Search Filters

Search Results

Found 89 results

510(k) Data Aggregation

    K Number
    K211865

    Validate with FDA (Live)

    Date Cleared
    2021-12-20

    (187 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    The subject device is disposable medical PVC gloves that made from PVC compound, Clear (non-colored), powder free and nonsterile. The device meets the specifications in ASTM D5250-06(Reapproved 2015) Standard specification for poly (vinyl chloride) gloves for medical application.

    AI/ML Overview

    This document is a 510(k) Premarket Notification summary for "Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)". It describes the device, its intended use, and the studies conducted to demonstrate substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test MethodologyPurposeAcceptance CriteriaReported Device Performance
    ASTM D5250-06 (Reapproved 2015)Dimension (mm) IL S-2 AQL4.0
    Length$\geq$ 230233-241
    Width (Small)80-9081-89
    Width (Medium)90-10092-99
    Width (Large)100-110102-109
    Width (X-Large)110-120112-119
    Thickness (Fingertip)$\geq$ 0.050.09-0.11
    Thickness (Palm)$\geq$ 0.080.10-0.11
    ASTM D5250-06 (Reapproved 2015)Physical Properties IL S-2 AQL4.0
    Tensile strength$\geq$ 14MPa (Before & After aging)17-25 MPa (implicitly for before & after, as it covers the range)
    Elongation (Before aging)$\geq$ 500%560-610%
    Elongation (After aging)$\geq$ 400%460-570%
    21 CFR 800.20, ASTM D5250-06 (Reapproved 2015), ASTM D5151-19Freedom from pinholesWaterleakage test: Inspection Level I, AQL2.5, and Accept/Reject criteria of 10/11. (5 noncompliance is allowed.)Pass
    ASTM D5250-06 (Reapproved 2015), ASTM D6124-06 (Reapproved 2017)Powder Residual$<$ 2mg/gloveMean: 0.1mg/pcs (Pass)
    ISO 10993-10: 2010-08-01Biocompatibility (Primary Skin Irritation in rabbits)Under the conditions of the study, the subject device is not a primary skin irritant.Passes
    ISO 10993-10: 2010-08-01Biocompatibility (Dermal sensitization in the guinea pig)Under the conditions of the study, the subject device is not a skin sensitizer.Passes
    ISO 10993-5: 2009Biocompatibility (Cytotoxicity - L929 cells measured by MTT assay)Under the conditions of this study, the test article was non-cytotoxicity to L-929 cells.Pass

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the specific sample sizes for each non-clinical test (e.g., number of gloves tested for dimensions, physical properties, pinholes). However, the tests are based on recognized standards (ASTM, ISO), which typically specify sample sizes or sampling plans (e.g., AQL levels).

    • Data Provenance: The studies are non-clinical (laboratory/bench testing, and animal studies for biocompatibility). The document does not specify the country of origin of the test data but implies it was conducted by or for the manufacturer. All tests are retrospective, meaning they were performed on already manufactured product or materials to demonstrate compliance with standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is not applicable to this type of device and study. The "ground truth" for non-clinical performance and biocompatibility relies on objective measurements according to standardized test methods, not expert consensus or clinical interpretation.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable. This is not a study involving human subjective assessment where adjudication would be necessary. The results are objective measurements from laboratory tests.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a medical glove, not an AI-powered diagnostic or assistive technology. Therefore, an MRMC study and AI assistance are irrelevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is a medical glove and does not involve an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the performance of these gloves is defined by the acceptance criteria specified in internationally recognized standards (ASTM and ISO). These standards establish objective parameters and methods for evaluating physical properties (dimensions, tensile strength, elongation), integrity (pinholes), material properties (powder residual), and biological safety (biocompatibility).

    8. The sample size for the training set:

    Not applicable. This is a physical medical device, not a machine learning algorithm, so there is no concept of a "training set."

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211177

    Validate with FDA (Live)

    Date Cleared
    2021-07-17

    (88 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Powder Free Vinyl Patient Examination Gloves is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Powder Free Vinyl Patient Examination Gloves Are Patient Examination Gloves, Disposable, single use only and non-sterile. The gloves are made of vinyl materials and are powder free. The physical and performance characteristics of the devices meet all requirements of ASTM D5250-19 Standard Specification for Poly (Vinyl Chloride) Gloves for Medical Application.

    AI/ML Overview

    The provided document describes the acceptance criteria and performance data for Powder Free Vinyl Patient Examination Gloves (K211177). This is a medical device, and the data provided relates to its physical performance and biocompatibility, not an AI/ML-based diagnostic or imaging device. Therefore, many of the requested fields (such as sample size for test/training sets, data provenance, number of experts, adjudication method, MRMC studies, standalone algorithm performance, and ground truth types related to AI/ML) are not applicable (N/A) for this type of medical device submission.

    Here's the breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicsStandard / Acceptance CriteriaDevice PerformanceResult
    Physical Properties (ASTM D 5250-19)
    Overall Length (mm)230 mm for all sizes minimumS: 235-241mm; M: 233-240mm; L: 234-241mm; XL: 236-242mmMeets
    Width (mm)S: 85±5; M: 95±5; L: 105±5; XL: 115±5S: 86-88 mm; M: 96-98 mm; L: 106-108 mm; XL: 116-118 mmMeets
    Palm Thickness (mm)0.08mm minimum0.08mmMeets
    Finger Thickness (mm)0.08mm minimum0.08-0.10mmMeets
    Tensile Strength (Mpa) Before aging11Mpa minimum13.0-16.1MpaMeets
    Tensile Strength (Mpa) After aging11Mpa minimum13.0-15.8MpaMeets
    Ultimate Elongation (%) Before aging300% minimum310-410%Meets
    Ultimate Elongation (%) After aging300% minimum310-400%Meets
    Pinhole (Freedom from)ASTM D 5250-19, ASTM D5151-19, 21CFR800.20 G-I, AQL2.5125 gloves sampled and 1 piece leaks (Meets AQL2.5 for G-I inspection level)Meets
    Residual PowderNot more than 2mg per glove0.35-0.73mgMeets
    Biocompatibility
    Primary Skin Irritation TestISO 10993-10: Not an irritantUnder conditions of the study, not an irritantMeets
    Dermal Sensitization AssayISO 10993-10: Not an irritantUnder conditions of the study, not an irritantMeets
    Cytotoxicity TestISO 10993-5: No cytotoxic potentialUnder the conditions of this study, no cytotoxic potentialMeets

