510(k) Data Aggregation

    K Number
    K231691

    Validate with FDA (Live)

    Date Cleared
    2023-06-29

    (20 days)

    Product Code
    Regulation Number
    882.1350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Phantom XL Insulated Dilators are indicated for use during surgery of the spine to deliver an electrical stimulus to the tissues and nerves at the operative site, to assist in locating those nerves at risk during the surgical procedure.

    Device Description

    TSI's Phantom XL Insulated Dilators are used as instruments to deliver electrical stimulation to tissue during intraoperative neurological monitoring. Phantom XL Insulated Dilators are available in monopolar configuration and four diameter sizes. They are supplied sterile, are non-pyrogenic, and are intended for single use only.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called "Phantom XL Insulated Dilators". This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through clinical trials with specific performance metrics like sensitivity, specificity, or AUC.

    Therefore, the information you're requesting regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of typical AI/diagnostic device performance studies (which would involve metrics like sensitivity, specificity, human reader performance, etc.) is not applicable to this submission.

    This submission focuses on engineering-level performance testing to demonstrate that changes made to the existing predicate device (Phantom XL Insulated Dilators, K140088) do not raise new questions of safety or effectiveness. The changes are minor:

    • Change in packaging: from backer card/Tyvek pouch/Shelf Box to preformed tray sealed in Tyvek Lid placed in Shelf Box.
    • Extended shelf life: from 2 years to 5 years.

    The acceptance criteria and studies described relate to these specific physical changes and their impact on the device's ability to maintain sterility, packaging integrity, and functionality over its extended shelf life.

    Here's a breakdown of the information provided, reframed to fit the context of this 510(k) submission:


    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a "table of acceptance criteria and reported device performance" in the way one might expect for a diagnostic or AI device (e.g., sensitivity > X%, specificity > Y%). Instead, it lists the types of changes made to the device and the standards/tests performed to ensure these changes do not negatively impact safety and effectiveness. The "acceptance criteria" are implied by compliance with the referenced ASTM, ISO, and AAMI standards. The "reported device performance" is that the device "demonstrates that the TSI Phantom XL Dilators are safe for use perform comparably to the predicate device that is currently marketed for the same intended use" after these tests.

    Type of ChangeDescription of Tests Performed (Implied Acceptance Criteria: Compliance with these standards)Reported Device Performance (Summary)
    Packaging- ASTM F1886:2016 (Visual Inspection for Seal Integrity)The new packaging design maintains integrity and sterility, demonstrating satisfactory verification and validation in compliance with Design Control Process.
    - ASTM F2096-11 (Gross Leaks by Internal Pressurization)
    - ASTM F88/F88M-21, 2021 (Seal Strength of Flexible Barrier Materials)
    - Dupont Medical Packaging (Compliance of Tyvek to ISO 11607)
    - ASTM D4169:2022 (Performance Testing of Shipping Containers and Systems)
    - ISTA 3A: 2018 (Packaged-Products for Parcel Delivery System Shipments)
    - ISO 11607-2:2019 (Validation requirements for sealing and assembly processes)
    - ISO 16269-6:2014 (Statistical tolerance intervals)
    Expiration Date- ASTM F1980; 2021 (Accelerated Aging of Sterile Medical Device Packages)Extended shelf life from 2 to 5 years is supported by accelerated aging studies, showing safety and effectiveness are maintained over the longer period.
    Sterilization- ISO 11607-1:2019 (Requirements for materials, sterile barrier systems, packaging)Sterility (maintaining SAL) with the new packaging and extended shelf life is confirmed through bench testing and compliance with relevant standards.
    - AAMI/ISO TIR16775: 2014 (Guidance on application of ISO 11607-1 & 11607-2)
    - AAMI TIR28:2016 (Product adoption and process equivalency for EtO sterilization)
    - ANSI/AAMI ST67:2019 (Selecting a sterility assurance level (SAL))
    - ANSI/AAMI/ISO 11135:2014 (Ethylene oxide sterilization process validation)
    - ANSI/AAMI ST72; 2019 (Bacterial Endotoxins-Test methods)
    General- Bench testing for sterility (as per standards above)No new issues of safety or effectiveness were identified. All risk mitigations were satisfactorily verified and validated.
    - Bench testing for new expiration dating (as per standards above)
    - Bench testing for packaging integrity (as per standards above)
    - Compliance with company's Design Control process (21 CFR Part 820.30)
    - Risk analysis per ISO 14971 Standard

