    K222381

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-04-27

    (265 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PeriPRO™ Femur Plating System is indicated for internal fixation of femur fractures and osteotomies in normal and osteopenic bone. This includes:
    · Non-unions, malunions and deformities
    · Fracture in the presence of intramedullary implants, including periprosthetic fractures

    The Variable Angle Fixation System is indicated for the internal fixation of fractures and osteotomies of the femur in normal and osteopenic bone, including:

    • · Non-unions, malunions and deformities
    • · Periprosthetic fractures

    AxSOS 3 Ti is intended for long bone fracture fixation. Indications include:
    · Diaphyseal, metaphyseal, epiphyseal, extra- and intra-articular fractures

    • Non-unions and malunions
    • · Normal and osteopenic bone
    • · Osteotomies
    • · Periprosthetic fractures of the femur and proximal tibia.
    • The AxSOS 3 Ti Waisted Compression Plates are also indicated for fracture fixation of:
    • · Periprosthetic fractures
    • · Diaphyseal and metaphyseal areas of long bones in pediatric patients.
      The 4 mm Waisted Compression Plate indications also include fixation of the scapula and the pelvis. Screws can also be used for arthrodesis.

    The Stryker Plating System, Small Frag Set is indicated for fractures of the metaphysis of the following:

    • · One third tubular plate: fibula, metatarsals, metacarpals
    • Fibular plate: fibula
    • · Compression plate: radius, ulna, distal tibia, fibula, distal humerus, clavicle
    • · Oblique T-plate: distal radius
    • · T-plate: distal radius, calcaneus, lateral clavicle
    • Cloverleaf plate: proximal humerus, distal tibia
    • · Calcaneal plate: calcaneus
      · Reconstructive plate: humerus, pelvis Screws are used either to fasten plates or similar devices onto bone, or, as lag screws, to hold together fragments of bone.
      The Basic Fragment Set is intended for use in long bone fracture fixation plates, wide and narrow straight and waisted compression plates are indicated for fixation of long bone fractures including but not limited: to fractures of the femur, the tibia, the humerus and the pelvis. T-buttress plates and L-buttress plates are indicated for fractures at the proximal or distal end of long bones including but not limited to: fractures of the femoral condyles, the tibial plateau, the distal tibia and the proximal humerus.
    Device Description

    This Traditional 510(k) submission is being supplied to the U.S. FDA to gain clearance to market the PeriPRO Femur and Variable Angle Fixation System and align labeling across the AxSOS 3 Ti and SPS System. This submission encompasses multiple systems (PeriPRO Femur, Variable Angle Fixation, AxSOS 3Ti and SPS Systems) that have similar intended use and/or will be used together during the surgical procedure. The PeriPRO™ Femur System consists of proximal, interprosthetic and distal femur plates. All devices in the system are provided sterile and non-sterile. The proximal femur plates range from 257 up to 371 mm in length. The interprosthetic femur plate ranges from 320 up to 420 mm in length. The distal femur plates range from 173 up to 396 mm in length. All plates are made of titanium alloy (Ti6AL4VELI) according to ASTM 136. The Variable Angle Fixation System consists of screws and cable plugs. All devices in the system are provided sterile and non-sterile. Screws are offered as non-locking or locking variants. The locking screws are available as Ø4 (L22-95 mm), Ø5 mm (L22-120 mm) and Ø5 mm flat tip (L10-20 mm) are made of CoCr alloy. The Cable plugs are available in a 5 mm and are made of titanium alloy. The PeriPRO™ Femur System and Variable Angle Fixation contains Class II instruments as targeters, frame fixators, handles, Those devices are part of the targeting system and as such class II instruments. They are manufactured of either PEI/Carbon or stainless steel. AxSOS 3 Ti System is an internal fixation system composed of sterile and non-sterile plates, screws, and complementary implantable devices, along with associated instruments. The plates are available in a variety of anatomical orientations and lengths, the screws in a variety of diameters and lengths. Implants of this system are available in titanium alloy, with Type II anodization. The Stryker Plating System (SPS System) contains the Basic Fragment Set and the Small Fragment Set. Both sets contain plates, screws, and washers available in titanium alloy (Ti-6A14V) per ASTM F136 as well as stainless steel (316L) in sterile and non-sterile forms. SPS plates come with holes that can accommodate non-locking screws and include holes for suture or Kwires. The screws are provided in 2 types (cortical and cancellous) and are available in several different lengths and diameters.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for bone fixation systems (PeriPRO™ Femur System, Variable Angle Fixation System, AxSOS 3Ti, Stryker Plating System).

    Based on the provided text, there is NO information regarding acceptance criteria or a study proving device performance in the context of an AI/algorithm-driven device that would typically have such criteria.

    The document solely addresses the substantial equivalence of these bone fixation systems to previously cleared predicate devices based on non-clinical performance data (mechanical testing and MRI compatibility).

    Therefore, I cannot provide the requested information. The document focuses on the mechanical and material properties of the bone plates and screws.

    Specifically, the document states under "Clinical Performance and Conclusions:" that "Clinical data and conclusions were not needed for this device." This further confirms that no clinical studies related to human reader performance, AI assistance, or standalone algorithm performance were conducted or reported for this submission.