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Pinhole Test: 125 gloves were sampled.
    • Other Physical and Biocompatibility Tests: Sample sizes are not explicitly stated for all tests but are implied by the standards referenced (e.g., "N=5" for residual powder).
    • Data Provenance: The manufacturer is Shijiazhuang Hongjun Plastic Manufacture Co., Ltd. located in Shijiazhuang City, Hebei Province, China. The testing was conducted to meet international and US standards (ASTM, ISO, CFR). Specific country of origin for data collection for these non-clinical tests is not detailed beyond the manufacturer's location, but the tests generally follow standardized laboratory protocols rather than clinical data collection. This would be considered prospective testing against established specifications, not retrospective clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • N/A. This is not relevant for the type of device and testing performed. The "ground truth" for these tests is defined by the objective physical and chemical properties measured against recognized industry standards (ASTM, ISO). No expert clinical assessment of "ground truth" is involved.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • N/A. Adjudication methods are typically for subjective assessments or discrepancy resolution in clinical studies or expert evaluations, which is not applicable here. The tests are objective measurements based on specified standard methods.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • N/A. This is a physical non-diagnostic product (gloves). MRMC studies are for evaluating diagnostic tools involving human interpretation, often in conjunction with AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • N/A. This device does not have an algorithm or AI component.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The "ground truth" is defined by established engineering and material science standards and specifications (e.g., ASTM D5250-19, ISO 10993-10, ISO 10993-5, ASTM D6124-06, ASTM D5151-19). These standards provide objective, measurable criteria for acceptable performance and biocompatibility.

    8. The sample size for the training set

    • N/A. This is a physical non-diagnostic product. There is no "training set" in the context of AI/ML.

    9. How the ground truth for the training set was established

    • N/A. This is a physical non-diagnostic product. No AI/ML training set or associated ground truth establishment method is applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K210780

    Validate with FDA (Live)

    Date Cleared
    2021-05-14

    (60 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Poly Vinyl Chloride (PVC) films form a barrier to prevent contamination between patient and examiner. The PVC rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure. The Poly (vinyl chloride) glove acts as a barrier to prevent contamination between patient and examiner. The glove is manufactured in accordance with the requirements of ASTM D5250 and ASTM D5151.

    AI/ML Overview

    This document is a 510(k) Pre-Market Notification for Powder Free Vinyl Patient Examination Gloves. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving a novel device meets specific acceptance criteria in a clinical setting.

    Therefore, many of the requested sections are not applicable in this context. However, I can extract information related to the device's technical specifications and the non-clinical tests performed to ensure it meets established safety and performance standards for gloves.

    Here's a breakdown of the relevant information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally defined by compliance with established ASTM and ISO standards for medical examination gloves. The reported device performance is presented as meeting these standards.

    CharacteristicAcceptance Criteria (Standard)Reported Device Performance
    DimensionWidth (mm)
    Width (Small)80-90 mm82-88
    Width (Medium)90-100 mm94-98
    Width (Large)100-110 mm103-109
    Width (X-Large)110-120 mm114-117
    Thickness (mm)
    Fingertip≥0.05 mm (ASTM D 5250-06)0.09-0.10
    Palm≥0.08 mm (ASTM D 5250-06)0.10-0.11
    Physical PropertiesASTM D 5250-06 (Reapproved 2015)
    Tensile Strength (Before & After aging)≥11 MPa15-22 MPa
    Elongation Rate (Before & After aging)≥300%350-420%
    Freedom from PinholesHoles at Inspection Level I AQL2.5 (21 CFR 800.20, ASTM D5250-06, ASTM D5151-19)Met Acceptance Criteria; Holes at Inspection Level I AQL2.5
    Powder Residual<2 mg/glove (ASTM D 5250-06, ASTM D6124-06)0.1 mg
    Biocompatibility
    Primary Skin IrritationPasses (ISO 10993-10: 2010-08-01)Not a primary skin irritant
    Dermal SensitizationPasses (ISO 10993-10: 2010-08-01)Not a skin sensitizer
    CytotoxicityPasses (ISO 10993-5: 2009)Non-cytotoxicity to L-929 cells

    2. Sample size used for the test set and the data provenance
    The document does not specify the exact sample sizes used for each individual non-clinical test (e.g., how many gloves were tested for pinholes, or how many rabbits for skin irritation). The data provenance is implied to be from the manufacturer (Shandong Shangwei Medical Products Co.,Ltd in China) through in-house testing or contracted labs, as these are non-clinical tests used for product compliance. The data is retrospective in the sense that these tests are performed on manufactured product samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    This information is not applicable. The "ground truth" for these tests are objective measurements against established international and national standards (ASTM, ISO, CFR). There isn't a subjective "expert ground truth" for attributes like glove dimensions or tensile strength.

    4. Adjudication method for the test set
    Not applicable, as the tests involve objective measurements against predefined standards, not expert adjudication of subjective assessments.

    5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This is not an AI-powered diagnostic device. It is a medical examination glove.

    6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done
    Not applicable. This is not an AI-powered diagnostic device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
    The "ground truth" for the performance of the gloves is based on objective measurements performed according to validated test methods defined in recognized international and national standards (e.g., ASTM D5250 for dimensions and physical properties, ASTM D5151 for pinholes, ASTM D6124 for powder residual, ISO 10993 for biocompatibility).

