510(k) Data Aggregation

    K Number
    K141965

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-07-02

    (346 days)

    Product Code
    Regulation Number
    870.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pleural Catheter Kits are indicated for intermittent, long term drainage of symptomatic, recurrent, pleural effusion, mcluding malignant pleural effusion and other recurrent effusions that do not respond to medical management of the underlying disease. The devices are indicated for 1) the palliation of dyspnea effusion and 2) providing pleurodesis (resolution of the pleural effusion).

    The Pleurx Drainage Kits and Drainage Line Set are indicated for use only with the Pleurx Catheter for intermittent drainage. The Drainage Line Kit is used to drain fluid using standard wall suction, water seal drainage system, vacuum bottle or other appropriate method.

    The Pleurx Catheter Access Kit is intended to provide access to the Pleurx Catheter to inject or withdraw fluids

    Device Description

    The Pleurx Pleural Catheter System provides patients with a convenient method to relieve pleural effusion symptoms at home. The primary components of the Pleurx Catheter System are the Pleurx Pleural Catheter and the Pleurx Drainage Kits.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Pleurx Pleural Catheter System. This document focuses on demonstrating substantial equivalence to a predicate device, rather than providing a study where the device performance is measured against specific acceptance criteria in a clinical setting.

    Therefore, the requested information regarding acceptance criteria, device performance, sample sizes, expert ground truth, adjudication methods, MRMC studies, standalone performance, and training set details cannot be fully extracted from this document. The submission relies on non-clinical (bench) testing to support its claims of substantial equivalence for a labeling update.

    Here's what can be inferred and what cannot:

    1. Table of acceptance criteria and the reported device performance:

    This 510(k) submission updates previously cleared device labeling to include the use of sclerosing agents (Talc and Bleomycin) with the Pleurx Pleural Catheter System. The acceptance criteria for this specific submission relate to the validation of this change through biocompatibility and bench testing. The "performance data" section lists standards and tests conducted, implying that the device "passed" or met the requirements of these standards. However, no specific quantitative acceptance criteria or corresponding reported device performance values are provided for the device's clinical function in this document.

    For example, for Biocompatibility, the acceptance criterion would be "meets ISO 10993-1, -3, -4, -5, -6, -10, -11 requirements" and the reported device performance is "passed" or "met." The specific metrics and thresholds for each test are not detailed in this summary.

    CharacteristicAcceptance Criteria (Implied)Reported Device Performance (Implied)
    BiocompatibilityMeets requirements of listed ISO 10993 standardsPassed
    ResidualsMeets requirements of AAMI/ANSI/ISO 10993-7Passed
    Performance (e.g., sterilization, packaging)Meets requirements of listed EN, ISO, ASTM, AAMI standardsPassed

    2. Sample size used for the test set and the data provenance:

    • Test set sample size: Not specified. The document refers to "non-clinical tests," which likely involve a number of samples for each bench test conducted according to the standards.
    • Data provenance: Non-clinical (bench) testing data. The country of origin is not specified, but the standards cited are international (ISO, EN, ASTM, AAMI).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable. Given this is a non-clinical (bench) testing submission, expert opinion for establishing a "ground truth" in the clinical sense is not relevant. The "ground truth" for these tests would be the established pass/fail criteria within the cited international standards by which the device was evaluated.

    4. Adjudication method for the test set:

    Not applicable. This type of adjudication is typically for clinical studies where subjective assessments might be involved. For bench testing, the results are typically objective measurements against a standard.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is not an AI/software device and no MRMC study was conducted. The document explicitly states: "N/A – No clinical tests were conducted for this submission."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is not an AI/software device.

    7. The type of ground truth used:

    For the non-clinical tests, the "ground truth" is defined by the specific test method and acceptance criteria outlined in the cited international standards (e.g., ISO 10993 for biocompatibility, ISO 11135 for ethylene oxide sterilization).

    8. The sample size for the training set:

    Not applicable. This is not an AI/software device, so there is no training set in that context.

    9. How the ground truth for the training set was established:

    Not applicable, for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121849

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-10-18

    (115 days)

    Product Code
    Regulation Number
    870.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pleurx Pleural Catheter Kits are indicated for intermittent, long term drainage of symptomatic, recurrent, pleural effusion, including malignant pleural effusion and other recurrent effusions that do not respond to medical management of the underlying disease. The devices are indicated for 1) the palliation of dyspnea due to pleural effusion and 2) providing pleurodesis (resolution of the pleural effusion).

    The Pleurx Drainage Kits and Drainage Line Set are indicated for use only with the Pleurx Catheter for intermittent drainage. The Drainage Line Kit is used to drain fluid using standard wall suction, water seal drainage system, vacuum bottle or other appropriate method.

    The Pleurx Dressing Kits are indicated for dressing of a catheter and exit site.

    The Pleurx Catheter Access Kit is intended to provide access to the Pleurx Catheter for aspiration and catheter maintenance.

    The Pleurx Catheter Insertion Stylet is intended to aid in the percutaneous insertion of the Pleurx Catheter into the pleural space.

    The Valved Peelable Introducers are intended for use in the percutaneous insertion of a catheter into the pleural space.

    Device Description

    The Pleurx Pleural Cather System provides patients with a convenient method to relieve pleural effusion symptoms at home. The primary components of the Pleurx Catheter System are the Pleurx Pleural Catheter and the Pleurx Drainage Kits.

