Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K172236

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-10-30

    (97 days)

    Product Code
    Regulation Number
    870.4875
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pantheris System is intended to remove plaque (atherectomy) from partially occluded vessels in the peripheral vasculature with a reference diameter of 3.0 mm, using OCT-assisted orientation and imaging. The system is an adjunct to fluoroscopy by providing images of vessel lumen, wall structures, and vessel morphologies. The Pantheris System is NOT intended for use in the iliac, coronary, cerebral, renal or carotid vasculature.

    Device Description

    The Pantheris System combines the use of Avinger's optical coherence tomography (OCT) technology with peripheral vascular atherectomy capabilities. The Pantheris System consists of the Pantheris catheter, Lightbox Sled with integrated Umbilical (referred to as "Sled") and the Lightbox HS Imaging Console (referred to as "Lightbox"). The Pantheris Catheter comes in both a 7Fr and an 8Fr size. Both sizes have a working length of 110 cm and are sterile, single-use devices that are compatible with 7 and 8F sheaths (respectively) and 0.014" guidewires. The Pantheris Catheter incorporates an Optical Fiber that allows for real-time OCT guided directional atherectomy during the procedure. The Pantheris catheter is connected to the Lightbox via the Sled. The Lightbox is an optical transceiver, transmitting light to the intraluminal environment through the optical fiber on the Pantheris Catheter and receiving and interpreting the signal from the tissue using a PCbased processing system. The Lightbox provides a visualization platform for the real-time OCT-assisted directional atherectomy. The Lightbox consists of a cart with two monitors, a PC-based processing system, software, an isolation transformer, and an OCT system.

    AI/ML Overview

    The provided document is a 510(k) Summary for the Pantheris System, which describes modifications to an already cleared device. This submission focuses on demonstrating that these modifications do not impact the safety and effectiveness of the device compared to its predicate. Therefore, the "acceptance criteria" and "device performance" primarily revolve around showing equivalence to the predicate and passing various safety and performance tests.

    Here's the information extracted and organized as requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance
    BiocompatibilityNo changes in patient-contacting materials; accessories (tweezers) without direct patient contact must pass: Cytotoxicity, Sensitization, Irritation, Systemic Toxicity.Patient-contacting materials unchanged. Stainless steel tweezers (accessory) passed Cytotoxicity, Sensitization, Irritation, and Systemic Toxicity tests.
    Electrical SafetyCompliance with IEC 60601-1 standard for safety.Complies with IEC 60601-1.
    Electromagnetic Compatibility (EMC)Compliance with IEC 60601-1-2 standard for EMC.Complies with IEC 60601-1-2.
    SoftwareVerification and validation testing, including regression testing, conducted per FDA Guidance for Medical Device Software (Moderate Level of Concern).Software V&V and regression testing conducted, documentation provided per FDA guidance. Device software determined to be "moderate" level of concern.
    Mechanical TestingDevice modifications must not negatively impact mechanical performance. Tests include: Sheath insertion cycle, Destructive balloon burst, Torque capacity, Leak evaluation, Bond joint tensile strength, Life-cycle tests (8-month accelerated shelf life, simulated use), Corrosion, Heat generation, Rotational speed, Dimensional verification, Balloon inflation/deflation cycle, Plaque removal efficiency, Flushing tool capacity.All listed mechanical tests were performed. (Specific pass/fail results for each test are not detailed in the summary, but the conclusion states "these modifications do not impact the safety and effectiveness").
    Plaque Removal EfficiencyMaintain equivalent plaque removal efficiency to the predicate device.Testing for "Plaque removal efficiency" was performed.
    Clinical EquivalenceThe modifications should not raise new questions of safety and effectiveness, maintaining substantial equivalence to the predicate (K163264)."The information submitted... confirms that the minor modifications... raise no new questions of safety and effectiveness and that the device is substantially equivalent to the predicate device."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify exact sample sizes for each mechanical or biocompatibility test. It mentions that "Testing was performed to demonstrate that these changes do not impact the safety and effectiveness of the final device."

