Found 2 results

510(k) Data Aggregation

    K Number
    K232223

    Validate with FDA (Live)

    Device Name
    PlaDuo System
    Manufacturer
    Date Cleared
    2023-10-24

    (90 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PlaDuo is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The Argon handpiece is intended for the removal and destruction of skin lesions and the coagulation of tissue.

    Device Description

    The PlaDuo System is an electro-surgical device for use in dermatological applications. The device utilizes nitrogen and argon gas to be used separately to generate the desired power level to patients. The effect of the device is achieved by heating the outer layer of the skin so that part or all of it becomes non-viable and there is controlled damage to the underlying skin.

    The PlaDuo system consists of a system console with wells for 2 gas tanks, footswitch, 2 handpieces, and tip with 3 interchangeable guides.

    AI/ML Overview

    This document describes the FDA's 510(k) clearance for the ShenB Co Ltd's PlaDuo System. The system is an electrosurgical device that uses nitrogen and argon gas to generate plasma for dermatological and general surgical procedures.

    Based on the provided text, the device meets acceptance criteria through various performance tests, including:

    1. IEC 60601-1:2005 + A1:2012: Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance.
    2. IEC 60601-1-2:2014: Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests.
    3. IEC 60601-2-2:2017: Medical Electrical Equipment - Part 2-2: Particular Requirements For The Basic Safety And Essential Performance Of High Frequency Surgical Equipment And High Frequency Surgical Accessories.
    4. IEC 62304:2006+A1:2015: Medical Device Software Life Cycle Processes.
    5. EN ISO 14971:2012: Medical Devices - Application Of Risk Management To Medical Devices.
    6. Biocompatibility testing (cytotoxicity, irritation, and sensitization) per ISO 10993-1:2018 for patient-contacting materials (space guides).
    7. Software verification and validation testing in accordance with FDA's guidance.
    8. Thermal testing on ex vivo tissue specimens to verify the performance of the argon gas.

    Table of Acceptance Criteria and Reported Device Performance

    CategoryAcceptance Criteria (Implied)Reported Device Performance
    Electrical Safety & PerformanceCompliance with relevant IEC standards (60601-1, 60601-1-2, 60601-2-2) for basic safety and essential performance of electrosurgical equipment."Verification and validation activities were successfully completed and establish that the PlaDuo System control unit performs as intended. Testing included the following: IEC 60601-1:2005 + A1:2012; IEC 60601-1-2:2014; IEC 60601-2-2:2017;"
    Software Safety & PerformanceCompliance with IEC 62304 for medical device software life cycle processes and FDA's software guidance."IEC 62304:2006+A1:2015; Medical Device Software Life Cycle Processes;" and "Software verification and validation testing was conducted, and documentation provided in accordance with FDA's Guidance or the Content of Premarket Submissions for Software Contained in Medical Devices."
    Risk ManagementCompliance with ISO 14971 for application of risk management to medical devices."EN ISO 14971:2012; Medical Devices - Application Of Risk Management To Medical Devices"
    BiocompatibilityAll patient-contacting materials must be biocompatible (non-cytotoxic, non-irritating, non-sensitizing)."Biocompatibility –The only patient contacting material in the device are the space guides attached to each handpiece. Biocompatibility testing for cytotoxicity, irritation, and sensitization per ISO 10993-1:2018 are included with the submission and show no concern for patient contacting materials."
    Thermal Performance (Argon Handpiece)The argon handpiece should achieve consistent and desirable thermal damage profiles comparable to predicate/reference devices for the intended clinical effect."In addition, thermal testing on ex vivo tissue specimens was conducted to verify the performance of the argon gas used with the device. The Test device at stacked pulses produced measurable thermal damage in skin and in liver models and traceable thermal effect in muscle. The study demonstrated the ability of the device to achieve consistent thermal damage profiles in line with the target treatment and comparable to the profiles produced by the predicate and reference devices. Thus, it can be concluded that treatment by such a device at the appropriate testings will possess a desirable clinical treatment effect."
    Substantial Equivalence (Overall)The modified PlaDuo System (with Argon handpiece) must be substantially equivalent to the predicate device(s) for the proposed indications for use."The PlaDuo System that is the subject of this application is a next-generation of the currently cleared PlaDuo with Nitrogen only. Besides the addition of the gas and handpiece tip for use with Argon, there have been no changes to the device. Performance testing shows that the newly added argon feature produces consistent thermal damage profiles similar to predicate and reference devices. The subject device is substantially equivalent to the predicate device for the requested indications for use."

