Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K200571

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-05-01

    (58 days)

    Product Code
    Regulation Number
    888.3075
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PERLA® posterior occipito-cervico-thoracic fixation system is intended to provide immobilization of spinal segments as an adjunct to following acute and chronic instabilities of the crania-cervical junction. the cervical spine (Cl to C7) and the thoracic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability.

    The PERLA® posterior occipito-cervico-thoracic fixation system is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

    The occipital bone screws are limited to occipital fixation only. The use of the multi-axial screws is limited to placement in the cervical spine (Cl to C7) and the thoracic spine (Tl to T3).

    In order to achieve additional levels of fixation, the PERLA® posterior occipito-cervico-thoracic fixation system may be connected to the ROMEO®2 and PERLA® TL Posterior Osteosynthesis Systems with rod connectors. Transition rods mav also be used to connect the PERLA® posterior occipito-cervico-thoracic fixation system to the ROMEO®2 and PERLA®TL Posterior Osteosynthesis Systems. Refer to the ROMEO®2 and PERLA®TL Posterior Osteosynthesis Systems packages inserts for a list of the ROMEO®2 and PERLA®TL Posterior Osteosynthesis Systems indications of use. PERLA® posterior occipito-cervico-thoracic fixation system is indicated for skeletally mature patients.

    Device Description

    The PERLA® system consists of a range of screws, rods, set screws, hooks, rod connectors and cross-connectors. These connecting components can be rigidly locked to the rod in a variety of configurations to be adapted for the individual case. The PERLA® system is manufactured from medical grade titanium alloy and medical grade cobalt chromium conforming respectively to standards ASTM F136 and ASTM F1537. The PERLA® implants are delivered either non sterile or sterile (gamma sterilization) and supplied with dedicated surgical instruments (reusable - provided non-sterile except for the drill supplied as sterile or not sterile). Bacterial endotoxin testing on final, finished devices as specified in USP standard is used for pyrogenicity testing to achieve the Endotoxin limit of 20 EU / device.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the PERLA® Posterior Occipito-Cervico-Thoracic Fixation System.

    Important Note: This document is a 510(k) summary for a medical device. It describes the device as "substantially equivalent" to predicate devices based on non-clinical testing. It does not contain information about software performance, AI algorithms, clinical study outcomes for human readers, or ground truth establishment in the ways typically expected for AI/ML device submissions. The questions asked are highly relevant for AI/ML medical devices, but the provided text describes a traditional orthopedic implant. Therefore, many of your questions cannot be answered from this document.

    Explanation of Device: The PERLA® Posterior Occipito-Cervico-Thoracic Fixation System is a surgical implant designed to provide immobilization and stabilization of spinal segments. It consists of various components like screws, rods, and connectors, made of medical-grade titanium alloy and cobalt chromium. The 510(k) submission is for a line extension of the PERLA® system to include components for occipital fixation.


    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this is a hardware orthopedic implant and not an AI/ML device, the "acceptance criteria" and "device performance" are related to mechanical and material properties, rather than diagnostic accuracy metrics (e.g., sensitivity, specificity).

    Acceptance Criteria (Study Type)Reported Device Performance
    Static Compression Bending per ASTM F2706-18Results demonstrate comparable mechanical properties to the predicate devices.
    Static Torsion per ASTM F2706-18Results demonstrate comparable mechanical properties to the predicate devices.
    Dynamic Compression Bending per ASTM F2706-18Results demonstrate comparable mechanical properties to the predicate devices.
    Dynamic Torsion per ASTM F2706-18Results demonstrate comparable mechanical properties to the predicate devices.
    Bacterial Endotoxin TestingAchieved Endotoxin limit of 20 EU / device (as specified in USP standard for pyrogenicity testing).
    Material CompositionMedical grade titanium alloy and medical grade cobalt chromium conforming to standards ASTM F136 and ASTM F1537, respectively.
    SterilizationDelivered non-sterile or sterile (gamma sterilization).

    2. Sample Size Used for the Test Set and Data Provenance

    The study described is non-clinical bench testing (mechanical and material testing), not a clinical study involving patient data. Therefore, the concepts of "test set sample size" and "data provenance" (country of origin, retrospective/prospective) are not applicable in the context of this device. The "test set" would refer to the physical samples of the device components subjected to the mechanical tests. The document does not specify the number of individual components tested but states that "non-clinical tests were conducted on the PERLA® components for occipital fixation."


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This question is not applicable to an orthopedic implant submission based on non-clinical bench testing. "Ground truth" in this context would be the measured physical properties of the materials and constructs, determined by laboratory equipment and standard mechanical testing protocols, not by expert human interpretation.


    4. Adjudication Method for the Test Set

    Not applicable. As explained above, the "test set" is physical device components, and the "adjudication method" does not apply to non-clinical mechanical testing.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    No, this type of study was not done. An MRMC study is relevant for AI/ML diagnostic devices where human readers interpret medical images or data. The PERLA® system is a surgical implant.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, this is not applicable. The device is a physical surgical implant, not an algorithm.


    7. The Type of Ground Truth Used

    For the non-clinical testing:

    • Mechanical Performance: The "ground truth" is the quantitative results (e.g., load to failure, fatigue life) obtained from standardized mechanical tests (ASTM F2706-18). This is based on objective physical measurements according to established engineering standards.
    • Material Composition: The "ground truth" is that the materials conform to specific ASTM standards (F136, F1537) for medical-grade titanium alloy and cobalt chromium, verified through material characterization tests.
    • Pyrogenicity: The "ground truth" is that the device meets the endotoxin limit of 20 EU/device, as measured by bacterial endotoxin testing following USP standards.

    8. The Sample Size for the Training Set

    Not applicable. There is no AI/ML model or "training set" for this physical implant device.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set, the establishment of ground truth for it is irrelevant.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1