510(k) Data Aggregation

    K Number
    K182559

    Validate with FDA (Live)

    Date Cleared
    2018-11-05

    (49 days)

    Product Code
    Regulation Number
    866.3110
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PREMIER Platinum HpSA PLUS enzyme immunoassay (EIA) is an in vitro qualitative procedure for the detection of Helicobacter pylori antigens in human stool. Test results are intended to aid in the diagnosis of H. pylori infection and to monitor response during and post-therapy in patients. Accepted medical practice recommends that testing by any current method, to confirm eradication, be done at least four weeks following completion of therapy.

    Device Description

    The PREMIER Platinum HpSA® PLUS test is a microwell-based enzyme immunoassay that detects H. pylori antigens present in human stool. The test utilizes a plurality (mixture) of monoclonal antibodies adsorbed to microwells. Diluted patient samples and an enzyme conjugate reagent are added to the microwells and incubated for one hour at room temperature. A wash is performed to remove unbound material. Substrate is added and incubated for 10 minutes at room temperature. Color develops in the presence of bound enzyme. Stop solution is added and the results are interpreted visually or spectrophotometrically. No calculations are required and the visual color change makes the interpretation of results objective and simple.

    In addition, the HpSA test permits assessment of established or novel anti-H. pylori treatment during and posttherapy to monitor for treatment effectiveness, relapse or eradication.

    PREMIER Platinum HpSA PLUS (K053335), as the predicate device for this submission, was a modification of PREMIER Platinum HpSA (K983255, K980076) that provided increased signal strengths with positive test results and better discrimination between low positive and negative tests. This submission is for modifications to the antibodies used in the microwells and conjugate reagent.

    AI/ML Overview

    The PREMIER Platinum HpSA PLUS assay is an in vitro qualitative procedure for the detection of Helicobacter pylori antigens in human stool, intended to aid in the diagnosis of H. pylori infection and to monitor response during and post-therapy.

    Here's the breakdown of the acceptance criteria and study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The modification being assessed is to the antibodies used in the microwells and conjugate reagent of the PREMIER Platinum HpSA® PLUS. The relevant performance characteristics for comparison are based on the predicate device.

    Performance MetricAcceptance Criteria (Predicate Device)Reported Device Performance (Modified Device)
    Analytical Sensitivity (LoD)≥ 4.67 ng H. pylori protein/mL of stool4.66 ng H. pylori protein/mL of stool (meets)
    Clinical Sensitivity (PPA)100%100%
    Clinical Specificity (NPA)94.8%100%
    ReproducibilityNot explicitly stated as a minimum, but implied to be high based on predicate acceptance100% (300/300) of results as expected
    Cross-ReactivityNone of the listed organisms affected positive or negative test resultsNone of the listed organisms affected positive or negative test results
    Interfering SubstancesNone of the listed substances interfered with positive or negative test resultsNone of the listed substances interfered with positive or negative test results

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Clinical Study: 159 archived, unpreserved stool samples.
    • Data Provenance: The samples were from symptomatic patients but no country of origin is specified. The samples were "archived," indicating a retrospective study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The documentation does not provide details on the number or qualifications of experts used to establish the ground truth for this specific clinical study comparing the modified device to the predicate. The "ground truth" for this comparison was the result obtained from the predicate device itself.

    4. Adjudication Method for the Test Set

    Not applicable. The study involved a direct comparison of the modified device's results against the predicate device's results. There was no independent adjudication of individual cases against an external ground truth in the traditional sense, as the predicate served as the reference.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This device is an in-vitro diagnostic (EIA) test for detecting antigens, not an imaging or diagnostic aid that involves human readers interpreting results in a variable manner. The interpretation of results is visual or spectrophotometric, implying a more objective, less reader-dependent outcome.

    6. Standalone Performance Study

    Yes, a standalone performance of the modified device was effectively demonstrated by comparing its results against the predicate device on the same set of samples. The performance metrics (sensitivity, specificity, reproducibility, analytical sensitivity, cross-reactivity, interfering substances) are all measures of the algorithm's (or device's) standalone performance under specified conditions.

