Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K181110

    Validate with FDA (Live)

    Device Name
    PROPHYflex 4
    Date Cleared
    2019-04-10

    (348 days)

    Product Code
    Regulation Number
    872.4200
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PROPHYflex 4 Handpiece and Accessories is a dental handpiece intended for perform supragingival and subgingival prophylaxis treatments. The prophylaxis treatments are obtained by the projection of water, air and appropriate dental prophylaxis powders, onto the treatment site. PROPHYflex 4 Handpiece and Accessories in supra-gingival configuration is intended for use in the cleaning and polishing of teeth by removal soft deposits and areas of discoloration as well as preparing teeth for fillings or veneers and cleaning fixed bands and brackets on orthodontic appliances. PROPHYflex 4 Handpiece and Accessories in subgingival configuration is indicated for the non-surgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment.

    Device Description

    The PROPHYflex 4 is an air polishing device/powder jet handpiece that projects a mixture of water and air with sodium bicarbonate, calcium carbonate or fine glycine power onto tooth surfaces. The device includes the handpiece main body with adjustable spray feature, two cannulas, a wrench for installing the cannula, two interchangeable gripping sleeves, powder container with cover, a cleaning burr and nozzle pin, and a set of spare o-rings. An optional adapter and perio tip can be installed for subgingival treatments.

    The powder is filled into the powder container by the user. When activated through foot control, air is pushed into the powder container, swirling the powder and pushing it through the device and out of the cannula. The spray is only mixed with the air and powder at the very tip of the cannula, providing an even and gentle cleaning effect. The amount of powder can be regulated via three different spray settings by the adjusting ring, helping the users to perform individual treatments such as subgingival biofilm removal, removal of discoloration and bacterial plaque, orthodontics, cleaning prior to fissure sealing, prosthetics, conservative and aesthetic dentistry by providing the possibility to adjust the cleaning effects to the clinical situation of the patient at hand.

    The components of the PROPHYflex 4, including the powder container, is to be completely reprocessed after each use.

    The three powder levels are as follows:

    • Highest Level: Suitable for supragingival treatment and offers full cleaning power .
    • Medium Level: Suitable for sub- and supragingival treatment and enables . gentle cleaning at a reduced powder quantity
    • . Lowest Level: Helps rinsing the powder off the tooth and blowing the powder from the system after the treatment

    During supraqinqival treatments (e.g. stain removal), the device doesn't touch the patient but will be placed 3 – 5 mm from the area that is to be cleaned. During subgingival treatments (biofilm management) with the perio tip, the tip is inserted into the pocket up to 5mm and angled towards the tooth.

    AI/ML Overview

    The provided text is a 510(k) summary for a dental handpiece, the PROPHYflex 4. It focuses on demonstrating substantial equivalence to predicate devices through technological characteristic comparison and non-clinical performance testing.

    Crucially, this document explicitly states: "Clinical data is not needed to characterize performance and establish substantial equivalence. The non-clinical test data characterize all performance aspects of the device based on well-established scientific and engineering principles. Clinical testing has not been conducted on this product."

    Therefore, a study demonstrating the device meets acceptance criteria via clinical performance (which would involve human readers, ground truth established by experts, MRMC studies, etc.) was explicitly not performed and is not required for this type of device according to the provided FDA submission.

    However, I can extract the acceptance criteria and reported performance based on the non-clinical (bench) testing, and other relevant information from the document, framing it within the context of a "study" as requested, even if it's a bench study rather than a clinical one.

    Here's a breakdown based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the PROPHYflex 4 are implicitly defined by its ability to perform as well as, or better than, its predicate devices across various technical specifications and to comply with relevant international standards. The study proving these criteria are met consists of non-clinical bench testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are essentially demonstrated equivalence or improvement over the predicate device (AIR-FLOW Handy 3.0 Plus (K151912)) and compliance with specified ISO standards.

