Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K172365

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-04-19

    (258 days)

    Product Code
    Regulation Number
    868.1400
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Panoramic Oxygen Mask (POM) is a single patient, disposable device intended for delivering supplemental oxygen and monitoring expired gases from the patient, with ports to allow the clinician to insert scopes, probes, or tubes. It is for non-intubated, spontaneously breathing patients greater than 30 kg.

    Device Description

    The Panoramic Oxygen Mask (POM) is a multi-port mask that serves several functions:

    • A standard oxygen mask for when a patient requires supplemental oxygen
    • Sampling of exhaled gases for monitoring, typically end-tidal CO2
    • Additional ports (membranes) to allow for most types of scopes, probes, and tubes to be inserted while still delivering supplemental O2 and sampling exhaled gases.
      The design of the POM's oral or nasal membranes allow access of a scope and are soft and pliable to help maintain the oxygen concentration to the patient while having scopes, etc. inserted through these oral or nasal membranes.
    AI/ML Overview

    The Panoramic Oxygen Mask (POM) is a single-patient, disposable device intended for delivering supplemental oxygen and monitoring expired gases from the patient, with ports to allow clinicians to insert scopes, probes, or tubes. It is designed for non-intubated, spontaneously breathing patients greater than 30 kg.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the POM device are primarily based on demonstrating substantial equivalence to a predicate device (K133806 – Monitored Mask, M1 Capnography mask) in terms of performance for CO2 accuracy, respiration rate (RR), and FiO2 delivery, across various simulated patient settings, as well as meeting biocompatibility, mechanical durability, and shelf-life standards.

    Acceptance Criteria CategorySpecific Criteria/Test DescriptionPOM Device PerformancePredicate Device Performance
    CO2 Accuracy & RREvaluation of EtCO2 and Respiration Rate at various simulated patient settings (adult and pediatric).Adult:BPM 12, TV 500ml, O2 Flow 1 Lpm, 5% CO2: EtCO2 7.37% (Closed)/6.57% (Inserted), RR 12/12BPM 12, TV 500ml, O2 Flow 8 Lpm, 5% CO2: EtCO2 5.17% (Closed)/4.78% (Inserted), RR 12/12BPM 12, TV 500ml, O2 Flow 1 Lpm, 1% CO2: EtCO2 1.40%, RR 12BPM 12, TV 500ml, O2 Flow 8 Lpm, 1% CO2: EtCO2 1.00%, RR 12BPM 20, TV 300ml, O2 Flow 1 Lpm, 5% CO2: EtCO2 8.23% (Closed)/7.07% (Inserted), RR 20/20BPM 20, TV 300ml, O2 Flow 8 Lpm, 5% CO2: EtCO2 6.17% (Closed)/4.40% (Inserted), RR 20/20BPM 20, TV 300ml, O2 Flow 1 Lpm, 1% CO2: EtCO2 1.63%, RR 20BPM 20, TV 300ml, O2 Flow 8 Lpm, 1% CO2: EtCO2 1.13%, RR 20Pediatric:BPM 12, TV 500ml, O2 Flow 1 Lpm, 5% CO2: EtCO2 6.40% (Closed)/6.13% (Inserted), RR 12/12BPM 12, TV 500ml, O2 Flow 8 Lpm, 5% CO2: EtCO2 3.40% (Closed)/3.17% (Inserted), RR 12/12BPM 12, TV 500ml, O2 Flow 1 Lpm, 1% CO2: EtCO2 1.10%, RR 12BPM 12, TV 500ml, O2 Flow 8 Lpm, 1% CO2: EtCO2 0.73%, RR 12BPM 20, TV 300ml, O2 Flow 1 Lpm, 5% CO2: EtCO2 6.40% (Closed)/6.23% (Inserted), RR 20/20BPM 20, TV 300ml, O2 Flow 8 Lpm, 5% CO2: EtCO2 3.10% (Closed)/2.23% (Inserted), RR 20/20BPM 20, TV 300ml, O2 Flow 1 Lpm, 1% CO2: EtCO2 1.23%, RR 20BPM 20, TV 300ml, O2 Flow 8 Lpm, 1% CO2: EtCO2 0.70%, RR 20Adult:BPM 12, TV 500ml, O2 Flow 1 Lpm, 5% CO2: EtCO2 6.43%, RR 12BPM 12, TV 500ml, O2 Flow 8 Lpm, 5% CO2: EtCO2 2.50%, RR 12BPM 12, TV 500ml, O2 Flow 1 Lpm, 1% CO2: EtCO2 1.20%, RR 12BPM 12, TV 500ml, O2 Flow 8 Lpm, 1% CO2: EtCO2 0.47%, RR 12BPM 20, TV 300ml, O2 Flow 1 Lpm, 5% CO2: EtCO2 6.77%, RR 20BPM 20, TV 300ml, O2 Flow 8 Lpm, 5% CO2: EtCO2 3.37%, RR 20BPM 20, TV 300ml, O2 Flow 1 Lpm, 1% CO2: EtCO2 1.27%, RR 20BPM 20, TV 300ml, O2 Flow 8 Lpm, 1% CO2: EtCO2 0.60%, RR 20Pediatric:BPM 12, TV 500ml, O2 Flow 1 Lpm, 5% CO2: EtCO2 6.17%, RR 12BPM 12, TV 500ml, O2 Flow 8 Lpm, 5% CO2: EtCO2 4.07%, RR 12BPM 12, TV 500ml, O2 Flow 1 Lpm, 1% CO2: EtCO2 1.07%, RR 12BPM 12, TV 500ml, O2 Flow 8 Lpm, 1% CO2: EtCO2 0.77%, RR 12BPM 20, TV 300ml, O2 Flow 1 Lpm, 5% CO2: EtCO2 6.03%, RR 20BPM 20, TV 300ml, O2 Flow 8 Lpm, 5% CO2: EtCO2 3.97%, RR 20BPM 20, TV 300ml, O2 Flow 1 Lpm, 1% CO2: EtCO2 1.17%, RR 20BPM 20, TV 300ml, O2 Flow 8 Lpm, 1% CO2: EtCO2 0.77%, RR 20
    FiO2 DeliveryEvaluation of FiO2 at various simulated patient settings (adult and pediatric).Adult:BPM 12, TV 400ml, O2 Flow 8 Lpm, 4% CO2: 77.6%BPM 12, TV 400ml, O2 Flow 10 Lpm, 4% CO2: 83.4%BPM 12, TV 400ml, O2 Flow 12 Lpm, 4% CO2: 87.0%Pediatric:BPM 12, TV 500ml, O2 Flow 1 Lpm, 5% CO2: 25.0% (Closed)/24.33% (Inserted)BPM 12, TV 500ml, O2 Flow 8 Lpm, 5% CO2: 48.67% (Closed)/49.0% (Inserted)BPM 20, TV 300ml, O2 Flow 1 Lpm, 5% CO2: 23.67% (Closed)/24.67% (Inserted)BPM 20, TV 300ml, O2 Flow 8 Lpm, 5% CO2: 56.0% (Closed)/56.33% (Inserted)Not provided for predicate in summary.
    BiocompatibilityCompliance with ISO 10993 for limited duration (< 24 hrs) external communicating/surface contact.Cytotoxicity (ISO 10993-5:2009), Sensitization (ISO 10993-10:2010), Intracutaneous/Irritation (ISO 10993-10:2010), Acute Systemic Toxicity (ISO 10993-11:2017) – All results acceptable.Similar testing conducted for predicate.
    Mechanical IntegrityDrop testPerformed, results indicate no adverse effect on performance.Not explicitly detailed in the comparative table for the predicate, but implied in general device safety.
    Shelf-Life3 years real-time aging and shelf-life.Testing supports the claim, no effects of aging on performance. Storage: -20° to +50°C.Not specified for predicate.
    Luer ConnectorISO 594 Luer connectorTesting performed, results acceptable.Not explicitly detailed for predicate, but similar connection type (Luer lock) mentioned.
    Internal VolumeComparison of internal volume between new device and predicate.Child: 93 ml; Adult: 198 ml. The difference in internal volume does not affect performance of oxygen delivery or monitoring of exhaled gases.Child: 73 ml; Adult: 159 ml.
    Entrainment VentsDesign of entrainment vents to prevent rebreathing.One-way valves to prevent rebreathing. Without one-way valves, exhaled gas monitoring performance is reduced. For intended use, designs are similar. This does not raise different questions of safety or effectiveness.Open vents allowing room entrainment.
    Access PortsAllowance for introduction of instruments while delivering supplemental oxygen and sampling expired gases.Oral port: 60 mm ID / 20 mm OD; Nasal: 36 mm ID / 12 mm OD. Access port allows introduction of instruments while delivering supplemental oxygen and sampling expired gases. Impact on monitoring is characterized. Labeling ensures compatibility of scopes.Not available for predicate, but predicate's "CapnoVue Scope model" implies similar functionality.

