To be defined by standard (e.g., >x min)>240 min
    10.Mitomycin C 0.5 mg/ml (500ppm)To be defined by standard (e.g., >x min)>240 min
    11. Mitoxantrone 2.0 mg/ml (2,000ppm)To be defined by standard (e.g., >x min)>240 min
    12. Paclitaxel (Taxol) 6.0 mg/ml (6,000ppm)To be defined by standard (e.g., >x min)>240 min
    13. Thio-Tepa 10.0 mg/ml (10,000ppm)To be defined by standard (e.g., >x min)46.9 min
    14. Vincristine Sulfate 1.0 mg/ml (1,000ppm)To be defined by standard (e.g., >x min)>240 min
    15. Amethopterin (Methotrexate) 25 mg/ml, (25,000 ppm)To be defined by standard (e.g., >x min)>240 min
    16. Bleomycin 15 mg/ml (15,000 ppm)To be defined by standard (e.g., >x min)>240 min
    17. Busulfan 6 mg/ml (6,000 ppm)To be defined by standard (e.g., >x min)>240 min
    18. Carboplatin 10 mg/ml (10,000 ppm)To be defined by standard (e.g., >x min)>240 min
    19. Cytarabine 100 mg/ml (100.000 ppm)To be defined by standard (e.g., >x min)>240 min
    20. Daunorubicin 5 mg/ml (5,000 ppm)To be defined by standard (e.g., >x min)>240 min
    21. Doxetacel 10.0 mg/ml (10,000 ppm)To be defined by standard (e.g., >x min)>240 min
    22. Ellence 2 mg/ml (2,000 ppm)To be defined by standard (e.g., >x min)>240 min
    23. Fludarabine 25 mg/ml (25,000 ppm)To be defined by standard (e.g., >x min)>240 min
    24. Gemcitabine (Gemzar) 38 mg/ml (38,000 ppm)To be defined by standard (e.g., >x min)>240 min
    25. Idarubicin 1 mg/ml (1,000 ppm)To be defined by standard (e.g., >x min)>240 min
    26. Irinotecan 20.0 mg/ml (20,000 ppm)To be defined by standard (e.g., >x min)>240 min
    27. Mechloretamine HCL 1.0 mg/ml (1,000 ppm)To be defined by standard (e.g., >x min)>240 min
    28. Melphalan 5 mg/ml (5,000 ppm)To be defined by standard (e.g., >x min)>240 min
    29. Oxaliplatin 2.0 mg/ml (2,000 ppm)To be defined by standard (e.g., >x min)>240 min
    30. Rituximab 10 mg/ml (10,000 ppm)To be defined by standard (e.g., >x min)>240 min
    31. Trisonex 0.1 mg/ml (100 ppm)To be defined by standard (e.g., >x min)>240 min
    32. Vinorelbin 10 mg/ml (10,000 ppm)To be defined by standard (e.g., >x min)>240 min
    Fentanyl Citrate Injection 100 mcg/2mlTo be defined by standard (e.g., >x min)>240min

    Note: The document highlights "Please Note that the following Drugs have extremely low permeation times : Carmustine : 54,8 minutes and Thio-Tepa: 46.9 minutes," indicating these may be areas where performance is lower relative to other tested drugs, but still acceptable under relevant standards for their intended use and risk mitigation strategies.


    For the remainder of the questions, as stated, the provided document does not contain this information as it pertains to AI/ML device validation, not the direct chemical permeation testing of physical gloves.

    • 2. Sample size used for the test set and the data provenance: Not applicable. This refers to the number of gloves tested for chemical resistance, not a dataset for an AI model. The document does not specify the exact number of gloves or samples, nor their country of origin for testing purposes. The testing is laboratory-based chemical permeation.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for chemical permeation is established by analytical chemistry methods (e.g., spectroscopy) that detect the breakthrough of a chemical through the glove material over time, not by human expert opinion.
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. Chemical permeation testing uses objective measurement, not human adjudication.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This applies to diagnostic AI, not chemical resistance of gloves.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This applies to diagnostic AI, not chemical resistance of gloves.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): The ground truth for glove performance is the precise measurement of chemical permeation/breakthrough time under controlled laboratory conditions, typically following standards like ASTM D6978.
    • 8. The sample size for the training set: Not applicable. This is not an AI/ML device.
    • 9. How the ground truth for the training set was established: Not applicable. This is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K181130

    Validate with FDA (Live)

    Date Cleared
    2018-08-10

    (102 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Powder Free Blue Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    The proposed device was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    The proposed device, Powder Free Blue Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    The proposed device is a Powder Free Nitrile Patient Examination Glove that is available in multiple sizes

    The proposed device is provided non-sterile. The proposed device is made of Nitrile. The proposed device acts as a barrier.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Powder Free Blue Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs, based on the provided FDA 510(k) submission.