    • Test Set Sample Sizes: Not explicitly stated for each test.
    • Data Provenance: The studies are described as "Performance Data" from various tests (Biocompatibility, Electrical Safety, EMC, Software, Mechanical). These are laboratory-based, non-clinical tests conducted by the manufacturer to demonstrate compliance with standards and equivalence. There is no mention of country of origin for the data or if it's retrospective or prospective, as these are device performance tests, not clinical studies on patients.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable (N/A) to this 510(k) submission. The submission describes a device with mechanical, electrical, and software components, not an AI or diagnostic application requiring expert review for ground truth in a test set. The performance tests (e.g., destructive balloon burst, torque capacity) rely on engineering specifications and physical measurements, not expert consensus on medical images or clinical outcomes.

    4. Adjudication Method for the Test Set

    This information is not applicable (N/A) for the reasons stated above. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or studies involving human readers interpreting medical data, not for engineering performance tests of a physical device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done, and this type of study is not applicable (N/A) to this 510(k) submission. This submission is for modifications to an atherectomy system that includes OCT imaging as an adjunctive visualization tool, not an AI-powered diagnostic system requiring human reader performance evaluation.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable (N/A). While the device contains software and OCT imaging, it is an interventional atherectomy system. The OCT provides real-time, surgeon-guided visualization, it's not a standalone diagnostic algorithm that would typically undergo "algorithm-only" performance testing in the context of a 510(k) for an AI device. The software update specifically mentioned is an attenuation of the laser when the drive is not rotating, a safety feature, not an AI algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the tests performed is based on engineering specifications, established physical properties, and compliance with recognized standards. For example:

    • Biocompatibility: Adherence to ISO standards for material compatibility.
    • Electrical/EMC Safety: Compliance with IEC standards (e.g., IEC 60601-1).
    • Mechanical Testing: Meeting specified tensile strengths, torque capacities, leak rates, burst pressures, cycle counts, dimensional accuracies, and plaque removal efficiency based on defined test methods.
    • Software Validation: Successful execution of test cases against documented requirements and expected outcomes.

    There is no mention of expert consensus, pathology, or outcomes data being used as ground truth for these device modification performance tests.

    8. The Sample Size for the Training Set

    This information is not applicable (N/A) as the device functionality described does not involve a "training set" in the context of machine learning or AI algorithms requiring such a dataset. The software update mentioned is a safety feature for laser attenuation, not a complex algorithm dependent on a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable (N/A) for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163264

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-12-21

    (30 days)

    Product Code
    Regulation Number
    870.4875
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pantheris System is intended to remove plaque (atherectomy) from partially occluded vessels in the peripheral vasculature with a reference diameter of 3.0mm to 7.0mm, using OCT-assisted orientation and imaging. The system as an adjunct to fluoroscopy by providing images of vessel lumen, wall structures and vessel morphologies. The Pantheris System is NOT intended for use in the iliac, coronary, cerebral, renal or carotid vasculature.

    Device Description

    The Pantheris System (cleared under K162326) combines the use of Avinger's Optical Coherence Tomography (OCT) technology with peripheral vascular atherectomy capabilities. The Pantheris System consists of the Pantheris Catheter, Lightbox Sled with integrated Umbilical (referred to as "Sled") and the Lightbox HS Imaging Console (referred to as "Lightbox"). The Pantheris Catheter is connected to the Lightbox via the Sled. The Lightbox is an optical transceiver, transmitting light to the intraluminal environment through the optical fiber on the Pantheris Catheter and receiving and interpreting the signal from the tissue using a PC-based processing system. The Lightbox provides a visualization platform for the real-time OCT-assisted directional atherectomy. The Lightbox consists of a cart with two monitors; a PC based processing system, software, an isolation transformer and an OCT system. This Special 510(k) introduces an optional vessel measurement feature to the Pantheris System software in the Lightbox HS Imaging Console.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called the Pantheris System, specifically focusing on a software modification that introduces an optional vessel measurement feature. It's important to note that this document is not a clinical study report but rather a regulatory submission demonstrating substantial equivalence to a previously cleared device. Therefore, it does not contain extensive clinical trial data or detailed performance metrics as one might find in a full clinical study publication. However, it does discuss verification and accuracy testing for the new feature.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document states that "the accuracy of the measurement feature was assessed and found to be within the specified acceptance criteria." However, it does not explicitly list the acceptance criteria or the specific reported device performance values for the measurement feature. It only broadly states they were met.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document refers to "software verification and measurement accuracy testing." This testing appears to be non-clinical in nature, as stated in the section "Non-Clinical Test Data." It does not mention a specific test set of patient data, but rather testing of the software's ability to measure simulated or in-vitro data. There is no information provided on the sample size for this test set, nor on data provenance (country of origin, retrospective/prospective), as it's not a clinical study involving patients.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    Given that the testing was described as "non-clinical" and focused on software verification and measurement accuracy, it's highly unlikely that human experts were used to establish ground truth in the traditional sense of clinical image interpretation. The ground truth would likely be established by known physical dimensions in simulated or in-vitro models. Therefore, this information is not applicable in the context of this document.