    8. The sample size for the training set
    Not applicable. This is a manufactured product, not an AI model requiring a training set.

    9. How the ground truth for the training set was established
    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210799

    Validate with FDA (Live)

    Date Cleared
    2021-05-14

    (59 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Powder Free Vinyl Patient Examination Glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The subject device is powder free vinyl patient examination gloves. The subject device is colorless. The design of subject device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D5250. The subject device is nonsterile.

    AI/ML Overview

    This document describes the premarket notification for the "Powder Free Vinyl Patient Examination Glove" (K210799). It is a Class I device and the submission focuses on demonstrating substantial equivalence to a predicate device (K163168).

    Here's an analysis of the acceptance criteria and study data:

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria are implicitly defined by compliance with specific ASTM and ISO standards, and by showing equivalence to the predicate device, particularly regarding physical properties and biocompatibility.

    Acceptance Criteria (Standard/Requirement)Reported Device Performance (Subject Device)
    Physical Dimensions:
    Length (ASTM D5250)230mm (XS, S, M, L, XL) min
    Width (ASTM D5250)75mm (XS), 85mm (S), 95mm (M), 105mm (L), 115mm (XL) $\pm$ 5mm
    Finger Thickness (ASTM D5250)0.08mm min
    Palm Thickness (ASTM D5250)0.08mm min
    Physical Properties:
    Tensile Strength Before Aging (ASTM D5250)11MPa, min
    Ultimate Elongation Before Aging (ASTM D5250)300% min
    Tensile Strength After Aging (ASTM D5250)11MPa, min
    Ultimate Elongation After Aging (ASTM D5250)300% min
    Freedom from Holes (ASTM D5151)Be free from holes when tested in accordance with ASTM D5151, AQL=2.5
    Powder Content (ASTM D6124)< 0.02 mg per glove
    Biocompatibility:
    Irritation (ISO 10993-10)Under the conditions of the study, not an irritant
    Sensitization (ISO 10993-10)Under conditions of the study, not a sensitizer.
    Cytotoxicity (ISO 10993-5)Under conditions of the study, did not show potential toxicity to L-929 cells.
    LabelingMeets FDA's Requirement (Single use, powder free, device color, device name, glove size and quantity, Vinyl Examination Gloves, Non-Sterile information stated)
    MaterialVinyl
    Intended UseDisposable device for medical purposes, worn on examiner's hands to prevent contamination between patient and examiner.

    2. Sample sized used for the test set and the data provenance

    The document does not specify the exact sample sizes used for each test (e.g., number of gloves tested for tensile strength, pinholes, etc.). It only references compliance with the standards, which typically define sampling plans. The data provenance is not explicitly stated (e.g., country of origin, retrospective/prospective). However, given the manufacturer's location in China, it is highly probable that the testing was conducted in China.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable. The device is a medical glove, and its performance is evaluated against engineering and biological standards, not through expert human interpretation of medical data (like radiology images).

    4. Adjudication method for the test set

    This information is not applicable. Adjudication methods like "2+1" or "3+1" are typically used for establishing ground truth in expert-dependent diagnostic studies, not for the physical and biological testing of medical devices like gloves.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This type of study is relevant for AI-powered diagnostic tools, not for medical gloves. The device is not an AI-assisted tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A standalone performance study was not applicable/conducted in the context of an algorithm or AI. The performance studies conducted were on the physical and biological properties of the glove itself.

    7. The type of ground truth used

    The "ground truth" for this device's performance is established by compliance with recognized industry standards and regulatory requirements. This includes:

    • Engineering Standards: ASTM D5250 (Standard Specification for Poly (vinyl chloride) Gloves for Medical Application), ASTM D5151 (Standard Test Method for Detection of Holes in Medical Gloves), ASTM D6124 (Standard Test Method for Residual Powder on Medical Gloves).
    • Biocompatibility Standards: ISO 10993-10 (Tests For Irritation And Skin Sensitization), ISO 10993-5 (Tests For In Vitro Cytotoxicity).
    • Predicate Device Equivalence: Demonstrating that the subject device performs similarly or better than a legally marketed predicate device (K163168) in key functional and safety attributes.

    8. The sample size for the training set

    This information is not applicable. The device is a physical product (medical glove), not an AI algorithm requiring a training set.

    9. How the ground truth for the training set was established

    This information is not applicable. As explained above, there is no AI algorithm or training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210774

    Validate with FDA (Live)

    Date Cleared
    2021-05-14

    (60 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Poly Vinyl Chloride (PVC) films form a barrier to prevent contamination between patient and examiner. The PVC rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure. The Poly (vinyl chloride) glove acts as a barrier to prevent contamination between patient and examiner. The glove is manufactured in accordance with the requirements of ASTM D5250 and ASTM D5151.

    AI/ML Overview

    This document pertains to the 510(k) premarket notification for Powder Free Vinyl Patient Examination Gloves, Clear (non-colored). The information provided heavily focuses on proving substantial equivalence to a predicate device through non-clinical performance data, primarily compliance with established ASTM and ISO standards for medical gloves.