    7. Type of Ground Truth Used

    For the clinical study, the ground truth was established by the predicate device (PREMIER Platinum HpSA® PLUS K053335). The study aimed to demonstrate substantial equivalence by showing agreement between the modified device and the predicate. For analytical studies, the ground truth was based on controlled experimental conditions (e.g., known concentrations of H. pylori protein for LoD, known presence/absence of microorganisms for cross-reactivity, known interfering substances).

    8. Sample Size for the Training Set

    The document does not specify a separate "training set" sample size. The PREMIER Platinum HpSA PLUS is an enzyme immunoassay, implying a deterministic chemical reaction, not a machine learning algorithm that requires a separate training phase. The development of the device would involve optimization and internal testing, which might be analogous to model training, but these details are not provided in terms of sample size or methodology.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, there's no mention of a traditional "training set" in the context of an EIA device. The "ground truth" during initial development and optimization would have been established through controlled experiments using known concentrations of H. pylori antigens and various other substances (cross-reactants, interferents) to characterize the assay's performance and specificity. This would involve laboratory standards and analytical methods to confirm the composition of samples.

    Ask a Question

    Ask a specific question about this device

    K Number
    K053335

    Validate with FDA (Live)

    Date Cleared
    2006-03-10

    (99 days)

    Product Code
    Regulation Number
    866.3110
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Premier Platinum HpSA PLUS enzyme immunoassay (EIA) is an in vitro qualitative procedure for the detection of Helicobacter pylori antigens in human stool. Test results are intended to aid in the diagnosis of H. pylori infection and to monitor response during and post-therapy in patients. Accepted medical practice recommends that testing by any current method, to confirm eradication, be done at least four weeks following completion of therapy.

    Device Description

    Premier Platinum HpSA PLUS is an in vitro diagnostic, microwell-based, enzymelinked immunoassay for the detection of Helicobacter pylori antigen in human stool. The assay is intended for use in clinical laboratories to test for bacterial colonization to aid diagnosis, or monitor a patient's response during therapy to eradicate infection. The assay consists of Microwells coated with specific antibodies (solid phase/capture antibodies), Enzyme Conjugate (detector antibodies), Sample Diluent, Premier 20X Wash Buffer I, Premier Substrate Solution I, Premier Stop Solution I and Positive Control. Sample Diluent also functions as the Negative Control reagent.

    No calibrators are needed to use this device.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Premier Platinum HpSA PLUS device:

    Device Acceptance Criteria and Performance Study Analysis

    The submission describes the Premier Platinum HpSA PLUS, an in vitro diagnostic enzyme immunoassay (EIA) for detecting H. pylori antigen in human stool. The primary goal of the submission is to demonstrate substantial equivalence to a predicate device, Premier Platinum HpSA.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as a separate section with predefined targets. Instead, the study aims to demonstrate that the new device performs "similarly" to the predicate, and that certain analytical performance characteristics are met. Based on the provided text, I've inferred the acceptance criteria from the reported performance and the comparisons made to the predicate or industry standards like EP12-A.

