Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K113163

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-11-23

    (28 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Penumbra System is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (in the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset.

    Device Description

    The Penumbra System MAX components are additional components of the currently available Penumbra System. The Penumbra System MAX components provide a larger lumen to assist in the efficient removal of thrombus from the brain. The devices are provided sterile, non-pyrogenic, and intended for single use only.

    AI/ML Overview

    The Penumbra System® MAX components are additional components of the currently available Penumbra System. The summary of non-clinical data details the testing that substantiates the safe and effective performance of the Neuron MAX System and its substantial equivalence to predicate devices. This includes Biocompatibility testing, Design Verification (Bench-Top Testing), and an Animal study.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and the Reported Device Performance

    The acceptance criteria for each test were "Pass," meaning the device met the established requirements for each attribute.

    Test / Test SubjectAttributeReported Device Performance (Result)
    Biocompatibility Testing
    In Vitro CytotoxicityNo evidence of cell lysis or toxicityNo evidence of cell lysis or toxicity
    Acute Intracutaneous Reactivity (Irritation)No evidence of irritationNo evidence of irritation
    Acute Systemic ToxicityNo evidence of systemic toxicityNo evidence of systemic toxicity
    Rabbit Pyrogen StudyNo evidence of material-mediated pyrogenicityNo evidence of material-mediated pyrogenicity
    SensitizationNon-SensitizingNon-Sensitizing
    Hemo-compatibility - In Vitro HemolysisNon-hemolyticNon-hemolytic
    Hemo-compatibility - In Vitro Coagulation (PT, PTT)Coagulation times are within the normal range; Non-ThrombogenicWithin normal range; Non-Thrombogenic
    Hemo-compatibility - Complement ActivationNo greater biological response than corresponding controlNo greater biological response than corresponding control
    Bench-top Testing (Reperfusion Catheters)
    Pouch SealPouch Seal StrengthPass
    Dimensional / Visual InspectionUnits meet all inspection criteria for release of finished goods (clinically acceptable) product.Pass
    Simulated Use [Intracranial Access & Thrombus Removal]Units meet all inspection criteria for release of finished goods (clinically acceptable) product.Pass
    Reperfusion Catheter 3MAX / Reperfusion Catheter 4MAXHub /Shaft & Hub / HypotubeTensile StrengthPass
    Reperfusion Catheter 3MAX / Reperfusion Catheter 4MAXJoint Tensile StrengthPass
    Reperfusion Catheter 3MAX / Reperfusion Catheter 4MAXHub Air AspirationPass
    Reperfusion Catheter 3MAX / Reperfusion Catheter 4MAXBurst TestPass
    Reperfusion Catheter 3MAX / Reperfusion Catheter 4MAXParticulate Testing (Hydrophilic Coating)Pass
    Reperfusion Catheter4MAX / Sheath or 8F Guide compatibility (Friction Force)Pass
    Reperfusion Catheter 4MAX / 0.014" Guidewire compatibility (Friction Force)Pass
    Reperfusion Catheter 3MAX / Reperfusion Catheter 054 (Friction Force)Friction ForcePass
    Reperfusion Catheter 3MAX / 0.014" Guide wire compatibility (Friction Force)Pass
    Reperfusion Catheter 3MAX / Reperfusion Catheter 4MAXFlow RatePass
    Reperfusion Catheter 3MAX / Reperfusion Catheter 4MAXElongation to FailurePass
    Reperfusion Catheter 3MAX / Reperfusion Catheter 4MAXCorrosionPass
    Reperfusion Catheter 3MAX / Reperfusion Catheter 4MAXTorsionPass
    Bench-top Testing (Separator 3MAX)
    Dimensional / Visual Inspection (all sizes)Units meet all inspection criteria for release of finished goods (clinically acceptable) product.Pass
    Simulated Use: Hub TransitionPass
    Reperfusion Catheter 3MAX / Separator 3MAX / Aspiration Tubing Assembly PerformanceTrackingPass
    Reperfusion Catheter 3MAX / Separator 3MAX / Aspiration Tubing Assembly PerformanceSeparator 3MAX & Reperfusion Catheter CompatiblePass
    Reperfusion Catheter 3MAX / Separator 3MAX / Aspiration Tubing Assembly PerformanceSeparator 3MAX / Reperfusion Catheter / Pump and accessories / Aspiration Tubing (Aspiration Remove Clot)Pass
    Reperfusion Catheter 3MAX / Separator 3MAX / Aspiration Tubing Assembly PerformanceSeparator 3MAX and Reperfusion Catheter Compatibility (Separator Advance / Retract)Pass
    Separator Bond Joint TestSeparator 054 and Wire Joint Break ForcePass
    Separator Bond Joint TestSeparator 041 and Wire Joint Break ForcePass
    Separator Bond Joint TestSeparator 032 and Wire Joint Break ForcePass
    Separator Bond Joint TestSeparator 026 and Wire Joint Break ForcePass
    Separator Bond Joint TestSeparator 3MAX and Wire Joint Break ForcePass
    Animal Study
    Safe use of Penumbra System MAX in swine modelNo vessel injury noted on final angiograms following the vessel response procedure. No abnormal gross or histology findings in test vessel segments. No significant vascular response.No vessel injury; No abnormal findings; No significant vascular response