    Here's the breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    CharacteristicsAcceptance CriteriaReported Device Performance
    DimensionsASTM D 5250-06 (Reapproved 2015)Met ASTM D 5250-06 (Reapproved 2015)
    Length≥230mmSubject Device: 231-241mm (Predicate: 231-242mm) - Stated as "Similar"
    Width (Small)80-90 mmSubject Device: 81-89mm (Predicate: 82-89mm) - Stated as "Similar"
    Width (Medium)90-100mmSubject Device: 93-99mm (Predicate: 93-99mm) - Stated as "Similar"
    Width (Large)100-110mmSubject Device: 102-110mm (Predicate: 103-109mm) - Stated as "Similar"
    Width (X-Large)110-120 mmSubject Device: 111-119mm (Predicate: 114-119mm) - Stated as "Similar"
    Thickness (Fingertip)≥0.05mmSubject Device: 0.09-0.10mm (Predicate: 0.05-0.10mm) - Stated as "Similar"
    Thickness (Palm)≥0.08mmSubject Device: 0.10-0.11mm (Predicate: 0.09-0.13mm) - Stated as "Similar"
    Physical PropertiesASTM D 5250-06 (Reapproved 2015)Met ASTM D 5250-06 (Reapproved 2015)
    Tensile Strength≥11MPa (Before & After aging)Subject Device: 15-22 MPa (Predicate: 16-20 MPa) - Stated as "Similar"
    Elongation Rate≥300% (Before & After aging)Subject Device: 350-420% (Predicate: 380-410%) - Stated as "Similar"
    Freedom from Pinholes21 CFR 800.20 ASTM D5250-06 (Reapproved 2015) ASTM D5151-19 Holes at Inspection Level I AQL2.5Subject Device: "Passed Standard Acceptance Criteria"Holes at Inspection Level I AQL2.5 (Same as predicate)
    Powder ResidualASTM D5250-06 (Reapproved 2015) ASTM D6124-06 (Reapproved 2017) <2mg/gloveSubject Device: 0.1mg (Predicate: 0.3mg) - Stated as "Similar" and "Meets <2mg/glove"
    BiocompatibilityISO 10993-5:2009 (Cytotoxicity) ISO 10993-10:2010 (Irritation & Sensitization)Primary Skin Irritation: Passed (Not a primary skin irritant). Met ISO 10993-10: 2010-08-01. Dermal Sensitization: Passed (Not a skin sensitizer). Met ISO 10993-10: 2010-08-01. Cytotoxicity: Passed (Non-cytotoxicity to L-929 cells). Met ISO 10993-5: 2009. All stated as "Same" or fully compliant with standards.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: The document does not explicitly state the sample sizes used for each specific test (e.g., how many gloves were tested for dimensions, pinholes, tensile strength). It refers to compliance with standards like ASTM D5250. These standards typically define the sampling plans.
    • Data Provenance: The tests were conducted by the manufacturer, Shandong Jieshi Medical Products Co.,Ltd, located in China. The data would be considered prospective, as it was generated to support the 510(k) submission for this specific device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This document concerns medical device performance testing, not diagnostic AI. Therefore, the concept of "experts establishing ground truth" in the context of clinical interpretation (like radiologists for imaging) is not applicable here. The "ground truth" for these tests is defined by established regulatory standards (ASTM, ISO) and the laboratory methods prescribed within them. The "experts" would be the qualified laboratory personnel performing these physical and chemical tests. Their specific qualification details are not provided in this regulatory summary but are presumed to be in compliance with good laboratory practices.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not a study involving human interpretation or subjective judgment that would require an adjudication process. The tests are objective measurements against defined physical and chemical criteria.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document is for a physical medical device (examination gloves), not an AI diagnostic algorithm. Clinical testing was explicitly stated as "not needed for this device."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is for a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" here is defined by established consensus standards from organizations like ASTM (American Society for Testing and Materials) and ISO (International Organization for Standardization). These standards specify physical properties, test methods, and acceptable limits for medical gloves. This includes:

    • ASTM D5250 (Standard specification for poly (vinyl chloride) gloves for medical application)
    • ASTM D5151 (Standard Test Method for Detection of Holes in Medical Gloves)
    • ASTM D6124 (Standard Test Method for Residual Powder on Medical Gloves)
    • ISO 10993-5 (Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity)
    • ISO 10993-10 (Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization)

    8. The sample size for the training set

    Not applicable. This is for a physical medical device, not an AI/machine learning model. There is no concept of a "training set" for this type of submission. The manufacturing process itself (Good Manufacturing Practices) would ensure consistent product quality, but there's no data training aspect.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for a physical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191292

    Validate with FDA (Live)

    Date Cleared
    2019-11-05

    (175 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    The proposed device is Powder Free Examination Gloves. The proposed device is Clear (non-colored). The proposed device is non-sterile.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided text, focusing on what's available and noting what's not:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test/CharacteristicAcceptance CriteriaReported Device Performance (K191292)
    Dimensions - LengthASTM D 5250-06 (Reapproved 2011): ≥ 230mm233-241mm
    Dimensions - WidthASTM D 5250-06 (Reapproved 2011):Small 80-90 mmMedium 90-100mmLarge 100-110mmX large 110-120 mm84-88 mm94-98 mm103-109 mm114-117 mm
    Dimensions - ThicknessASTM D 5250-06 (Reapproved 2011):Fingertip ≥0.05mmPalm ≥0.08mmFingertip: 0.08-0.11mmPalm: 0.10-0.11mm
    Physical Properties (Tensile Strength)ASTM D 5250-06 (Reapproved 2011): ≥11MPa (Before & After aging)15-17 MPa (for "Physical Properties")
    Physical Properties (Elongation Rate)ASTM D 5250-06 (Reapproved 2011): ≥300% (Before & After aging)350-390% (for "Physical Properties")
    Freedom from PinholesASTM D5151-06 (Reapproved 2011): Holes at Inspection Level I, AQL2.5, Act/Re: 10/11Water leakage test: 3 noncompliance is allowed (This is a different test/criterion reported in the comparison table, but within the "similar" category for pinholes). The detailed result for the ASTM D5151-06 specific criterion is not explicitly stated as "Pass" or "Fail" in the result column, but the overall "similar" comparison implies conformity.
    Powder ResidualASTM D6124-06 (Reaffirmation 2011): <2mg/gloveMean: 0.1mg/pcs
    Biocompatibility - Primary Skin IrritationISO 10993-10: 2010-08-01: Non-irritant or non-sensitizer (from predicate)PassesUnder the conditions of the study, the subject device is not a skin sensitizer. (This result is specifically for dermal sensitization, not explicitly primary skin irritation. However, the predicate's acceptance criteria covers "non-irritant or non-sensitizer".)
    Biocompatibility - Dermal SensitizationISO 10993-10: 2010-08-01: Dermal sensitization in the guinea pigsPassesUnder the conditions of the study, the subject device is not a skin sensitizer.
    Biocompatibility - CytotoxicityISO 10993-5: 2009: Non-cytotoxicity to L-929 cells (from predicate, implied)PassUnder the conditions of this study, the test article was non-cytotoxicity to L-929 cells.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated for each test, but quantities like "Checked on 5pcs sub-samples (N=5)" are mentioned for "Residual Powder." "Inspection Level I AQL2.5" is an AQL (Acceptance Quality Limit) sampling plan, implying a specific sample size based on the lot size, but the exact number of units tested is not provided.
    • Data Provenance: The studies were non-clinical tests performed by Anhui Powerguard Technology Co.,Ltd, likely in China (country of origin of the manufacturer). The studies are retrospective in the sense that they were conducted specifically for this 510(k) submission, not ongoing clinical trials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • This device is a medical glove undergoing non-clinical performance testing. The "ground truth" is established by adherence to recognized international standards (e.g., ASTM, ISO). Therefore, "experts" in the context of clinical interpretation or diagnosis are not relevant here. The "experts" would be the scientists and engineers conducting the tests according to the standard protocols. Their qualifications are not detailed.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This is not a study involving human interpretation or adjudication of results in the way described (e.g., radiologists reviewing images). Performance is measured against physical and chemical standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC study was not done. This device is a medical glove, not an AI or diagnostic imaging device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This is not an algorithmic device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • The "ground truth" for the device's performance is defined by the acceptance criteria of recognized international standards (ASTM D5250, ASTM D5151, ASTM D6124, ISO 10993-5, ISO 10993-10). These standards specify the test methods and the acceptable range of performance for each characteristic (e.g., tensile strength, elongation, freedom from pinholes).

