Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K132848

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-11-29

    (79 days)

    Product Code
    Regulation Number
    888.4580
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Piezosurgery FLEX is an ultrasonic surgical system consisting of handpieces and associated tips for cutting bone, osteotomy, osteoplasty and drilling in variety of surgical procedures, including but not limited to otolaryngological, oral/maxillofacial, hand, foot, neurosurgical, spine, and paalstic/reconstructive surgery. It may also be used with endoscopic visual assistance to perform the above listed procedures.

    Device Description

    Piezosurgery FLEX is a piezoelectric ultrasonic device that uses ultrasonic energy to generate mechanical micro-vibration of tip inserts intended for bone cutting. The device consists of a base control unit with intearated peristaltic pump, a sterilizable handpiece with connecting cord, a range of sterile, single use tip inserts, a torque wrench, a foot-pedal, an irrigation set, and other minor attachments.

    The console is equipped with a touch panel that allows the surgeon to control device operation. The control panel also provides monitoring information, by means icons, in case of malfunction of the device. Inside the console are located the ultrasonic generator, the electrical power supply module and the micro-processor electronic board that controls and supervises the functional parameters of the device. The console is connected to mains power by an electrical cord, and includes connectors for the handpiece and for the foot-pedal. The console incorporates a peristaltic pump which provides, through the irrigation tubing set, a sterile fluid supply to the surgical site. Ultrasonic power and irrigation flow to the handpiece are simultaneously activated by pressing the foot-pedal. The handpiece contains a piezoelectric ultrasonic transducer which attaches to the generator (inside the console) by a cable at one end of the handpiece. Tip inserts are attached to the other end of the handpiece.

    The touch panel of the console allows the operator to set the treatment parameters, as follows:

    • Seven levels of power: .
    • Six levels of irrigation: .
    • Three ultrasonic output modes; .
    • · The pump function.

    The Piezosurgery FLEX uses piezoelectric ultrasonic technology to generate mechanical microvibrations of the tip insert connected to the handpiece, the piezoelectric transducer converting the electrical voltage supplied by the ultrasonic generator into mechanical energy that induces vibration of the tip insert at the resonant frequency of the tip insert.

    AI/ML Overview

    This 510(k) summary describes a traditional submission for the Mectron Spa Piezosurgery FLEX, an ultrasonic surgical system.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Non-Clinical Testing)Reported Device Performance
    Compliance with safety requirements of standard IEC 60601-1Confirmed
    Compliance with EMC requirements of standard EN 60601-1-2Confirmed
    Similar intended use/indications of use to predicate devicesConfirmed (demonstrated through submission)
    Similar technology to predicate devicesConfirmed (demonstrated through submission)
    No significant effect on safety and effectiveness due to technological differences from predicate devicesConfirmed (stated in submission)

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document describes non-clinical testing focused on compliance with electrical safety and electromagnetic compatibility (EMC) standards. It does not contain information about a "test set" in the context of clinical performance evaluation (e.g., patient data). The testing performed was related to device characteristics.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The document describes compliance testing against established engineering standards (IEC 60601-1, EN 60601-1-2), not a study requiring expert-established ground truth for a clinical dataset.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies involving human interpretation or performance, which is not described here.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

    Not applicable. This device is an ultrasonic surgical system, not an AI-based diagnostic or assistive technology that would involve human "readers" or an MRMC study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device is a surgical instrument. "Standalone" performance testing would relate to its engineering specifications and functionality, which are covered by the non-clinical testing mentioned (IEC and EMC standards compliance).

    7. The Type of Ground Truth Used

    For the non-clinical testing: The "ground truth" was defined by the requirements and specifications outlined in the international standards IEC 60601-1 (medical electrical equipment safety) and EN 60601-1-2 (electromagnetic compatibility).

    8. The Sample Size for the Training Set

    Not applicable. The Piezosurgery FLEX is a hardware surgical device, not a machine learning or AI model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As noted above, this device does not utilize a "training set" in the context of machine learning.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1