    Study Details:

    1. Sample sizes used for the test set and data provenance:

      • The document mentions "thermal testing on ex vivo tissue specimens." It specifies "skin and in liver models" and "muscle" as the tissue types. However, the exact sample size (number of specimens or tests) is not provided.
      • Data provenance: "ex vivo tissue specimens." The country of origin is not specified. The study's nature is likely prospective as it involves new testing for the modified device.
    2. Number of experts used to establish the ground truth for the test set and qualifications of those experts:

      • The document does not specify the number of experts or their qualifications used to establish ground truth or interpret the thermal testing results. The conclusion about "desirable clinical treatment effect" based on thermal damage profiles implies expert assessment, but details are absent.
    3. Adjudication method for the test set:

      • Since the thermal testing is described as direct measurement of physical effects ("measurable thermal damage," "traceable thermal effect"), a human adjudication method in the sense of consensus reading (2+1, 3+1) is not applicable in the same way it would be for image interpretation. The method of assessing "consistent thermal damage profiles" and comparability to predicate devices would likely rely on quantitative measurements and engineering/biomedical expertise, but no specific adjudication method is mentioned.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC study was done. This device is an electrosurgical tool, not an AI interpreting medical images. The evaluation focuses on the safety and performance of the electrosurgical function itself, particularly its thermal effects, rather than how it assists human interpretation or decision-making in a diagnostic context.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • This question is not applicable as the device is a piece of hardware with integrated software for control, not an AI algorithm to be evaluated in a standalone diagnostic capacity. The "software verification and validation testing" confirms the software's functional performance, which is a component of the device's overall standalone performance.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the thermal testing, the ground truth was observed thermal damage profiles on ex vivo tissue, measured directly. The comparison was made against profiles produced by predicate and reference devices, implying their established effectiveness provides a benchmark. It's a physical/biophysical ground truth.
    7. The sample size for the training set:

      • The document describes a 510(k) submission for a physical medical device, not an AI/machine learning model that typically has a "training set." Therefore, the concept of a training set sample size does not apply in this context.
    8. How the ground truth for the training set was established:

      • As this is not an AI/ML device, the question of establishing ground truth for a training set is not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K201735

    Validate with FDA (Live)

    Device Name
    PlaDuo System
    Manufacturer
    Date Cleared
    2020-12-14

    (172 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PlaDuo is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis.

    Device Description

    The ShenB Co Ltd PlaDuo is an electro-surgical device for use in dermatological applications. The effect of the device is achieved by heating the outer layer of the skin so that part or all of the epidermis becomes non-viable and there is controlled damage to the underlying dermis.

    The PlaDuo system consists of a system console, footswitch, handpiece, and tip with 3 interchangeable guides.

    AI/ML Overview

    The provided text is a 510(k) summary for the PlaDuo System, an electrosurgical device. It outlines non-clinical tests conducted to support its substantial equivalence to predicate devices, but no clinical studies or acceptance criteria based on clinical performance are mentioned. Therefore, I cannot provide details on acceptance criteria or studies that prove the device meets them in the context of clinical performance.

    However, based on the non-clinical tests performed, I can infer acceptance criteria for those tests and describe the study that proves the device met them.