    8. The sample size for the training set:

    • Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. This is not an AI/machine learning device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K180381

    Validate with FDA (Live)

    Date Cleared
    2018-11-07

    (268 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is disposable non-sterile device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Powder Free Vinyl Patient Examination Gloves are Patient Examination Gloves, Disposable, single use only and non-sterile. The gloves are made of vinyl materials and are powder free. The physical and performance characteristics of the devices meet all requirements of ASTM D5250-06 (Reapproved 2015) Standard Specification For Poly(Vinyl Chloride) Gloves For Medical Application.

    AI/ML Overview

    This document is a 510(k) summary for "Powder Free Vinyl Patient Examination Gloves," a Class I medical device. The information provided pertains to the substantial equivalence of the subject device to a predicate device, focusing on non-clinical performance characteristics to demonstrate safety and effectiveness.

    Here's the breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria are generally established by the ASTM D 5250-06 (Reapproved 2015) Standard Specification for Poly(Vinyl Chloride) Gloves For Medical Application. The predicate device and the subject device both aim to meet these requirements.

    CharacteristicsAcceptance Criteria (ASTM D 5250-06 (2015) requirements)Reported Device Performance (Subject Device K180381)Result of Comparison (Predicate vs Subject)Conclusion (Subject Device performance)
    Physical Properties:
    Tensile Strength (Before aging)11 Mpa minimum14.4-17.1 MpaSame (Met/Exceeded)Pass (Meets/Exceeds criteria)
    Tensile Strength (After aging)11 Mpa minimum14.1-16.9 MpaSame (Met/Exceeded)Pass (Meets/Exceeds criteria)
    Elongation (Before aging)300% minimum390-480%Same (Met/Exceeded)Pass (Meets/Exceeds criteria)
    Elongation (After aging)300% minimum380-470%Same (Met/Exceeded)Pass (Meets/Exceeds criteria)
    Freedom from PinholeG-I, AQL 2.5 (Meets ASTM D 5250-06 (2015) requirements)Meets AQL 2.5 requirementsSamePass (Meets criteria)
    Dimensions:Meets ASTM D 5250-06 (2015) requirementsSamePass (Meets criteria)
    - Overall Length (mm)230 for all sizes minimumS: 233-241mm, M:231-242mm, L:230-244mm, XL: 234-245mmN/A (Directly compared to standard)Pass (Meets/Exceeds criteria)
    - Width (mm)S: 85 ± 5, M: 95 ± 5, L: 105±5, XL: 115±5S: 87-88 mm, M: 97-98 mm, L: 107-108 mm, XL: 118-119 mmN/A (Directly compared to standard)Pass (Meets criteria)
    - Palm Thickness (mm)0.08mm minimum0.08mmN/A (Directly compared to standard)Pass (Meets criteria)
    - Finger Thickness (mm)0.05mm minimum0.09-0.11mmN/A (Directly compared to standard)Pass (Meets/Exceeds criteria)
    Residual PowderNot more than 2 mg per glove (Meets Applicable Definition for Powder Free)0.20 mg/gloveSame (Met/Exceeded)Pass (Meets/Exceeds criteria)
    Biocompatibility:
    - Primary Skin IrritationUnder conditions of the study, not an irritantUnder conditions of the study, not an irritantSamePass (Not an irritant)
    - Dermal SensitizationUnder conditions of the study, not an irritantUnder conditions of the study, not an irritantSamePass (Not an irritant)
    - Cytotoxicity TestN/A (Predicate had no data)Under the conditions of this study, not a cytotoxic potentialDifferent (Subject had data)Pass (Not a cytotoxic potential)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the specific numerical sample sizes for each test. Instead, it refers to "Inspection Level and AQL" (Acceptable Quality Level) from the ASTM standards for some tests (e.g., S-2, AQL 4.0 for dimensions and tensile strength; G-I, AQL 2.5 for pinhole). These AQLs imply a sampling plan rather than a fixed sample size.

