Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K242104

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-02-13

    (210 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Penumbra Reperfusion Catheters and Separators

    As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.

    Penumbra 3D Revascularization Device

    As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.

    Penumbra Aspiration Tubing

    As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.

    Penumbra Aspiration Pump

    The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

    Device Description

    The Penumbra System is comprised of the following devices:

    • Penumbra Reperfusion Catheter
    • . Penumbra 3D Revascularization Device
    • Penumbra Aspiration Pump
    • . Penumbra Aspiration Pump Canister/Tubing
    • Penumbra Aspiration Tubing
    • Penumbra Separator .

    The Penumbra System is designed to remove thrombus and restore blood flow in the neurovasculature using aspiration. The Reperfusion Catheter delivers aspiration from the pump directly to the site of occlusion to remove the clot. The 3D Revascularization Device is used with Reperfusion Catheters to facilitate aspiration and removal of the thrombus when needed. Alternatively, a Separator may be used to clear the lumen of the Reperfusion Catheter should it become blocked with thrombus. The Reperfusion Catheter is introduced through a guide catheter or long femoral sheath and into the intracranial vasculature and guided over a neurovascular guidewire to the site of the primary occlusion. The Penumbra Reperfusion Catheter is used with the Penumbra Aspiration Pump to aspirate thrombus from an occluded vessel. The Penumbra Reperfusion Catheter is connected to the Penumbra Aspiration Pump using the Penumbra Aspiration Tubing and the Penumbra Aspiration Pump Canister. The Penumbra Reperfusion Catheter is provided with a steam shaping mandrel, rotating hemostasis valve (RHV), peelable sheath, and optionally, SENDit Technology (cleared under K191946). The Penumbra 3D Revascularization Device is provided with an introducer sheath. The Penumbra Separator is provided with an introducer and torque device. The devices are visible under fluoroscopy.

    AI/ML Overview

    The provided text describes specific acceptance criteria and performance data for the Penumbra System (Reperfusion Catheter RED 72). However, it notes that no animal or clinical studies were conducted for this particular device submission (K242104). The evidence for meeting acceptance criteria comes solely from bench and biocompatibility testing. Therefore, information regarding sample sizes for test sets, data provenance, number and qualifications of experts, adjudication methods, MRMC studies, standalone performance, and training set details, which are typically associated with clinical or AI/algorithm performance studies, are not applicable or available in this document.

