Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K112150

    Validate with FDA (Live)

    Date Cleared
    2011-11-23

    (119 days)

    Product Code
    Regulation Number
    888.3330
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PROFEMUR® GLADIATOR® HA Hip Stems are intended for use in uncemented total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients.

    Indications for Use:

    1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
    2. inflammatory degenerative joint disease such as rheumatoid arthritis;
    3. correction of functional deformity; and,
    4. revision procedures where other treatments or devices have failed.

    The PROFEMUR® Gladiator HA Hip Stem is intended for cementless hip arthroplasty.

    Device Description

    The PROFEMUR® GLADIATOR® HA stems are hydroxyapatite coated hip stems that couple with modular necks. Design features of the stems are summarized below:

    • Cementless stem with hydroxyapatite plasma spray coating
    • Available in 10 sizes
    • Manufactured from Ti alloy (Ti-6Al-4V) conforming to ASTM F620
    • Hydroxyapatite coating conforming to ASTM F1185
    • Available with and without collar
    AI/ML Overview

    I am sorry, but based on the provided text, there is no information about an AI/ML device, its acceptance criteria, or a study proving that it meets those criteria. The document describes the PROFEMUR® GLADIATOR® HA Hip Stems, which are a mechanical medical device (a hip implant). The evaluation for this device was based on mechanical testing (fatigue, fretting, and distraction evaluation) and a comparison to predicate devices, not on AI/ML performance metrics.

    Therefore, I cannot provide a response to the requested subsections as they pertain to AI/ML device evaluation, which is not present in the given text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1