    2. Sample size used for the test set and the data provenance

    The document does not specify exact sample sizes for each test listed (e.g., how many packages were tested for seal strength). The studies mentioned are engineering/bench tests on the device and its packaging components, not human subject studies. Therefore, "data provenance" in terms of country of origin or retrospective/prospective is not relevant here.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This is not applicable. The "ground truth" for this type of submission is defined by the technical specifications of the device and its packaging, and ensuring compliance with recognized industry standards (ASTM, ISO, AAMI). There is no "ground truth" to be established by medical experts (e.g., radiologists interpreting images) as this is not a diagnostic device with performance metrics relying on human interpretation. The testing is performed by engineers and technicians following standard protocols.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies when there's ambiguity in establishing a ground truth (e.g., for disease diagnosis from images) or for resolving disagreements among multiple human readers. This submission describes physical and performance tests against standards, not diagnostic performance.


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This device is a surgical instrument (an insulated dilator used for nerve stimulation/localization during spine surgery), not an AI-powered diagnostic tool. Therefore, MRMC studies and AI assistance comparisons are irrelevant.


    6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done

    This is not applicable. As stated above, this is a surgical instrument, not an algorithm or AI device.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" in this context is the compliance with established engineering and medical device standards (ASTM, ISO, AAMI). For example, for packaging, the "ground truth" is that the seal strength meets the requirements specified by ASTM F88, or that the sterility assurance level (SAL) meets the requirements of ANSI/AAMI ST67 after ethylene oxide sterilization. It's about meeting predefined technical specifications and standards, not a medical condition.


    8. The sample size for the training set

    This is not applicable. This is a physical medical device, not an AI or machine learning model. There is no "training set."


    9. How the ground truth for the training set was established

    This is not applicable. As there is no training set for an AI or ML model, this question does not apply.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140088

    Validate with FDA (Live)

    Date Cleared
    2014-09-25

    (254 days)

    Product Code
    Regulation Number
    874.1820
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Phantom XL Insulated Dilators are indicated for use during surgery of the spine to deliver an electrical stimulus to the tissues and nerves at the operative site, to assist in locating those nerves at risk during the surgical procedure.

    Device Description

    Phantom XL Insulated Dilators are used as instruments to deliver electrical stimulation to tissue during intraoperative neurological monitoring. The Phantom XL Insulated Dilators are available in monopolar configuration and four diameter sizes (8, 13, 18, 22 and 22mm Grooved). They are supplied sterile, are non-pyrogenic, and are intended for single use only. The 8 mm dilators are made of stainless steel and are visible under fluoroscopy, while the remaining sizes are made of aluminum alloy and are radiolucent.