    Acceptance Criteria (Implied/Direct)Reported Device Performance (Premier Platinum HpSA PLUS)
    Clinical Performance (vs. Predicate)
    Overall Agreement with Predicate96.5%
    Positive Agreement with Predicate100%
    Negative Agreement with Predicate94.8%
    Improvement over Predicate for Indeterminate ResultsClarified 7 out of 7 indeterminate results from predicate as definite positive using other tests (CLO/Histology/UBT) by HpSA PLUS, and 1 indeterminate result as definite negative.
    Analytical Performance
    Precision/Reproducibility (intra-assay)97% (compared to predicate's 100% – this is for intra-assay, not overall reproducibility which is stated as 100%)
    Overall Reproducibility100% (for samples above or below the limit of analytical sensitivity)
    Performance of "High negative" samples at cutoffProduced weakly positive results 42 out of 162 times (expected ~50% based on EP12-A)
    Limit of Detection (H. pylori flagellar antigen in stool)≥ 4.67 ng/mL
    Limit of Detection (H. pylori bacterial strain in stool)≥ 1.0 % organisms/mL (assuming the missing exponent is 10^?)
    Assay Cutoff0.100 at OD450/630
    Absence of Indeterminate RangeAll results definitive (no 0.100-0.120 OD range needed, unlike predicate)
    Absence of Interference from Drugs/Nonmicrobial SubstancesNone had a significant effect on negative test results; positive results correlated closely with unspiked samples.
    Absence of Interference from Microbial/Viral OrganismsNone adversely affected final positive or negative test results.
    Therapeutic Monitoring PerformanceEradication curves were substantively the same as predicate for strongly positive samples; differed at low positive states initially but identical by week four post-treatment.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Test Set (Clinical Comparison): 291 samples from symptomatic patients.
      • Data Provenance: Not explicitly stated, but collected from "symptomatic patients" suggesting a clinical setting. It is retrospective as 33 of these samples were originally evaluated in an earlier trial for the predicate device.
    • Test Set (Reproducibility): 2 high positive, 2 low negative, 1 low positive, and 1 high negative specimen in a reference panel, with 9 replicates each of the low positive and high negative samples, bringing the total cohort to 22 reference specimens.
      • Data Provenance: Prepared from "archival specimens" and tested in "three different laboratories." This indicates a controlled, retrospective, and multi-site analytical study.
    • Test Set (Interference Testing): 5 known H. pylori-positive and 5 known negative samples for drug interference; an unspecified number of samples for microbial/viral interference.
      • Data Provenance: Laboratory-controlled spike-in experiments, prospective in nature for the study itself, using known positive/negative samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    There were no experts used to establish ground truth in the traditional sense for the clinical comparison. Instead, "other conventional tests such as CLO, Histology, or UBT" were used to arbitrate discordant results between the new device and the predicate. The qualifications of those performing these conventional tests are not specified.

    4. Adjudication Method for the Test Set

    For the clinical comparison, when results between Premier Platinum HpSA PLUS and the predicate were discordant, they were adjudicated against "test data from other conventional tests such as CLO, Histology, or UBT to determine the trueness of the results." This implies a form of external reference standard adjudication, rather than a consensus among human readers of the device output. It's not a 2+1 or 3+1 method directly related to human interpretation of the device results, but rather a comparison to established diagnostic methods.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    No MRMC study was done, nor is this an AI/human-in-the-loop device. It is a laboratory-based enzyme immunoassay. The comparison is between two similar lab tests.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Yes, this is a standalone device. Its performance characteristics are reported as the output of the assay itself, interpreted quantitatively (OD readings) against a predefined cutoff. There is no human interpretation component in the sense of a radiologist reading an image or a pathologist reviewing a slide that would then be assisted by AI.

    7. The Type of Ground Truth Used

    • Clinical Comparison (for discordant results): A combination of other conventional diagnostic tests for H. pylori infection (CLO, Histology, UBT). These serve as the de facto "ground truth" to determine if the device's results are true positives/negatives.
    • Analytical Sensitivity (Limit of Detection): Purified antigens and bacterial strains, where known concentrations were spiked, serving as a controlled, known standard.
    • Interference Testing: Known H. pylori-positive and negative samples spiked with various substances/microbes, representing a known state (positive/negative for H. pylori, plus presence of interferent).
    • Reproducibility: A reference panel with defined states (high positive, low positive, high negative, low negative), serving as a known standard.

    8. The Sample Size for the Training Set

    No specific training set is mentioned as this device is not an AI/machine learning algorithm that requires a training phase. It is a traditional immunoassay.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K980076

    Validate with FDA (Live)

    Date Cleared
    1998-05-12

    (124 days)

    Product Code
    Regulation Number
    866.3110
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Premier Platinum HpSA enzyme immunoassay (EIA) is an in vitro qualitative procedure for the detection of Helicobacter pylori antigens in human stool. Test results are intended to aid in the diagnosis of H. pylori infection in adult symptomatic patients.

    Device Description

    The Premier Platinum HpSA test utilizes polyclonal anti-H. pylori capture antibody adsorbed to microwells. Diluted patient samples and a peroxidase conjugated polyclonal antibody are added to the wells and incubated for one hour at room temperature. A wash is performed to remove unbound material. Substrate is added and incubated for ten minutes at room temperature. Color develops in the presence of bound enzyme. Stop solution is added and the results are interpreted visually or spectrophotometrically.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the Premier Platinum HpSA device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the comparison to the predicate device and the overall performance evaluation. They aim to demonstrate that the Premier Platinum HpSA test is "substantially equivalent" to existing methods for detecting H. pylori.