    Here's a summary of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Dimensional/Visual TestUnits meet all dimensional and visual product specifications.Acceptance Criteria Met
    Friction TestUnits meet product specification related to friction.Acceptance Criteria Met
    Radiopacity TestMarkerband is fluoroscopically visible.Acceptance Criteria Met
    Simulated Use TestFunctionality of units using clinically relevant benchtop model.Acceptance Criteria Met
    Particulate and Coating Integrity TestParticulates generated during simulated use and coating integrity after simulated use (including multiple deployment cycles) were evaluated.Acceptance Criteria Met
    Hub/Air TestUnits have no leaks when tested.Acceptance Criteria Met
    Tensile TestUnits meet product specification related to tensile strength.Acceptance Criteria Met
    Pressure TestUnits meet product specification related to pressure.Acceptance Criteria Met
    Elongation TestUnits meet product specification related to elongation.Acceptance Criteria Met
    Corrosion Resistance TestNo visible corrosion on the units when tested.Acceptance Criteria Met
    Torque Strength TestUnits have sufficient torque strength.Acceptance Criteria Met
    Burst Pressure TestUnits can withstand sufficient pressure.Acceptance Criteria Met
    Distal Tip Stiffness TestUnits have distal tip stiffness comparable to the reference device.Acceptance Criteria Met
    Kink Resistance TestUnits meet product specification related to kink resistance.Acceptance Criteria Met
    Simulated Use Physician Usability TestUsability of units in clinically relevant benchtop model.Acceptance Criteria Met
    Shelf LifeExpiration date based on accelerated aging test studies.Acceptance Criteria Met
    Packaging ValidationPackaging of the units meets all product specifications.Acceptance Criteria Met
    Sterilization TestUnits are sterilized in accordance with ISO 11135+A1 and ISO 10993-7.Acceptance Criteria Met
    Biocompatibility Tests
    Cytotoxicity: MEM ElutionSample extracts must have a cytotoxic reactivity score of Grade 2 or lower.Pass (Non-cytotoxic)
    Sensitization: Magnusson-Kligman MethodTest group shall yield Grade < 1 score on Magnusson and Kligman scale (provided control group yields Grade < 1).Pass (Non-sensitizing)
    Irritation: Intracutaneous ReactivityThe difference between the average scores for the extract of the test article and the control is ≤ 1.0.Pass (Non-irritating)
    Systemic Toxicity: Acute Systemic InjectionSample extracts must not cause significant biological reaction greater than control. That is: Death in 2 or more animals. Signs of toxicity in 2 or more animals (i.e. convulsions, prostration). Weight loss > 10% in 3 or more animals.Pass (Non-toxic)
    Systemic Toxicity: Material-Mediated PyrogenSample extracts must not cause a total rise in body temperature of ≥ 0.5 °C.Pass (Non-pyrogenic)
    Hemocompatibility: In-vitro ThrombogenicityDevice must not be thrombogenic in vitro when compared to a predicate device.Pass (Comparable to the predicate)
    Hemocompatibility: Partial Thromboplastin Time (PTT)Clotting times of test article must be similar to predicate values.Pass (Comparable to the predicate)
    Hemocompatibility: Complement ActivationThe concentration of SC5b-9 of test article must be similar to predicate values.Pass (Non-activator of the complement system)
    Hemocompatibility: Hemolysis (indirect contact)Sample extracts must be non-hemolytic (≤ 2% hemolytic index).Pass (Non-hemolytic)
    Hemocompatibility: Hemolysis (direct contact)Sample must be non-hemolytic (≤ 2% hemolytic index).Pass (Non-hemolytic)
    Hemocompatibility: Platelet and Leukocyte CountSample must meet the requirements of ASTM F2888-19, Standard Practice for Platelet Leukocyte Count.Pass (Similar to the predicate)

    2. Sample size used for the test set and the data provenance: Not applicable. The document states that "No animal or clinical study was conducted as bench and biocompatibility testing were determined sufficient for verification and validation purposes." The tests listed are primarily bench (in-vitro) tests on manufactured units or material samples. The specific number of units or samples tested for each bench test is not provided in this summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. The "ground truth" for these engineering and biocompatibility tests is based on established scientific principles, industry standards, and specified material properties, not expert consensus on medical images or diagnoses.

    4. Adjudication method for the test set: Not applicable. Adjudication methods are typically employed in clinical studies or studies involving human assessment of data, which were not performed in this case.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No MRMC study was performed as this device is a physical medical device (catheter system) and not an AI-powered diagnostic or assistive tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used:
    * Bench Testing: Engineering specifications, material properties, and functionality performance standards.
    * Biocompatibility Testing: Established biological safety standards (ISO 10993 series, USP standards, 21 CFR Part 58 GLP).

    8. The sample size for the training set: Not applicable. This is a physical medical device, not an algorithm that requires a training set.

    9. How the ground truth for the training set was established: Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211654

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-08-16

    (80 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Penumbra Reperfusion Catheters and Separators

    As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA therapy are candidates for treatment.

    Penumbra 3D Revascularization Device

    As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    Penumbra Aspiration Tubing

    As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.