    Here's a summary of the non-clinical acceptance criteria and the studies performed:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategorySpecific TestAcceptance CriteriaReported Device Performance
    BiocompatibilityCytotoxicity (ISO 10993-05:2009)Device components (electrode tips) must show no cytotoxic effects in vitro.Pass
    Sensitization (ISO 10993-10:2010, Guinea Pig Maximization Test)Device components (electrode tips) must not induce sensitization (allergic reaction).Pass
    Irritation or Intracutaneous Reactivity (ISO 10993-10:2010, Animal Intracutaneous Test)Device components (electrode tips) must not cause irritation or intracutaneous reactivity.Pass
    Electrical Safety & EMCIEC 60601-1:2005 (Third Edition) + CORR. 1:2006 + CORR. 2:2007 + A1:2-12 (General requirements for basic safety and essential performance)Device must meet general requirements for basic safety and essential performance for medical electrical equipment.Not explicitly stated as "Pass" but implied by submission to standards.
    IEC 60101-2-2:2017 (Particular requirements for the basic safety and essential performance of high-frequency surgical equipment)Device must meet particular requirements for basic safety and essential performance of high-frequency surgical equipment.Not explicitly stated as "Pass" but implied by submission to standards.
    IEC 60601-1-2:2014 (General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances - Requirements and Tests)Device must meet electromagnetic compatibility (EMC) requirements and relevant tests for medical electrical equipment.Not explicitly stated as "Pass" but implied by submission to standards.
    Software Verification & ValidationFDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", May 11, 2005 (moderate level of concern)Software must be verified and validated according to FDA guidance for a moderate level of concern, ensuring proper functionality and safety.Documentation provided. Not explicitly stated as "Pass" but implied by submission and acceptance of documentation.
    Bench TestingOutput EnergyOutput energy must be within predefined design specifications.Operates safely and within predefined design specifications.
    Durability of the tipTip must demonstrate adequate durability according to predefined design specifications.Operates safely and within predefined design specifications.
    Thermal testing (in accordance with FDA's "Guidance for Industry and FDA Staff: Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery”)Thermal effects must be within acceptable limits as outlined by FDA guidance to ensure patient safety and device efficacy.Operates safely and within predefined design specifications.

    Important Note: The document explicitly states "No Clinical study was conducted as part of this submission." Therefore, the following points address the questions in the context of the available non-clinical information.


    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Biocompatibility: The sample sizes for these tests are specific to the biological assay standards (e.g., number of cells for cytotoxicity, number of guinea pigs for sensitization, number of animals for intracutaneous reactivity). These specific numbers are not detailed in the provided summary, but the tests were performed in accordance with international ISO standards.
    • Electrical Safety & EMC: These are typically laboratory tests performed on a single or a few units of the device. The specific number of units tested is not stated.
    • Software Verification & Validation: This involves reviewing and testing the software, often through simulated scenarios or on development units. Sample size for actual "test cases" within software testing is not specified.
    • Bench Testing: This involved testing of the device and its components (e.g., tips). The specific number of devices or tips tested is not provided.

    The provenance of this non-clinical data is from the manufacturer, ShenB Co. Ltd., which is based in South Korea (Seongdong-Gu, Seoul). These were prospective tests conducted to demonstrate compliance with the relevant standards.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This question is not directly applicable as the studies were non-clinical, involving engineering, materials science, and software experts.

    • Biocompatibility: The "ground truth" (i.e., whether a material is biocompatible) is established by the results of standardized laboratory tests interpreted by trained toxicologists/biocompatibility specialists.
    • Electrical Safety & EMC: The "ground truth" (i.e., whether the device meets electrical safety and EMC standards) is established by measurements against predefined limits by qualified electrical engineers and testing laboratories.
    • Software Verification & Validation: The "ground truth" (i.e., whether the software functions as intended and is safe) is established by software engineers and quality assurance specialists comparing software behavior against its requirements and design specifications.
    • Bench Testing: The "ground truth" (i.e., performance against specifications) is established by engineers performing measurements and observations against design requirements.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This concept is primarily relevant for clinical studies involving human interpretation or subjective endpoints. For these non-clinical tests, adjudication methods like 2+1 or 3+1 are not typically used. The results of these tests are objectively measured against established standards and specifications. If there were any ambiguous results, internal quality processes or re-testing would be employed.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was performed as no clinical studies were conducted, and the device is an electrosurgical tool, not an AI-assisted diagnostic or interpretative system.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable as the PlaDuo System is an electrosurgical device, not an algorithm. The "standalone" performance relates to the device's functional integrity as an electrosurgical unit, which was assessed through bench testing and compliance with electrical safety standards.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical tests:

    • Biocompatibility: The ground truth is established by standardized laboratory test results (e.g., cell viability, immune response observed in animal models) compared against established safety thresholds defined by ISO standards.
    • Electrical Safety & EMC: The ground truth is established by objective measurements of electrical characteristics (e.g., leakage current, electromagnetic emissions) against the limits defined in IEC standards.
    • Software Verification & Validation: The ground truth is established by the software requirements specification and design documents, against which the functionality and performance of the software are tested.
    • Bench Testing: The ground truth is established by the device's predefined design specifications and engineering requirements for parameters like output energy, tip durability, and thermal effects.

    8. The sample size for the training set

    This question is not applicable as there is no mention of an algorithm or AI model in the submission that would require a training set.


    9. How the ground truth for the training set was established

    Not applicable, as no training set for an AI/algorithm was mentioned.