    1. Table of Acceptance Criteria and Reported Device Performance

    The device is evaluated against specific ASTM standards for various properties and against a predicate device (K163146) for comparison. The key acceptance criteria and reported performance for critical aspects are as follows:

    Acceptance Criteria (Standard / Property)Specific Requirement (as per standard)Proposed Device PerformanceMet Criteria?
    Chemotherapy Drug PermeationBreakthrough Detection Time: > 240 minutes for most drugs, except where noted by predicateVaries per drug (see table below)Met for most, not for Carmustine and Thiotepa (but disclosed)
    Dimension (Length)>= 230 mm (min)230 mm (all sizes)Yes
    Dimension (Width)Varies per size (e.g., 70-120 mm) ±10Matches requirementYes
    Dimension (Thickness - Finger)0.07 mm ±0.02 (for proposed)0.07 mm ±0.02 (specified)Yes (meets own spec)
    Dimension (Thickness - Palm)0.05 mm (min, for proposed)0.05 mm (specified)Yes (meets own spec)
    Dimension (Thickness - Cuff)0.05 mm ±0.02 (for proposed)0.05 mm ±0.02 (specified)Yes (meets own spec)
    Tensile Strength (Before Aging)15 MPa, min (ASTM D6319)15 MPa, minYes
    Ultimate Elongation (Before Aging)500% min (ASTM D6319)500% minYes
    Tensile Strength (After Aging)14 MPa, min (ASTM D6319)14 MPa, minYes
    Ultimate Elongation (After Aging)400% min (ASTM D6319)400% minYes
    Detection of HolesNot detected (ASTM D5151)Not detectedYes
    Powder ContentMax. 0.35 mg per glove (specified for proposed, meets ASTM D6319)Max. 0.35 mg per gloveYes (meets own spec)
    Biocompatibility (Irritation)Not an irritant (ISO 10993-10)Not an irritantYes
    Biocompatibility (Sensitization)Not a sensitizer (ISO 10993-10)Not a sensitizerYes
    Biocompatibility (In Vitro Cytotoxicity)Not cytotoxic (ISO 10993-5)Not cytotoxicYes

    Chemotherapy Drug Permeation Performance and Acceptance Criteria (from tables on pages 2 and 5-6):

    The acceptance criterion for chemotherapy drug permeation is generally a breakthrough detection time of >240 minutes, as demonstrated by most drugs. However, for Carmustine and Thiotepa, lower breakthrough times are explicitly noted and acknowledged. The device performance for these drugs is compared against the general expectation and specific values are provided.

    Chemotherapy DrugConcentrationBreakthrough Detection Time in Minutes (Reported Device Performance)Acceptance Criteria Met? (Based on general >240 min, or specific disclosure for lower times)
    Bleomycin15 mg/ml (15,000 ppm)>240Yes
    Busulfan6 mg/ml (6,000 ppm)>240Yes
    Carboplatin10 mg/ml (10,000ppm)>240Yes
    Carmustine (BCNU)3.3 mg/ml (3,300ppm)8.5 (12.7, 13.4, 8.5) (Noted as having low permeation time)Yes (Disclosed and acceptable with warning)
    Cisplatin1.0 mg/ml (1,000ppm)>240Yes
    Cyclophosphamide(Cytoxan)20.0 mg/ml (20,000ppm)>240Yes
    Cytarabine100 mg/ml (100,000ppm)>240Yes
    Cytovene10 mg/ml (10,000ppm)>240Yes
    Dacarbazine(DTIC)10.0 mg/ml (10,000ppm)>240Yes
    Daunorubicin5 mg/ml (5,000ppm)>240Yes
    Docetaxel10.0 mg/ml(10,000 ppm)>240Yes
    Doxorubicin Hydrochloride2.0 mg/ml (2,000ppm)>240Yes
    Ellence2 mg/ml (2,000ppm)>240Yes
    Etoposide(Toposar)20.0 mg/ml(20,000ppm)>240Yes
    Fludarabine25 mg/ml(25,000ppm)>240Yes
    Fluorouracil50 mg/ml(50,000ppm)>240Yes
    Gemcitabine (Gemzar)38 mg/ml(38,000ppm)>240Yes
    Idarubicin1 mg/ml (1,000ppm)>240Yes
    Ifosfamide50.0 mg/ml (50,000ppm)>240Yes
    Irinotecan20.0 mg/ml (20,000ppm)>240Yes
    Mechlorethamine HCI1.0 mg/ml (1,000ppm)>240Yes
    Melphalan5 mg/ml (5,000ppm)>240Yes
    Methotrexate25mg/ml (25,000ppm)>240Yes
    Mitomycin C0.5 mg/ml (500 ppm)>240Yes
    Mitoxantrone2.0 mg/ml(2,000ppm)>240Yes
    Oxaliplatin2.0 mg/ml(2,000ppm)>240Yes
    Paclitaxel (Taxol)6.0 mg/ml(6,000ppm)>240Yes
    Rituximab10 mg/ml(10,000ppm)>240Yes
    Thiotepa10.0 mg/ml (10,000ppm)36.1 (51.2, 36.1, 45.6) (Noted as having low permeation time)Yes (Disclosed and acceptable with warning)
    Trisenox0.1 mg/ml (100ppm)>240Yes
    Vincristine Sulfate1.0 mg/ml (1,000ppm)>240Yes
    Vinorelbine10 mg/ml(10,000ppm)>240Yes

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the exact sample size for each specific test. However, it indicates that "Bench tests were conducted to verify that the proposed device met all specifications."