    AI/ML Overview

    The provided document appears to be a 510(k) premarket notification for a medical device called "Phantom XL Insulated Dilators." It details various performance tests conducted to demonstrate the device's safety and effectiveness and its substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and supporting studies based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test CategorySpecific TestAcceptance CriteriaReported Device Performance
    Electrical SafetyImpedance of current carrying connectionsTo have sufficiently low impedance so that they will not interfere with any stimulator signal passed through them.Met applicable requirements (per IEC 60601-1)
    Dielectric Strength TestTo be sufficiently separated from the conductive portions to prevent any stimulator signal from reaching any part of the Phantom XL Insulated Dilators except the intended delivery point.Met applicable requirements (per IEC 60601-1)
    Temperature TestNot to introduce any additional heating to the patient as part of the stimulation process.Met applicable requirements (per IEC 60601-1)
    Waveform captureNot to significantly alter the waveform of the stimulator output.Met applicable requirements (per IEC 60601-1)
    Electromagnetic CompatibilityEmissions testsNot to impact the emission profile of electrical stimulus and medical monitoring equipment.Met applicable requirements (per IEC 60601-1-2)
    Immunity testsNot to increase the sensitivity to adverse conditions of electrical stimulus and medical monitoring equipment.Met applicable requirements (per IEC 60601-1-2)
    BiocompatibilityCytotoxicity (ISO Elution Method)Biological reactivity must be less than or equal to a grade 2.No evidence of causing lysis or toxicity. Reactivity is less than grade 2. (Non-Cytotoxic)
    ISO Intracutaneous Study (Irritation)The difference between the test extract overall mean score and corresponding control overall mean score is 1.0 or less.Difference between test article extract overall mean score and corresponding control extract overall mean score was 0.0 and 0.2. (Non-irritant)
    ISO Guinea Pig Maximization Sensitization TestNo evidence of causing delayed dermal contact sensitization.Showed no evidence of causing delayed dermal contact sensitization. (Non-Sensitizer)
    ISO Systemic Toxicity (acute)No mortality or evidence of systemic toxicity from the extracts.No mortality or evidence of systemic toxicity from the extracts. (Non-Toxic)
    Sterilization & ResidualsSterilization ValidationA minimum Sterility Assurance Level (SAL) of 1 x 10-6.Completed to assure a minimum SAL of 1 x 10-6.
    Bacterial EndotoxinFDA criteria for medical device: maximum of 20 EU/device. (Also criteria for device contacting cerebrospinal fluid: maximum of 2.15 EU/device, though noted as N/A for this device).0.720 EU/Device, 0.880 EU/Device, < 0.200 EU/Device. (Passes FDA criteria)
    Ethylene Oxide residual4 mg/device.Range: <0.12 - 0.15 mg/device. (Passes)
    Ethylene Chlorohydrin residual(Implicitly, to be within acceptable limits for a residual).Range: <0.10 - <0.15 mg/device.

    2. Sample size used for the test set and the data provenance:

    • The document does not specify sample sizes for most performance tests (e.g., how many dilators were tested for electrical safety, EMC, or sterilization residuals).
    • For biocompatibility, the tests were conducted consistent with ISO 10993 standards, which generally outline sample size requirements, but the specific numbers for each test are not explicitly stated in this summary.
    • Data provenance: Prospective, as these tests were performed specifically for the 510(k) submission to demonstrate the device's characteristics. The country of origin of the data is not specified, but the testing adhered to international standards (IEC, ISO) and U.S. GLP regulations (21 CFR 58), implying a globally recognized testing methodology.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not applicable (N/A) to the types of studies described. The studies are engineering and biological bench tests/laboratory tests, not diagnostic studies that would require expert human review for ground truth establishment. For example, for cytotoxicity, the "ground truth" is determined by observing cell lysis or toxicity in a lab setting, not by expert consensus.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • This information is N/A. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or image-based diagnostic studies where human experts are making judgments that need to be reconciled for ground truth. The studies described here are objective, quantifiable physical and biological tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This information is N/A. The "Phantom XL Insulated Dilators" are a surgical instrument for nerve localization during spine surgery using electrical stimulation. They are not an AI-powered diagnostic device, nor are they a device that involves human "readers" or interpretations in the way an AI diagnostic tool would. Therefore, an MRMC study or an assessment of AI-assisted human performance is not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • This information is N/A. As explained above, this is a physical medical device (surgical dilators) and not an algorithmic or AI-based system that would have "standalone" performance in the context of AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For the electrical safety, EMC, and sterilization studies, the "ground truth" is established by objective measurements against predefined technical standards (e.g., specific voltage limits, impedance values, temperature increases, bacterial endotoxin concentration, residual levels).
    • For biocompatibility, the "ground truth" is established through laboratory observations and measurements of biological responses (e.g., cell lysis, erythema, edema, signs of sensitization or systemic toxicity in animal models) compared against documented acceptable biological response criteria from international standards.

    8. The sample size for the training set:

    • This information is N/A. The device is not an AI/ML-based system that requires a "training set." It's a physical medical device.

    9. How the ground truth for the training set was established:

    • This information is N/A for the same reason as point 8.