Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K241586

    Validate with FDA (Live)

    Date Cleared
    2025-02-27

    (269 days)

    Product Code
    Regulation Number
    862.1675
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PIVO™ Pro Needle-free Blood Collection Device attaches to a peripheral IV catheter system for use to obtain venous blood specimens into a vacuum tube, blood culture bottle, or syringe from adult and pediatric patients, including those with difficult intravenous access who may have small, fragile, and/or non-palpable veins. For blood culture, the device can be used at initial placement of the peripheral IV catheter.

    Device Description

    The PIVOTM Pro Needle-free Blood Collection Device is a sterile, single use needle-free collection device that attaches to a peripheral intravascular (IV) catheter system for use to obtain venous blood specimens. The device is comprised of an inner flow tube with a slider, proximal flexible tube with female Luer, outer housing and winged clip on the distal end. The winged clip attaches to the catheter system The female Luer attaches to an evacuated tube holder or syringe. The inner flow tube is then advanced to collect a blood sample. Once complete, the inner flow tube is retracted, and the device is removed from the IV. The device is also suitable for use for blood culture collection at initial placement of peripheral IV catheters. The device is available in three sizes: 20GA, 22GA and 24GA IV compatible. The device is compatible with the corresponding IV gauge and larger IV catheters. The PIVOTM Pro Needle-free Blood Collection Device is compatible with peripheral IV catheters and extension sets with NearPortTM IV Access.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for a medical device called the "PIVO™ Pro Needle-free Blood Collection Device." This document details the device's indications for use, technological characteristics, and comparison to a predicate device, along with a summary of performance tests. However, it does not describe an AI/ML powered device or a study involving human experts for ground truth establishment, MRMC studies, or training/test set sample sizes in the context of AI/ML.

    The document primarily focuses on establishing substantial equivalence for the device itself (a needle-free blood collection device), specifically for the addition of blood culture collection and an extended shelf life. The performance tests mentioned (e.g., Insertion Test, Blood Leak Test, Package Integrity) are related to the physical and functional aspects of the hardware device, not the performance of an AI/ML algorithm.

    Therefore, I cannot extract the requested information about acceptance criteria for an AI/ML powered device, the study proving its performance (in terms of AI metrics), sample sizes for AI test/training sets, expert involvement in ground truth, or MRMC studies, as this information is not present in the provided text.

    The text does state:

    • "Clinical studies are not required to demonstrate substantial equivalence to the predicate device."
    • "Summary of Performance tests completed on the subject devices were limited to those tests required Performance to support a determination of substantial equivalence to the predicate device."
    • "Per design control requirements specified in 21 CFR 820.30, the subject device met all predetermined acceptance criteria for the above-listed performance test, demonstrating substantial equivalence to the predicate device."

    This indicates that internal performance tests were conducted against pre-determined acceptance criteria, but these are for the physical device's functionality and safety, not for an AI/ML algorithm's diagnostic or predictive performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230865

    Validate with FDA (Live)

    Date Cleared
    2023-09-28

    (183 days)

    Product Code
    Regulation Number
    862.1675
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PIVO™ Pro Needle-free Blood Collection Device attaches to a peripheral IV catheter system for use to obtain venous blood specimens into a vacuum tube or syringe from adult and pediatric patients, including those with difficult intravenous access who may have small, fragile, and/or non-palpable veins.

    Device Description

    The PIVO™ Pro Needle-free Blood Collection Device is a sterile, single use needle-free collection device that attaches to a peripheral intravascular catheter (IV) system. The device is comprised of an inner flow tube with a slider, proximal flexible tube with female Luer, outer housing, and winged clip on the distal end. The winged clip attaches to the IV system. The female Luer attaches to an evacuated tube holder or syringe. The inner flow tube is then advanced to collect a blood sample. Once complete, the inner flow tube is retracted, and the device is removed from the IV.