    Feature / Acceptance Criterion (Implicit)Acceptance Criterion (Specific Value from Predicate or Standard)Reported Device Performance (PROPHYflex 4)Met?
    Indications for Use (Subgingival)Non-surgical removal of subgingival plaque in pockets up to 4 mm (Predicate H3.0 Plus), Up to 5mm (Reference Predicate H3.0 PERIO)Non-surgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment.Yes
    Max Mode OperationPresent (Predicate H3.0 Plus)PresentYes
    Internal SprayPresent (Predicate H3.0 Plus)PresentYes
    Coupling DimensionsComplies with ISO 7785-1, ISO 14457, and ISO 9168 (Predicate H3.0 Plus)Complies with ISO 7785-1, ISO 14457, and ISO 9168. (Specifically mentioning KaVo MULTIflex-Coupling, Sirona Coupling compatibility)Yes
    BiocompatibilityAccording to ISO 10993-1 (Predicate H3.0 Plus)Biocompatible according to ISO 10993-1 (Biological evaluation of medical devices), ISO 10993-5 (In vitro cytotoxicity), ISO 10993-10 (Irritation/sensitization), ISO 10993-18 (Chemical characterization).Yes
    Noise Level74.9 dB(A) (Predicate H3.0 Plus)56.6 - 69.0 dB(A) (Quieter than predicate)Yes
    Operating Temperature (Grip Sleeve)19.93°C - 23.33°C (Predicate H3.0 Plus)18.0°C - 25.1°C (Wider range, but within acceptable dental operating temperatures)Yes
    Maximum Air Pressure2.7 – 3.5 bar (39 – 51 psi) (Predicate H3.0 Plus)3.2 - 5 bar (46 - 73 psi) (Higher pressure capability)Yes
    Air and water supply10.3 – 13.5 Nl/min (Predicate H3.0 Plus)6.4 – 14.3 Nl/min (Comparable range)Yes
    Maximum Water Pressure1.0 – 2.2 bar (15 – 32 psi) (Predicate H3.0 Plus)1.0 - 2.5 bar (15 - 36 psi) (Higher pressure capability)Yes
    Spray Water Flow106 ml/min (Predicate H3.0 Plus)28 – 54 ml/min (Lower flow, noted as a difference but not raising new concerns)Yes
    Compliance with Powder Jet Handpiece StandardDIN 19402 (Predicate H3.0 Plus)DIN EN ISO 20608 (Dentistry - Powder jet handpieces and powders), DIN 19402 (Powder Jet Handpieces)Yes
    Additional Operational ModesN/A (Predicate H3.0 Plus only has Max Mode)Middle Mode (reduced powder), Spray-function Mode (water only)N/A (Improvement over predicate)
    Device Dimensions (Length)200 mm (Predicate H3.0 Plus)Up to 190 mm (10 mm shorter)Yes
    Device Dimensions (Largest Diameter)43 mm (Predicate H3.0 Plus)Up to 40 mm (3 mm thinner)Yes
    Reprocessing MethodN/A (Standard dental practice)Components are to be completely reprocessed after each use (standard for dental handpieces)Yes

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify exact "sample sizes" in terms of number of devices tested for each bench test. These are typically covered under the manufacturer's internal quality system documentation and test protocols, which are summarized rather than detailed in a 510(k). The tests are performed on representative units of the device.
    • Data Provenance: The testing was non-clinical (bench testing).
      • Country of Origin: The manufacturer is Kaltenbach & Voigt GmbH, located in Biberach, Germany. Therefore, the testing would likely have been conducted in Germany or by approved testing facilities.
      • Retrospective or Prospective: This refers to the collection of patient data, not bench testing. Bench testing is inherently prospective in its execution, demonstrating performance against specifications.

    3. Number of Experts Used to Establish Ground Truth and Qualifications of Those Experts

    • This question is not applicable as the study was purely non-clinical bench testing. There were no "experts" establishing a clinical "ground truth" for a test set, as no clinical test set involving human subjects was used. The "ground truth" for bench testing is derived from engineering specifications, established medical device safety and performance standards (e.g., ISO standards), and comparative performance against legally marketed predicate devices.

    4. Adjudication Method for the Test Set

    • This question is not applicable as there was no test set requiring human expert adjudication (e.g., image interpretation). Bench testing results are typically objectively measured against defined criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, an MRMC comparative effectiveness study was explicitly NOT done. The document states: "Clinical data is not needed to characterize performance and establish substantial equivalence. The non-clinical test data characterize all performance aspects of the device based on well-established scientific and engineering principles. Clinical testing has not been conducted on this product." Therefore, there is no effect size reported for human readers improving with AI vs. without AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • This question is not applicable in the context of a dental handpiece. The device is a physical instrument, not an AI algorithm. Its "performance" is its mechanical, fluidic, and material characteristics demonstrated through bench testing.

    7. The Type of Ground Truth Used

    • The "ground truth" for this device's performance validation is based on:
      • International Standards: Compliance with ISO standards (e.g., ISO 7785-1, ISO 14457, ISO 9168, ISO 10993 series, ISO 20608).
      • Predicate Device Performance: Demonstrating that the PROPHYflex 4's performance characteristics (dimensions, noise, pressure, flow, etc.) are comparable to or improved over the legally marketed predicate device (AIR-FLOW Handy 3.0 Plus (K151912)) and other reference devices.
      • Engineering Specifications: Internal design specifications and performance targets set by the manufacturer for the device.

    8. The Sample Size for the Training Set

    • This question is not applicable as this is a physical medical device, not an AI/ML algorithm that requires a "training set" of data.

    9. How the Ground Truth for the Training Set Was Established

    • This question is not applicable as this is a physical medical device, not an AI/ML algorithm; therefore, there is no "training set" or "ground truth" establishment for it in this context.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1