    AI/ML Overview

    The provided document describes the 510(k) submission for the CareFusion Pleurx Pleural Catheter System. This submission focuses on expanding the indications for use and adding accessories to an existing legally marketed device. Therefore, the information provided primarily concerns non-clinical testing for substantial equivalence, rather than a clinical study evaluating the device's performance against specific acceptance criteria for efficacy or accuracy in the way an AI/ML device would be assessed.

    Here's a breakdown of the requested information based on the provided text, with notes where the information is not applicable (N/A) due to the nature of the device and submission:

    1. Table of Acceptance Criteria and Reported Device Performance

    For this device, the "acceptance criteria" are compliance with various recognized standards for medical device safety and performance, and the "reported device performance" is the conclusion that the device meets or exceeds these standards, demonstrating substantial equivalence to predicate devices. There are no specific quantitative clinical performance metrics (like sensitivity, specificity, or accuracy) reported that would be typical for an AI/ML device.

    CharacteristicAcceptance Criteria (Standard/Test/FDA Guidance)Reported Device Performance
    BiocompatibilityAAMI/ANSI/ISO 10993-1:2009 Biological evaluation of Medical Devices Part 1: Evaluation and TestingMet requirements
    BiocompatibilityAAMI/ANSI/ISO 10993-3: 2003 (R) 2009 Biological Evaluation of Medical Devices - Part 3 Tests for Genotoxicity, Carcinogenicity and Reproductive ToxicityMet requirements
    BiocompatibilityAAMI/ANSI/ISO 10993-4: 2002 (R) 2009 Biological Evaluation of Medical Devices - Part 4 Selection of Tests for Interaction with BloodMet requirements
    BiocompatibilityAAMI/ANSI/ISO 10993-5: 2009 Biological Evaluation of Medical Devices - Part 5 Tests for In Vitro CytotoxicityMet requirements
    BiocompatibilityAAMI/ANSI/ISO 10993-6: 2007 (R) 2010 Biological Evaluation of Medical Devices - Part 6 Tests for Local Effects After ImplantationMet requirements
    ResidualsAAMI/ANSI/ISO 10993-7:2008 Biological evaluation of Medical Devices Part 7: Ethylene Oxide Sterilization ResidualsMet requirements
    BiocompatibilityAAMI/ANSI/ISO 10993-10: 2010 Biological Evaluation of Medical Devices - Part 10 Tests for Irritation and Skin SensitivityMet requirements
    BiocompatibilityAAMI/ANSI/ISO 10993-11: 2006 Biological Evaluation of Medical Devices - Part 11 Tests for Systemic ToxicityMet requirements
    BiocompatibilityUSP <661> Containers - Plastics, Physiochemical TestsMet requirements
    PerformanceEN 1617:1997 Sterile Drainage Catheters and Accessory Devices for Single UseMet requirements
    PerformanceEN 1618:1997 Catheters Other Than Intravascular Catheters - Test Methods for Common PropertiesMet requirements
    PerformanceISO 11070 Sterile, Single-use Intravascular CathetersMet requirements
    PerformanceANSI/AAMI/ISO 11607-1,2:2006 Packaging for Terminally Sterilized Medical DevicesMet requirements
    PerformanceASTM F1980-07 Accelerated Aging of Sterile Barrier SystemsMet requirements
    PerformanceISO 11138-1,2:2006 Sterilization of healthcare products - Biological IndicatorsMet requirements
    PerformanceISO 11737-1,2:2006 Sterilization of Medical Devices - Microbiological Methods Part 1 & 2Met requirements
    PerformanceISO 11135:2007 Medical Device, Validation and Routine Control of Ethylene Oxide SterilizationMet requirements
    PerformanceAAMI TIR28:2009 Product Adoption and Process Equivalency for Ethylene Oxide SterilizationMet requirements
    Overall ConclusionAll performance requirements based on recognized standards for safety and effectivenessThe results of the non-clinical tests show that the CareFusion Pleurx Pleural Catheter System meets or exceed all performance requirements, and are substantially equivalent to the predicate devices.

    2. Sample Size Used for the Test Set and the Data Provenance

    This information is N/A. The document states, "N/A - No clinical tests were conducted for this submission." The testing described is non-clinical (biocompatibility, performance characteristics like sterilization, packaging integrity, etc.) and refers to compliance with standards rather than a test set of patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is N/A as no clinical test set requiring expert ground truth was conducted.

    4. Adjudication Method for the Test Set

    This information is N/A as no clinical test set requiring adjudication was conducted.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance.

    This information is N/A. This is not an AI/ML device, and no MRMC study was conducted.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done.

    This information is N/A. This is not an AI/ML device.

    7. The Type of Ground Truth Used

    This information is N/A as no clinical ground truth was established for this submission. The "ground truth" for this submission is adherence to recognized non-clinical performance standards and substantial equivalence to legally marketed predicate devices.

    8. The Sample Size for the Training Set

    This information is N/A. This is not an AI/ML device, and therefore no training set was used.

    9. How the Ground Truth for the Training Set Was Established

    This information is N/A. This is not an AI/ML device, and therefore no training set or its associated ground truth was established.

    Summary of the Study Proving Acceptance Criteria:

    The study proving the device meets its "acceptance criteria" is a series of non-clinical bench tests and evaluations demonstrating compliance with a comprehensive list of national and international standards related to biocompatibility, sterilization, packaging, and general catheter performance. The submission concludes that these non-clinical tests collectively show the CareFusion Pleurx Pleural Catheter System meets or exceeds all performance requirements and is substantially equivalent to existing predicate devices. No clinical studies were conducted for this specific 510(k) submission, as it focused on expanding indications and adding accessories to an already legally marketed device by demonstrating equivalence to previously cleared devices and adherence to established non-clinical performance and safety standards.