Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K132701

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-09-26

    (28 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Primi™ Support Catheter is intended to be used in conjunction with guidewires in order to access discreet regions of the peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices. The support catheter also may be used to selectively infuse/deliver physician specified diagnostic and/or therapeutic agents, such as sterile saline, contrast agents, vaso-dilators, anti-blastics, lytics, and anti-thrombotics, with or without a guidewire in position.

    Device Description

    The ThermopeutiX Support Catheter that is the submission is a single lumen catheter with a radiopaque marker that aids in positioning the catheter and a tapered tip to facilitate entry. The catheter utilizes an integral spine wire. The catheter is available in several diameters and lengths which are 0.014", 0.018", or 0.035" guidewire compatible. The working lengths are 90 cm thru 170 cm. Reference Table 1 for model number availability. The distal portion of the catheter shaft has a hydrophilic coating to facilitate insertion. The device has a luer hub connector to allow infusion of diagnostic and/or therapeutic agents. The device is supplied sterile and is intended for single use only. Reference Table 2 below for physical and performance characteristics.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the ThermopeutiX Primi™ Support Catheter with an advanced hydrophilic coating. This submission aims to demonstrate substantial equivalence to a previously cleared version of the same catheter (K130850) by ThermopeutiX, Inc., with the primary difference being the advanced hydrophilic coating.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the physical and performance characteristics of the predicate device and the new device. The study demonstrates that the new device's performance is either identical or comparable, and meets internal protocol requirements and product specifications.

    Element/FeatureAcceptance Criteria (from Predicate Device)Reported Device Performance (New Device)
    Working Length (s)90 cm – 170 cm; polymer shaft with integral spine wire90 cm – 170 cm; polymer shaft with integral spine wire
    Outside Diameter (s)Distal:0.045" – 0.064"Proximal: 0.060" – 0.080"Distal:0.045" – 0.064"Proximal: 0.060" – 0.080"
    LumenOneOne
    Guide wire Compatibility0.014"-0.035"0.014"-0.035"
    Guide wire UseOver the WireOver the Wire
    Distal End ConfigurationTapered tip for ease of crossingTapered tip for ease of crossing
    Fluid connectionLuer hub connectorLuer hub connector
    Flow rate @ 150 psi w/Saline0.014" wire models > 0.80 cc/sec0.035" wire models > 0.50 cc/sec0.014" wire models > 0.80 cc/sec0.035" wire models > 0.50 cc/sec
    Flow rate @ 150 psi w/Contrast0.014" wire models > 0.30 cc/sec0.035" wire models > 0.15 cc/sec0.014" wire models > 0.30 cc/sec0.035" wire models > 0.15 cc/sec
    Maximum Injection PressureDo not exceed 300 psiDo not exceed 300 psi
    CoatingYes; hydrophilic, distal portionYes; advanced hydrophilic, distal portion (provides comparable lubricity with a reduction in particulates)
    Coating Length65 cm, 115 cm or 140 cm65 cm, 115 cm or 140 cm
    Coating ThicknessProprietarySame as predicate (considered equivalent)
    Radiopaque MarkersOneOne
    Sheath Compatibility5 Fr and 6 Fr5 Fr and 6 Fr
    Materials UsedIdentical list of materials as the new device, except for the coating.Hub-PebaxOuter Shaft-PebaxInner Shaft-HDPEDistal Shaft-Pebax w/radiopaque fillerCore Wire-Stainless Steel tapered wireCoating-Advanced hydrophilic proprietary coatingMarker Band Platinum/Iridium

    Study Proving Device Meets Acceptance Criteria:

    The study conducted is a non-clinical performance and safety study comparing the new device with the advanced hydrophilic coating to the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify an exact numerical sample size for each individual test. It states "The non-clinical tests performed on sample devices are listed below," and "All performance testing was repeated on samples with a minimum of 10 month aging." This indicates that multiple samples were used for testing, including aged samples, but the exact number isn't quantified.

    • Sample Size for Test Set: Not explicitly stated as a number, but "sample devices" were used, including aged samples.
    • Data Provenance: The data is from in-vitro (bench) testing performed by the manufacturer, ThermopeutiX, Inc., in the United States. This is prospective testing performed for the purpose of this 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This type of submission (510(k) for a medical device with minor modification) typically relies on engineering and biological testing against established standards and product specifications, rather than expert interpretation of a test set in the way AI/ML algorithms use ground truth. Therefore, the concept of "experts establishing ground truth for the test set" is not applicable here. The "ground truth" is the objective physical and performance measurements determined by standardized testing protocols.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, this study is based on objective physical and performance measurements via bench testing, not subjective assessment requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging AI, not for performance testing of a physical medical device like a catheter. The aim here is to show "substantial equivalence" of device characteristics and safety, not to measure human reader improvement with AI.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not applicable. This is a physical medical device, not an algorithm or AI system. The "standalone" performance is the device's measured performance during bench testing without human intervention in the device's function.

    7. The Type of Ground Truth Used

    The ground truth used for this study is based on objective physical and performance measurements derived from standardized engineering and biocompatibility tests (e.g., flow rate, tensile strength, lubricity, cytotoxicity, etc.) and comparison to the known characteristics of the predicate device. It's essentially test results against predetermined specifications and accepted standards (e.g., ISO, ASTM).

    8. The Sample Size for the Training Set

    Not applicable. This is a physical medical device, not an AI/ML algorithm that requires a training set. The device itself is the product, and its performance is being evaluated, not an algorithm's learning capability.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1