    4. Adjudication Method for the Test Set:

    As no human experts or patient data are described for the test set, an adjudication method is not applicable.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size:

    The document does not describe an MRMC comparative effectiveness study, nor does it mention any human reader performance or improvements with AI assistance. The modification is described as providing "an optional confirmation of their initial vessel diameter/size assessment" and "additional information." It's not presented as an AI-powered diagnostic tool that directly aids human readers in their primary diagnosis.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    The document describes "measurement accuracy testing" for the new vessel measurement feature as part of "software verification." This implies a standalone performance evaluation of the algorithm's measurement accuracy against known values (ground truth) in a non-clinical setting. The results are summarized as "found to be within the specified acceptance criteria," though the criteria and specific numbers are not provided.

    7. The Type of Ground Truth Used:

    For the "measurement accuracy testing," the ground truth would have been established by known physical dimensions or precisely measured values in the non-clinical test environment (e.g., phantoms, simulated vessels, or in-vitro models). The document doesn't explicitly state the type of ground truth but implies it's a quantitative comparison to a known standard.

    8. The Sample Size for the Training Set:

    This document describes a software modification to an existing device, not a new algorithm developed from scratch. Therefore, it does not provide information on a training set sample size. The focus is on verifying and validating the new feature's performance, not on a machine learning model's training process.

    9. How the Ground Truth for the Training Set Was Established:

    As there is no mention of a training set, there is no information on how its ground truth would have been established.

    In summary:

    This 510(k) submission addresses a minor software modification (addition of a vessel measurement feature) to an already cleared device. The "study" described is a non-clinical software verification and measurement accuracy testing. It demonstrates that the new feature meets its internal specifications and acceptance criteria, without raising new safety or effectiveness concerns. It is not a clinical trial and thus lacks the detailed clinical performance metrics (e.g., sensitivity, specificity, reader performance) typically associated with such studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162326

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-10-18

    (60 days)

    Product Code
    Regulation Number
    870.4875
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pantheris System is intended to remove plaque (atherectomy) from partially occluded vessels in the peripheral vasculature with a reference diameter of 3.0mm to 7.0mm, using OCT-assisted orientation and imaging. The system is an adjunct to fluoroscopy by providing images of vessel lumen, wall structures and vessel morphologies. The Pantheris System is NOT intended for use in the iliac, coronary, cerebral, renal or carotid vasculature.

    Device Description

    The current Pantheris System (K160827), listed as the primary predicate device, consists of the Pantheris Catheter, Lightbox Sled with integrated Umbilical (referred to as "Sled") the Lightbox HS Imaging Console (referred to as "Lightbox") and the accessories that are packaged with the Pantheris Catheter: flush fixture, tweezers, syringes and stopcocks, and a Sterile Drape accessory that is packaged separately. The subject Pantheris System submitted as part of this 510(k) submission does not introduce any changes to the device specifications, materials or manufacturing processes and is identical to the current Pantheris System.

    The Pantheris System combines the use of Avinger's Optical Coherence Tomography (OCT) technology (identical to the OCT Technology used in the Ocelot System (K140185)) with peripheral vascular atherectomy capabilities.

    The Pantheris Catheter comes in both a 7Fr and 8Fr size. Both sizes have a working length of 110 cm and are sterile, single-use devices that are compatible with 7Fr and 8Fr sheaths (respectively) and 0.014" guidewires. The Pantheris Catheter incorporates an Optical Fiber that allows for real-time OCT guided directional atherectomy during the procedure.

