    Intended Use

    The Pleur-evac Sahara Plus Model S-1150 Continuous Reinfusion Autotransfusion System, is a sterile, non-pyrogenic, single use, three chamber collection/reinfusion device that is intended for collection and continuous reinfusion of autologous blood.

    The Pleur-evac Sahara Model S-100 Autotransfusion Bag is a sterile, non-pyrogenic, single-use device, used for post-surgical collection and reinfusion of autologous blood from the thoracic cavity when attached to a Pleur-evac Sahara Plus Continuous Reinfusion Autotransfusion System.

    Device Description

    The Pleur-evac Sahara Plus Model S-1150 Continuous Reinfusion Autotransfusion System is a chest drainage system that incorporates features from the Thora-Klex Model 0077000, the Pleur-evac Model A-6000, and the Pleur-evac Plus Model A-9150 systems. It includes a dry one-way seal, negative pressure indicator, automatic high negative pressure relief valve, dry suction regulator, air leak meter, positive pressure relief valve, manual high negative pressure relief valve, floorstand, hangers, and a patient drainage tube configuration. It also has a reinfusion port, reinfusion tube, and spike port for continuous reinfusion. The collection/reinfusion chamber is shorter and deeper than the Pleur-evac Plus Model A-9150 to increase capacity. The patient drainage tube components are latex-free. The system can be used with a Pleur-evac Sahara Model S-100 Autotransfusion Bag or a blood transfer bag for reinfusion. The Model S-100 Autotransfusion Bag is a flexible vinyl bag with a rigid top plate and a wire support frame, designed for collection and reinfusion of blood. It includes an Easy-Link Adaptor and tubing clamps.

    AI/ML Overview

    This 510(k) submission (K963850) is for a medical device seeking clearance through substantial equivalence to previously cleared devices, rather than establishing de novo safety and effectiveness through clinical trials with defined acceptance criteria and statistical performance metrics. Therefore, the information typically requested in your prompt regarding acceptance criteria, study design for proving performance against those criteria, and AI-specific details is not applicable in this context.

    Here's why and what information can be extracted based on the provided text:

    Key Takeaways from a 510(k) of this Nature:

    • Substantial Equivalence: The primary goal of this submission is to demonstrate that the new device (Pleur-evac Sahara Plus Model S-1150 Continuous Reinfusion Autotransfusion System and Pleur-evac Sahara Model S-100 Autotransfusion Bag) is as safe and effective as already legally marketed predicate devices. This typically involves showing similar design, materials, intended use, and functioning, often supported by bench testing, rather than extensive clinical studies with human subjects.
    • No AI Component: The device described is a mechanical medical device (chest drainage and autotransfusion system), not an AI/ML-driven device. Therefore, questions about AI assistance, standalone algorithm performance, training sets, or expert ground truthing for an AI model are not relevant.
    • Bench Testing over Clinical Trials for Performance: The "study" mentioned is a series of bench tests designed to compare functional aspects and material safety with predicate devices. This is a common approach for 510(k) submissions where the technology is well-established, and the device represents modifications or combinations of existing cleared technologies.

    Based on the provided text, here is an articulation of the closest equivalent information to your request, with an emphasis on why certain aspects are not applicable:

    1. A table of acceptance criteria and the reported device performance

    Performance Metric Tested (Acceptance Criteria Implied: "met specification" or "comparable")Reported Device Performance (Pleur-evac Sahara Plus S-1150 / S-100)
    Suction Control Accuracy (at each set point)Met specification on each of the units.
    Airflow CapacityComparable to currently marketed Thora-Klex and Pleur-evac Chest Drainage Systems.
    Response to Patient Air LeakComparable to currently marketed Pleur-evac Chest Drainage Systems.
    System Cracking Pressure PerformanceFunctionally comparable to currently marketed Thora-Klex Chest Drainage System.
    High Negative Pressure Relief Valve PerformanceFunctionally comparable to currently marketed Thora-Klex Chest Drainage System.
    Negative Pressure Indicator PerformanceFunctionally comparable to currently marketed Thora-Klex Chest Drainage System.
    Autotransfusion Bag PerformanceSubstantial equivalence demonstrated between Pleur-evac Model A-6000 with A-1500 Bag and Pleur-evac Sahara Plus Model S-1150 with S-100 Bag (when subjected to simulated patient pressure).
    Carrying Handle StrengthTested and compared (specific result not detailed, but implied satisfactory for equivalence).
    Floorstand PerformanceTested and compared (specific result not detailed, but implied satisfactory for equivalence).
    Hanger to Post StrengthTested and compared (specific result not detailed, but implied satisfactory for equivalence).
    Biocompatibility (patient tube and injection site materials)Met the requirements of ISO 10993 Part-1. Suitable for use.
    Biocompatibility (other blood-contacting materials)Relying on results from predicate device 510(k) #K911656A and #K854301, as materials are identical.
    Blood Pathway and Blood Flow ReinfusionEquivalent to Pleur-evac Plus Model A-9150 (same mesh filter, port size, tubing, spike port). No additional testing performed.

    Study Proving Acceptance Criteria:

    The "study" was a series of bench tests (laboratory-based performance evaluations) comparing the new device's functional aspects to its predicate devices. The objective was to demonstrate substantial equivalence, meaning the new device performs similarly and is as safe and effective as the predicate devices that are already legally marketed.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: The document does not specify the number of units/samples tested for each performance metric (e.g., how many S-1150 units were tested for suction control accuracy). For bench testing of this nature, usually, a defined number of production units are subjected to each test, but specific quantities are often not detailed in the public 510(k) summary.
    • Data Provenance: The data is from prospective bench testing conducted by the manufacturer, Deknatel DSP Worldwide Inc., likely at their facilities (implied, no specific country mentioned, but the submitter is based in Fall River, MA, USA). The data is not clinical data from patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. This is a mechanical device, and its performance is assessed against engineering specifications and comparison to predicate devices, not against a "ground truth" derived from expert clinical assessment in the way an AI/diagnostic device would be. The "ground truth" here is the established performance characteristics and safety profile of the predicate devices, as determined by previous FDA clearances.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. As there are no human-readout interpretations or diagnostic classifications involved in this type of device's performance testing, adjudication methods like 2+1 or 3+1 are not relevant. Performance is measured objectively via instruments and engineering standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This device does not involve human readers, diagnostic imaging, or AI assistance. It is a mechanical chest drainage and autotransfusion system.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not Applicable. There is no AI algorithm in this device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For functional performance, the "ground truth" is implied by the established specifications and validated performance of the predicate devices (Thora-Klex Model 0077000, Pleur-evac Model A-6000, Pleur-evac Plus Model A-9150). The new device's performance is gauged against these known benchmarks for comparability.
    • For biocompatibility, the "ground truth" is adherence to ISO 10993 Part-1 standards and the previously cleared biocompatibility of identical materials in predicate devices.

    8. The sample size for the training set

    • Not Applicable. There is no AI component, and thus no "training set." The device's design is based on the features of existing, cleared predicate devices.

    9. How the ground truth for the training set was established

    • Not Applicable. As there is no AI component or training set, this question is not relevant.