    2. Sample Size Used for the Test Set and the Data Provenance

    • Bench-top Testing: Sample sizes for individual tests ranged from 3 to 30 units (e.g., N=30 for Pouch Seal Strength, N=10 for Particulate Testing, N=3 for Flow Rate). The data provenance is internal laboratory testing ("All studies were conducted using good scientific practices and statistical sampling methods as required by the Penumbra Design Control procedures. All testing was performed using units which were 2x sterilized and met finished goods release requirements."). This is considered retrospective data from a manufacturing/design verification process.
    • Biocompatibility Testing: The document does not specify exact sample sizes for each biocompatibility test but states that "All studies were conducted pursuant to 21 CFR, Part 58, Good Laboratory Practices," which implies standard laboratory animal testing where applicable (e.g., Rabbit Pyrogen Study, Acute Intracutaneous Reactivity, Acute Systemic Toxicity, Sensitization). This data is from controlled laboratory experiments.
    • Animal Study: The document mentions "An animal study was conducted to evaluate the safe use of the Penumbra System MAX in a swine model." The exact sample size for the animal study is not explicitly provided in the summary, but it refers to "experimental conditions" and "test vessel segments," suggesting a controlled study within a lab setting. This is prospective animal data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of those Experts

    Not applicable. The studies described are non-clinical (bench-top, biocompatibility, animal studies) and do not involve human diagnostic or treatment decision-making where expert consensus would establish ground truth for a test set. The "ground truth" here is determined by objective physical, mechanical, biological, or physiological measurements against predefined engineering and biological safety standards.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, these are non-clinical studies. The results are objective measurements/observations (e.g., "Pass," "No evidence of toxicity," "No vessel injury") rather than interpretations requiring adjudication among human experts.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document describes the safety and effectiveness testing of a physical medical device (catheter system) for mechanical thrombus removal, not an AI or imaging-based diagnostic device that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This document describes a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for these non-clinical studies was established by:

    • Predefined engineering specifications and performance parameters for mechanical and physical properties.
    • Established international and national standards for biocompatibility (e.g., ISO 10993, USP, ASTM, 21 CFR Part 58).
    • Physiological observations, angiographic evaluations, gross pathology, and histology in the animal model.

    8. The Sample Size for the Training Set

    Not applicable. This document refers to the testing of a physical medical device. There is no "training set" in the context of machine learning or AI. The design verification and biocompatibility testing involved various "test units" or "samples" for specific tests, as detailed in the table, to ensure the device met its design specifications.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, for the same reason as #8. Ground truth in this context refers to the defined specifications and standards the device was designed to meet and was tested against.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1