    The Pantheris Catheter is connected to the Lightbox via the Sled. The Lightbox is an optical transceiver, transmitting light to the intraluminal environment through the optical fiber on the Pantheris Catheter and receiving and interpreting the signal from the tissue using a PCbased processing system. The Lightbox provides a visualization platform for the real-time OCT-assisted directional atherectomy. The Lightbox consists of a cart with two monitors; a PC based processing system, an isolation transformer and an OCT system.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the Avinger Pantheris System. It states that the "subject Pantheris System submitted as part of this 510(k) submission does not introduce any changes to the device specifications, materials or manufacturing processes and is identical to the current Pantheris System (K160827)." Therefore, the document primarily relies on previous submissions (K160827 and K152275) for evidence of safety and effectiveness, including clinical testing.

    As a result, much of the requested information regarding detailed acceptance criteria, specific device performance metrics, sample sizes for test and training sets, and expert details for ground truth establishment is not directly present in this specific 510(k) submission. This document asserts substantial equivalence based on the lack of changes to the device and the use of the same technology as previously cleared devices.

    However, I can extract information about the device's intended use and the general assertion of meeting specifications based on prior testing.


    1. Table of acceptance criteria and the reported device performance:

    Specific numerical acceptance criteria and reported device performance (e.g., sensitivity, specificity, accuracy, or other quantifiable metrics) for the OCT-assisted imaging aspect of the Pantheris System are not provided in this document. The document primarily focuses on demonstrating substantial equivalence by stating the device is identical to a prior cleared device and uses the same OCT technology.

    The "acceptance criteria" can be inferred as successful completion of the non-clinical tests and the established specifications referenced in previous submissions (K160827 and K152275).

    Acceptance Criteria (Inferred from non-clinical testing)Reported Device Performance (Summary from Section 5.0)
    Design verification and bench validation studies met established specifications."The collective results of the non-clinical testing demonstrate that the subject Pantheris System meets the established specifications necessary for consistent performance for its intended use."
    Packaging and shelf-life validated.(No specific performance reported in this document, but implied successful validation.)
    Software verification and validation completed.(No specific performance reported in this document, but implied successful validation.)
    Electrical safety, electromagnetic compatibility, and laser safety testing passed.(No specific performance reported in this document, but implied successful passing.)
    Biocompatibility demonstrated.(No specific performance reported in this document, but implied successful demonstration.)
    In-vivo animal validation study successfully conducted.(No specific performance reported in this document, but implied successful validation.)
    Sterilization validated.(No specific performance reported in this document, but implied successful validation.)
    OCT image comparisons confirmed substantial equivalence of imaging functionality."Use of the same OCT technology between the subject and the predicate devices, along with successful OCT image comparisons between the subject Pantheris System, the Ocelot System (K140185) and images from the submitted compendium, confirm a determination of substantial equivalence between the subject device and the predicate devices regarding the updates to the subject Pantheris System labeling."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • For the non-clinical studies: The document refers to "Design verification and bench validation studies," "Packaging and shelf-life," "Software verification and validation," "Electrical safety, electromagnetic compatibility, and laser safety testing," "Biocompatibility," "In-vivo animal validation study," and "Sterilization." The specific sample sizes for these tests are not detailed in this document, as it relies on prior submissions. The provenance is not specified.

    • For clinical testing: The document explicitly states: "Clinical testing of the Pantheris System was included for the secondary predicate, the initial Pantheris System original 510(k) submission (K152275). Due to the equivalence between the Systems no new clinical testing is included in this 510(k) submission." Therefore, details of sample size and provenance for clinical testing would need to be found in the K152275 submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This information is not provided in this document. It would likely be contained within the clinical study reports referenced in the K152275 submission if such studies established ground truth for performance metrics. The "OCT image comparisons" mentioned imply expert review, but details are not given.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not provided in this document. It would depend on the design of any clinical studies for K152275.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    The document describes the Pantheris System as an atherectomy device that uses OCT-assisted orientation and imaging, an "adjunct to fluoroscopy by providing images of vessel lumen, wall structures and vessel morphologies." It's a device with an imaging component, not primarily an AI-driven diagnostic tool that assists human readers in interpreting images for improved accuracy like in a typical MRMC study setting for AI comparison. Therefore, an MRMC comparative effectiveness study in the context of "human readers improve with AI vs without AI assistance" is not explicitly mentioned or implied as the primary mode of effectiveness testing for this historical device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The device is described as "OCT-assisted orientation and imaging" and an "adjunct to fluoroscopy." This implies that the imaging component is an integral part of a human-in-the-loop interventional procedure, not a standalone diagnostic algorithm. Therefore, "standalone (algorithm only)" performance in the typical sense of AI diagnostics is not applicable or discussed for this device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    This specific detail for the clinical testing (referenced from K152275) is not provided in this document. However, for a device intended to remove plaque and provide images of vessel morphology, ground truth might typically involve histopathology (pathology) from excised tissue, angiography, or other validated imaging modalities.