    • Pinhole: Inspection Level G-I, AQL 2.5.
    • Dimensions, Tensile Strength, Ultimate Elongation: Inspection Level S-2, AQL 4.0.
    • Residual Powder: N=5 (5 gloves were tested).
    • Biocompatibility Tests (Primary Skin Irritation, Dermal Sensitization, Cytotoxicity Test): No sample sizes are specified for these in the table; the results are qualitative ("not an irritant," "not a cytotoxic potential").

    The data provenance is from Shanxi Hongjin Plastic Technology Co., Ltd in China, as indicated by their address on the document. The studies are non-clinical lab tests conducted to support the 510(k) submission, implying a prospective data generation process for the purpose of regulatory approval.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This section is not applicable as the device is a medical glove, and the "ground truth" is based on objective physical, chemical, and biological performance characteristics measured by laboratory tests and adherence to industry standards (ASTM D5250-06), not expert medical interpretation of images or clinical outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in studies involving subjective human interpretation (e.g., radiologists reading medical images) to establish consensus ground truth. The tests for these gloves are objective measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This type of study relates to AI-assisted diagnostic devices and human reader performance, not physical medical devices like examination gloves.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. This concept applies to AI/software as a medical device, where the algorithm performs a task independently. The tests performed here are direct measurements of a physical product's attributes.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance is established by objective measurements against pre-defined specifications and compliance with recognized industry standards, specifically ASTM D 5250-06 (Reapproved 2015). For biocompatibility, it's based on standard biological evaluation tests determining if the material is an irritant or cytotoxic.

    8. The sample size for the training set

    This is not applicable. The concept of a "training set" refers to data used to train machine learning models. This document describes the testing and performance of a physical product according to established standards, not an AI/ML device.

    9. How the ground truth for the training set was established

    This is not applicable for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180260

    Validate with FDA (Live)

    Date Cleared
    2018-06-12

    (133 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is disposable non-sterile device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Powder Free Vinyl Patient Examination Gloves (Yellow, White) are Patient Examination Gloves, Disposable, single use only and non-sterile. The gloves are made of vinyl materials and are powder free. The physical and performance characteristics of the devices meet all requirements of ASTM D5250-06 (Reapproved 2015) Standard Specification For Poly (Vinyl Chloride) Gloves For Medical Application.

    AI/ML Overview

    This document is a 510(k) Pre-Market Notification for Powder Free Vinyl Patient Examination Gloves. It focuses on demonstrating substantial equivalence to a predicate device, rather than proving the device meets acceptance criteria through clinical studies in the typical sense of AI/software devices. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" here refer to compliance with recognized standards and performance thresholds for examination gloves.

    Here's a breakdown of the requested information based on the provided text, interpreted in the context of a medical glove submission:

    1. A table of acceptance criteria and the reported device performance

    Testing ItemsAcceptance Criteria (FDA-recognized Standard Requirements)Reported Device Performance (Actual Testing Results)Conclusion
    Overall Length (mm)230 mm for all sizes minimumS: 233-244mm, M: 231-241mm, L: 230-243mm, XL: 234-242mmPass
    Width (mm)S: 85±5, M: 95±5, L: 105±5, XL: 115±5S: 87-88 mm, M: 97-98 mm, L: 107-108 mm, XL: 118-119 mmPass
    Palm Thickness (mm)0.08mm minimum0.08mmPass
    Finger Thickness (mm)0.05mm minimum0.09-0.12mmPass
    Tensile Strength (Mpa) - Before aging11Mpa minimum15.1-18.1MpaPass
    Tensile Strength (Mpa) - After aging11Mpa minimum14.9-17.7MpaPass
    Ultimate Elongation (%) - Before aging300% minimum390-470%Pass
    Ultimate Elongation (%) - After aging300% minimum380-490%Pass
    PinholeASTM D 5250-06 (2015) requirements / Meet AQL 2.5 requirementsMeets AQL 2.5 requirementsPass
    Residual PowderNot more than 2mg per glove0.18 mg/glovePass
    Primary Skin Irritation TestUnder conditions of the study, not an irritantUnder conditions of the study, not an irritantPass
    Dermal Sensitization StudyUnder conditions of the study, not a sensitizerUnder conditions of the study, not a sensitizerPass
    Cytotoxicity TestUnder the conditions of this study, not a cytotoxic potentialUnder the conditions of this study, not a cytotoxic potentialPass
    Overall PerformanceMeets ASTM D5250-06 (Reapproved 2015) Standard Specification For Poly (Vinyl Chloride) Gloves For Medical ApplicationConforms fully to ASTM D 5250-06 (2015) standard, applicable 21 CFR references, and FDA requirementsOverall Compliance Demonstrated

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: The sample sizes are indicated by the "Inspection Level and AQL" (Acceptable Quality Limit) for most physical and performance tests, such as S-2, AQL4.0, and G-I, AQL2.5. For Residual Powder, the sample size was N=5. These AQL levels refer to sampling plans (e.g., ISO 2859-1 or ANSI/ASQ Z1.4), where a specific number of units are inspected from a lot, and the lot is accepted or rejected based on the number of defects found. The exact number of units tested isn't explicitly stated but would be determined by the AQL tables for the lot size.
    • Data Provenance: The manufacturing entity is Shanxi Hongjin Plastic Technology Co., LTD, located in China. The testing would have been conducted by or on behalf of this company. The data is prospective in the sense that it's generated for the purpose of demonstrating compliance for this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This product is a medical device (patient examination gloves) tested against recognized physical, chemical, and biological standards. "Ground truth" in this context is established by the accepted scientific methodologies and criteria outlined in the ASTM and ISO standards, rather than expert human interpretation in the way it applies to diagnostic AI. Therefore, there are no "experts" in the sense of clinicians establishing ground truth for individual cases. The ground truth is the objective measurement against the specified standard requirements.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 are used for resolving discrepancies in expert interpretations (e.g., in medical image reading). For device performance testing against quantitative standards, the measurements are objective.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a physical product (medical glove), not an AI or software algorithm. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm or AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the performance tests in this submission is the objective measurement against scientifically established and recognized standards (e.g., ASTM D5250-06 for physical properties, specific thresholds for residual powder, and standard biocompatibility tests for irritation and sensitization). It is not pathology, outcomes data, or expert consensus on a diagnostic interpretation.