    The device is available in three sizes: 20GA, 22GA, and 24GA IV compatible. The device is compatible with the corresponding IV gauge and larger IV catheters. The PIVOTM Pro Needle-free Blood Collection Device is compatible with peripheral IV catheters and extension sets with NearPort™ IV Access.

    AI/ML Overview

    The provided document is a 510(k) summary for the PIVOTM Pro Needle-free Blood Collection Device, which seeks to demonstrate substantial equivalence to a predicate device (PIVOTM Needle-free Blood Collection Device). This type of submission relies on demonstrating that the new device is as safe and effective as a legally marketed predicate, rather than proving absolute safety and effectiveness from scratch. Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are framed within the context of demonstrating substantial equivalence, primarily through non-clinical performance testing and a comparison of technological characteristics.

    Here’s a breakdown of the information requested, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance results. Instead, it lists the performance tests conducted and states that the device "met all predetermined acceptance criteria for the above-listed performance tests, demonstrating substantial equivalence to the predicate devices."

    However, we can infer some "acceptance criteria" based on the tests performed:

    Acceptance Criteria (Inferred from tests)Reported Device Performance
    Safety and Biocompatibility:
    Biocompatibility (ISO 10993-1)Met all requirements
    Packaging Integrity (ASTM D4169 & ISO 11607-1)Met all requirements
    Sterilization Validation (ISO 11137-1)Met all requirements
    HemolysisMet all requirements
    Blood LeakMet all requirements
    Pressure LeakMet all requirements
    Vacuum LeakMet all requirements
    Device Functionality:
    Catheter Gauge CompatibilityMet all requirements
    Extension LengthMet all requirements
    Catheter PerforationMet all requirements
    InsertionMet all requirements
    AlignmentMet all requirements
    Atraumatic TipMet all requirements
    Flow RateMet all requirements
    Pusher Pull ForceMet all requirements
    Complete RetractionYes (inherent design feature, verified through testing)
    Usability:
    Usability testing per FDA GuidanceMet all requirements

    2. Sample sizes used for the test set and the data provenance

    The document does not specify the exact sample sizes (N values) for each individual performance test (e.g., how many devices were tested for catheter perforation, how many for flow rate). It also does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective. Given that these are bench and lab tests for device clearance, they are inherently prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable in the context of this 510(k) summary. The "ground truth" for this device is based on objective measurements and compliance with recognized standards and internal design requirements, not expert consensus on interpretations like with diagnostic imaging. Usability testing would involve users, but the number and qualifications of those users are not detailed.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This concept is not applicable here as these are performance and engineering tests, not studies requiring human interpretation and adjudication (e.g., clinical trials for diagnostic algorithms).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC study was not conducted. This type of study applies to AI/CADe (Computer-Assisted Detection) or CADx (Computer-Assisted Diagnosis) devices where human readers (e.g., radiologists) use AI assistance to improve their diagnostic performance. The PIVOTM Pro is a physical blood collection device, not an AI software. The submission explicitly states: "Clinical studies are not required to demonstrate substantial equivalence to the predicate device."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This concept is not applicable as the device is a physical medical device, not an algorithm or AI system. Its performance is evaluated through physical and functional testing.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For this device, the "ground truth" relates to meeting engineering specifications, material properties, sterility, and functional performance benchmarks derived from industry standards (e.g., ISO, ASTM), internal design requirements, and prior knowledge from the predicate device. It is primarily based on:

    • Physical measurements and criteria: e.g., acceptable flow rates, specified dimensions, absence of leaks, specified pull force.
    • Compliance with recognized standards: e.g., ISO 10993-1 for biocompatibility, ISO 11137-1 for sterilization, ASTM D4169 and ISO 11607-1 for packaging.
    • Functional tests: verifying proper insertion, alignment, and absence of catheter perforation.

    8. The sample size for the training set

    This is not applicable. This is a physical device, not an AI/machine learning model, so there is no "training set."

    9. How the ground truth for the training set was established

    This is not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1