    8. The sample size for the training set:

    The document makes no mention of a "training set" in the context of an AI/machine learning algorithm. The OCT technology is described as "using a PC-based processing system" to interpret signals, but it does not specify if this involves a machine learning model that requires a distinct training set. If the "PC-based processing system" contains a machine learning component, information about its training set is not included in this document.

    9. How the ground truth for the training set was established:

    As no "training set" is discussed in this document, this information is not provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160827

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-04-21

    (27 days)

    Product Code
    Regulation Number
    870.4875
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pantheris System is intended to remove plaque (atherectomy) from partially occluded vessels in the peripheral vasculature with a reference diameter of 3.0mm to 7.0 mm, using OCT-assisted orientation as an adjunct to fluoroscopy. The Pantheris System is NOT intended for use in the iliac, coronary, cerebral, renal or carotid vasculature.

    Device Description

    The Pantheris Catheter received clearance under K153460 to remove plaque (atherectomy) from partially occluded vessels in the peripheral vasculature using OCTassisted orientation. This Special 510(k) submission is intended to add Tweezers and two 3-way Stopcocks to the Pantheris Catheter packaging. The Tweezers are provided to assist in the extraction of tissue that is collected during the atherectomy procedure from the distal end of the catheter once the catheter is removed from the patient. The catheter can then be re-inserted into the vessel to continue plaque removal using the same technique. The Tweezers are not intended to come into direct contact with the patient.

    The Stopcocks are standard, off the shelf accessories, used to facilitate the delivery of CO2 and saline to the catheter. The Syringes are connected to the Flush and Balloon Inflation Lumens (components on the catheter's handle assembly) via the Stopcocks. The Stopcocks are not intended to come into direct contact with the patient.

    Both the Tweezers and the Stopcocks are provided in order to make readily available these commonly used accessories and for user's convenience.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Pantheris Catheter, focusing on acceptance criteria and study details.

    Based on the provided document, the submission (K160827) is a Special 510(k) for adding accessories (Tweezers and 3-way Stopcocks) to an already cleared device (Pantheris Catheter, K153460). Therefore, the "device" in question for this specific submission is primarily the addition of these accessories and their impact on the overall system's safety and effectiveness, rather than a fundamental re-evaluation of the core Pantheris Catheter's performance.

    The document explicitly states: "There are no modifications to the indications for use, intended use, material, design, functionality or performance of the catheter, flush fixture and syringes cleared under K153460." This means the performance of the Pantheris Catheter itself was established in a prior submission (K153460), and this current submission focuses on ensuring the accessories do not negatively impact safety or effectiveness.

    Given this context, the acceptance criteria and supporting studies are framed around the accessories, not the catheter's primary function of plaque removal.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    BiocompatibilityTweezers and Stopcocks verified per ISO 10993-1:2009 and related collateral standards for patient contacting materials.
    Packaging and Shelf-life ValidationPerformed to test the packaging tray configuration in accordance with ISO 11607-1:2006 and ISO 11607-2:2006.
    Sterilization Assurance Level (SAL)Maintained a SAL of 10^-6 using Electron Beam sterilization per Method VDMax25, in accordance with ISO 11137-1:2006 and ISO 11137-2:2006.
    No new questions of safety and effectivenessThe non-clinical testing collectively demonstrated that the Pantheris Catheter (with new accessories) meets established specifications, and the accessories raise no new questions of safety and effectiveness.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the traditional sense of patient data or clinical trials for evaluating the performance of the atherectomy aspect of the device. Instead, the testing described is focused on the non-clinical performance of the new accessories and their impact on the existing system.