    Penumbra Aspiration Pump

    The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

    Device Description

    The Penumbra System® is comprised of the following devices:

    • Penumbra Reperfusion Catheter
    • Penumbra Aspiration Pump
    • Penumbra Aspiration Pump/Canister Tubing
    • Penumbra Aspiration Tubing
    • Penumbra Separator
    • Penumbra 3D Revascularization Device

    The Penumbra System is designed to remove thrombus from the vasculature using continuous aspiration. The Reperfusion Catheter targets aspiration from the pump directly to the thrombus. The 3D Revascularization Device is used with Reperfusion Catheters to facilitate aspiration and removal of the thrombus when needed. The Separator may be used to clear the lumen of the Reperfusion Catheter should it become blocked with thrombus. The use of the Separator may not be necessary when using a Reperfusion Catheter with an ID of 0.054 in. or larger. The Reperfusion Catheter is introduced through a guide catheter or long femoral sheath and into the intracranial vasculature and guided over a neurovascular guidewire to the site of the primary occlusion. The Penumbra Reperfusion Catheter is used with the Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, a Penumbra Separator may be deployed from the Reperfusion Catheter to assist with thrombus removal. The Penumbra Separator is advanced and retracted through the Penumbra Reperfusion Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the Reperfusion Catheter tip. For the aspiration source, the Penumbra Reperfusion Catheter is used in conjunction with the Aspiration Pump, which is connected using the Penumbra Aspiration Tubing and the Penumbra Pump/Canister Tubing. The Penumbra Reperfusion Catheter is provided with a steam shaping mandrel and rotating hemostasis valve, and a peelable sheath. The Penumbra 3D Revascularization Device is provided with an introducer sheath. The Penumbra Separator is provided with an introducer and torque device. The Penumbra Reperfusion Catheters, 3D Revascularization Device and Separators are visible under fluoroscopy.

    AI/ML Overview

    This document describes the acceptance criteria and study proving the performance of the Penumbra System® (Reperfusion Catheter RED™ 72). It is a 510(k) summary, indicating that the device is being cleared based on substantial equivalence to a legally marketed predicate device, rather than requiring a full premarket approval (PMA) with extensive clinical trials.

    It's important to note that because this is a 510(k) submission based on substantial equivalence to an existing device (the Penumbra System (Reperfusion Catheter JET 7)), the studies primarily focus on engineering and biocompatibility performance data against established requirements, rather than clinical efficacy studies often seen with novel AI devices. There are no AI/ML components described in this document, thus the requested information regarding MRMC studies, standalone AI performance, human expert ground truth for imaging, and training/test set sample sizes for AI models are not applicable.

    Here's the breakdown of the acceptance criteria and study details provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document provides a series of design verification tests and biocompatibility tests. The acceptance criteria for each test are stated, and the conclusion is consistently that the "Acceptance Criteria Met" or "Pass."

    Design Verification Testing:

    TestTest Method SummaryAcceptance Criteria Met (Reported Performance)
    Dimensional/Visual TestConfirms the units meet all dimensional and visual product specifications.Yes
    Friction TestConfirms units meet product specification related to friction.Yes
    Fluoroscopy TestConfirms the markerband is fluoroscopically visible.Yes
    Simulated Use TestConfirms the functionality of units using clinically relevant benchtop model.Yes
    Particulate TestParticulates generated during simulated use (including multiple deployment cycling) were evaluated.Yes
    Hub/Air TestConfirms units have no leaks when tested.Yes
    Tensile TestConfirms units meet product specification related to tensile strength.Yes
    Pressure TestConfirms units meet product specification related to pressure.Yes
    Elongation TestConfirms units meet product specification related to elongation.Yes
    Corrosion Resistance TestConfirms there is no visible corrosion on the units when tested.Yes
    Torque Strength TestConfirms units have sufficient torque strength.Yes
    Burst Pressure TestConfirms units can withstand sufficient pressure.Yes
    Shelf-LifeConfirms expiration date based on accelerated aging test studies.Yes
    Sterilization TestConfirms the units are sterilized in accordance with ISO 11135 and ISO 10993-7.Yes

    Biocompatibility Testing (Based on predicate/reference devices):

