Search Results
Found 1 results
510(k) Data Aggregation
(448 days)
Play and Joy Water-Based Lubricant is a water-based personal lubricant, for penile and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex, polyurethane, and polyisoprene condoms.
Play and Joy Water-Based Lubricant is a non-sterile, clear, semi-viscous water-based personal lubricant that provides lubrication during intimate sexual activity. The subject device is compatible with natural rubber latex, polyisoprene, and polyurethane condoms. Its formulation consists of water, xantham gum, glycerine, Chondrus crispus extract, sodium hyaluronate, saccharide isomerate, hydroxyethylcellulose, citric acid, sodium citrate, lactic acid, 1,2-pentanediol, phenoxyethanol, iodopropynyl butylcarbamate, hydroxyacetophenone, gluconolactone, and sodium benzoate. Play and Joy Water-Based Lubricant is for over-the-counter use and is provided in volumes of 50 mL and 100 mL in polyethylene tubes closed with polypropylene caps.
The provided document is a 510(k) Summary for a medical device called "Play and Joy Water-Based Lubricant." It is a submission to the FDA demonstrating substantial equivalence to a predicate device.
This document does not contain the type of acceptance criteria and study information (such as AI performance metrics, sample sizes for test/training sets, expert qualifications, or adjudication methods) typically found for AI/ML-driven medical devices.
The product described is a personal lubricant, and the performance testing focuses on its physical characteristics, biocompatibility, and condom compatibility, not on an AI algorithm's diagnostic performance. Therefore, I cannot extract the information requested in your prompt regarding AI device acceptance criteria and study data from this document.
However, I can provide the acceptance criteria and performance data for the mechanical/physical performance of the lubricant as reported in the document:
1. A table of acceptance criteria and the reported device performance
| Physical Specification | Acceptance Criteria (Specifications) | Reported Device Performance |
|---|---|---|
| Appearance | Clear, semi-viscous liquid | Clear, semi-viscous liquid |
| Color | Colorless | Colorless |
| Odor | Odorless | Odorless |
| PH@25° C (per USP <791>) | 4.92-5.38 | 4.92-5.38 (within range) |
| Viscosity (cps, per USP <912>) | 1727-3463 cps | 1727-3463 cps (within range) |
| Specific Gravity (per USP <841>) | 1.04-1.08 | 1.04-1.08 (within range) |
| Osmolality (mOsm/Kg, per USP <785>) | (944-984) mOsm/kg | (944-984) mOsm/kg (within range) |
| Antimicrobial effectiveness (per USP <51>) | Category 2 product: bacteria should show not less than 2.0 log reduction at 14 days and no increase from 14-day count at the 28-day count. Yeast and molds should show no increase from the initial calculated count at 14 and 28 days | Met criteria (demonstrated antimicrobial effectiveness) |
| Total yeast and mold count (TYMC, per USP <61>) | <10 cfu/g | <10 cfu/g (met criteria) |
| Total aerobic microbial count (TAMC, per USP <61>) | <100 cfu/g | <100 cfu/g (met criteria) |
| Absence of Pathogenic Organisms (P. aeruginosa, S. aureus, Salmonella/Shigella, E. coli, and C. albicans) | Absent | Absent (met criteria) |
| Biocompatibility: Cytotoxicity (ISO 10993-5:2009) | Non-cytotoxic | Non-cytotoxic |
| Biocompatibility: Sensitization (ISO 10993-10:2010) | Non-sensitizing | Non-sensitizing |
| Biocompatibility: Vaginal Irritation (ISO 10993-10:2010) | Non-irritating | Non-irritating |
| Biocompatibility: Acute Systemic Toxicity (ISO 10993-11:2017) | Non-systemically toxic | Non-systemically toxic |
| Condom Compatibility (ASTM D7661-10/(R)2017 -Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms) | Compatible with natural rubber latex, polyisoprene, and polyurethane condoms (no detrimental effect on condom integrity/performance) | Compatible with natural rubber latex, polyisoprene, and polyurethane condoms (demonstrated compatibility) |
| Shelf Life (Accelerated Aging Study) | Maintain all device specifications (as listed above in Table 1) for the stated duration. The predicate had 36 months, the subject device initially aimed for this but demonstrated 6 months. | 6 months (all device specifications from Table 1 met over this duration). The shelf life is reported as 6 months, which is lower than the predicate (36 months), but the document states this difference does not raise different questions of safety and effectiveness (S&E). |
Regarding the other points you requested, as they relate to AI/ML device studies, the document does not contain this information because the device is a physical product (personal lubricant), not an AI/ML diagnostic or therapeutic tool.
- Sample size used for the test set and the data provenance: Not applicable. For physical property testing and biocompatibility, samples of the lubricant itself were tested.
- Number of experts used to establish the ground truth... and qualifications: Not applicable. Performance was measured against objective chemical/physical standards and biological endpoints using validated lab tests, not expert interpretation of AI output.
- Adjudication method: Not applicable.
- Multi reader multi case (MRMC) comparative effectiveness study... effect size of human readers improve with AI vs without AI assistance: Not applicable.
- Standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
- The type of ground truth used: For physical/chemical properties, the ground truth is established by the specified reference standards (e.g., USP monographs). For biocompatibility, it's defined by the biological response as per ISO 10993 standards. For condom compatibility, it's the lack of degradation of condoms as per ASTM D7661.
- The sample size for the training set: Not applicable.
- How the ground truth for the training set was established: Not applicable.
Ask a specific question about this device
Page 1 of 1