510(k) Data Aggregation

    K Number
    K133645

    Validate with FDA (Live)

    Date Cleared
    2014-01-03

    (37 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PERIMETER® C Spinal System is intended to be used for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc who have had six weeks of non-operative treatment. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. Additionally, the PERIMETER® C Spinal System implants are to be used with autogenous bone graft and supplemental fixation and implanted via an open, anterior approach.

    The CAPSTONE® Spinal System is indicated for interbody fusion with autogenous bone graft in patients with Degenerative Disc Disease (DDD) at one or two levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Additionally, the CAPSTONE® Spinal System is indicated in the setting of spinal deformity as a supplement to pedicle screw fixation in patients diagnosed with degenerative scoliosis. These implants may be implanted via an open or a minimally invasive posterior approach. Alternatively, these implants may also be implanted via an anterior and/or transforaminal approach. These implants are to be used with autogenous bone graft. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the FDA for use in the lumbar spine.

    The CAPSTONE® L Spinal System is designed to be used with autogenous bone graft to facilitate interbody fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine. The CAPSTONE® L Spinal System is used for patients diagnosed with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via a minimally invasive lateral approach.

    The PERIMETER® Interbody Fusion Device is indicated for interbody fusion with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via a variety of open or minimally invasive approaches. These approaches include anterior, lateral and oblique. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared for use in the lumbar spine.

    Device Description

    The PERIMETER® C Spinal System consists of PEEK spacers with tantalum markers of various widths and heights, which can be inserted between two cervical vertebral bodies to give support and correction during cervical interbody fusion procedures. Additionally, the superior and inferior surfaces of the implant are designed with teeth which interact with the surface of the vertebral endplates to aid in resisting expulsion. The hollow geometry of the implants allows them to be packed with autogenous bone graft and is to be used with supplemental fixation in all procedures.

    The CAPSTONE® PEEK Spinal System consists of PEEK cages with tantalum markers of various widths and heights, which can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. Additionally, the superior and inferior surfaces of the implant are designed with teeth which interact with the surface of the vertebral endplates to aid in resisting expulsion. The hollow geometry of the implants allows them to be packed with autogenous bone graft.

    The CAPSTONE® L Spinal System consists of PEEK cages of various widths and heights, which include Tantalum markers. These devices can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. Additionally, the superior and inferior surfaces of the implant are designed with teeth which interact with the surface of the vertebral endplates to aid in resisting expulsion. The hollow geometry of the implants allows them to be packed with autogenous bone graft.

    The PERIMETER® Interbody Fusion Device consists of PEEK cages with tantalum markers of various widths and heights which can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. Additionally, the superior and inferior surfaces of the implant are designed with teeth which interact with the surface of the vertebral endplates to aid in resisting expulsion. The hollow geometry of the implants allows them to be packed with autogenous bone graft. The PERIMETER® Interbody Device is to be used with supplemental fixation instrumentation.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for Medtronic Sofamor Danek MRI Update for PEEK Interbody Fusion Devices. The primary purpose of this submission is to obtain MRI safety labeling for these devices and provide instructions on how to perform MRI scans on them. It asserts that the devices are "MR Conditional" according to ASTM F2503.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Standard Reference)Device Performance as Reported
    ASTM F2052: Measurement of magnetically induced displacement force on passive implants in the magnetic resonance environment.Testing completed on the worst-case implants, determining the devices are MR Conditional.
    ASTM F2213: Measurement of magnetically induced torque on medical devices in the magnetic resonance environment.Testing completed on the worst-case implants, determining the devices are MR Conditional.
    ASTM F2119: Evaluation of MR image artifacts from passive implants.Testing completed on the worst-case implants, determining the devices are MR Conditional.
    ASTM F2182: Measurement of radio frequency induced heating on or near passive implant during magnetic resonance imaging.Testing completed on the worst-case implants, determining the devices are MR Conditional.
    ASTM F2503: Standard practice for marking medical devices and other items for safety in the magnetic resonance environment.Testing completed, and devices are labeled as MR Conditional in accordance with this standard.
    Substantial Equivalence: No significant changes to overall design, material, or indications from predicate devices (K122037).The subject devices have the "same fundamental technology," are "manufactured from the same PEEK and tantalum materials," have "no changes to the overall design, to the material or the indications," and are therefore considered substantially equivalent to their predicates for MRI safety labeling.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: The document repeatedly mentions "worst case implants" were used for testing. However, it does not specify the exact number of implants constitutive of this "worst case" test set for any of the ASTM standards.
    • Data Provenance: The document does not specify the country of origin of the data. The testing was non-clinical and conducted in accordance with FDA Guidance "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment." It's reasonable to infer the testing was conducted in a laboratory setting, likely in the USA where Medtronic Sofamor Danek is based. The data would be prospective for the purpose of this 510(k) submission, as it was specifically generated to demonstrate MRI safety for these devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • This information is not provided in the document. The testing described focuses on non-clinical, objective measurements defined by ASTM standards. These standards themselves establish the "ground truth" for MR safety parameters through their methodologies. There is no mention of human experts establishing a ground truth for the test set data in the context of MRI safety.