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes non-clinical comparative performance testing under simulated patient settings. Thus, there are no human patient test sets.

    • Sample Size for Test Set: Not applicable as real patient data was not used. Instead, in-vitro bench testing was performed with simulated patient settings. Detailed quantitative results are provided for various scenarios (e.g., adult/pediatric, different BPM, Tidal Volumes, O2 flows, CO2 concentrations).
    • Data Provenance: The data is generated from bench testing under controlled, simulated conditions. This is not retrospective or prospective patient data, nor is it tied to a specific country of origin in terms of patient population.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The ground truth for the performance parameters (EtCO2, RR, FiO2) was established by the precise, controlled input parameters of the simulated breathing model (e.g., 5% CO2 gas mixture, set BPM). This is a physical or engineering ground truth, not one based on expert medical interpretation.

    4. Adjudication Method for the Test Set

    Not applicable. There was no human interpretation or adjudication required as the "ground truth" was the known input parameters of the test setup, and the device output was measured directly.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    No. A MRMC comparative effectiveness study was not done. The study was non-clinical bench testing comparing the device's performance to that of a predicate device in a controlled, simulated environment, not human readers with or without AI assistance.

    6. Standalone Performance

    Yes, a standalone (algorithm only without human-in-the-loop performance) study was done in the sense that the device's physical performance characteristics (e.g., CO2 accuracy, FiO2 delivery) were measured independently of human involvement during the testing itself. The data presented in Tables 2, 3, and 4 represent the performance of the device in these controlled bench environments.

    7. Type of Ground Truth Used

    The ground truth used was physical parameters/engineering ground truth. Specifically:

    • For CO2 accuracy and respiration rate: known gas concentrations (e.g., 5% CO2, 1% CO2) and programmed breathing rates (e.g., 12 BPM, 20 BPM) and tidal volumes (e.g., 500ml, 300ml) in a simulated lung model.
    • For FiO2: known oxygen flow rates and simulated breathing parameters.
    • For biocompatibility, mechanical testing, and shelf-life: adherence to established international standards (e.g., ISO 10993) and internal protocols.

    8. Sample Size for the Training Set

    Not applicable. This device is a physical medical device (mask), not an AI/ML algorithm that requires a training set. The term "training set" is typically used in the context of machine learning model development.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there was no training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1