    Performance MetricAcceptance Criteria (Implied/Predicate)Reported Premier Platinum HpSA Performance (Compiled Data)
    Sensitivity~95.4% (89.6-98.5%) (from predicate CLOtest) or demonstrating high clinical agreement96.1% (90.4-98.9%)
    Specificity~100.0% (96.4-100.0%) (from predicate CLOtest) or demonstrating high clinical agreement95.7% (89.5-98.8%)
    Correlation~97.6% (94.5-99.2%) (from predicate CLOtest) or demonstrating high clinical agreement (95.9% desired)95.9% (92.2-98.2%)

    Note: The document explicitly states, "The differences in technology does not raise additional concerns regarding safety and effectiveness. Safety and effectiveness are demonstrated to be substantially equivalent." This indicates that the reported performance metrics for the Premier Platinum HpSA were considered sufficient to meet the acceptance criteria by demonstrating equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 200 symptomatic adults were evaluated in the clinical study. Nine patients were considered unevaluable, meaning the effective sample size for performance calculations was 191 (99 infected, 90 not infected, plus 2 'Equ' and 1 'Equ' at site 3, 2 'Equ' at site 4, which were excluded from the main table's infected/not infected counts but are implicitly part of the overall 200 evaluated).
    • Data Provenance: The data was collected prospectively from four sites:
      • One midwestern United States location
      • One site in Canada
      • Two sites in Italy

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not specify the number of experts or their qualifications directly. However, it states that the diagnosis of H. pylori infection (ground truth) was "judged by objective reference methods." These reference methods are:

    • Culture
    • Rapid Urease Test (RUT)
    • Histology
    • Urea Breath Test (UBT)

    It's implied that these objective reference methods were interpreted by qualified personnel (e.g., microbiologists for culture, pathologists for histology, and trained technicians for RUT and UBT), but no specific number or detailed qualifications are provided.

    4. Adjudication Method for the Test Set

    Patients were considered H. pylori positive if:

    • Culture was positive, OR
    • Two or more of the other three tests (Rapid Urease, Histology, UBT) were positive.

    Patients were considered unevaluable if:

    • They had negative or no culture results, AND
    • Only one other test (RUT, Histology, UBT) was positive.

    This acts as a consensus-based adjudication method for establishing the ground truth.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The study compares the Premier Platinum HpSA test (an in vitro diagnostic device) against established diagnostic reference methods for H. pylori infection (the ground truth), not against human readers with and without AI assistance. The device in question is a laboratory assay, not an AI-powered image analysis tool for human readers.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    The Premier Platinum HpSA test is a standalone in vitro diagnostic device. Its performance metrics (Sensitivity, Specificity, Correlation) are reported for the device itself, functioning independently, with results interpreted visually or spectrophotometrically. There is no "human-in-the-loop" component in the direct processing of the sample by the device; rather, it's a lab test where technicians perform the procedure and interpret the final result based on established cutoffs.

    7. The Type of Ground Truth Used

    The ground truth used was a composite reference standard based on objective laboratory and pathological evidence. Specifically, it relied on a combination of:

    • Culture
    • Rapid Urease Test (RUT)
    • Histology
    • Urea Breath Test (UBT)

    A patient was considered positive if culture was positive, or if two or more of the other three tests were positive. This is referred to as expert consensus (albeit on objective data) or composite clinical diagnosis.

    8. The Sample Size for the Training Set

    The document does not specify a separate training set or its sample size. This device is an immunoassay, not a machine learning model that typically requires a distinct training phase. Therefore, the concept of a "training set" in the context of AI models does not apply here. The performance evaluation is based on a clinical test set.

    9. How the Ground Truth for the Training Set Was Established

    As noted in point 8, there is no mention of a "training set" for this immunoassay device. The ground truth for the test set was established using a composite reference standard (Culture, Rapid Urease, Histology, UBT) as detailed in point 7.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1