Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K023639

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2003-01-24

    (86 days)

    Product Code
    Regulation Number
    862.1705
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Piccolo Triglycerides Test System used with the Piccolo Point-of-Care Chemistry Analyzer is intended for the in vitro quantitative determination of triglycerides in heparinized whole blood, heparinized plasma, or serum in a clinical laboratory setting or point-of-care location.

    Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, and other diseases involving lipid metabolism, or various endocrine disorders.

    Device Description

    The Piccolo® Lipid Panel Reagent Disc (which contains the Piccolo® Triglycerides Test System) is designed to separate a heparinized whole blood sample into plasma and blood cells. The disc meters the required quantity of plasma and diluent, mixes the plasma with diluent, and delivers the mixture to the reaction cuvettes along the disc perimeter. The diluted plasma mixes with the reagent beads, initiating the chemical reactions that are then monitored by the analvzer. Alternately, the disc mav also be used with serum.

    AI/ML Overview

    This document describes the regulatory submission for the Abaxis Piccolo® Triglycerides Test System. The device is a laboratory test system designed to measure triglyceride levels in blood samples.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and the Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" as a separate, pre-defined set of thresholds. Instead, it presents performance characteristics of the new device and compares them to those of a legally marketed predicate device (Bayer Triglycerides-RA Assay). The underlying acceptance criterion for regulatory approval in this context is "substantial equivalence" to the predicate device, meaning the new device performs as safely and effectively. The tests conducted (linearity, precision, sample type comparison, and method comparison) are standard ways to demonstrate this equivalence for in vitro diagnostic devices.

    Below is a table summarizing the reported device performance and comparing it to the characteristics of the predicate device, which implicitly defines the 'acceptance' for a substantially equivalent device.

    Performance CharacteristicAcceptance Criteria (Implied by Predicate Device)Reported Device Performance (Piccolo® Triglycerides Test System)
    Intended UseQuantitative analysis of triglyceridesQuantitative analysis of triglycerides
    MethodologyEnzymatic endpoint reactionEnzymatic endpoint reaction
    Sample TypeHeparinized plasma and serumHeparinized whole blood, heparinized plasma, and serum
    Sensitivity0.99 mA per mg/dL or 0.088 A per mmol/L3.82 mA per mg/dL or 0.338 A per mmol/L; 20 mg/dL
    ReagentsDry reagents (reconstituted by user prior to use) and liquid reagentsDry test-specific reagent beads and liquid diluent; reconstitution performed by analyzer
    Temperature of Reaction37°C37°C
    CalibrationCalibrated periodically using calibrators supplied by vendorBar code with factory calibrated lot specific data
    Assay Range0 - 500 mg/dL20 - 500 mg/dL
    Testing EnvironmentProfessional useProfessional use
    Sample Size2 µLApprox 100 µL
    Linearity (Slope)(Implied to be close to 1)1.000
    Linearity (Intercept)(Implied to be close to 0)3.0
    Linearity (Correlation Coefficient 'r')(Implied to be close to 1)1.000
    Precision (Serum 1, Mean)(Not specified as a target, but precision is expected)206.8 mg/dL (Within-Run & Total)
    Precision (Serum 1, SD)(Not specified as a target, but precision is expected)4.7 (Within-Run), 5.5 (Total)
    Precision (Serum 1, %CV)(Not specified as a target, but precision is expected)2.3 (Within-Run), 2.6 (Total)
    Precision (Serum 2, Mean)(Not specified as a target, but precision is expected)163.7 mg/dL (Within-Run & Total)
    Precision (Serum 2, SD)(Not specified as a target, but precision is expected)1.8 (Within-Run), 2.4 (Total)
    Precision (Serum 2, %CV)(Not specified as a target, but precision is expected)1.1 (Within-Run), 1.5 (Total)
    Method Comparison (Correlation Coefficient 'r')(Implied to be close to 1 for substantial equivalence)0.999
    Method Comparison (Slope)(Implied to be close to 1 for substantial equivalence)0.983 (Linear), 0.984 (Deming)
    Method Comparison (Intercept)(Implied to be close to 0 for substantial equivalence)8.2 (Linear), 8.1 (Deming)

    2. Sample Size Used for the Test Set and the Data Provenance:

    • Linearity Study: The sample size for the linearity study is not explicitly stated in terms of number of patient samples, but the results for slope, intercept, and correlation coefficient are provided.
    • Precision Study:
      • Sample Size: n = 160 for both within-run and total precision for each serum level tested (Serum 1 and Serum 2). This appears to be 160 replicates per serum level, not 160 distinct patient samples.
      • Data Provenance: Not specified, but generally, such studies use commercially available control sera or pooled patient samples.
    • Sample Type Comparison Study:
      • Sample Size: Samples from 20 different patients. Each sample type (serum, heparinized plasma, heparinized whole blood) was tested in quadruplicate.
      • Data Provenance: Not specified, presumed to be clinical samples.
    • Method Comparison Study:
      • Sample Size: n = 172 data points. The document clarifies: "number of patient samples = n/2", meaning 86 unique patient samples were run in duplicate.
      • Data Provenance: Not specified, but these are patient samples tested against a predicate device, presumably from a clinical setting. Retrospective or prospective is not specified, but given the nature of the study, it's likely a mix or prospective collection for comparison purposes. No country of origin is mentioned.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    This information is not provided in the document. For in vitro diagnostic devices like this, the "ground truth" for the test set (patient samples) is typically established by the reference method (the predicate device, in the case of method comparison) or by highly accurate laboratory methods, rather than by human expert consensus or adjudication.

    4. Adjudication Method for the Test Set:

    This information is not applicable and therefore not provided. Adjudication methods (like 2+1, 3+1) are common in image-based diagnostic studies where human readers interpret results which are then reconciled. For quantitative laboratory tests, the result is a numerical value, and "adjudication" in the traditional sense is not performed on the test results themselves. Method comparison studies statistically compare the new device's numerical results to those of the predicate device.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable and therefore not provided. MRMC studies are specific to diagnostic tools that involve human interpretation, often in imaging. The Abaxis Piccolo® Triglycerides Test System is an automated in vitro diagnostic device for quantitative chemical analysis, not an AI-assisted diagnostic tool that involves human reader interpretation for primary diagnosis.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Yes, the studies presented (Linearity, Precision, Sample Type Comparison, Method Comparison) represent the standalone performance of the Abaxis Piccolo® Triglycerides Test System. The device is designed to operate as an automated system producing a quantitative result without direct human interpretation influencing the measurement itself (beyond sample collection and loading). The studies assess the device's inherent analytical performance.

    7. The Type of Ground Truth Used:

    For the method comparison study, the predicate device's measurement (Bayer Triglycerides-RA Assay) served as the de facto "ground truth" for comparison to establish substantial equivalence. For linearity, precision, and sample type comparison, the "ground truth" is typically established by the inherent analytical capabilities of the device itself and its ability to consistently and accurately measure known concentrations or produce reproducible results.

    8. The Sample Size for the Training Set:

    This information is not applicable and therefore not provided. The Abaxis Piccolo® Triglycerides Test System is a chemical analyzer, not a machine learning or AI-based system that requires a "training set" in the computational sense. Its reagents and methods are developed through biochemical and analytical chemistry principles, not by training an algorithm on a dataset.

    9. How the Ground Truth for the Training Set was Established:

    This information is not applicable due to the nature of the device (see point 8).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1