Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K022177

    Validate with FDA (Live)

    Date Cleared
    2002-12-17

    (167 days)

    Product Code
    Regulation Number
    866.6010
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bayer ADVIA® IMS PSA assay is an in vitro diagnostic device intended to quantitatively measure prostate specific antigen (PSA) in human serum. This assay is indicated for the measurement of serum PSA as an aid in management (monitoring) of prostate cancer patients. PSA values obtained using the Bayer ADVIA IMS assay method must be interpreted in conjunction with all other available clinical and laboratory data before a medical decision is determined.

    Device Description

    Not Found

    AI/ML Overview

    The provided text describes the acceptance criteria and the study for the Bayer ADVIA® IMS PSA assay.

    Here's the information extracted and organized as requested:

    Acceptance Criteria and Device Performance for Bayer ADVIA® IMS PSA Assay

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly state pre-defined acceptance criteria in a table format, but rather reports performance characteristics. Based on the data presented and common analytical testing standards, the implicit acceptance criteria are inferred to be comparable or superior to the predicate device (Immuno 1 PSA Assay).

    Performance CharacteristicImplicit Acceptance Criteria (Inferred)Reported Device Performance (ADVIA IMS)
    Imprecision (Total CV%)Comparable to or better than predicate device Immuno 1. Generally, lower CV% is better.Level 2.6 ng/mL: 2.7% Level 6.5 ng/mL: 1.8% Level 24.1 ng/mL: 1.7% Level 99.1 ng/mL: 2.0% (Predicate Immuno 1: Level 3 ng/mL: 3.1%, Level 23 ng/mL: 3.4%, Level 96 ng/mL: 2.5%)
    Correlation (R-value)R-value close to 1 (indicating strong linear correlation with predicate).Serum (N=54): R = 0.999 Serum (N=44): R = 0.999
    Regression SlopeClose to 1.0 (indicating similar proportionality to predicate).Serum (N=54): 0.97 Serum (N=44): 0.97 95% Confidence Interval (Slope): 0.951 - 0.981
    Regression InterceptClose to 0.0 (indicating similar bias to predicate).Serum (N=54): -0.18 Serum (N=44): -0.06 95% Confidence Interval (Intercept): -0.49 - 0.37
    Sy.x (Std Error of Est.)Low value, indicating low variability around the regression line.Serum (N=54): 0.93 ng/mL Serum (N=44): 0.78 ng/mL 95% Confidence Interval (Sy.x): 0.53 - 1.27
    Interfering Substances (Effect %)Recovery bias found to be of no clinical significance.Hemoglobin: 2.8% Lipids (Triglycerides): 0.1% Bilirubin: 0.7% Immunoglobulin (IgG): -5.3% Albumin: -1.3% Cross-reactants (PAP, Kallikrein): 0.013%-0.018% Various Drugs: -1.8% to +2.4% (All reported as "of no clinical significance")
    Recovery (%)Recovery within a specified acceptable range (e.g., 90-110%).Overall Mean Recovery: Sample #1: 99%, Sample #2: 101%, Sample #3: 100%, Sample #4: 101% (Individual dilution recoveries ranged from 97% to 104%).
    Analytical RangeClearly defined and clinically appropriate range.0.01 – 100 ng/mL
    Minimum Detectable Concentration (ng/mL)Comparable to or better than predicate device.ADVIA IMS: 0.01 ng/mL (Predicate Immuno 1: 0.03 ng/mL)

    2. Sample sizes used for the test set and the data provenance

    • Correlation Studies (Test Set):

      • N=54 serum samples (0.37-79.70 ng/mL)
      • N=44 serum samples (0.07-62.93 ng/mL)
    • Imprecision Study: Six calibrator levels and commercial controls were tested. The exact number of individual samples (replicates for each level) is not specified but implied to be sufficient for robust CV calculation.

    • Interfering Substances Study: Serum pool with PSA concentration of 4.0 ng/mL was spiked. Specific number of samples beyond "serum pool" is not detailed.

    • Recovery Study: Four samples were used, each with several dilutions (5 dilution points per sample).

    • Data Provenance: Not explicitly stated (e.g., country of origin, specific clinical sites). The study type is retrospective as it involves analyzing existing serum samples and comparing the new assay results to those obtained from a predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This type of in vitro diagnostic device (immunoassay) does not typically involve human expert interpretation for establishing ground truth in the same way as imaging or pathology studies. The "ground truth" for the test set is established by the measurements from the predicate device (Immuno 1 PSA Assay), which is itself a legally marketed and presumably validated diagnostic tool. Therefore:

    • Number of experts: Not applicable in the context of interpretation.
    • Qualifications of experts: Not applicable.

    4. Adjudication method for the test set

    Not applicable. The ground truth for comparative studies of quantitative assays is typically the result from the established predicate method, not a consensus of human interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is an in vitro diagnostic assay, not an imaging or AI-assisted diagnostic tool that involves human readers interpreting cases.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, the studies presented (Imprecision, Correlation, Interfering Substances, Recovery, Analytical Range, Minimum Detectable Concentration) are all standalone performance evaluations of the ADVIA IMS PSA assay system. There is no human intervention in the measurement process once the sample is loaded and the assay runs. The results are quantitative values generated directly by the device.

    7. The type of ground truth used

    The primary ground truth used for performance validation is:

    • Predicate Device Measurements: For correlation studies, the results obtained from the Immuno 1 PSA Assay on the same samples serve as the comparison (reference) "ground truth" for the new ADVIA IMS assay.
    • Known Concentrations/Spikes: For imprecision, recovery, and interfering substance studies, known concentrations of PSA or spiked substances are used as the "ground truth" against which the device's measurements are evaluated.

    8. The sample size for the training set

    This document does not describe a "training set" in the context of machine learning or AI models. For an immunoassay, the concept of a training set is not directly applicable. The assay is developed and optimized during its R&D phase, and the studies presented here are for validation.

    9. How the ground truth for the training set was established

    Not applicable, as there is no mention of a training set for an AI model in this document. The "ground truth" for developing an immunoassay would be established through a combination of chemical principles, antibody specificity, detection methods, and extensive optimization using known calibrators and reference materials.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1