    4. Adjudication Method for the Test Set:

    • None specified. The testing involves physical measurements adhering to established ASTM standards, not subjective evaluation requiring adjudication by multiple experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    • No, an MRMC comparative effectiveness study was not done. This submission is for MRI safety labeling of interbody fusion devices, which are passive implants. An MRMC study would be relevant for diagnostic imaging interpretation or active devices with AI assistance, neither of which applies here.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    • Not applicable. This device is a passive implant (interbody fusion device), not an algorithm or an AI-powered diagnostic tool. The performance evaluated is the physical interaction of the implant with MRI fields, not an algorithm's output.

    7. The Type of Ground Truth Used:

    • The ground truth relies on objective physical measurements and adherence to established industry standards (ASTM F2052, F2213, F2119, F2182, F2503) for assessing MRI safety of passive metallic medical implants. These standards provide the methodology and criteria for evaluating displacement force, torque, image artifacts, and RF-induced heating.

    8. The Sample Size for the Training Set:

    • Not applicable. This is a non-clinical evaluation of a physical device's interaction with an MRI environment, not a machine learning model. Therefore, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable. As there is no training set, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    K Number
    K132584

    Validate with FDA (Live)

    Date Cleared
    2013-12-04

    (110 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PERIMETER® C Spinal System is intended to be used for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc who have had six weeks of nonoperative treatment. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. Additionally, the PERIMETER® C Spinal System implants are to be used with autogenous bone graft and supplemental fixation and implanted via an open, anterior approach.

    Device Description

    The PERIMETER® C Spinal System consists of spacers/cages of various widths and heights, which can be inserted between two cervical bodies to give support and correction during cervical interbody fusion procedures. Additionally, this implant has six degrees of lordosis and the superior and inferior surfaces of the implant are designed with teeth which interact with the surface of the vertebral endplates to aid in resisting expulsion. The hollow geometry of the implants allows them to be packed with autogenous bone graft and is to be used with supplemental fixation in all procedures.

    The purpose of this submission is to include additional interbody cages manufactured from medical grade titanium alloy (Ti-6A1-4V ELI) and designed with lateral windows. The lateral windows allow for visibility of bone graft placement. The subject device is offered in a non-sterile form.

    AI/ML Overview

    The Medtronic PERIMETER® C Spinal System is a medical device for anterior cervical interbody fusion procedures. The study demonstrates substantial equivalence to existing predicate devices based on non-clinical mechanical testing and Finite Element Analysis (FEA), following FDA guidance document, "Class II Special Controls Guidance Document: Intervertebral Body Fusion Devices".

    1. Acceptance Criteria and Device Performance:

    Test PerformedApplicable StandardAcceptance Criteria (Implicitly: Meet/Exceed Predicate Performance)Reported Device Performance
    Static Torsion TestingASTM F2077 (Test Methods for Intervertebral Body Fusion Devices)Met/Exceeded performance of predicate devicesSuccessfully met all predetermined acceptance criteria.
    Static Compression Bending TestingASTM F2077 (Test Methods for Intervertebral Body Fusion Devices)Met/Exceeded performance of predicate devicesSuccessfully met all predetermined acceptance criteria.
    Static Compression Shear TestingASTM F2077 (Test Methods for Intervertebral Body Fusion Devices)Met/Exceeded performance of predicate devicesSuccessfully met all predetermined acceptance criteria.
    Dynamic Torsion TestingASTM F2077 (Test Methods for Intervertebral Body Fusion Devices)Met/Exceeded performance of predicate devicesSuccessfully met all predetermined acceptance criteria.
    Dynamic Compression Bending TestingASTM F2077 (Test Methods for Intervertebral Body Fusion Devices)Met/Exceeded performance of predicate devicesSuccessfully met all predetermined acceptance criteria.
    Dynamic Compression Shear TestingASTM F2077 (Test Methods for Intervertebral Body Fusion Devices)Met/Exceeded performance of predicate devicesSuccessfully met all predetermined acceptance criteria.
    Subsidence TestingASTM F2267 (Standard Test Method for Measuring Load Induced Subsidence of the Intervertebral Body Fusion Device under Static Axial Compression)Met/Exceeded performance of predicate devicesSuccessfully met all predetermined acceptance criteria.
    Expulsion TestingDRAFT ASTM F-04.25.02.02Met/Exceeded performance of predicate devicesSuccessfully met all predetermined acceptance criteria.

    The study concludes that "Based on the results the subject intervertebral devices demonstrated that they are as safe and effective as the predicate device(s)."

    2. Sample Size and Data Provenance for Test Set:

    No specific sample sizes for mechanical testing are provided in the document. The testing was non-clinical (laboratory-based) and not based on patient data, so "data provenance" in terms of country of origin or retrospective/prospective is not applicable.

    3. Number of Experts and Qualifications for Ground Truth (Test Set):

    Not applicable. This was a non-clinical, mechanical engineering study, not involving human expert assessment for ground truth.