    • Sample Size for Test Set: Not applicable or not specified in terms of patient data. The non-clinical tests (biocompatibility, packaging, sterilization) would involve specific numbers of samples/units for each test, but these are not detailed in the summary.
    • Data Provenance: Not applicable for clinical data. The tests are non-clinical (laboratory/bench testing) of the device components.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not applicable. The ground truth for biocompatibility, packaging, and sterilization standards is defined by international standards (ISO series) and accepted regulatory practices, not by expert consensus on clinical data in this context.

    4. Adjudication Method for the Test Set

    Not applicable. There's no indication of clinical data requiring expert adjudication in this submission. The non-clinical tests rely on meeting predefined specifications and standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance.

    Not applicable. This device is an atherectomy catheter, not an AI-powered diagnostic device. The submission does not mention any AI component or MRMC study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done.

    Not applicable. This device is not an algorithm.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for this submission's acceptance criteria primarily relies on:

    • International Standards and Regulations: ISO 10993-1, ISO 11607-1, ISO 11607-2, ISO 11137-1, ISO 11137-2.
    • Design Specifications: The device (with accessories) meeting its established performance specifications.
    • Predicate Device Equivalence: The primary "ground truth" for the overall system's safety and effectiveness is its substantial equivalence to the previously cleared Pantheris Catheter (K153460), as no changes were made to the core functionality.

    8. The Sample Size for the Training Set

    Not applicable. There is no mention of a "training set" as this is not an AI/machine learning device.

    9. How the Ground Truth for the Training Set was Established

    Not applicable.


    Summary of the Document's Focus:

    This FDA 510(k) submission (K160827) is a "Special 510(k)" for minor modifications (adding accessories) to an already cleared medical device. The key takeaway is that the safety and effectiveness of the original Pantheris Catheter for its intended use were established previously. This submission's purpose is to demonstrate that the addition of Tweezers and Stopcocks does not introduce new risks or modify the fundamental performance of the catheter. Therefore, the "acceptance criteria" and "studies" are focused on standard non-clinical evaluations related to these new components' biocompatibility, packaging, and sterilization, ensuring they do not compromise the system's previous clearance basis. The document explicitly states: "The performance testing conducted establishes that the addition of Tweezers and Stopcocks does not raise new questions of the safety and efficacy for the Pantheris Catheter cleared under K153460."

    Ask a Question

    Ask a specific question about this device

    K Number
    K153460

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-03-01

    (91 days)

    Product Code
    Regulation Number
    870.4875
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pantheris System is intended to remove plaque (atherectomy) from partially occluded vessels in the peripheral vasculature with a reference diameter of 3.0mm to 7.0 mm, using OCT-assisted orientation as an adjunct to fluoroscopy. The Pantheris System is NOT intended for use in the iliac, coronary, cerebral, renal or carotid vasculature.

    Device Description

    The Pantheris System consists of the Pantheris Catheter, Lightbox Sled (referred to as Sled), the Lightbox HS Imaging Console and the Sterile Drape (accessory).

    AI/ML Overview

    The Pantheris System and its modified version (K153460) are intended to remove plaque from partially occluded peripheral vasculature. The device uses OCT-assisted orientation as an adjunct to fluoroscopy.

    Here's an analysis of the provided information regarding acceptance criteria and the study that proves the device meets those criteria:

    1. Table of Acceptance Criteria and the Reported Device Performance

    The document does not explicitly present a "table of acceptance criteria" with corresponding "reported device performance" in a quantitative manner that would typically be seen for diagnostic device accuracy (e.g., sensitivity, specificity, AUC). Instead, the studies performed are primarily focused on demonstrating the safety and effective function of the device's modifications and ensuring substantial equivalence to a previously cleared predicate device.

    The acceptance criteria for this device seem to be implicitly defined by the successful completion of various non-clinical engineering and biological tests, and the modified device's performance matching or exceeding that of its predicate. The "reported device performance" is qualitative and indicates "meets established specifications necessary for consistent performance for its intended use."