    • Sample Size: Not explicitly stated as a numerical value for each test. The adherence to standards like ASTM D6319, ASTM D5151, ASTM D6124, ASTM D6978, ISO 2859-1, ISO 10993-10, and ISO 10993-5 implies that appropriate sample sizes, as dictated by these standards, were used for testing various properties (e.g., AQL for holes, specific number of samples for mechanical properties and chemical permeation). ISO 2859-1 specifically relates to "Sampling Procedures for Inspection by Attributes – Part I: Sampling Plans Indexed by Acceptable Quality Level (AQL) for Lot-by-Lot Inspection," indicating a quality control sampling approach for certain attributes.
    • Data Provenance: The tests were non-clinical bench tests conducted to evaluate the physical, performance, and biocompatibility characteristics of the gloves. The country of origin for the manufacturing company is China (Hebei Titans Hongsen Medical Technology Co., LTD.). The submission is for a medical device (gloves), and the testing appears to be prospective in nature, conducted specifically for this 510(k) submission to demonstrate compliance with standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable to this submission. The device is a medical glove, and its performance is evaluated against established physical, chemical, and biological standards (ASTM and ISO). The "ground truth" here is the pass/fail criteria defined by these objective engineering and material science standards, not subjective human interpretations or expert consensus on medical images or diagnoses.

    4. Adjudication Method for the Test Set

    This is not applicable as the evaluation is based on objective, standardized test methods with clear pass/fail criteria, not on subjective expert review requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    This is not applicable. This submission is for a medical device (gloves) and does not involve AI or human readers for diagnostic or interpretive tasks.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This is not applicable. This submission is not for an algorithm or AI device.

    7. The Type of Ground Truth Used

    The "ground truth" in this context refers to the established scientific and engineering standards and their predefined pass/fail criteria. These include:

    • ASTM D6319-10: Standard Specification for Nitrile Examination Gloves for Medical Application (for dimensions, physical properties).
    • ASTM D5151-06: Standard Test Method for Detection of Holes in Medical Gloves.
    • ASTM D6124-06: Standard Test Method for Residual Powder on Medical Gloves.
    • ASTM D6978-05: Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs (for chemical permeation).
    • ISO 2859-1:1999: Sampling Procedures for Inspection by Attributes – Part I: Sampling Plans Indexed by Acceptable Quality Level (AQL) for Lot-by-Lot Inspection.
    • ISO 10993-10: 2010: Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization.
    • ISO 10993-5:2009: Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity.

    The performance values obtained from the actual testing of the gloves are compared directly against the quantitative limits set forth in these standards.

    8. The Sample Size for the Training Set

    This is not applicable. The device is a physical product (gloves) and does not involve a "training set" in the context of machine learning or AI models.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153732

    Validate with FDA (Live)

    Date Cleared
    2016-07-29

    (214 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner

    Device Description

    Powder Free Blue Nitrile Examination Glove

    AI/ML Overview

    I am sorry, but the provided text from the FDA letter is a 510(k) premarket notification for a Class I medical device (Powder Free Blue Nitrile Examination Glove). This document confirms the device's substantial equivalence to predicate devices and outlines general controls and regulatory requirements.

    The document does NOT contain any information regarding acceptance criteria, performance studies, sample sizes, expert qualifications, or ground truth establishment for an AI/CAD/imaging device.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria based on the provided text. The questions you've asked are typically relevant for more complex devices, especially those involving AI/machine learning for diagnosis or prognosis, which require rigorous performance evaluation. This document is for a very low-risk physical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153562

    Validate with FDA (Live)

    Date Cleared
    2016-04-01

    (109 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner

    Device Description

    POWDER FREE BLUE NITRILE EXAMINATION GLOVE

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for a 'Powder Free Blue Nitrile Examination Glove'. It primarily deals with the regulatory approval of a medical device (a glove), determining its substantial equivalence to previously marketed devices.

    The request asks for information related to acceptance criteria and a study proving a device meets those criteria, specifically within the context of an AI/algorithm-based medical device study.

    This document does not contain the information requested.

    Specifically:

    • It is not about an AI/algorithm-based medical device.
    • It does not describe acceptance criteria for an algorithm's performance (e.g., sensitivity, specificity, AUC). Instead, it's about the regulatory acceptance of a physical medical device (gloves).
    • It does not describe a study involving test sets, training sets, ground truth establishment by experts, or MRMC studies. The documentation for mechanical properties, biocompatibility, or other performance characteristics of the glove would be in the 510(k) submission itself, not in the clearance letter.

    Therefore, I cannot extract the requested information from this document.