    TestsAcceptance CriteriaPass/Conclusion (Reported Performance)
    Cytotoxicity: MEM Elution (ISO 10993-5)Sample extracts must have a cytotoxic reactivity score of grade 2 or lower.Pass (Non-cytotoxic)
    Sensitization: Magnusson-Kligman Method (ISO 10993-10)Test Group shall yield Grade < 1 score on Magnusson and Kligman scale (provided Control Group yields Grade < 1).Pass (Non-sensitizing)
    Irritation: Intracutaneous Reactivity (ISO 10993-10)The difference between the average scores for the extract of the test article and the control is ≤ 1.0.Pass (Non-irritating)
    Systemic Toxicity: Acute Systemic Injection (ISO 10993-11)Sample extracts must not cause significant biological reaction greater than control. That is: • Death in 2 or more animals. • Signs of toxicity in 2 or more animals (i.e. convulsions, prostration). • Weight loss > 10% in 3 or more animals.Pass (Non-toxic)
    Systemic Toxicity: Material Mediated Pyrogen (ISO 10993-11)Sample extracts must not cause a total rise in body temperature of ≥ 0.5°C.Pass (Non-pyrogenic)
    Hemocompatibility: Thrombogenicity (ISO 10993-4)The test article must have similar or less thrombus formation than predicate.Pass (Non-thrombogenic)
    Hemocompatibility: Partial Thromboplastin Time (PTT) (ISO 10993-4)Clotting times of test article must be similar to predicate values.Pass (Hemocompatible)
    Hemocompatibility: Complement Activation (ISO 10993-4)The concentration of SC5b-9 of test article must be similar to predicate values.Pass (Hemocompatible)
    Hemocompatibility: Hemolysis, indirect contact (ISO 10993-4)Sample extracts must be non-hemolytic (≤ 2% hemolytic index).Pass (Non-hemolytic)
    Hemocompatibility: Hemolysis, direct contact (ISO 10993-4)Sample must be non-hemolytic (≤ 2% hemolytic index).Pass (Non-hemolytic)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes (number of units) used for each design verification bench test, as is typical for 510(k) summaries of mechanical devices. However, it indicates that units were tested. The data provenance would be from internal testing conducted by Penumbra, Inc. (the manufacturer). This is a retrospective evaluation of the device's physical and functional properties. There is no indication of country of origin for specific test data beyond the manufacturer being in Alameda, California, USA.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. This submission is for a mechanical medical device (catheter system), not an AI/ML device that requires human expert annotation for ground truth in image analysis. The "ground truth" for this device's performance is established by objective physical and chemical testing against engineering specifications and international standards (e.g., ISO, USP).

    4. Adjudication Method for the Test Set

    Not applicable. As there are no human expert readings or interpretations involved in the performance data presented (it's benchmark and material testing), there is no need for an adjudication method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No. An MRMC study is for evaluating the impact of AI on human reader performance, typically in diagnostic imaging. This device is a mechanical catheter system for stroke intervention, not an AI/ML diagnostic tool. Therefore, no such study was performed or required.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    No. This refers to the standalone performance of an AI algorithm. Since this is a mechanical device, this concept is not applicable. The standalone "performance" refers to the device meeting its physical and functional specifications during bench testing.

    7. The Type of Ground Truth Used

    The "ground truth" used for this device's performance evaluation comes from engineering specifications, international standards (e.g., ISO 10993 series), and validated test methods. This includes:

    • Pre-defined dimensional and visual product specifications.
    • Measured physical properties (friction, tensile strength, pressure, elongation, torque, burst pressure).
    • Chemical and biological properties (cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility) determined through laboratory assays adhering to GLP (Good Laboratory Practices) and ISO standards.
    • Functionality in simulated clinical use benchtop models.

    8. The Sample Size for the Training Set

    Not applicable. This refers to a training set for an AI model. This device is a mechanical medical device, not an AI/ML product.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As there is no AI training set, there is no ground truth establishment for it. The "ground truth" for the device's design and manufacturing is based on established engineering principles, materials science, and medical device regulations.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1