    4. Adjudication Method for Test Set:

    Not applicable. As this was a non-clinical mechanical test, there was no adjudication method involving human interpretation of results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No MRMC comparative effectiveness study was done, as this was a non-clinical mechanical engineering study and did not involve human readers or AI assistance.

    6. Standalone Performance Study:

    Yes, a standalone performance study was done for the device in the sense of mechanical testing of the device itself (algorithm only without human-in-the-loop performance). The tests listed above are all standalone tests of the device's mechanical properties.

    7. Type of Ground Truth Used:

    The "ground truth" for the non-clinical tests was based on the performance requirements specified in the FDA guidance document "Class II Special Controls Guidance Document: Intervertebral Body Fusion Devices" and the applicable ASTM standards (ASTM F2077, ASTM F2267, and DRAFT ASTM F-04.25.02.02). The device's performance was compared against these established engineering standards and implicitly, against the performance of predicate devices.

    8. Sample Size for Training Set:

    Not applicable, as this was a non-clinical mechanical engineering study, not involving machine learning or a "training set" of data in the conventional sense.

    9. How Ground Truth for Training Set Was Established:

    Not applicable, as there was no training set in the context of this non-clinical mechanical study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K100967

    Validate with FDA (Live)

    Date Cleared
    2011-08-05

    (484 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PERIMETER® C Spinal System is intended to be used for anterior cervical interbody fusion procedures in skeletally mature patients with cervical disc disease at one level from the C2-C3 disc to the C7-T1 disc who have had six weeks of non-operative treatment. Cervical disc disease is defined as intractable radiculopathy and/or myelopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. Additionally, the PERIMETER® C Spinal System implants are to be used with autogenous bone graft and supplemental fixation and implanted via an open, anterior approach.

    Device Description

    The PERIMETER® C Spinal System consists of spacers of various widths and heights, which can be inserted between two cervical vertebral bodies to give support and correction during cervical interbody fusion procedures. Additionally, this implant has six degrees of lordosis and the superior and inferior surfaces of the implant are designed with teeth which interact with the surface of the vertebral endplates to aid in resisting expulsion. The hollow geometry of the implants allows them to be packed with autogenous bone graft and is to be used with supplemental fixation in all procedures. This device is offered in PEEK-OPTIMA® (i.e., POLYETHERETHERKETONE) with tantalum markers. Medical grade titanium implants and medical grade PEEK implants may be used together. Never use titanium or titanium alloy implants with stainless steel in the same construct.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the PERIMETER® C Spinal System, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriterionReported Device Performance
    Mechanical Test Standards:Met:
    Static Torsion (per ASTM F2077-03)Results indicate acceptance criteria were met.
    Static Compression (per ASTM F2077-03)Results indicate acceptance criteria were met.
    Dynamic Compression (per ASTM F2077-03)Results indicate acceptance criteria were met.
    Dynamic Torsion (per ASTM F2077-03)Results indicate acceptance criteria were met.
    Static Compression-Shear (per ASTM F2077-03)Results indicate acceptance criteria were met.
    Subsidence (per ASTM F2267-04)Results indicate acceptance criteria were met.
    Expulsion TestingResults indicate acceptance criteria were met.
    Overall Goal:Met:
    Demonstrate substantial equivalence to predicate devicesSubstantial equivalence demonstrated through mechanical tests and engineering rationale.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided 510(k) summary focuses on non-clinical mechanical testing for the PERIMETER® C Spinal System. Therefore, there is no "test set" in the context of patient data (e.g., medical images, clinical outcomes).

    • Sample Size for Test Set: Not applicable, as this refers to mechanical test specimens, not patient data. The document does not specify the number of individual test articles (implants) used for each mechanical test.
    • Data Provenance: Not applicable. The data comes from in-vitro mechanical testing, not a country of origin for patient data. This is a retrospective analysis of device performance against established ASTM standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. For mechanical testing, the "ground truth" is defined by the established ASTM standards and engineering principles. The results are compared against these predetermined criteria, not against expert human interpretations of patient data.

    4. Adjudication Method for the Test Set

    Not applicable. Mechanical test results are measured and directly compared to quantitative acceptance criteria defined by the ASTM standards. There is no need for adjudication as you would have for subjective interpretations of clinical findings.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    Not applicable. This submission is for a medical implant (spinal system), not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study comparing human reader performance with and without AI is not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is not an algorithm-based device.

    7. The Type of Ground Truth Used

    The ground truth used for demonstrating the device's performance against its acceptance criteria is:

    • Standardized Mechanical Test Criteria: Specifically, the requirements outlined in ASTM F2077-03 (Standard Test Methods for Intervertebral Body Fusion Devices) and ASTM F2267-04 (Standard Test Method for Evaluating the Subsidence of an Intervertebral Body Fusion Device Under Static Axial Compression).
    • Engineering Rationale: Complementary engineering analysis to support the mechanical test results.

    8. The Sample Size for the Training Set

    Not applicable. There is no "training set" as this is not a machine learning or AI-driven device requiring data for model training.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set, there is no ground truth to be established for it.