    8. The sample size for the training set

    Not applicable. This is not a machine learning/AI device, so there is no training set in that context. Statistical sampling is used for quality control during manufacturing and for conformity testing.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for a machine learning model, this question is not relevant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160675

    Validate with FDA (Live)

    Date Cleared
    2016-08-02

    (145 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Powder free vinyl patient examination glove, colored (yellow) is a non-sterile, disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Powder free vinyl patient examination glove, colored (yellow) are patient examination gloves available in S, M, L, XL. They are provided non-sterile and meet the entire requirement of ASTM standard D5250-06(reapproved), except for sterility requirements.

    AI/ML Overview

    This document describes the 510(k) premarket notification for the "Powder free vinyl patient examination glove, colored (yellow)" (K160675). The submission aims to demonstrate substantial equivalence to a predicate device, the "White vinyl patient examination glove" (K143277) from Life safety products (Hui Zhou) Co., Ltd.

    Here's the breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (Subject Device, K160675)Predicate Device Performance (K143277)Comparison Comment
    Dimensions
    Length$\geq$ 230mm243mmSubstantially Equivalent
    Width (S Size)85±5mm85mmSubstantially Equivalent
    Width (M Size)95±5mm95mmSubstantially Equivalent
    Width (L Size)105±5mm105mmSubstantially Equivalent
    Width (XL Size)115±5mm115mmSubstantially Equivalent
    Finger thickness$\geq$ 0.05mm0.05mmSubstantially Equivalent
    Palm thickness$\geq$ 0.08mm0.08mmSubstantially Equivalent
    Overall dimensionMeets ASTM D5250-06Meets ASTM D5250-06Substantially Equivalent
    Physical Properties
    Tensile strength$\geq$ 11MPa15.7MPaSubstantially Equivalent
    Elongation$\geq$ 300%385%Substantially Equivalent
    Overall physical propertiesMeets ASTM D5250-06Meets ASTM D5250-06Substantially Equivalent
    Freedom from PinholePinhole = 0/200; AQL 2.5; Meets ASTM D5151-06Pinhole = 10/500; AQL 1.5; Meets ASTM D5151-06Substantially Equivalent
    Powder ResidualPowder residual = 0.5mg per glove; Meets ASTM D5250-06 and D6124-06Powder residual = 1mg; Meets ASTM D5250-06 and D6124-06Substantially Equivalent
    CytotoxicityPer ISO10993-5, under the conditions tested, the subject glove is non-cytotoxic.Per ISO10993-5, under the conditions tested, the subject glove is non-cytotoxic.Substantially Equivalent
    BiocompatibilityPer ISO10993-10, under the conditions tested, the subject glove was non-sensitizing and non-irritating.Per ISO10993-10, under the conditions tested, the subject glove was non-sensitizing and non-irritating.Substantially Equivalent

    Note: The "acceptance criteria" for the subject device are generally presented as meeting relevant ASTM and ISO standards, with specific values reported for the subject device's performance which are then compared to the predicate device's reported performance.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a test set from a clinical study for the device, as this is a device clearance based on substantial equivalence to a predicate, heavily relying on non-clinical performance data and conformity to recognized standards.

    • For Freedom from pinholes, the subject device reported 0/200, implying a sample size of 200 gloves tested for pinholes. The predicate device reported 10/500, implying a sample size of 500 gloves.
    • For Powder residual, the measurement is "0.5mg per glove," which would typically be an average or maximum over a sampled number of gloves, but the specific sample size for this test is not provided.
    • Biocompatibility testing (Cytotoxicity, Sensitization, Irritation) involved testing samples of the subject device. The specific number of samples tested for these biocompatibility evaluations is not mentioned in the provided text.

    Data Provenance: The document does not specify the country of origin for the testing data, but the manufacturer is based in China. The data would be retrospective relative to the 510(k) submission, as it represents tests already conducted to support the filing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable as the device is a medical glove, and the evaluation relies on objective physical and chemical testing against established standards, not on expert interpretation of results like in diagnostic imaging. Ground truth for these types of devices is based on laboratory measurements and adherence to specified performance metrics.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, the evaluation is based on objective laboratory testing against performance standards, which does not involve subjective adjudication or consensus methods typically seen in clinical studies with human readers.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC study was done. This device is a patient examination glove, not an AI-assisted diagnostic tool. Therefore, the concept of human readers improving with or without AI assistance is irrelevant and not part of the submission for this type of device.

    6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study

    Not applicable. This is a physical medical device (glove), not an algorithm or software. Therefore, a standalone performance study in the context of AI algorithms is not relevant.

    7. Type of Ground Truth Used

    The "ground truth" for this device's performance is established by:

    • Standardized laboratory measurements: Adherence to physical properties (length, width, thickness, tensile strength, elongation), freedom from pinholes, and powder residual, as defined by ASTM standards (D5250-06, D5151-06, D6124-06).
    • Biocompatibility testing results: Based on ISO 10993-5 (Cytotoxicity) and ISO 10993-10 (Sensitization and Irritation), assessing the biological response of the device material.

    8. Sample Size for the Training Set

    Not applicable. This is a physical medical device, not a machine learning model. Therefore, there is no "training set" in the context of AI development.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set for an AI model, the concept of establishing ground truth for it does not apply here.

    Ask a Question

    Ask a specific question about this device

    K Number
    K152692

    Validate with FDA (Live)

    Date Cleared
    2015-12-21

    (91 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Powder Free Vinyl Patient Examination Gloves (Green) is disposable non-sterile device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Powder Free Vinyl Patient Examination Gloves (Green) are Patient Examination Gloves, Disposable, single use only and non-sterile. The gloves are made of vinyl materials and are powder free. The physical and performance characteristics of the devices meet all requirements of ASTM D5250-06 (Reapproved 2011)--Standard Specification For Poly (Vinyl Chloride) Gloves For Medical Application.