    The identified non-clinical tests and their implicit acceptance criteria are:

    Acceptance Criteria (Implied by Test Name)Reported Device Performance (Qualitative)
    Design Verification
    Working LengthMeets specifications
    Catheter FlushMeets specifications
    OCT Image Generation and Sled InterfaceMeets specifications
    Catheter Field of ViewMeets specifications
    Distal Tip RotationMeets specifications
    Guidewire compatibilityMeets specifications
    Distal Tip Max ODMeets specifications
    Insertion ForceMeets specifications
    Force to Overcome User Slide LockMeets specifications
    Force to Remove Sled BagMeets specifications
    Force to Remove Sled Bag ringMeets specifications
    Full 360° ImageMeets specifications
    Inflation CyclesMeets specifications
    Cut/Pack CyclesMeets specifications
    Packed Position Life CycleMeets specifications
    Active position Life CycleMeets specifications
    Balloon BurstMeets specifications
    Torque Shaft Torque Proof loadingMeets specifications
    Drive Shaft Torque Proof loadingMeets specifications
    Inflation Luer Tensile StrengthMeets specifications
    Flush Lumen Tensile StrengthMeets specifications
    Distal Catheter Joints Tensile StrengthMeets specifications
    Proximal Catheter Joints Tensile StrengthMeets specifications
    Simulated Use (Design Validation)
    Setup (Ease of Use & Sterility)Meets specifications
    Functionality (Visualization, Flushing, Device Tracking, Target Sites, Treatment, Markings, Safety)Meets specifications
    Insertion/Retraction forceMeets specifications
    Number of InsertionsMeets specifications
    Post-Retraction (Visualization, Device Tracking, Treatment, Markings, Safety)Meets specifications
    Sterilization ValidationMeets specifications
    Packaging ValidationMeets specifications
    Product Shelf LifeMeets specifications
    BiocompatibilityMeets specifications
    CytotoxicityMeets specifications
    SensitizationMeets specifications
    Intracutaneous ReactivityMeets specifications
    Systemic ToxicityMeets specifications
    Hemolysis extractMeets specifications
    Hemolysis directMeets specifications
    Complement ActivationMeets specifications
    Material Mediated (Rabbit) PyrogenicityMeets specifications
    Preclinical Animal Testing
    Porcine Study (acute and chronic histopathological assessments, in vivo performance comparison)Acceptable histopathological assessments; in vivo performance deemed comparable to predicate
    Functionality (Radiopacity, OCT marker visibility, Insertion/Retraction, Advance to target location, Ability to orient device to a landmark)Demonstrated
    Minimum Effective Flow Rate (MEFR) for clear OCT imagingMeets specifications
    ThrombogenicityMeets specifications
    Safety AssessmentAcceptable safety profile

    2. Sample Size Used for the Test Set and the Data Provenance

    The document mentions a "Porcine Study" as a preclinical animal test. The sample size for this study is not explicitly stated, only that it involved "acute and chronic histopathological assessments." The data provenance is from an animal model (porcine), and it is a prospective study design to evaluate the modified device. No human test set data is provided or referenced in this 510(k) summary for this specific modification. The study is a "comparison to a predicate device."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    For the preclinical porcine study, the documentation refers to "histopathological assessments." This implies that experts (likely veterinary pathologists) were involved in establishing the ground truth regarding tissue response and safety. However, the exact number of experts and their specific qualifications (e.g., "veterinary pathologist with X years of experience") are not provided in this document.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for the test set, as it does not present a clinical study with human outcomes that would typically require independent expert adjudication. For the animal study, the pathological assessments would likely follow standard practices for veterinary pathology reporting, but no specific adjudication process (e.g., 2+1, 3+1 consensus) is detailed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study, nor is there any discussion of AI assistance or human reader improvement. The Pantheris System (and its modification) is a medical device for atherectomy, and while it uses "OCT-assisted orientation," this is not presented as an AI-driven diagnostic aid for human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This document describes a medical device rather than an algorithm. Its performance is assessed through engineering tests, biocompatibility, and preclinical animal studies. Therefore, the concept of "standalone algorithm performance" is not applicable here. The device's OCT imaging component is an "adjunct to fluoroscopy" and is part of a human-operated system for guided intervention.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the preclinical animal study, the ground truth was primarily established through pathology (histopathological assessments). The functional aspects of the device in vivo were also evaluated. For the extensive non-clinical and design validation testing, the ground truth was based on pre-defined engineering specifications and performance standards.

    8. The Sample Size for the Training Set

    This document does not describe the development of an artificial intelligence or machine learning algorithm. Therefore, the concept of a "training set" in that context is not applicable. The device's design and modifications are based on engineering principles and iterative design, not on machine learning training data.

    9. How the Ground Truth for the Training Set was Established

    As there is no mention of an AI/ML algorithm or a training set in this document, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1