510(k) Data Aggregation

    K Number
    K092313

    Validate with FDA (Live)

    Date Cleared
    2009-11-04

    (97 days)

    Product Code
    Regulation Number
    880.5725
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Perfusor® Space Infusion Syringe Pump System is an electrical, external, syringe infusion pump system indicated for use with adults, pediatrics and neonates and is intended to provide infusions of parenteral fluids/medications, blood and blood products indicated for infusion through FDA approved routes of administration.

    Device Description

    The Perfusor® Space Infusion Syringe Pump System includes an external transportable electronic syringe infusion pump and pump accessories. The Perfusor Space Infusion Syringe Pump System utilizes a swivel-drive pumping mechanism and is intended to provide infusions of parenteral fluids. The Perfusor® Space Infusion Syringe Pump System is intended to be used by trained healthcare professionals in healthcare facilities, home care, outpatient, and medical transport environments. A trained Biomedical Technician must perform a complete set-up of the pump prior to use in a clinical setting. The system is intended to provide intermittent or continuous flow of parenteral fluids to the patient. Parenteral fluids may include standard fluids and/or medications indicated for infusion as well as blood and blood products.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the B. Braun Medical Inc. Perfusor® Space Infusion Syringe Pump System. The document focuses on establishing substantial equivalence to previously cleared devices rather than providing detailed acceptance criteria and study results for device performance in the way a clinical trial for a diagnostic AI device would.

    Therefore, much of the requested information (acceptance criteria, device performance, sample sizes, ground truth establishment, expert qualifications, adjudication methods, MRMC study, standalone performance) is not present in the provided text.

    Specifically, the document addresses new functionalities added to an existing device: barcode, out-bound wireless data transmission, and PCA (Patient-Controlled Analgesia) functionality. The primary claim is "substantial equivalence" to predicate devices, meaning it performs similarly and presents similar safety and effectiveness.

    Here's what can be extracted and what is missing:


    1. Table of Acceptance Criteria and Reported Device Performance

    Not available in the provided text. The document focuses on substantial equivalence based on comparable type, indications for use, and technological characteristics rather than specific performance metrics against pre-defined acceptance criteria.


    2. Sample Size Used for the Test Set and Data Provenance

    Not available. The document does not describe a "test set" or a study designed to measure performance metrics. It's a regulatory submission for device modification.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable. There is no mention of a ground truth being established by experts for a test set in this regulatory submission.


    4. Adjudication Method for the Test Set

    Not applicable. No test set or adjudication method is described.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. No MRMC study is mentioned. The document primarily discusses the technical features and intended use of the infusion pump system and its new functionalities. This is a medical device, not a diagnostic AI system, so MRMC studies are not typically conducted for this type of product.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is an infusion pump system, not an AI algorithm designed to interpret data or diagnose conditions. Its performance is inherent in its mechanical and software functions for drug delivery.


    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    Not applicable. The concept of "ground truth" as used in evaluating AI or diagnostic systems does not apply to this medical device submission. The "truth" here relates to the reliable and accurate functioning of the pump in delivering fluids/medications as intended.


    8. The Sample Size for the Training Set

    Not applicable. There is no mention of a "training set" as this is a traditional medical device, not an AI/machine learning device that would typically involve a training phase on data.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable. See point 8.


    Summary of the document's content relevant to evaluation:

    The document establishes that the Perfusor® Space Infusion Syringe Pump System, with its added barcode, out-bound wireless data transmission, and PCA functionalities, is substantially equivalent to predicate devices:

    • Predicate device for the original pump system: The Perfusor® Space Infusion Syringe Pump System (K062699).
    • Predicate device for barcode and out-bound wireless data transmission: B. Braun Medical Inc. Outlook® Safety Infusion System (K011975).
    • Predicate device for PCA functionality: MEDLEY™ PCA Module (K032233) by ALARIS Medical Systems Inc.

    The argument for substantial equivalence is based on:

    • Comparable type: All are computer-controlled electrical, external, syringe infusion pumps.
    • Similar indications for use: For infusion of parenteral fluids/medications, blood, and blood products for adults, pediatrics, and neonates.
    • Similar technological characteristics: Implementing advanced features (barcode, wireless, PCA) in a similar manner, using communication protocols (CAN bus) for data exchange with hospital information systems. All data transmitted uses B. Braun's proprietary information protocol to ensure data integrity.
    • Similar accessories: Providing clinician flexibility, organizing the bedside, central power supply, tubing retainers, ease of transportation, and customization of menu/drug library parameters.
    Ask a Question

    Ask a specific question about this device

    K Number
    K062699

    Validate with FDA (Live)

    Date Cleared
    2006-10-26

    (45 days)

    Product Code
    Regulation Number
    880.5725
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Perfusor Space Infusion Syringe Pump System includes an external transportable electronic infusion syringe pump and pump accessories. The system is intended for use on adults, pediatrics, and neonates for the intermittent or continuous delivery of parenteral fluids through clinically accepted routes of administration. These routes include, but are not limited to intravenous, intra-arterial, subcutaneous, epidural, irrigation/ablation, and enteral. The system is used for the delivery of medications indicated for infusion therapy including but not limited to drugs like anesthetics, sedatives, analgesics, catecholamines, anticoagulants etc., blood and blood components, Total Parenteral Nutrition (TPN), lipids, and enteral fluids. The Perfusor® Space Infusion Syringe Pump System is intended to be used by trained healthcare professionals in healthcare facilities, home care, outpatient, and medical transport environments (only road ambulances).

