510(k) Data Aggregation

    K Number
    K190911

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-11-26

    (232 days)

    Product Code
    Regulation Number
    888.3560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Physica total knee system is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions:

    • Non-inflammatory degenerative joint disease: including osteoarthritis, traumatic arthritis, or . avascular necrosis:
    • Inflammatory degenerative joint disease including rheumatoid arthritis; ●
    • Correction of functional deformity; ●
    • Revision procedures where other treatments or devices have failed; and
    • Treatment of fractures that are unmanageable using other techniques.

    Physica knee system is intended for cemented fixation

    The SMR Shoulder System is intended for partial or total, primary or revision shoulder joint replacement.

    The SMR Anatomic Shoulder System is indicated for partial or total, primary or revision shoulder joint replacement in patients suffering from disability due to:

    • non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis; ●
    • inflammatory degenerative joint disease such as rheumatoid arthritis;
    • . treatment of acute fractures of the humeral head that cannot be treated with other fracture fixation methods:
    • revision of a failed primary implant;
    • . cuff tear arthropathy (CTA Heads only);
    • glenoid arthrosis without excessive glenoid bone loss: A1. A2 and B1 according to Walch . classification (SMR Hybrid Glenoid only).

    The SMR Reverse Shoulder System is indicated for primary, fracture or revision total shoulder replacement in a grossly rotator cuff deficient joint with severe arthropathy (disabled shoulder). The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.

    The Hybrid Glenoid Reverse Baseplate must not be used in cases of excessive glenoid bone loss and/or when bone graft is needed.

    The Modular SMR Shoulder System allows the assembly of components in various humeral and glenoid constructs. The constructs are intended for cemented use as specified in the following table.

    In the Anatomic shoulder the humeral construct consists of the humeral stem, the humeral body, the adaptor taper and the humeral head. In the Reverse shoulder the humeral consists of the humeral stem, the reverse humeral body and the reverse liner. On the humeral side the fixation of the humeral stem determines if the construct is cemented or uncemented.

    The Anatomic glenoid construct consists of an all polyethylene glenoid, a polyethylene glenoid with metal peg or a metal back assembled with a liner, the Reverse glenoid consists of a metal back/connector/glenosphere construct or of a peg/baseplate/glenosphere construct.

    On the glenoid side, the fixation of the all polyethylene glenoid, the polyethylene glenoid with metal peg or the metal back determines if the construct is cemented or uncemented.

    Device Description

    The Physica System is a total knee system and the SMR Reverse is a reverse shoulder system.

    The purpose of the Special 510(k) is to add an alternative sterilization method for the Physica KR Liner and SMR Reverse Humeral Liners. Both systems utilize conventional (non-crosslinked) UHMWPE liners (ISO 5834-2 ASTM F648). The liners can be sterilized using a minimum gamma radiation dose of 25 kGy.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for artificial joint components (knee and shoulder liners). It focuses on demonstrating substantial equivalence to previously cleared devices, primarily through preclinical testing and does not describe a clinical study or a study involving AI/software.

    Therefore, most of the requested information regarding acceptance criteria, study design, ground truth, expert involvement, and AI performance cannot be extracted from this document.

    Here's what can be gathered, addressing the relevant parts of your request:

    1. A table of acceptance criteria and the reported device performance:

    The document states: "The Physica KR Liner and SMR Reverse Humeral Liner is substantially equivalent to the predicate devices cited on the previous page with respect to indications, design, function, and performance." It then lists the following preclinical tests:

    Acceptance Criteria CategoryReported Device Performance (Summary)Details Provided in Document
    Sterilization ValidationMet established sterilization requirements."Sterilization validation" was performed, indicating the new sterilization method (minimum gamma radiation dose of 25 kGy) was validated. Implies acceptance criteria related to sterility assurance level (SAL) were met.
    UHMWPE CharacterizationMaterials meet industry standards."UHMWPE Characterization" was performed. This would involve tests to confirm material properties (e.g., molecular weight, density, mechanical properties) meet the specifications for conventional UHMWPE (ISO 5834-2 ASTM F648). Implies acceptance criteria related to material specifications were met.
    Mechanical TestingDevice mechanically performs as intended and is comparable to predicate."Mechanical Testing" was performed. This would have involved tests such as wear, fatigue, static strength, etc., to demonstrate the device's mechanical integrity and performance is equivalent to the predicate devices under simulated physiological conditions. Implies acceptance criteria related to biomechanical performance were met.

    Important Note: The document does not provide specific numerical acceptance criteria (e.g., "wear rate must be < X mg") or the exact numerical results for each test. It only broadly states that the device is "substantially equivalent" based on these tests.

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: Not explicitly stated. The tests are preclinical (sterilization, material characterization, mechanical testing), not clinical trials involving human subjects or large data sets. The sample size would refer to the number of physical liners or material samples tested in a lab.
    • Data Provenance: The preclinical tests were performed by LimaCorporate (Italy). The document doesn't specify if external labs were used, but the manufacturer is based in Italy. The data is from preclinical laboratory testing, not human data (retrospective or prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable: This filing describes preclinical performance testing of a physical medical device. It does not involve "ground truth" derived from expert consensus on medical images or clinical outcomes. The "ground truth" for these tests would be the established scientific and engineering principles, material standards (e.g., ISO, ASTM), and validated test methods.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not Applicable: Adjudication methods like 2+1 or 3+1 are used in clinical trials or studies involving expert readers for subjective interpretations. This document describes objective preclinical laboratory tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable: No MRMC study was done, as this submission is for a physical medical device (joint liners), not an AI/software device. The document explicitly states: "Clinical testing was not necessary to support equivalence."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable: This device is not an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Preclinical Test Benchmarks/Standards: The "ground truth" for the preclinical tests would be defined by established engineering and material standards (e.g., ISO, ASTM), recognized test methodologies (e.g., for wear testing, fatigue testing, sterilization validation), and historical performance data of predicate devices. For example, for "UHMWPE Characterization," the ground truth would be the known, acceptable material properties defined by ASTM F648 or ISO 5834-2.

    8. The sample size for the training set:

    • Not Applicable: There is no "training set" as this is not an AI/machine learning device.

    9. How the ground truth for the training set was established:

    • Not Applicable: See point 8.