510(k) Data Aggregation

    K Number
    K100826

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-07-13

    (111 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Penumbra Catheter 025 is intended to assist in the delivery of diagnostic agents, such as contrast media, and therapeutic agents, such as occlusion coils to the peripheral and neuro vasculature.

    Device Description

    The Penumbra Catheter 025 is a variable stiffness, coil reinforced catheter with a maximum distal outer diameter of 0.040" and a maximum proximal outer diameter of 0.045". It is available with an inner diameter of 0.025". The Penumbra Catheter 025 has a PTFE-lined lumen, is coil re-enforced, flexible, and hydrophilically coated. The Penumbra Catheter 025 is inserted through a guide catheter or vascular sheath, provides access to the target site and once in place, provides a reinforcing conduit for other intravascular devices. The device is provided sterile and includes a rotating hemostasis valve, tip shaping mandrel and introducer sheath. The Penumbra Catheter 025 will be available in various tip configurations including straight, 45°, 90° and J, to allow physician ease of device tracking to the target site. The Penumbra Catheter 025 is sterile, non-pyrogenic and intended for single use only.

    AI/ML Overview

    The Penumbra Catheter 025 is a medical device and is not an AI/ML device. Therefore, the questions related to AI/ML development (e.g., training set, ground truth experts, effect on human readers) are not applicable.

    The acceptance criteria and the study proving the device meets these criteria are detailed in the "Summary of Non-clinical Data" section, which includes Biocompatibility and Design Verification (Bench-Top Testing). The studies conducted for this device were for 510(k) clearance, establishing substantial equivalence to predicate devices, rather than a clinical trial demonstrating efficacy against a specific medical condition.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of "acceptance criteria" alongside specific numerical "reported device performance." Instead, it lists the tests performed and states that "All tests performed passed successfully." For biocompatibility, specific results are given qualitative descriptions.

    Test CategorySpecific TestMethodAcceptance Criteria (Implied) & Reported Device Performance
    BiocompatibilityCytotoxicityL929 MEM Elution TestNon-cytotoxic / Slight cell lysis or reactivity (Acceptable)
    SensitizationKligman MaximizationNon-sensitizing
    Intracutaneous Reactivity (Irritation)Intracutaneous Injection TestNon-irritant
    Systemic Toxicity (Acute)ISO Acute Systemic Injection TestNon-toxic
    Hemocompatibility - Complement ActivationComplement Activation TestNo greater biological response than corresponding control
    Hemocompatibility - HemolysisHemolysis TestNon-hemolytic
    Hemocompatibility - In vivo thrombogenicityIn vivo thrombogenicity TestNon-thrombogenic
    PyrogenicityUSP Material Mediated Rabbit Pyrogen TestNon-pyrogenic
    Design VerificationDimensional / Visual InspectionNot specified (Standard QC)Passed successfully
    TrackabilityNot specified (Bench-top simulation)Passed successfully
    FrictionNot specified (Bench-top test)Passed successfully
    TorsionNot specified (Bench-top test)Passed successfully
    Joint Tensile StrengthNot specified (Bench-top test)Passed successfully
    Catheter BurstNot specified (Bench-top test)Passed successfully
    Catheter Flow RateNot specified (Bench-top test)Passed successfully
    Catheter ElongationNot specified (Bench-top test)Passed successfully
    Kink ResistanceNot specified (Bench-top test)Passed successfully
    CorrosionNot specified (Bench-top test)Passed successfully
    GLP Simulated UseNot specified (Bench-top simulation)Passed successfully

    In summary, the statement "In summary, non-clinical testing found the Penumbra Catheter 025 to be non-cytotoxic, non-mutagenic, non-reactive (short and long-term implantation), non-sensitizing, a negligible irritant, non-pyrogenic, and non-toxic (acute systemic)" acts as the overarching reported performance for biocompatibility. For design verification, "All tests performed passed successfully" serves as the reported performance, implying the device met the established acceptance criteria for each test.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: The document states that "statistical sampling methods as required by the Penumbra Design Control procedures" were used, but it does not specify the exact sample sizes for each test (e.g., how many catheters were tested for burst strength or kink resistance). For biocompatibility, standard animal models or cell cultures are typically used, but specific numbers are not provided.
    • Data Provenance: The studies were non-clinical (bench-top and in vitro/in vivo biocompatibility tests, likely animal studies for some bio-tests). No information is provided regarding the country of origin of the data, but the manufacturer is based in Alameda, CA, USA, and the submission is to the US FDA, implying testing was conducted to US regulatory standards. These are prospective tests designed specifically for the device's 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This question is not applicable. The Penumbra Catheter 025 is a physical medical device, not an AI/ML diagnostic or prognostic tool. "Ground truth" in the context of expert medical interpretation or diagnosis is not relevant for this type of device. The "ground truth" for this device would be established by the objective results of the physical and chemical tests (e.g., did it burst at X pressure, did it cause a cytotoxic reaction).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not applicable for a physical medical device. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or for establishing ground truth for AI/ML algorithms where expert consensus on interpretations is required. For this device, the tests have objective pass/fail criteria.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. MRMC studies and the concept of "human readers improving with AI assistance" are relevant for AI/ML-driven diagnostic tools, not for a physical medical catheter.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. This refers to AI/ML algorithm performance, which is not relevant to the Penumbra Catheter 025.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device is established by objective physical, mechanical, and chemical test results. For example:

    • Biocompatibility: Laboratory standard assays (e.g., L929 MEM Elution for cytotoxicity, ISO Acute Systemic Injection Test for toxicity), which have defined biological responses or lack thereof as "ground truth."
    • Design Verification: Engineering specifications and performance metrics (e.g., a catheter burst at or above a certain pressure, trackability within defined parameters). The "ground truth" is measured and compared against established engineering and safety standards, often derived from predicate devices or industry guidelines.

    8. The sample size for the training set

    This question is not applicable. There is no "training set" in the context of a physical medical device. Training sets are used in AI/ML model development.

    9. How the ground truth for the training set was established

    This question is not applicable as there is no "training set."