    Device Description

    The Perfusor® Space Infusion Syringe Pump System includes an external transportable electronic volumetric infusion pump and pump accessories. The Perfusor® Space is a 12V DC or battery powered external, transportable, infusion syringe pump. The Perfusor® Space utilizes a swivel-drive pumping mechanism is intended to provide infusions of parenteral and enteral fluids. The system is intended to provide intermittent or continuous flow of parenteral and enteral fluids to the patient. Parenteral fluids may include all standard fluids and/or medications indicated for infusion as well as blood and blood products. The Ferfusor Space uses standard, single-use, disposable syringes (with luer-lock connectors) designed for use on syringe pumps and validated on the Perfusor® Space. The B. Braun Space Station is a 12V DC or battery powered flexible docking and communication system for the medical workplace, in particular the intensive medical care. It serves the perspicuous accommodation of the infusion and infusion-syringe pumps Infusomat® Space and Perfusor® Space. SpaceCover Standard or SpaceCover Comfort forms the top of each column. The DrugListEditor Space is a programm to create a drug pool with drug names including the corresponding parameter.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for the B. Braun Medical Inc. Perfusor® Space Infusion Syringe Pump System. It details the device's description, intended use, and argues for its substantial equivalence to previously cleared predicate devices.

    This document describes a medical device, specifically an infusion syringe pump system, and demonstrates its substantial equivalence to existing devices for market clearance. It does NOT contain information about acceptance criteria or a study that proves the device meets those criteria in the context of performance metrics that would typically apply to AI/ML or diagnostic devices (e.g., sensitivity, specificity, AUC). Instead, the "acceptance criteria" here implicitly refers to meeting regulatory requirements for substantial equivalence to predicates.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance in the format usually requested for AI/ML studies, nor can I answer questions 2 through 9, as they pertain to detailed study design and ground truth establishment for performance evaluation, which are not present in this regulatory submission for a medical pump.

    The document's primary "study" is the comparison against predicate devices to establish substantial equivalence under the 510(k) pathway, which is a regulatory assessment, not a performance study in the AI/ML sense.

    Here's a breakdown of what can be extracted or inferred based on the provided text's context:


    1. Table of "Acceptance Criteria" and "Reported Device Performance" for Substantial Equivalence

    Acceptance Criteria (for Substantial Equivalence to Predicates)Reported Device Performance (vs. Predicates)
    Intended Use: Same as predicate devices.The Perfusor® Space Infusion Syringe Pump System has the "same intended use" as the Perfusor® compact with fm-System (K022575) and MEDLEY™ System with Medication Management System (K030459) together with MEDLEY™ Syringe Pump Module (K023264). Intended for adults, pediatrics, and neonates for intermittent or continuous delivery of parenteral and enteral fluids via various routes (intravenous, intra-arterial, subcutaneous, epidural, irrigation/ablation, enteral). Used by trained HCPs in various environments.
    Operation & Function: Similar to predicate devices.The subject device and predicates are all "computer controlled electrical, external, swivel-driven infusion syringe pumps composed of injection molded thermoplastic components." They control infusion rate of parenteral and enteral fluids, use standard, single-use, disposable syringes (with luer-lock connectors) validated with the pumps. Provide possibilities to clean up/organize patient bedside, central power, tubing retainers, transportation, and customization of menu items/drug library.
    Technology/Design: Similar to predicate devices.The components of the subject and predicate pumps are "all non solution contact parts." The accessories are similar in offering clinician flexibility and providing filing and communication systems.
    Safety and Effectiveness: No new questions of safety or effectiveness.Not explicitly stated in a quantitative manner, but implied by the successful 510(k) clearance, indicating the FDA found no new questions of safety or effectiveness compared to the predicate devices.

    Clarification: In the context of a 510(k) for a medical pump, "acceptance criteria" and "reported device performance" are judged against the predicate devices for establishing substantial equivalence, rather than a specific numerical diagnostic performance for an AI algorithm. The FDA's letter explicitly states: "We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent...".


    The following questions cannot be answered from the provided text as they relate to performance studies, especially those relevant to AI/ML or diagnostic devices, which are not covered in this 510(k) summary for an infusion pump system.

    1. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable/Provided. This document does not describe a performance study with a test set in the diagnostic sense.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable/Provided.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable/Provided.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable/Provided. This is not an AI-assisted diagnostic device.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable/Provided. This is an infusion pump, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable/Provided.
    7. The sample size for the training set: Not applicable/Provided.
    8. How the ground truth for the training set was established: Not applicable/Provided.