Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K111640

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-10-07

    (116 days)

    Product Code
    Regulation Number
    868.5450
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Precision Flow® - Heliox is intended to warm and add moisture to breathing gases from an external source for administration to neonate/infant, pediatric and adult patients in the hospital and sub-acute institutions. It adds heat and moisture to a blended medical heliox (79% helium. 21% oxygen)/oxygen mixture and assures the integrity of the precise heliox (79% helium, 21% oxygen)/oxygen mixture via an integral oxygen analyzer. The flow rates may be from 1 to 40 liters per minute via nasal cannula.

    Device Description

    The Precision Flow® - Heliox consists of two parts: The main unit which contains all the electrical and electronic components including the electronic blender and flow controllers. All the sensors are located in the main unit. The main unit has no water pathways and the gas pathway contains only dry gas at room temperature, and consequently does not need internal cleaning or disinfection. The disposable components comprising the disposable water module, vapor transfer cartridge and heated delivery tube. Conditions in the circulating water and gas streams are sensed remotely via the interface between the main unit and the disposable module.

    AI/ML Overview

    The Vapotherm Precision Flow® - Heliox is a medical device intended to warm and humidify breathing gases, specifically a heliox/oxygen mixture, for neonate/infant, pediatric, and adult patients in hospital and sub-acute settings.

    Here's a breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document lists various standards that the Precision Flow® - Heliox met, with some exceptions. It also includes functional and design verification tests. The "Reported Device Performance" for these tests is generally stated as: "In all instances, the Precision Flow® - Heliox functioned as intended and the results observed were as expected." Specific quantitative performance data is limited within this summary but is generally implied by compliance with the referenced standards.

    Acceptance CriteriaReported Device Performance
    I. Compliance with Standards:"In all instances, the Precision Flow® - Heliox functioned as intended and the results observed were as expected." (Implied compliance with the intent of the standard, with specific exceptions noted)
    ISO 62304: Medical Device Software - Software Life Cycle ProcessMet
    ISO 8185:2007: Respiratory tract humidifiers for medical useMet
    ISO 10993-5:2009: Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityMet
    ISO 10993-10:2010: Tests for Irritation and SensitizationMet
    ISO 11195:1995: Gas mixers for medical use - Stand-alone gas mixersMet
    ISO 14971:2007: Medical devices - Application of risk management to medical devicesMet
    ANSI/AAMI/ISO 15223-1:2007/A1:2008: Medical Devices - Symbols to be used with Medical Device Labels, Labeling and information to be supplied - Part 1: General RequirementsMet
    ANSI/AAMI ES60601-1:2005: Medical electrical equipment – Part 1: General requirements for basic safety and essential performanceMet
    ANSI/AAMI/IEC 60601-1-2:2007 3rd edition: Medical Electrical Equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic Compatibility - Requirements and testsMet (with exceptions: Section 5: Device cannot produce hazardous radiation; Section 6: Anesthetic mixtures not applicable)
    EN60601-1-4:1999: Medical Electrical Equipment: Part 1: General Requirements for Safety 4. Collateral Standard: Programmable Electrical Medical SystemsMet
    IEC 60601-1-8 2006-10: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systemsMet
    IEC 60529 IPX1: Drip ProofMet
    AAMI TIR 32:2004: Software Risk ManagementMet
    AAMI TIR 36:2007: Validation of SoftwareMet
    ISTA -1A:2001: Procedure 1A: Packaged-Products weighing 150 LB (68 KG) or LessMet
    EPA/625/R-96/010b (VOC) Method TO-15 (1999): Compendium Method TO-15 Determination of Volatile Organic Compounds (VOCs) in Air collected in specially-Prepared canisters and analyzed by chromatography/mass spectrometry (GC/MS)Met
    NIOSH Method 0500 (1994): NIOSH manual of Analytical methods (NMNM) fourth edition August 1994 (Particulate Matter Testing)Met
    II. Design Verification Plan (Required Tests):"In all instances, the Precision Flow® - Heliox functioned as intended and the results observed were as expected."
    Functional Performance TestMet
    Alarms & Fault Conditions TestMet
    Patient Connector Attachment TestMet
    Water Connector Attachment TestMet
    Extended Life TestMet
    Membrane Switch and Adjustment Dial TestMet
    Software Verification TestMet
    Fluid & Gas Pathways Leak TestMet
    Cartridge Insertion TestMet
    Door TestMet
    IPX TestMet
    Environmental Temperature/Humidity Extremes TestMet
    III. Performance Characteristics (Implied by Comparison for Substantial Equivalence):
    Oxygen % Range21 to 100% (Equivalent to predicates)
    FiO2 Accuracy2 to 40 lpm: +/-2%; 1 to 2 lpm: +/-5% (Equivalent to predicates)
    Max Flow Range40 lpm (Equivalent to predicates)
    Alarm Sound Level (Medium Priority)47 dB @ 1m (Equivalent to predicate)
    Alarm Sound Level (Low Priority)45 dB @ 1m (Equivalent to predicate)

    2. Sample Size Used for the Test Set and Data Provenance

    The 510(k) summary does not provide specific sample sizes for particular performance tests or explicitly mention a "test set" in the context of clinical or image-based studies. The performance data section broadly states: "In all instances, the Precision Flow® - Heliox functioned as intended and the results observed were as expected." This implies that the design verification tests listed were conducted on the device.

    • Sample Size for Test Set: Not explicitly stated for each test, but implied to be sufficient for demonstrating compliance with cited standards and internal design verification plans.
    • Data Provenance (Country of Origin, Retrospective/Prospective): Not specified. The tests are likely laboratory or bench testing, not clinical studies involving patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • This information is not applicable to this type of device and study. The testing for the Precision Flow® - Heliox involves engineering and performance validation against objective technical standards (e.g., flow rates, temperature, alarm function, electrical safety), not subjective interpretation requiring expert consensus.

    4. Adjudication Method for the Test Set

    • This information is not applicable as the evaluation relies on objective measurements against engineering specifications and regulatory standards, not subjective interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of AI vs. Without AI Assistance

    • No, an MRMC comparative effectiveness study was not done. This device is a medical gas humidifier/blender, not an AI-powered diagnostic or decision-support tool. It does not involve human readers interpreting cases with or without AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • No, this is not applicable. The device is hardware with integrated software for control, not a standalone algorithm. Its performance is inherent to its physical operation and control systems.

    7. The Type of Ground Truth Used

    • The "ground truth" for this device's performance is established by objective engineering specifications, technical standards (e.g., ISO, ANSI/AAMI, IEC), and manufacturer-defined design requirements. This includes accuracy of gas blending, flow rates, temperature control, alarm thresholds, electrical safety, and biocompatibility.

    8. The Sample Size for the Training Set

    • This is not applicable in the context of a traditional "training set" for machine learning or AI models. The device's software is likely developed using standard software engineering practices and verified accordingly, rather than trained on a dataset.

    9. How the Ground Truth for the Training Set Was Established

    • This is not applicable for the reasons stated above. The "ground truth" for the device's operational parameters is derived from engineering design, relevant medical standards, and intended physiological effects of humidified gas delivery.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1