    AI/ML Overview

    The provided text describes the submission for a 510(k) premarket notification for "Powder Free Vinyl Patient Examination Gloves (Green)". This document type does not include information about AI/ML device performance or large-scale comparative effectiveness studies involving human readers, as it pertains to a Class I medical device (patient examination gloves). The "acceptance criteria" and "study" described herein are related to the physical and chemical properties of the gloves, demonstrating their substantial equivalence to a predicate device.

    Therefore, many of the requested fields regarding AI/ML device performance, human reader studies, and AI-specific ground truth establishment cannot be answered from the provided text.

    Here's a breakdown of the available information regarding the glove's acceptance criteria and the study proving it meets these criteria:


    Device Name: Powder Free Vinyl Patient Examination Gloves (Green)
    Device Type: Patient Examination Glove (Class I Medical Device)
    Regulation Number: 21 CFR 880.6250


    1. Table of Acceptance Criteria and Reported Device Performance:

    The document presents a comparison between the subject device and a predicate device (Tangshan Hongyun Plastic Products Company Limited Powder-Free Yellow Vinyl Patient Examination Gloves, K141878) against various standards.

    CharacteristicStandardAcceptance Criteria (Predicate Device)Reported Device Performance (Subject Device)Result of Comparison
    General
    Product CodeLYZLYZSame
    Intended UseNon-sterile disposable for medical purposes, worn on hand/finger to prevent contamination.Non-sterile disposable for medical purposes, worn on hand/finger to prevent contamination.Same
    SizeS/M/L/XLS/M/L/XLSame
    Materials UsedPVCPVCSame
    Dusting or Donning PowderPUPUSame
    Single Patient Use/Single patient useSingle patient useSimilar
    Physical Properties
    Dimensions: LengthASTM D 5250-06 (2011)Meets 230 mm min for all sizesMeets 230 mm min for all sizesSimilar
    Dimensions: WidthASTM D 5250-06 (2011)S: 83-86 mm, M: 94-97 mm, L: 104-107 mm, XL: 113-115 mmS: 85±5 mm, M: 95±5 mm, L: 105±5 mm, XL: 115±5 mmSimilar
    Dimensions: ThicknessASTM D 5250-06 (2011)Palm: 0.08 mm min, Finger: 0.05 mm minPalm: 0.08 mm min, Finger: 0.05 mm minSimilar
    Tensile Strength (before/after aging)ASTM D 5250-06 (2011)≥11 MPa≥11 MPaSimilar
    Elongation (before/after aging)ASTM D 5250-06 (2011)≥300%≥300%Similar
    Freedom from PinholesASTM D 5250-06 (2011), ASTM D5151-06 (Reapproved 2011)G-I, AQL 2.5G-I, AQL 2.5Similar
    Residual PowderASTM D 5250-06 (2011), ASTM D6124-06 (Reaffirmation 2011)Below 2mg of residual powderNot more than 2mg per gloveSimilar
    Biocompatibility
    Skin IrritationISO 10993-10Not an irritantNot an irritantSame
    SensitizationISO 10993-10Not a sensitizerNot a sensitizerSame
    Performance Data
    ASTM Standards MetASTM D5151-06(2011), ASTM D 5250-06 (2011), ASTM D6124-06(2011)ASTM D5151-06(2011), ASTM D 5250-06 (2011), ASTM D6124-06(2011)Same

    The acceptance criteria are generally "Meets" the specified ASTM standards and biocompatibility results (e.g., "not an irritant"). The reported device performance indicates that the subject device also meets these criteria, supporting substantial equivalence.


    2. Sample Size Used for the Test Set and the Data Provenance:

    The document states that the physical and performance characteristics meet "all requirements of ASTM D5250-06 (Reapproved 2011)". This implies that standard testing methodologies as defined by these ASTM standards were followed. However, the specific sample sizes used for each test (e.g., how many gloves were tested for pinholes, tensile strength, etc.) are not explicitly provided in this summary.

    • Data Provenance: The testing was conducted by Supermax Plastic Products Co., Ltd. in China (Hebei Province). The data is non-clinical testing performed to support substantial equivalence. It is inherently retrospective in the context of the 510(k) submission, as these tests would have been completed prior to submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    Not applicable. This is a physical product testing scenario, not a study requiring expert clinical labeling or ground truth establishment in the context of AI/ML. The "ground truth" here is determined by objective measurements against established ASTM and ISO standards by qualified testing personnel.


    4. Adjudication Method for the Test Set:

    Not applicable. This is not a study requiring adjudication of expert opinions. Device performance is measured against objective, standardized criteria.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This document pertains to patient examination gloves, not an AI-assisted diagnostic device.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This document pertains to patient examination gloves, not an algorithm.


    7. The Type of Ground Truth Used:

    The "ground truth" for this device is based on objective, standardized measurements against the specifications outlined in relevant industry standards, specifically:

    • ASTM D 5250-06 (2011) - Standard Specification For Poly (Vinyl Chloride) Gloves For Medical Application
    • ASTM D5151-06 (2011) - Standard Test Method For Detection Of Holes In Medical Gloves
    • ASTM D6124-06 (2011) - Standard Test Method For Residual Powder On Medical Gloves
    • ISO 10993-10 (third edition 2010-08-01) - Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
    • 21 CFR 800.20 - General requirements for device labeling

    8. The Sample Size for the Training Set:

    Not applicable. This is not an AI/ML device, so there is no training set in that context. The "training" for manufacturing would be industrial process control, not data training.


    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no AI/ML training set. The manufacturing processes would be subject to quality control and good manufacturing practices (GMP) to ensure the product meets specifications.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 9