510(k) Data Aggregation

    K Number
    K980703

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1998-05-04

    (70 days)

    Product Code
    Regulation Number
    878.5010
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Premilene® sutures are indicated for use in all types of general soft tissue approximation and ligation, including use in cardiovascular, ophthalmic, microsurgery and neural tissue.

    Device Description

    The subject device is a nonabsorbable, flexible monofilament suture thread which is supplied sterile. It is composed of isotactic polypropylene polymer, and is indicated for soft tissue approximation. It will be offered undyed, and dyed with the FDA approved colorant [Phthalocyaninati(2-)] copper in accordance with Title 21 CFR, §74.3045. It will be available with and without standard needles attached.

    AI/ML Overview

    The provided text describes the Premilene Nonabsorbable Polypropylene Surgical Suture and its substantial equivalence to predicate devices, focusing on non-clinical testing. This document is a 510(k) summary for a medical device and thus relates to regulatory approval rather than a study proving the device meets specific acceptance criteria in the typical sense of a clinical trial for AI/software devices. The acceptance criteria here are based on physicochemical equivalence to existing devices.

    Here's a breakdown of the requested information based only on the provided text, recognizing that many items common in AI/software device studies are not applicable or detailed in this type of submission:


    Description of Acceptance Criteria and Proving Device Meets Criteria

    The "acceptance criteria" for the Premilene Nonabsorbable Polypropylene Surgical Suture are primarily based on demonstrating substantial equivalence to existing predicate devices (Prolene®, Surgilene®, Sharpoint). This equivalence is established through adherence to USP XXIII standards and comparative non-clinical testing of physical, chemical, and biological properties.

    The study proving the device meets these acceptance criteria is presented as a series of non-clinical tests and validations, as summarized in Section G of the 510(k) summary.


    1. Table of Acceptance Criteria and Reported Device Performance

    Given the nature of this 510(k) submission, the "acceptance criteria" are compliance with specified standards and "device performance" is expressed as meeting these standards and being substantially equivalent to predicates. No specific numerical thresholds or performance metrics (like sensitivity, specificity, etc.) are provided for this type of medical device which is a physical suture.

    Acceptance Criteria (Based on Substantial Equivalence to Predicates)Reported Device Performance
    Material Composition: Isotactic polypropylene polymerComposed of the same material (isotactic form of polyolefin polymer polypropylene).
    Colorant: [Phthalocyaninato(2-)] copper, ≤0.5% by suture weight (if dyed)Offered undyed, and dyed with the same colorant at ≤0.5% by suture weight, in accordance with 21 CFR §74.3045.
    Design: Sterile, flexible monofilament thread, various lengths/diameters, with/without needlesSame design as predicate devices; sterile, flexible monofilament thread; offered in variety of lengths and range of diameters conforming to U.S. Pharmacopeia (U.S.P.) XXIII, with or without standard needles attached.
    Physical Properties: Conformance to U.S.P. XXIII for nonabsorbable surgical suture, including: - <861> Sutures -- Diameter - <871> Sutures -- Needle Attachment - <881> Tensile Strength - Fiber diameter - Knot pull tensile strength - Straight pull tensile strength - Flexibility - Elongation - Elasticity - Needle attachment strengthConforms in all respects to the requirements of the Official Monograph for Nonabsorbable Surgical Suture in U.S.P. XXIII for specified parameters. Physical properties are substantially equivalent to predicate devices for all identified parameters.
    Manufacturing Process: Same as at least one predicate deviceManufactured in the same manner as at least one predicate device (supplier of finished suture fiber is manufacturer of said predicate suture). Suture fiber has same chemical characteristics, biocompatibility profile, and in vivo performance properties as that predicate device.
    Packaging & Sterilization: Same or equivalent manner as predicatesPackaged and sterilized in the same or equivalent manner.
    Labeling Claims: Same as predicates (indications, contraindications, warnings, cautions, precautions)Has the same labeling claims as predicate devices.
    Chemical Assays: Identity and purity for finished suture fiberTesting conducted by manufacturer/supplier included chemical assays for identity and purity.
    Biosafety Studies: In vitro and in vivoTesting conducted by manufacturer/supplier included in vitro and in vivo biosafety studies.
    Tensile Strength Retention (in vivo): Over 1 yearOne (1) year implant studies in animals to demonstrate retention of tensile strength.
    Sterilization Validation & Residues: Validation and evaluationSterilization validation and evaluation of sterilant residues performed.
    Shelf-life Testing: Stability and integrity over timeShelf-life testing performed.

    2. Sample Size Used for the Test Set and the Data Provenance

    The text does not specify sample sizes (e.g., number of sutures, animals used) for the non-clinical tests. This information is typically proprietary and contained in the full 510(k) submission, not the summary.

    Data Provenance: The tests were conducted by the manufacturer and supplier to Aesculap®. The country of origin for the data is not specified. The tests are prospective in the sense that they were conducted specifically for this 510(k) submission to demonstrate equivalence.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable. The "ground truth" for this type of medical device is established through adherence to recognized pharmacopeial standards (U.S.P. XXIII) and direct comparative physical/chemical testing, not through expert consensus in a diagnostic sense. The experts involved would be quality control engineers, chemists, and materials scientists, but their number and specific qualifications are not detailed in this summary.


    4. Adjudication Method for the Test Set

    Not applicable. As this is not a diagnostic device involving interpretation errors, there is no adjudication method in the context of expert review. Test results are objective measurements against established standards.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC study was not done. This type of study is for evaluating observer performance with diagnostic imaging or similar outputs, which is not relevant to a surgical suture. The effectiveness is demonstrated through mechanical and biological equivalence to predicate devices and adherence to standards.


    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not applicable. This is not an AI/algorithm-based device.


    7. The Type of Ground Truth Used

    The "ground truth" consists of established, objective physical, chemical, and biological standards (e.g., U.S.P. XXIII monographs for sutures, recognized chemical and biocompatibility testing protocols, and the established properties of the predicate devices).


    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device that uses a "training set."


    9. How the Ground Truth for the Training Set Was Established

    Not applicable. See point 8.