Found 1 results

510(k) Data Aggregation

    K Number
    K961221

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1996-07-30

    (132 days)

    Product Code
    Regulation Number
    866.3110
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Wampole Pylori IgG ELISA assay may be used as an aid in the diagnosis of Helicobacter pylori infection in persons with gastrointestinal symptoms. For In Vitro Diagnostic Use Only.

    Device Description

    The Pylori IgG ELISA kit is an Enzyme-Linked Immunosorbent Assays (ELISA) for the qualitative determination of IgG antibodies in human serum to Helicobacter pylori antigen. The Pylori IgG ELISA test is an enzyme linked immunosorbent assay to detect IgG antibodies to Helicobacter pylori. Purified Helicobacter pylori antigen is attached to a solid phase microtiter well. Diluted test sera is added to each well. If the antibodies are present that recognize the antigen, they will bind to the antigen in the well. After incubation the wells are washed to remove unbound antibody. An enzyme labeled antihuman IgG is added to each well. If antibody is present it will bind to the antibody attached to the antigen on the well. After incubation the wells are washed to remove unbound conjugate. A substrate solution is added to each well. If enzyme is present the substrate will undergo a color change. After an incubation period the reaction is stopped and the color intensity is measured photometrically, producing an indirect measurement of specific antibody in the patient specimen.

    AI/ML Overview

    Here's an analysis of the provided information, structured according to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" as a set of predefined thresholds. Instead, it presents the performance characteristics of the Pylori IgG ELISA and the previously evaluated Pylori Stat, with the underlying implication that the Pylori IgG ELISA's performance is deemed acceptable relative to the Pylori Stat and biopsy.

    Based on the provided data, we can infer the achieved performance:

    Performance MetricPylori IgG ELISA (Reported)Pylori Stat (for comparison)
    Sensitivity96.4% (95% CI: 94.1%-98.8%)98.5% (95% CI: 97.0%-100%)
    Specificity96.1% (95% CI: 92.3%-99.9%)98.1% (95% CI: 95.5%-100%)
    Agreement96.4% (95% CI: 94.4%-98.3%)98.4% (95% CI: 97.1%-99.7%)
    Precision (CV - range for interassay)6.06% - 38.8% (for different sera)(Not provided for direct comparison)

    2. Sample Size Used for the Test Set and Data Provenance

    • Pylori IgG ELISA Test Set Sample Size: 371 serum samples
    • Pylori Stat (predicate device) Test Set Sample Size: 386 serum samples
    • Data Provenance (Both devices):
      • Country of Origin: Not explicitly stated, but mentioned as "five geographically different areas."
      • Retrospective or Prospective: Not explicitly stated. However, the use of "serum from patients" and existing biopsy results suggests a retrospective analysis of collected samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not provided in the document. The ground truth was established by "biopsy with stain or culture results for H. pylori." The number of pathologists/microbiologists interpreting these biopsies and their qualifications are not mentioned.

    4. Adjudication Method for the Test Set

    This information is not provided in the document. It only states that equivocal results were not included in the sensitivity and specificity calculations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    This is not applicable. The device is an immunoassay (ELISA kit) for detecting antibodies, not an AI-powered diagnostic tool that assists human readers. Therefore, an MRMC study related to AI assistance is not relevant to this submission.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone performance evaluation of the Pylori IgG ELISA was done. The device directly produces results (qualitative determination of IgG antibodies), and its performance metrics (sensitivity, specificity, agreement) were calculated independently against the biopsy ground truth. This is the inherent nature of an ELISA test.

    7. The Type of Ground Truth Used

    The ground truth used for performance evaluation was: Biopsy with stain or culture results for H. pylori. This combines histological and microbiological evidence.

    8. The Sample Size for the Training Set

    This information is not provided. The document describes the evaluation of the Pylori IgG ELISA and its predicate, Pylori Stat, which implies they are already developed products. There is no mention of a "training set" in the context of developing the ELISA assay, as it's a biochemical test, not a machine learning algorithm.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable/provided for the reasons stated in point 8.