Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K101853

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-01-03

    (186 days)

    Product Code
    Regulation Number
    866.5600
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PLAC® Test Reagent Kit is a turbidimetric immunoassay for the quantitative determination of Lp-PLA2 (lipoprotein-associated phospholipase A2) in human plasma or serum on automated clinical chemistry analyzers, to be used in conjunction with clinical evaluation and patient risk assessment as an aid in predicting risk for coronary heart disease, and ischemic stroke associated with atherosclerosis.

    Device Description

    The PLAC® Test Reagent Kit consists of separately packaged reagents, calibrators and controls for the measurement of Lp-PLA2 in serum or plasma on automated clinical chemistry analyzers.

    PLAC® Test Reagent Kit

    • R1 Buffer solution with protein stabilizers
    • R2 Suspension of polymeric microparticles coated with mouse monoclonal antibodies specific to Lp-PLA2 (2C10 and 4B4).

    The PLAC® Test Reagent Kit is based on turbidimetric immunoassay technology utilizing two Lp-PLA3-specific monoclonal antibodies (2C10 and 4B4) coated to polymeric microparticles. A set of Lp-PLA2 calibrators is used to plot a standard curve of absorbance (y-axis) versus Lp-PLA2 concentration in ng/mL (x-axis) from which the Lp-PLA2 concentration in the test sample can be determined. The concentration of Lp-PLA, in each sample and control is then interpolated from the standard curve using a spline curve fit with appropriate calibration curve fitting software. The kit expiration date and storage conditions are indicated on the package.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study that proves the device meets those criteria for the PLAC® Test Reagent Kit.

    Here's the breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Methods Comparison Data

    Acceptance CriteriaReported Device Performance (PLAC® Test Reagent Kit vs. PLAC® Test ELISA Kit)
    Slope: 0.9 to 1.11.09
    Intercept: ≤ 50 ng/mL-1.7 ng/mL
    Correlation Coefficient (R): ≥ 0.900.92

    Imprecision Testing (Acceptance criteria are implied by the reported low %CVs, indicating good precision. Specific numerical acceptance criteria for %CV are not explicitly stated but are generally understood to be low for assays.)

    Sample/ControlMean Concentration Lp-PLA2 (ng/mL)Intra-assay %CV (n=40)Inter-assay %CV (n=40)Total Assay %CV (n=80)
    Serum 156.22.5%4.6%5.3%
    Serum 2250.11.0%2.7%2.9%
    Buffer Control 1191.20.7%0.9%1.4%
    Buffer Control 2370.81.0%0.7%1.6%

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set:
      • Methods Comparison: n = 742 (number of samples used for the comparison between the predicate and modified device).
      • Imprecision Testing: n = 40 for intra-assay and inter-assay, leading to a total of n = 80 for total assay %CV (per sample/control type).
    • Data Provenance: The document does not specify the country of origin of the data. It also does not explicitly state whether the data was retrospective or prospective, though method comparison and imprecision studies are typically prospective tests performed during device development and validation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This device is an in vitro diagnostic reagent kit for measuring a biomarker. The "ground truth" for such devices is typically established by reference methods or clinical outcomes. Experts are not usually involved in establishing ground truth in the same way as they would be for image-based diagnostics.

    In this case, the predicate device (PLAC® Test ELISA Kit) serves as the reference method for the methods comparison study, effectively establishing the "ground truth" against which the new device's measurements are compared.

    4. Adjudication Method for the Test Set

    Not applicable for this type of in vitro diagnostic device study. Adjudication methods are typically used in studies where human interpretation of data (e.g., medical images) is being assessed, often to resolve discrepancies between multiple readers. Here, the comparison is quantitative between two analytical methods.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of AI on human reader performance, usually in diagnostic imaging. The PLAC® Test Reagent Kit is an automated quantitative immunoassay, not an AI-powered diagnostic imaging tool that assists human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the performance data presented (Methods Comparison and Imprecision Testing) represents the standalone performance of the PLAC® Test Reagent Kit as an automated turbidimetric immunoassay. This is an "algorithm only" in the sense that the device itself performs the quantitative measurement without direct human-in-the-loop interpretative steps during the assay run.

    7. The Type of Ground Truth Used

    The "ground truth" in this context is the measurements obtained from the legally marketed predicate device, the PLAC® Test ELISA Kit. For the imprecision studies, the ground truth is the inherent concentration of the analyte in the tested samples/controls.

    8. The Sample Size for the Training Set

    The document does not specify a separate training set. For in vitro diagnostic assays, particularly those based on established immunoassay principles, a dedicated "training set" in the machine learning sense is not typically used. The development and optimization of the assay (e.g., antibody selection, reagent formulation, calibration curve fitting algorithm) would precede the validation studies, but these development activities don't usually involve a distinct "training set" with established ground truth in the same way an AI algorithm does.

    9. How the Ground Truth for the Training Set was Established

    As no explicit training set is mentioned for an AI/algorithm-based device, this question is not directly applicable. The "training" or optimization of such an assay involves laboratory development and analytical characterization, with the "ground truth" reflecting accurate and precise measurements of Lp-PLA2 concentration, often established using highly characterized reference materials or established laboratory methods during the assay development phase. The information provided focuses on the validation against a predicate device and within-device performance.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1